K Number
K212623

Validate with FDA (Live)

Date Cleared
2022-03-16

(210 days)

Product Code
Regulation Number
866.3740
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Strep A Rapid Test Strip (Throat Swab) is a rapid chromatographic immunoassy for the qualitative detection of Streptococcus pyogenes (Group A B-hemolytic Streptococcus, Strep A) antigen from throat swab specimens of symptomatic patients to aid in the diagnosis of Group A Streptococcus bacterial infection.

All negative test results should be confirmed by bacterial culture because negative results do not prection with Group A Streptococcus and should not be used as the sole basis for treatment.

Device Description

Healgen Strep A Rapid Test Strip (Throat Swab) is a qualitative, lateral flow immunoassay for the detection of Strep A antigen directly from a throat swab sample.

In this test, antibody specific to Strep A carbohydrate antigen is coated on the test line region of the test. During testing, the extracted throat swab specimen reacts with an antibody to Strep A that is coated onto particles. The mixture migrates up the membrane to react with the antibody to Strep A on the membrane and generate a color line in the test line region. The presence of this color line in the test line region indicates a positive result, while its absence indicates a negative result. To serve as a procedural control, a colored line will always appear in the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Healgen Strep A Rapid Test Strip (Throat Swab):

1. Table of Acceptance Criteria and Reported Device Performance

Device Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Analytical Performance
Precision/ReproducibilityResults should be reproducible across different users, sites, lots, and days.Overall Positive Detection: - True negative sample: 0% (0/180) - Moderate positive sample (1.8x10^4 CFU/mL): 100% (180/180) - LoD sample (7.2x10^3 CFU/mL): 95.6% (172/180) - Low negative sample (3.6x10^3 CFU/mL): 44.4% (80/180) (Concluded: No significant differences between users, sites, lots, and days; results are reproducible with good precision.)
Limit of Detection (LoD)The LoD should be clearly established for both clinical matrix and saline solution.LoD in clinical matrix: 7.2x10^3 CFU/mL (equivalent to 360 bacteria on the swab) based on 95.2% detection (20/21) at 7.2x10^4 CFU/mL. LoD in saline solution: 7.2x10^3 CFU/mL (equivalent to 360 bacteria on the swab) based on 95.2% detection (20/21) at 7.2x10^3 CFU/mL.
InterferenceNo false positive or false negative results with common interfering substances (blood, mucus, saliva, medications).No false positive or false negative results observed with various interfering substances (blood, mucin, OTC mouthwashes, lozenges, throat sprays, cough syrups, active ingredients such as acetaminophen, ibuprofen, etc.) at tested concentrations (e.g., 20% vol/vol for liquids, 5mg/mL for solids).
Analytical SpecificityNo cross-reactivity with other common respiratory tract organisms (bacteria and viruses).No cross-reactivity found for a comprehensive list of organisms (e.g., Arcanobacterium haemolyticum, Bordetella pertussis, Candida albicans, Enterococcus faecalis, Escherichia coli, various Streptococcus species, Adenovirus, Cytomegalovirus, HSV, etc.) at tested concentrations.
Clinical Performance
Clinical SensitivityPerformance comparable to the legally marketed predicate device (Predicate: 95% CI (88-98%)).Overall Clinical Sensitivity: 97.1% (200/206) with 95% CI (93.7-98.8%) - Age 0-5: 97.4% (74/76) - Age 5-21: 96.7% (119/123) - Age 21+: 100% (7/7) (No statistical differences between age groups.)
Clinical SpecificityPerformance comparable to the legally marketed predicate device (Predicate: 98% CI (96-99%)).Overall Clinical Specificity: 99.4% (161/162) with 95% CI (96.2-100.0%) - Age 0-5: 98.1% (52/53) - Age 5-21: 100% (88/88) - Age 21+: 100% (21/21) (No statistical differences between age groups.)
ConsistencyPerformance across different age groups should be consistent.No statistical differences in performance between age groups.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size:
    • Clinical Study: 368 subjects (206 culture-positive, 162 culture-negative).
    • Analytical Precision: 180 determinations per sample type (60 determinations per site across 3 sites).
    • Analytical LoD: 21 results per dilution (7 operators x 3 lots).
    • Analytical Interference: Multiple tests across 3 lots for each interfering substance, for both positive and negative specimens.
    • Analytical Specificity (Cross-reactivity): Multiple tests across 3 lots for each organism by 3 professional users.
  • Data Provenance:
    • The document does not explicitly state the country of origin for the clinical data.
    • The clinical study appears to be prospective/concurrent as it describes testing "subjects...exhibiting symptoms of pharyngitis by both the Healgen Strep A Rapid Test Strip (Throat Swab) and the culture studies." This implies collection of samples and testing using both methods at the time of study.
    • The analytical studies (precision, LoD, interference, specificity) were laboratory-based, performed internally or by designated personnel.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Clinical Study Ground Truth: The ground truth for the clinical study was established by bacterial culture. The document implies that the culture results were considered the reference standard. It does not explicitly state the number of experts or their qualifications involved in performing or interpreting these cultures. However, bacterial culture is a standard clinical laboratory method typically performed by trained medical technologists or microbiologists.
  • Analytical Studies:
    • Precision/Reproducibility: 6 professional operators (2 at each of 3 sites) participated. Their specific qualifications are not detailed beyond "professional operators."
    • LoD: 7 operators performed the testing. Their specific qualifications are not detailed beyond "operators."
    • Interference: 3 laboratory assistants with relevant experience performed the test.
    • Analytical Specificity: 3 professional users performed the test.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • The document does not describe an adjudication method (like 2+1 or 3+1) for establishing the ground truth in the clinical study. The reference standard (bacterial culture) appears to have been used directly.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done in the context of AI assistance. This device is a manual rapid diagnostic test strip, not an AI-assisted diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not applicable here. The "operators" or "users" in the analytical studies are performing the manual test according to instructions.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This question is not applicable as the device is a manual rapid test strip, not an algorithm or AI-driven system. It does not operate without human interaction.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The ground truth for the clinical study was bacterial culture, which is considered the gold standard for diagnosing Group A Streptococcus bacterial infection.
  • For the analytical studies (LoD, interference, specificity), the ground truth was based on known concentrations of target analytes (S. pyogenes) or known presence/absence of interfering/cross-reacting organisms.

8. The sample size for the training set

  • The document does not mention a training set in the context of machine learning or AI. This is a traditional rapid diagnostic test, not a learning algorithm. The "training" of the device refers to its manufacturing and validation process, not data-driven machine learning.

9. How the ground truth for the training set was established

  • This question is not applicable as there is no mention or indication of a "training set" for an AI or machine learning model in this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 16, 2022

Healgen Scientific, LLC % Joe Shia Director LSI International Inc. 504 East Diamond Ave. Suite I Gaithersburg, Maryland 20877

Re: K212623

Trade/Device Name: Healgen Strep A Rapid Test Strip (Throat Swab) Regulation Number: 21 CFR 866.3740 Regulation Name: Streptococcus Spp. Serological Reagents Regulatory Class: Class I Product Code: GTY Dated: August 16, 2021 Received: August 18, 2021

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar. Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212623

Device Name Strep A Rapid Test Strip (Throat Swab)

Indications for Use (Describe)

The Strep A Rapid Test Strip (Throat Swab) is a rapid chromatographic immunoassy for the qualitative detection of Streptococcus pyogenes (Group A B-hemolytic Streptococcus, Strep A) antigen from throat swab specimens of symptomatic patients to aid in the diagnosis of Group A Streptococcus bacterial infection.

All negative test results should be confirmed by bacterial culture because negative results do not prection with Group A Streptococcus and should not be used as the sole basis for treatment.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY
K212623
1.Date:March 7, 2022
2.Submitter:HEALGEN SCIENTIFIC LLC
5213 Maple St
Bellaire, TX77401
3.Contact person:Joe Shia
LSI International Inc.
504 East Diamond Ave.
Gaithersburg, MD 20877
Telephone: 240-505-7880
Email:shiajl@yahoo.com
4.Device Name:Healgen Strep A Rapid Test Strip (Throat Swab)

Class I

Product CodeCFR #Panel
GTY866.3740 streptococcus spp serological reagentsMicrobiology
    1. Predicate Devices: K133343 Wondfo Strep A Rapid Test Guangzhou Wondfo Biotech Co., Ltd.
      Class:

6. Intended Use/Indications for Use

The Strep A Rapid Test Strip (Throat Swab) is a rapid chromatographic immunoassay for the qualitative detection of Streptococcus pyogenes (Group A B-hemolytic Strep A) antigen from throat swab specimens of symptomatic patients to aid in the diagnosis of Group A Streptococcus bacterial infection.

All negative test results should be confirmed by bacterial culture because negative results do not preclude infection with Group A Streptococcus and should not be used as the sole basis for treatment.

    1. Device Description
      Healgen Strep A Rapid Test Strip (Throat Swab) is a qualitative, lateral flow immunoassay for the detection of Strep A antigen directly from a throat swab sample.

In this test, antibody specific to Strep A carbohydrate antigen is coated on the test line region of the test. During testing, the extracted throat swab specimen reacts with an antibody to Strep A that is coated onto particles. The mixture migrates up the membrane to react with the antibody to Strep A on the membrane and

{4}------------------------------------------------

generate a color line in the test line region. The presence of this color line in the test line region indicates a positive result, while its absence indicates a negative result. To serve as a procedural control, a colored line will always appear in the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

    1. Substantial Equivalence Information
      A summary comparison of features of the Healgen Strep A Rapid Test Strip (Throat Swab) and the predicate device is provided in Table 1.
Similarities
ItemDevicePredicate (K133343)
Intended UseFor the qualitative detectionof group A streptococcalantigen directly from throatswabs.Same
SpecimenThroat swabSame
Assay technicalImmunochromatographicSame
Control AntibodiesGoat polyclonal anti-Rabbit IgGSame
Test AntibodyRabbit Polyclonal Anti-Strep ASame
Indication for UsePrescription UseSame
Differences
ItemDevicePredicate (K133343)
Test formatStripCassette
Analytical sensitivity$7.2×10^3$ CFU/mL1.5 x $10^5$ organisms/mL
Clinical Sensitivity97.1%: 95% CI (93.7-98.8%)95%: 95% CI (88-98%)
Clinical Specificity99.4%: 95% CI (96.2-100.0%)98%: 95% CI (96-99%)
Results Reading Time5 minutes10 minutes
    1. Test Principle
      Group A Streptococcus reacts with the antibody conjugated to the gold particle. The complex is then bound by the anti-Strep A capture antibody and a visible red test line appears, indicating a positive result. To serve as an onboard procedural control line observed at the control site prior to running the assay will turn red, indicating that the test has been performed properly.
    1. Performance Characteristics
      1. Analytical Performance
      • a. Precision/Reproducibility

A test panel consists of a true negative sample (diluent only), a low negative sample (3.6x102 CFU/mL), a moderate positive sample (1.8×10* CFU /mL) and a LoD sample (7.2×103 CFU/mL) were tested. Three lots

{5}------------------------------------------------

of the device are used. The study is performed at two runs per day in 5 different days at three different sites. Six professional operators who don't know the sample number code participated in the study (two operators at each site). Each operator tests two runs per day at each concentration with three lots of Healgen Strep A Rapid Test Strip (Throat Swab). A total of 30 determinations by each operator at each concentration are made. The obtained results are shown in the following table.

ResultsPositive Agreement % (Positive/Total tested)Overall PositiveDetection
SamplesSite ASite BSite C
True negative sample(Diluent only)0%(0/60)0%(0/60)0%(0/60)0%(0/180)
Moderate positive sample,$1.8\times10^4$ CFU/mL100%(60/60)100%(60/60)100%(60/60)100%(180/180)
LoD sample,$7.2\times10^3$ CFU/mL98.3%(59/60)93.3%(56/60)95.0%(57/60)95.6%(172/180)
Low negative sample,$3.6\times10^3$ CFU/mL48.3%(29/60)43.3%(26/60)41.7%(25/60)44.4%(80/180)

It's concluded that there are no significant differences of the test results obtained between different users, different sites and different lots in different days. The obtained results are reproducible in good precision.

  • b.Linearity
    • Not applicable
  • c. Stability

Stable at 4-30°C for 24 months based on real time stability determination at both 4°C and 30°C.

d. LoD

The limit of detection (LoD) for the Healgen Strep A Rapid Test Strip was established using limiting dilutions of Streptococcus pyogenes and tested by spiking with clinical matrix. A concentrated stock (3.6 × 107 CFU/mL) of inactivated S. pyogenes ATCC 19615 was serially diluted in saline solution. 5 uL of each contrived dilution sample was pipetted onto negative throat swab clinical matrix for testing. Each dilution was tested by seven operators with three lots of Healgen Strep A Rapid Test Strip (Throat Swab), for a total of 21 results for each dilution. The test results are shown in the following table.

Table: Determining Limit of Detection (LoD) for Healgen Strep A Rapid Test in clinicalmatrix
DilutionsNumber of bacteria loaded*(5 uL per swab)Positive/Tested% Detection
1.8×105 CFU/mL90021/21100%
7.2×104 CFU/mL36020/2195.2%
3.6×104 CFU/mL18012/2157.6%
1.8×104 CFU/mL901/214.8%
4.5×103 CFU/mL230/210%

*Calculated values based on dilution and volume of sample loaded on the swab per test.

{6}------------------------------------------------

The LoD was determined to be 7.2×10* CFU/mL when 5 uL of sample was pipetted onto a negative clinical matrix (equivalent to 360 bacteria on the swab).

The LoD for the Healgen Strep A Rapid Test Strip was also examined by using limiting dilutions of S. pyogenes ATCC 19615 in saline and pipetting 50 uL of each dilution with equivalent bacterial amounts as those used with the clinical matrix onto a swab.

DilutionsNumber of bacteria loaded*(50 µL per swab)Positive/Tested% Detection
1.8×104 CFU/mL90021/21100%
7.2×103 CFU/mL36020/2195.2%
3.6×103 CFU/mL18011/2147.6%
1.8×103 CFU/mL901/214.8%
4.5×103 CFU/mL230/210%

Table: Determining Detection of Limit (LoD) for Healgen Strep A Rapid Test in Saline Solution

*Calculated values based on dilution and volume of sample loaded on the swab per test.

The LoD was determined to be 7.2×103 CFU/mL when 50 µL of sample was pipetted onto a swab (equivalent to 360 bacteria on the swab) which is consistent with the LoD established using negative throat swab clinical matrix.

e. Interference

The potentially interfering substances of blood, mucus, saliva, and medications used to relieve a sore throat, such as over-the-counter cough drops, lozenges, cough syrups, throat sprays, mouth wash etc. were tested with Healgen Strep A Rapid Test Strip (Throat Swab). Each potentially interfering substance was diluted and splitted into two aliquot was spiked with S. pyogenes to a final concentration of 1.44 x 104 CFU/ml. The second aliquot contained no bacteria. These aliquot samples were tested by three batches of Healgen Strep A Rapid Test Strip (Throat Swab). Three laboratory assistants with relevant experience performed the test. The obtained results are shown in the following table.

Interfering SubstanceConcentration Tested$1.44 \times 10^4$ CFU/mL S. pyogenes Positive specimen (2×LOD)S. pyogenes Negative Specimen
Lot1Lot2Lot3Lot1Lot2Lot3
Blood (human)20% (vol/vol)+++---
Mucin1mg/mL+++---
OTC Mouthwashes
Listerine Antiseptic20%(vol/vol)+++---
Cool Mint20%(vol/vol)+++---
Crest Pro-Health Clean Mint20%(vol/vol)+++---
Crest Pro Health20%(vol/vol)+++---
Multi Protection Clean Mint
Colgate Total Pro-Shield,Spearmint20%(vol/vol)+++---
OTC Lozenges
Sucrets Sore Throat & CoughLozenges, Honey Lemon,5mg/mL+++---
Sucrets Sore ThroatLozenges Cherry5mg/mL+++---
Halls Mentho-Lyptus DropsCherry5mg/mL+++---
Halls Cough Suppressant CherryTriple Soothing Action5mg/mL+++---
Cepacol Extra Strength SoreThroat & Cough Drop Lozenges,Cherry5mg/mL+++---
OTC Throat Sprays
Cepacol Dual Relief20%(vol/vol)+++---
Chloraseptic Max20%(vol/vol)+++---
OTC Cough Syrups
Tylenol Cough and Sore Throat10%(vol/vol)+++---
Basic Care Tussin DM, CoughSuppressant & Expectorant10%(vol/vol)+++---
Robitussin (Guaifenesin Syrup)10%(vol/vol)+++---
Robitussin Nighttime Cough10%(vol/vol)+++---
Children's Dimetapp Cold & Flu10%(vol/vol)+++---
Children's Dimetapp Cold &Cough10%(vol/vol)+++---
Active Ingredients
Acetaminophen (Tylenol)5mg/mL+++---
Brompheniramine Maleate5mg/mL+++---
Chlorpheniramine Maleate5mg/mL+++---
Dextromethorphan HBr5mg/mL+++---
Diphenhydramine HCI5mg/mL+++---
Doxylamine Succinate5mg/mL+++---
Guaifenesin(Guaiacol Glyceryl)5mg/mL+++---
Ibuprofen (Advil)5mg/mL+++---
Phenylephrine HCI5mg/mL+++---

{7}------------------------------------------------

{8}------------------------------------------------

Neither false positive nor false negative results are shown in the Healgen Strep A Rapid Test Strip (Throat Swab) at the concentrations listed.

  • f. Analytical Specificity
    Analytical specificity (cross-reactivity) of Healgen Strep A Rapid Test Strip (Throat Swab) was carried out for organisms likely to be found in the respiratory tract. It was tested by three lots of Healgen Strep A Rapid Test Strip (Throat Swab). Three professional users performed the test. The obtained results are summarized in the following table.
Test results
OrganismsConcentrationTestedLot1#Lot2#Lot3#
OperatorAOperatorBOperatorCOperatorAOperatorBOperatorCOperatorAOperatorBOperatorC
Arcanobacteriumhaemolyticum2.6×108 CFU/mL---------
Bordetella pertussis7.5×108 CFU/mL---------
Candida albicans9.5×108 CFU/mL---------
Corynebacteriumdiphtheria5.37×108 CFU/mL---------
Enterococcus faecalis2.3×108 CFU/mL---------
Enterococcus faecium4.4×108 CFU/mL---------
Enterovirus (VR-28HumanCoxsackievirus)1.6×108 TCID50/mL---------
Escherichia coli1.1×108 CFU/mL---------
Fusobacteriumnecrophorum7.3×108 CFU/mL---------
Haemophilusparahaemolyticus1.3×108 CFU/mL---------
Haemophilus influenzae4.5×108 CFU/mL---------
Haemophilusparainfluenzae1.6×108 CFU/mL---------
Humanmetapneumovirus(HMPV-27 A2)3.55×105 TCID50/mL---------
Human coronavirusOC431.7×105 TCID50/mL---------
Klebsiella pneumoniae3.1×108 CFU/mL---------
Legionella pneumophila1×104 bacteria/mL---------

{9}------------------------------------------------

Lactobacillus sp.(Lactobacillus casei)6.5×108 CFU/mL-------
Mycobacteriumtuberculosis1×103 bacteria/mL-------
Moraxella lacunata1.95×108 CFU/mL-------
Moraxella(Branhamella)catarrhalis4.8×108 CFU/mL-------
Mycobacteriumtuberculosis(avirulent strain)2.3×108 CFU/mL-------
Neisseria gonorrhoeae3.8×108 CFU/mL-------
Neisseria lactamica1.19×108 CFU/mL-------
Neisseria meningitides7.5×108 CFU/mL-------
Neisseria mucosa3.25×108 CFU/mL-------
Neisseria sicca8.5×108 CFU/mL-------
Neisseria subflava3.27×108 CFU/mL-------
Proteus vulgaris2.9×108 CFU/mL-------
Pseudomonasaeruginosa5.1×108 CFU/mL-------
Serratia marcescens2.1×108 CFU/mL-------
Staphylococcus aureus3.2×108 CFU/mL-------
Staphylococcusepidermidis2.1×108 CFU/mL-------
Staphylococcusmarcescens1.5×108 CFU/mL-------
Staphylococcushaemolyticus1.58×108 CFU/mL-------
Streptococcus agalactiae(Group B)7.9×107 CFU/mL-------
Streptococcusdysgalactiae (Group C)1.43×105CFU/mL-------
Streptococcus sp. (bovisII) Group D5.6×108 CFU/mL-------
Streptococcus sp. Strain1×106 CFU/mL-------

{10}------------------------------------------------

H60R (Group F)
Streptococcus anginosus(Group G)4.2×107 CFU/mL---------
Streptococcuspneumoniae4.2×106 CFU/mL---------
Streptococcus salivarius8.7×108 CFU/mL---------
Streptococcus mitis5.9×108 CFU/mL---------
Streptococcus mutans4.7×108 CFU/mL---------
Streptococcus oralis6.4×108 CFU/mL---------
Streptococcus sanguis1.5×108 CFU/mL---------
Yersinia enterocolitica2.0×108 CFU/mL---------
Adenovirus Type I3.09×108 TCID50/mL---------
Adenovirus Type II3.9×107 TCID50/mL---------
Adenovirus 31.5×108 TCID50/mL---------
Adenovirus 72.8×106 TCID50/mL---------
Cytomegalovirus1.6×105 TCID50/mL---------
Epstein Barr Virus7.85×107 copies/mL---------
HSV Type 1 MacIntyrestrain1.6×105 TCID50/mL---------
Human parainfluenzaType 11.6×105 TCID50/mL---------
Human parainfluenzaType 21.6×105 TCID50/mL---------
Human parainfluenzaType 31.6×105 TCID50/mL---------
Human rhinovirus 265×106 TCID50/mL---------
Measles Virus8.9×105 TCID50/mL---------
Mumps virus1.38×107 TCID50/mL---------
Respiratory syncytialvirus Type A5.5×107 PFU/mL---------
Respiratory syncytialvirus Type B2.8×105 TCID50/mL---------

No cross reactivity was found for the above organisms at the concentrations tested.

{11}------------------------------------------------

    1. Comparison Studies
      NA

3. Clinical Studies

A total of 368 subjects were tested from patients exhibiting symptoms of pharyngitis by both the Healgen Strep A Rapid Test Strip (Throat Swab) and the culture studies. Of the 368 total subjects, 162 were found to be negative (-) by culture and 206 were found to be positive (+) by culture. These test results are summarized in the following tables.

ALL AGES
Strep A Rapid Test Strip(Throat Swab)Culture ResultTotal
+-
+2001201
-6161167
Total206162368
AgeSensitivitySensitivity(95%CI)SpecificitySpecificity(95%CI)
0~597.4% (74/76)90.4% - 99.8%98.1% (52/53)89.1% - 100.0%
5+~2196.7% (119/123)91.7% - 99.0%100% (88/88)95.0% - 100.0%
21+100% (7/7)59.6% - 100.0%100% (21/21)81.8% - 100.0%
All97.1% (200/206)93.7% - 98.8%99.4% (161/162)96.2% - 100.0%

There were no statistical differences in the Healgen Strep A Rapid Test Strip (Throat Swab) performance between the age groups. The overall Clinical Sensitivity is 97%. The overall clinical specificity is 99%.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, LoD, interference, specificity and clinical study of the device, it's concluded that Healgen Strep A Rapid Test Strip (Throat Swab) is substantially equivalent to the predicate.

§ 866.3740

Streptococcus spp. serological reagents.(a)
Identification. Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identifyStreptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.