The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.
The Flexible HD Cysto-Urethroscope System (Part Number: 11272VH(U)-TL) is intended to be used with the IMAGE1 S CCU consisting of Connect / Connect II and X-Link modules (cleared in K201135). Identical to the predicates, the scope cannot be operated on its own because it produces image data output in the form of video signals which require external processing and display. When the scope is used with the compatible CCU, it becomes the Flexible HD Cysto-Urethroscope System, which can then be operated as intended and indicated.
The Flexible HD Cysto-Urethroscope System consists of the following key components:
11272VH-TL / 11272VHU-TL – Cysto-Urethroscopes with positive (VH-TL) / contra-positive (VHU-TL) deflection control and permanent T-Luer affixed to working channel
TC200US / TC201US / TC301US – Connect/Connect II/Link camera control unit (CCU) modules for image processing
The subject device is a lower cost alternative to the primary predicate and incorporates the following differences. Housing Plug: Due to lack of clinical need, the subject device replaces the suction channel of the primary predicate device with a permanent stainless steel housing plug. Fixed LUER: The subject device uses a removable T-Luer whereas the primary predicate supported a removable T-Luer or Stopcock.
The FDA 510(k) summary for the KARL STORZ Flexible HD Cysto-Urethroscope System (K212476) describes the acceptance criteria and supporting studies as follows:
This submission is a Special 510(k), indicating a modification to a previously cleared device. Therefore, the focus is on demonstrating that the modifications do not raise new questions of safety and effectiveness compared to the predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
For this Special 510(k), the acceptance criteria are largely based on adherence to recognized consensus standards and demonstrating that the modified device maintains the same technological characteristics and performance as the predicate devices, or that any differences do not impact safety or effectiveness.
| Acceptance Criterion | Reported Device Performance / Supporting Data |
|---|---|
| Intended Use / Indications for Use | Identical to predicate devices: "The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys." Minor modifications do not change the intended use or indications for use. |
| Physical Characteristics (e.g., Insertion Shaft Diameter, Length, Working Channel Diameter, Deflection) | Same as primary predicate (K191357) for: - Type of scope (Flexible) - Insertion Shaft Diameter (5.5 mm) - Insertion Shaft Length (37 cm) - Working Channel Diameter (2.3 mm) - Deflection (Up: 210°, Down: 140°) |
| Optical Characteristics (e.g., Type of Imager, Field of View, Direction of View, Depth of Field, On-axis Resolution, Light Source) | Same as primary predicate (K191357) for: - Type of Imager (CMOS) - Field of View (100°) - Direction of View (0°) - Depth of Field (3-50 mm) - On-axis Resolution (40 lp/mm @ 3 mm, 2.5 lp/mm @ 50 mm) - Light Source (Internal LED) |
| Material / Design Characteristics (Differences from Predicate) | Suction Port / Channel: Proposed device has "No" (due to permanent stainless steel housing plug), while primary predicate has "Yes". This difference is described as due to "lack of clinical need" for the suction channel. Luer Type: Proposed device has "Permanent T-Luer", primary predicate has "Bayonet with removable T-Luer or Stopcock assembly". This is a minor modification. |
| Cleaning, Disinfection and Sterilization Methods | Same as primary predicate for Manual Cleaning, and High-Level Disinfection (Revital-Ox RESERT). Sterilization: Proposed device supports STERRAD 100NX (FLEX and DUO cycles), STERRAD NX (Advanced Cycle), V-PRO maX (Flexible Cycle), V-PRO 60. The primary predicate supported these with additional support for SSIE (Standard Cycle). The proposed device's methods are a recognized subset. |
| Electrical Safety and EMC | Compliance with recognized consensus standards: - IEC 60601-1:2005/(R)2012, Ed. 3.1 - IEC 60601-1-2:2014, 4th Ed. - IEC 60601-2-18:2009, Ed. 3.0 |
| Optical/Mechanical Performance | Compliance with recognized consensus standards: - IEC 62471:2006 - ISO 8600 |
| Biocompatibility | Compliance with recognized consensus standards: - ISO 10993-1:2009/(R) 2013 - ISO 10993-5:2009/(R) 2014 - ISO 10993-10:2010 - ISO 10993-11:2006/(R) 2010 |
| Reprocessing | Compliance with recognized consensus standards: - AAMI TIR 12:2010 - ISO 15883-5:2005 - AAMI TIR 30:2011 - AAMI/ANSI/ISO 11737-1:2006/ (R)2011 - ASTM E1837-96:2014 |
| Substantial Equivalence to effectiveness | Supported by comparison of images and standard image quality characteristics, including the performance testing listed. The differences do not raise new or different questions of safety and effectiveness. |
2. Sample size used for the test set and the data provenance:
The document does not detail specific "test set" sample sizes or data provenance in the context of clinical or image-based studies for the current 510(k). This submission relies on a comparison to predicate devices and adherence to established performance standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. The submission focuses on engineering specifications and adherence to standards, not a clinical study requiring expert ground truth for image interpretation or diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. No clinical or image-based test set with adjudication is described.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is an endoscope system for visualization and operative access; it does not involve AI assistance for human readers in diagnostic interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is not an algorithm-only device. It is a medical instrument.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
Not applicable in the context of diagnostic performance. The "ground truth" for this submission are the established performance parameters of the predicate device and the requirements of recognized consensus standards for medical devices.
8. The sample size for the training set:
Not applicable. As a medical device, and specifically a modification through a Special 510(k), there is no "training set" in the context of machine learning.
9. How the ground truth for the training set was established:
Not applicable. See point 8.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 3, 2021
KARL STORZ Endoscopy America, Inc. Thomas Ostrowski Senior Regulatory Affairs Specialist 2151 E. Grand Ave El Segundo, CA 90245
Re: K212476
Trade/Device Name: Flexible HD Cysto-Urethroscope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ, FBO Dated: August 5, 2021 Received: August 6, 2021
Dear Thomas Ostrowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K212476
Device Name Flexible HD Cysto-Urethroscope System
Indications for Use (Describe)
The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for KARL STORZ–ENDOSKOPE. The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ–ENDOSKOPE" are in smaller, blue letters.
8. 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Applicant: | KARL STORZ Endoscopy-America, Inc2151 E. Grand AvenueEI Segundo, CA 90245 |
|---|---|
| Contact: | Thomas OstrowskiSenior Regulatory Affairs Specialist781-910-8508 (phone) |
| Date of Preparation: | August 27th, 2021 |
| Type of 510(k)Submission: | Special |
| Device Identification: | Trade Name: Flexible HD Cysto-Urethroscope System |
| Common Name: | Cystoscope and accessories, Flexible/Rigid (FAJ)Cystourethroscope (FBO) |
| Product Code: | FAJ, FBO |
| Regulation Name /Number: | Endoscope and Accessories (21 CFR 876.1500) |
| Predicate Device(s): | Flexible HD Cysto-Urethroscope System (K191357) – PrimaryFlexible Video Cysto-Urethroscope (C-View) (K202957) - SecondaryThe above predicate devices have not been subject to any recall |
{4}------------------------------------------------
| Device Description: | The Flexible HD Cysto-Urethroscope System (Part Number: 11272VH(U)-TL) is intended to be used with the IMAGE1 S CCU consisting of Connect / Connect II and X-Link modules (cleared in K201135). Identical to the predicates, the scope cannot be operated on its own because it produces image data output in the form of video signals which require external processing and display. When the scope is used with the compatible CCU, it becomes the Flexible HD Cysto-Urethroscope System, which can then be operated as intended and indicated. | ||||
|---|---|---|---|---|---|
| The Flexible HD Cysto-Urethroscope System consists of the following key components:11272VH-TL / 11272VHU-TL – Cysto-Urethroscopes with positive (VH-TL) / contra-positive (VHU-TL) deflection control and permanent T-Luer affixed to working channel TC200US / TC201US / TC301US – Connect/Connect II/Link camera control unit (CCU) modules for image processing | |||||
| The subject device is a lower cost alternative to the primary predicate and incorporates the following differences. Housing Plug: Due to lack of clinical need, the subject device replaces the suction channel of the primary predicate device with a permanent stainless steel housing plug. Fixed LUER: The subject device uses a removable T-Luer whereas the primary predicate supported a removable T-Luer or Stopcock. | |||||
| Intended Use and Indications for use: | The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys. The minor modifications to the Flexible HD Cysto-Urethroscope System do not change the intended use and indications for use. The intended use / indications for use are identical between the proposed device and the predicate devices. | ||||
| Technological Characteristics: | The subject device is a modification of the primary predicate, 11272VH/VHU, cleared via K191357 and also incorporates design features of the secondary predicate device, 11272VE/VUE, cleared via K202957. A comparison between the proposed device and primary predicate device is included in the table below: | ||||
| Proposed Device | Primary PredicateK191357 | Secondary PredicateK202957 | |||
| Physical Characteristics | |||||
| Type of scope | Flexible | Same | Same | ||
| Insertion Shaft Diameter | 5.5 mm | Same | 5.2 mm | ||
| Insertion ShaftLength | 37 cm | Same | Same | ||
| Working ChannelDiameter | 2.3 mm | Same | Same | ||
| Deflection (°) | Up: 210°Down: 140° | Same | Same | ||
| Optical Characteristics | |||||
| Type of Imager | CMOS | Same | Same | ||
| Field of View | 100° | Same | Same | ||
| Direction of View | 0° | Same | Same | ||
| Depth of Field | 3-50 mm | Same | 5-50 mm | ||
| On-axis Resolution(minimal) | 40 lp/mm @ 3 mm2.5 lp/mm @ 50 mm | Same | 16 lp/mm @ 5 mm1.8 lp/mm @ 50 mm | ||
| Light Source | Internal LED | Same | Same | ||
| Material / Design Characteristics | |||||
| Suction Port /Channel | No | Yes | No | ||
| Luer Type | Permanent T-Luer | Bayonet withremovable T-Lueror Stopcockassembly | Same as proposeddevice | ||
| Cleaning, Disinfection and Sterilization Methods | |||||
| Cleaning | Manual | Same | Same | ||
| Sterilization | STERRAD 100NX(FLEX and DUOcycles),STERRAD NX(Advanced Cycle),V-PRO maX(Flexible Cycle),V-PRO 60 | Same withadditional supportfor SSIE (StandardCycle) | Same as proposeddevice | ||
| High LevelDisinfection | Revital-OxRESERT | Same | Same | ||
| Non-ClinicalPerformance Data: | There are no performance standards or special controls developed under Section514 of the FD&C Act for endoscopes, however, the subject device follows FDArecognized consensus standards identified in the primary and secondarypredicate device submissions:• Electrical Safety and EMCo IEC 60601-1:2005/(R)2012, Ed. 3.1o IEC 60601-1-2:2014, 4th Ed.o IEC 60601-2-18:2009, Ed. 3.0• Optical/Mechanical Performanceo IEC 62471:2006o ISO 8600 |
{5}------------------------------------------------
{6}------------------------------------------------
| Biocompatibility ISO 10993-1:2009/(R) 2013 ISO 10993-5:2009/(R) 2014 ISO 10993-10:2010 ISO 10993-11:2006/(R) 2010 Reprocessing AAMI TIR 12:2010 ISO 15883-5:2005 AAMI TIR 30:2011 AAMI/ANSI/ISO 11737-1:2006/ (R)2011 ASTM E1837-96:2014 | |
|---|---|
| SubstantialEquivalence: | The intended use, operating principles, technological characteristics and featuresare similar, if not identical, between that subject device and the predicatedevices, Flexible HD Cysto-Ureteroscope System (K191357) and FlexibleVideo Cysto-Uretheroscope (C-View) (K202957). The minor differencebetween the subject and predicate devices that does not raise new or differentquestions or safety and effectiveness are listed above in TechnologicalCharacteristics section.As proven by the comparisons, the differences do not raise different questions ofsafety and effectiveness because the intended use, operating principles,technological characteristics, and features are similar, if not identical. Bothsystems also comply with identical standards and safety testing, whereapplicable.Substantial equivalence to the effectiveness of the subject device is supported bythe comparison of the images and standard image quality characteristicsincluding, but not limited to the performance testing listed above. |
| Clinical PerformanceData: | Clinical performance is not required to demonstrate substantial equivalence tothe predicate devices. Comparative analysis is sufficient to establish substantialequivalence. |
| Conclusion: | The Flexible HD Cysto-Urethroscope System is substantially equivalent to itspredicate devices. The comparative analysis demonstrates that the device is assafe and effective as the legally marketed devices. |
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.