K Number
K212723

Validate with FDA (Live)

Device Name
BTL-995-rTMS
Date Cleared
2022-03-04

(189 days)

Product Code
Regulation Number
882.5805
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BTL-995-rTMS is indicated to be used for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

Device Description

The BTL-995-rTMS is a non-invasive therapeutic device that produces and delivers magnetic field to induce electrical currents targeting specific regions of the human cerebral cortex. BTL-995-rTMS consists of a main unit and applicator(s). The main unit consists of a master unit and a generator unit. The main unit is equipped with a color touch screen that makes the device easy to use. The on-screen information guides the Operator through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen. During therapy, the screen displays information about the remaining therapy time and other therapy parameters. The device functions with two applicator types: figure-of-8 coil applicator and circular coil applicator. The therapy is provided with the applicator attached to the applicator arm.

AI/ML Overview

The provided document is a 510(k) Summary for the BTL-995-rTMS device. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical study results and acceptance criteria for algorithmic performance.

Based on the document, here's an analysis of the requested information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria for device performance in the form of metrics (e.g., sensitivity, specificity, accuracy) that would typically be associated with an AI/machine learning device. This is because the device is a therapeutic medical device (rTMS system), not a diagnostic or AI-powered analytical tool. The "performance" discussed primarily relates to electrical safety and electromagnetic compatibility.

  • Acceptance Criteria (Implicit from Non-Clinical Testing): Compliance with specific international standards for medical electrical equipment.
  • Reported Device Performance:
    • Electrical Safety: Complies with IEC 60601-1
    • Electromagnetic Disturbances: Complies with IEC 60601-1-2
    • Usability: Complies with IEC 60601-1-6
    • Software Life Cycle Processes: Complies with IEC 62304
    • Risk Management: Complies with ISO 14971
    • Biological Evaluation (Cytotoxicity): Complies with ISO 10993-5
    • Biological Evaluation (Irritation & Skin Sensitization): Complies with ISO 10993-10

For the therapeutic parameters, the device reports the following parameters, which are comparable to the predicate device, implying that the device "performs" by delivering these specified outputs:

Therapeutic ParameterReported Performance (BTL-995-rTMS)
Magnetic Field Intensity120% of patient's measured hand Motor Threshold (MT)
Frequency10 Hz
Treatment train duration4 sec
Inter-train interval11-26 sec
Number of pulses administered per session3000

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "No clinical testing was performed for this device." Therefore, there is no test set, sample size, or data provenance to report in the context of clinical or AI performance. The evaluation was based on non-clinical (bench) testing and comparison to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. As no clinical testing or AI algorithm performance evaluation was conducted, there was no ground truth to establish by experts for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical testing or AI algorithm performance evaluation, so no adjudication method was used for a test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The BTL-995-rTMS is a therapeutic device, not a diagnostic imaging or AI assistance tool for human readers. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a Repetitive Transcranial Magnetic Stimulation (rTMS) system, a physical medical device, not an AI algorithm. Its function is to deliver magnetic pulses for therapeutic purposes, not to provide an independent diagnosis or analysis.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable in the context of AI/diagnostic evaluation. The "ground truth" for this device's performance is its compliance with safety and performance standards (bench testing) and its ability to deliver the specified therapeutic parameters as designed.

8. The sample size for the training set

Not applicable. The device is not an AI/machine learning system that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth establishment for it.

{0}------------------------------------------------

Build Correspondence

Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 4, 2022

BTL Industries Inc. David Chmel VP of Operations 362 Elm Street Marlborough, Massachusetts 01752

Re: K212723

Trade/Device Name: BTL-995-rTMS Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: February 9, 2022 Received: February 9, 2022

Dear David Chmel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Pamela Scott, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212723

Device Name BTL-995-rTMS

Indications for Use (Describe)

BTL-995-rTMS is indicated to be used for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three nested squares with the letters "BTL" inside the innermost square. To the right of the logo is the text "BTL Industries" in a simple, sans-serif font.

510(k) Summary

General Information

Sponsor:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Applicant:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Contact Person:David ChmelBTL Industries, Inc.chmel@btlnet.com
Summary PreparationDate:February 1, 2022

Device Name

Trade/Proprietary Name:BTL-995-rTMS
Primary Classification Name:Transcranial Magnetic Stimulator
Classification Regulation:882.5805, Class II
Classification Product Code:OBP

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interlocking blue squares on the left side of the image. To the right of the squares is the text "BTL Industries" in a dark gray sans-serif font.

Legally Marketed Predicate Device

The BTL-995-rTMS is a state-of-the-art device electromagnetic device with accessories and is substantially equivalent to the current product that is already cleared for distribution in the USA under the following 510(k) Premarket Notification number:

  • . MaqVita TMS Therapy System (K150641)

Product Description

The BTL-995-FTMS is a non-invasive therapeutic device that produces and delivers magnetic field to induce electrical currents targeting specific regions of the human cerebral cortex.

BTL-995-rTMS consists of a main unit and applicator(s). The main unit consists of a master unit and a generator unit. The main unit is equipped with a color touch screen that makes the device easy to use. The on-screen information guides the Operator through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen. During therapy, the screen displays information about the remaining therapy time and other therapy parameters.

The device functions with two applicator types: figure-of-8 coil applicator and circular coil applicator. The therapy is provided with the applicator attached to the applicator arm.

Indications for Use

BTL-995-rTMS is indicated to be used for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

Clinical Testing

No clinical testing was performed for this device.

Non-clinical Testing (Performance, Bench Testing)

The BTL-995-rTMS device has been thoroughly evaluated for electrical safety. The device has been found to comply with the following applicable medical device safety standards:

IEC 60601-1Medical electrical equipment – Part 1: General requirements for basic safetyand essential performance
IEC 60601-1-2Medical electrical equipment – Part 1-2: General requirements for basic safetyand essential performance - Collateral standard: Electromagnetic disturbances– Requirements and tests
IEC 60601-1-6Medical electrical equipment – Part 1-6: General requirements for basic safetyand essential performance – Collateral standard: Usability
IEC 62304Medical device software - Software life cycle processes

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a blue geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three squares, each containing the letters B, T, and L respectively.

ISO 14971Medical devices – Application of risk management to medical devices
ISO 10993-1Biological evaluation of medical devices – Part 1: Evaluation and testing withina risk management process
ISO 10993-5Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
ISO 10993-10Biological evaluation of medical devices – Part 10: Tests for irritation and skinsensitization

Technological Characteristics

The BTL-995-rTMS device has the same intended use and principles of operation to its predicate device. The BTL-995-rTMS device and its predicate are comprised of a system console, arm for applicator positioning and applicator(s).

The mechanism of action and technological similarities and differences between the BTL-995rTMS device and the predicate device are described below in the comparison table. The differences do not raise any new types of safety or effectiveness questions.

510(k) numberK212723K150641
Device nameCompany nameBTL-995-rTMSBTL Industries, Inc.MagVita TMS Therapy SystemTonica Elektronik A/S
Product Codeand RegulationNeurology21 CFR 882.5805OBP - Transcranial MagneticStimulatorNeurology21 CFR 882.5805OBP - Transcranial MagneticStimulator
Indications for UseBTL-995-rTMS is indicated to beused for the treatment of MajorDepressive Disorder in adultpatients who have failed to achievesatisfactory improvement from priorantidepressant medication in thecurrent episode.The MagVita TMS Therapy System isindicated for the treatment of MajorDepressive Disorder in adult patientswho have failed to receive satisfactoryimprovement from priorantidepressant medication in thecurrent episode.
Clinical usePrescription usePrescription use
Area of brain to be stimulatedDorsolateral prefrontal cortexDorsolateral prefrontal cortex
Applicator:
Configurationbiphasicbiphasic
Core materialairair

Comparison with the Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interconnected blue squares, each containing the letters "B", "T", and "L" respectively. To the right of the square logo is the text "BTL Industries" in a dark gray sans-serif font.

510(k) numberK212723K150641
Device nameCompany nameBTL-995-rTMSBTL Industries, Inc.MagVita TMS Therapy SystemTonica Elektronik A/S
MajorDepressiveDisorderMagnetic Field Intensity: 120% ofpatient's measured hand MotorThreshold (MT).Magnetic Field Intensity: 120% MT(MT=Motor Threshold intensity)
Frequency: 10 HzFrequency: 10 Hz
Treatment train duration: 4 secTreatment train duration: 4 sec
Inter-train interval: 11-26 secInter-train interval: 26 sec
Number of pulses administered persession: 3000Number of pulses administered persession: 3000
Mobile consoleyesyes
System software with GUIyesyes
Coil positioning systemyesyes
CoolingAir cooledUsed for both MT determination andtreatment.Liquid cooledUsed for both MT determination andtreatment.
Electrical safetyComplies with IEC60601-1 and IEC60601-1-2Complies with IEC60601-1 and IEC60601-1-2

Substantial Equivalence

The BTL-995-rTMS device has the same intended use as its predicate device. The technological characteristics of the predicate device are similar to the BTL-995-rTMS device. Any differences between the predicate device and BTL-995-rTMS have no significant influence on safety and effectiveness of the BTL-995-rTMS device. Therefore, the BTL-995-rTMS is substantially equivalent to the predicate device.

Conclusion

Based upon the intended use and the known technical data provided in this pre-market notification, the BTL-995-rTMS device has been shown to be substantially equivalent to the currently marketed predicate device.

§ 882.5805 Repetitive transcranial magnetic stimulation system.

(a)
Identification. A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.(b)
Classification. Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.