K Number
K212561

Validate with FDA (Live)

Device Name
MTX-C1
Manufacturer
Date Cleared
2022-09-27

(410 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MTX-C1 is intended for the following medical condition; using three types of handpieces for delivery of non-thermal RF combined with massage:

  • · Relief of minor muscle aches and pain, relief of muscle spasm
  • Temporary improvement of local blood circulation
  • · Temporary reduction in the appearance of cellulite
Device Description

The MTX-C1 main unit is a RF(Radio-frequency) energy generator employed for a variety of aesthetic applications. The Main unit output is set and monitored via touchscreen and controlled by a footswitch.
The main unit can be used with 3 different types of handpieces for treatment.
The system consists of ;

  1. Main unit
  2. 3 type treatment handpiece
  3. Footswitch
    1. Touch screen(User interface)
      The three treatment handpieces differ in size and configuration and are indicated for the treatment of various size areas.
      The operator can adjust treatment parameters such as the power level and treatment time from the user interface on the Main Unit.
      The handpiece is applied with a rubbing/massaging technique and the applicator should be moved continuously on the skin.
AI/ML Overview

This document is a 510(k) summary for the MTX-C1 device, seeking a substantial equivalence determination to a predicate device (Venus Legacy CX). It primarily focuses on demonstrating equivalence through comparison of technological characteristics and adherence to safety standards, rather than proving diagnostic performance against a specific set of acceptance criteria in a clinical study.

Therefore, the requested information regarding acceptance criteria and a study proving device performance (including sample sizes, expert ground truth, adjudication, MRMC studies, standalone performance, and ground truth establishment for training and test sets) is not present in this document.

The document details the following about the device and its testing:

  • Device Name: MTX-C1
  • Intended Use: Relief of minor muscle aches and pain, relief of muscle spasm; temporary improvement of local blood circulation; temporary reduction in the appearance of cellulite.
  • Predicate Device: Venus Legacy CX (K143554)
  • Testing Performed:
    • Biocompatibility testing: Patient contact components and materials tested and validated according to ISO10993-1.
    • Non-Clinical testing (Safety Standards):
      • IEC 60601-1 (General requirements for basic safety and essential performance)
      • IEC 60601-1-2 (Electromagnetic compatibility)
      • IEC 60601-2-2 (Particular requirements for high frequency surgical equipment and accessories)
    • Performance bench testing: The three handpieces of MTX-C1 were tested to achieve superficial skin temperature (40 - 45°C) and maintain it for the required therapy time. This demonstrated that the device "can achieve therapeutic parameters substantially equivalent to the currently cleared predicate device for the intended use."

The document states that "The technological differences... do not alter the device's core technology or performance and have been addressed by the manufacturer through the applicable safety standards... and through non-clinical performance bench testing (Skin temperature)." This strongly suggests that a clinical performance study with human subjects, as would be typical for an AI/CAD device, was not required or conducted for this 510(k) submission, as the substantial equivalence was established through technical and safety testing.

In summary, none of the specific criteria related to clinical study design, ground truth establishment, or human/AI performance metrics (points 1-9 in your request) can be extracted from this document, as it outlines a 510(k) submission primarily based on technological equivalence and adherence to safety standards for an electrosurgical device, not a diagnostic AI algorithm.

{0}------------------------------------------------

September 27, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.

ILOODA Co., Ltd % Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St. Ste 610 Houston, Texas 77054

Re: K212561

Trade/Device Name: Mtx-c1 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX Dated: July 22, 2022 Received: August 1, 2022

Dear Dave Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212561

Device Name MTX-C1

Indications for Use (Describe)

The MTX-C1 is intended for the following medical condition; using three types of handpieces for delivery of non-thermal RF combined with massage:

  • · Relief of minor muscle aches and pain, relief of muscle spasm
  • Temporary improvement of local blood circulation
  • · Temporary reduction in the appearance of cellulite

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for ilooda. The logo consists of the word "ilooda" in lowercase letters, with a stylized graphic above it. The graphic is made up of vertical lines of varying lengths, arranged in a pattern that resembles sound waves or a stylized cityscape. The logo is simple and modern, with a focus on clean lines and a minimalist design.

Traditional 510(k) Summary K212561

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: 9/19/2022

I. SUBMITTER

Submitter's Name :Ilooda Co.,Ltd
Submitter's HQ Address:120, Jangan-ro 458 Beon-gil, Jangan-Gu,
Suwon-Si Gyeonggido, KOREA, 16200
Submitter's Telephone:+82-31-278-4660
Contact person:Yun-Jung HA (yjha@ilooda.com) / RD Manager
Official Correspondent:Dave Kim (davekim@mtech-inc.net)
Address:7505 Fannin St, Suite 610, Houston, TX 77054
Telephone:+713-467-2607
PROPOSED DEVICE
Trade/proprietary name:MTX-C1
Common or Usual Name:Electrosurgical cutting and coagulation device and accessories
Regulation Number:21 CFR 878.4400 (Product Code: PBX)
Regulatory Class:Class II
Prescription Use.
PREDICATE DEVICE
Trade/proprietary name:Venus Legacy CX
510k number:K143554
Regulation Number:21 CFR 878.4400 (Product Code: PBX)
Regulatory Class:Class II
Prescription Use.
This predicate has not been subject to a design-related recall.

II. DEVICE DESCRIPTION

The MTX-C1 main unit is a RF(Radio-frequency) energy generator employed for a variety of aesthetic applications. The Main unit output is set and monitored via touchscreen and controlled by a footswitch.

The main unit can be used with 3 different types of handpieces for treatment.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for ilooda. The logo consists of a stylized graphic above the word "ilooda". The graphic is made up of vertical lines of varying lengths, arranged in a symmetrical pattern. The lines are dark gray, and the word "ilooda" is in a sans-serif font.

The system consists of ;

  1. Main unit

  2. 3 type treatment handpiece

  3. Footswitch

    1. Touch screen(User interface)
      The three treatment handpieces differ in size and configuration and are indicated for the treatment of various size areas.

The operator can adjust treatment parameters such as the power level and treatment time from the user interface on the Main Unit.

The handpiece is applied with a rubbing/massaging technique and the applicator should be moved continuously on the skin.

III. INDICATIONS FOR USE:

The MTX-C1 is intended for the treatment of the following medical condition; using three types of handpieces for delivery of non-thermal RF combined with massage:

  • · Relief of minor muscle aches and pain, relief of muscle spasm
  • Temporary improvement of local blood circulation
  • · Temporary reduction in the appearance of cellulite

{5}------------------------------------------------

IV. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Proposed devicePredicated device
Model nameMTX-C1Venus Legacy CX
ManufacturerIlooda Co.,LtdVenus Concept Ltd., Weston FL, USA
510(k)numberK212561K143554
Product codePBXPBX
Intended useThe MTX-C1 is intended for thetreatment of the following medicalcondition; using three types ofhandpieces for delivery of non-thermal RF combined with massage:• Relief of minor muscle achesand pain, relief of muscle spasm• Temporary improvement oflocal blood circulation• Temporary reduction in theappearance of celluliteThe Venus Legacy CX device is intendedfor the treatment of the followingmedical conditions; using the LB2 andLF2 applicators for delivery of non-thermal RF combined with massage andmagnetic field pulses:• Relief of minor muscle aches and pain, reliefof muscle spasm• Temporary improvement of local bloodcirculation• Temporary reduction in the appearance ofcellulite
Type of EnergyDeliveredRF energyRF energy
ComponentsMain unit3 type handpieces(Small, Medium, Large)FootswitchTouch screenMain unit2 type applicators(LB2, LF2)Footswitch / finger switchTouch screen
User interfaceTouchscreen on front of thegenerator allows the user to selectmode and outputTouchscreen on front of the generator allowsthe user to select mode and output
Safey features- Indicator(Ready/Standby)- An independent electroniccircuit stops the operationof the system in case of asoftware errorApplicators are equipped with a treatmentarea temperature monitoring system
Frequency1MHz ± 10%1MHz ± 10%
Max powerMax 55W @ medium , LargeMax 7.5W @ smallUp to 150W
RF mode ofoperationBipolarBipolar
MaterialsMaterials are biocompatibleMaterials are biocompatible
Powerrequirements100-240V~, 50/60Hz100-120 VAC / 60Hz220-240 VAC / 50Hz

As described in the comparison tables above, the MTX-C1 subject device has the same intended use and indications for use, principles of operation and similar technological characteristics such as type of energy delivered, user interface, frequency, RF mode of operations, as its predicate

{6}------------------------------------------------

510(k) Submission -Non invasive RF system

device.

The technological differences, for example, the max power outputs between the subject device and its predicate, are based on the same heating using radio frequency for the same indications for use.

The design and components in the subject device, including the main unit and the handpieces, are similar to the design and components found in the predicate device.

The technological differences do not alter the device's core technology or performance and have been addressed by the manufacturer through the applicable safety standards (General controls and mitigation measures) and through non-clinical performance bench testing (Skin temperature)

Thus, this does not raise any additional issues of safety and efficacy.

V. PERFORMANCE DATA

The following performance data was provided in support of the substantial equivalence determination.

Biocompatibility testing:

The patient contact components and materials are tested and validated according to ISO10993-1. They are identical to the predicate device.

Non Clinical testing:

IEC 60601-1 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance. The requirements of specified standards were fulfilled.

IEC 60601-1-2 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance (collateral standards: electromagnetic compatibility.

IEC 60601-2-2 Medical electrical equipment

Part 2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

The requirements of specified standards were fulfilled.

Performance bench testing :

The three handpieces of MTX-C1 were tested to achieve superficial skin temperature (40 - 45°C) and maintain it for the required therapy time.

The testing data demonstrated that the device can achieve therapeutic parameters substantially equivalent to the currently cleared predicate device for the intended use.

VI. CONCLUSIONS

The intended use of the MTX-C1 is within the scope of the predicate device. MTX-C1 system, from both a design and clinical perspective, uses similar or identical technology as the cited predicate device and has the same intended uses.

{7}------------------------------------------------

510(k) Submission –Non invasive RF system

Based upon the predicted overall performance characteristics for the MTX-C1, Ilooda Co.,Ltd ., believes that no significant differences exist in the usage of its underlying technological principles between MTX-C1 and the cited predicate device.

Image /page/7/Picture/4 description: The image shows the logo for ilooda. The logo consists of the word "ilooda" in a sans-serif font, with a stylized graphic above it. The graphic is made up of vertical bars of varying heights, arranged in two rows. The bars are all the same color, which appears to be a dark gray or black.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.