K Number
K212434

Validate with FDA (Live)

Date Cleared
2022-05-05

(274 days)

Product Code
Regulation Number
868.2775
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TOF3D is used to objectively monitor the level of neuromuscular transmission by measuring muscle contraction following stimulation. The TOF3D can also be used as a peripheral nerve stimulator (without the objective measuring function) for subjective monitoring. The device is intended for use for adolescents greater than 18 through 21 years of age, and adults.

Device Description

The Neuromuscular Transmission Monitor TOF3D is capable to monitor the level of neuromuscular transmission (NMT) during surgery of in the intensive care unit by stimulating different nerves and measuring the response of the respective muscles to the stimulation. The different locations for monitoring are for instance the ulnar nerve/adductor pollicis muscle, the posterior tibial nerve/flexor halluces brevis muscle and the facial nerve/orbicularis oculi muscle. The Neuromuscular Transmission Monitor TOF3D uses acceleromyography (AMG) measurement for recording of evoked muscle responses. The results are shown on the LCD display of the device and shall aid qualified medical staff to maintain the proper level of neuromuscular block and to determine the level of recovery from neuromuscular block.

AI/ML Overview

The provided text is a 510(k) summary for the Neuromuscular Transmission Monitor TOF3D. It describes the device, its intended use, and its substantial equivalence to a predicate device (TOF-WATCH SX).

However, the document explicitly states that no animal or clinical testing was required to demonstrate substantial equivalence (Section 5.7, "Animal and Clinical Studies"). This means there is no study described in this document that proves the device meets acceptance criteria through clinical performance metrics, effect sizes, or expert evaluations. The acceptance criteria and performance data provided are related to non-clinical testing, such as electrical safety, EMC, software verification, validation, and biocompatibility.

Therefore, I cannot provide a table of acceptance criteria and reported device performance related to clinical efficacy as requested in the prompt, nor can I answer questions about sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, or ground truth types for clinical use.

The available information about "acceptance criteria" and "device performance" in this document pertains to engineering and safety standards, not clinical effectiveness.

Here's what can be extracted regarding the non-clinical acceptance criteria and "performance":

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):

Acceptance Criteria CategoryStandard MetReported Device Performance/Compliance
Electrical Safety & EMCIEC 60601-1 (Safety & Performance general)Complies
IEC 60601-1-6 (Usability)Complies
IEC 60601-2-10 (Requirements for Nerve and Muscle Stimulators)Complies
IEC 60601-1-2 (Electromagnetic Compatibility)Complies
Software Verification & ValidationFDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices"Conducted and documentation provided
BiocompatibilityISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process"Virological safety risks of TOF3D sensors with accessories are acceptable for surface skin contact parts.

2. Sample sizes for the test set and data provenance:

  • Not applicable for clinical performance studies. The testing was non-clinical (electrical safety, software, biocompatibility).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable, as no clinical ground truth was established for performance testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable, as no clinical ground truth was established for performance testing.

5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

  • No, explicitly stated that "No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate."

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable, as the device is a hardware monitor, not an AI algorithm.

7. The type of ground truth used:

  • For non-clinical testing, the "ground truth" was compliance with established engineering and safety standards (e.g., IEC standards, FDA software guidance, ISO biocompatibility standards).

8. The sample size for the training set:

  • Not applicable, as this is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an abstract design. To the right of the seal, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in a bold, sans-serif font. The word "FDA" is in a blue square, and the rest of the text is in a smaller font size.

May 5, 2022

MIPM Mammendorfer Institut für Physik und Medizin GmbH Lukas Spögler Regulatory Affairs Oskar-von-Miller Str. 6 Mammendorf, Bavaria 82291 Germany

Re: K212434

Trade/Device Name: Neuromuscular Transmission Monitor TOF3D Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: KOI Dated: April 6, 2022 Received: April 11, 2022

Dear Lukas Spögler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Todd Courtnev Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212434

Device Name

Neuromuscular Transmission Monitor TOF3D

Indications for Use (Describe)

The TOF3D is used to objectively monitor the level of neuromuscular transmission by measuring muscle contraction following stimulation. The TOF3D can also be used as a peripheral nerve stimulator (without the objective measuring function) for subjective monitoring. The device is intended for use for adolescents greater than 18 through 21 years of age, and adults.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo featuring a stylized depiction of Leonardo da Vinci's Vitruvian Man, enclosed within a circle and a square. Encircling the top portion of the image is the text "HOMINUM SALUTI MEDICINAM EXERCENS", written in a serif font. Below the image is the acronym "MIPM", written in a bold, sans-serif font.

5. 510(k) Summary

The following information is provided in accordance with 21 CFR 807.92 for the Permarket 510(k) Summary:

5.1 Submitter Information

Company:Jennifer RosenheimerManaging Director
MIPM Mammendorfer Institut für Physik und MedizinGmbH
Oskar-von-Miller Str. 6
Mammendorf, Bavaria 82291, Germany
Telephone: +49 8145 / 92 09 0
Fax: +49 8145 / 92 09 33
quality@mipm.com
Contact:Lukas Spögler
Regulatory Affairs
MIPM Mammendorfer Institut für Physik und MedizinGmbH
Oskar-von-Miller Str. 6
Mammendorf, Bavaria 82291, Germany
Telephone: +49 8145 / 92 09 0
Fax: +49 8145 / 92 09 33
regulatory@mipm.com
Date Summary Prepared:April 27, 2022

5.2 Name of the Device

Trade Name:Neuromuscular Transmission Monitor TOF3D
Common Name:Neuromuscular Transmission Monitor
Classification Name:Electrical peripheral nerve stimulator
Review Panel:Anesthesiology (AN)
Regulation:868.2775
Class:Class II
Product Code:KOI

5.3 Equivalence Claimed to Predicate Device

The Neuromuscular Transmission Monitor TOF3D is equivalent to the TOF-WATCH SX (K992598), manufactured by ORGANON TEKNIKA CORP..

5.4 Device Description

The Neuromuscular Transmission Monitor TOF3D is capable to monitor the level of neuromuscular transmission (NMT) during surgery of in the intensive care unit by stimulating different nerves and measuring the response of the respective muscles to the stimulation. The different locations for monitoring are for instance the ulnar nerve/adductor pollicis muscle, the posterior tibial nerve/flexor halluces brevis muscle and the facial nerve/orbicularis oculi muscle.

The Neuromuscular Transmission Monitor TOF3D uses acceleromyography (AMG) measurement for recording of evoked muscle responses. The results are shown on the LCD display of the device and shall aid qualified medical staff to maintain the proper level of neuromuscular block and to determine the level of recovery from neuromuscular block.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo with a blue border. Inside the border is a drawing of a human figure in the style of Leonardo da Vinci's Vitruvian Man. The figure is surrounded by a circle and a square. Above the figure, the text "HOMINUM SALUTI MEDICINAM EXERCENS" is written in a circular pattern. Below the figure, the letters "MIPM" are written in blue.

ર.ર Indications for Use

The TOF3D is used to objectively monitor the level of neuromuscular transmission by measuring muscle contraction following stimulation. The TOF3D can also be used as a peripheral nerve stimulator (without the objective measuring function) for subjective monitoring. The device is intended for use for adolescents greater than 18 through 21 years of age, and adults.

5.6 Substantial Equivalence Discussion

FeaturesTOF3D(subject device)TOF-WATCH SX(predicate device)
Stimulation patterns
TOFYesYes
PTCYesYes
1 HzYesYes
0.1 HzYesYes
DBSYesYes
TETYesYes
Nerve locationNoYes
Maximum simulation voltage300V (60mA, 5kΩ)300V (60mA, 5kΩ)
Stimulation current range constantcurrent0 - 60 mA0 - 60 mA
Stimulation pulse widthMonophasic, 200 μs or 300 μsMonophasic, 200 μs or 300 μs
Surface temperature sensorYesYes

The Neuromuscular Transmission Monitor TOF3D shares the same indications for use, device operation, overall technical and functional capabilities with its predicate device TOF-WATCH SX (K992598). Both devices rely on the use of acceleromyograph to measure muscle response to electrical stimulation to objectively monitor the level of neuromuscular transmission. The additional functionality of the TOF-WATCH SX (nerve location device - localize nerves for loco-regional anesthesia) is based on the same technical principle used for the nerve stimulation functionality, therefore additional risks in clinical application are not to be expected.

The Neuromuscular Transmission Monitor TOF3D is substantially equivalent to the predicate device TOF-WATCH SX (K992598).

5.7 Performance Data

Electrical safety and electromagnetic compatibility (EMC)

The Neuromuscular Transmission Monitor TOF3D complies with the IEC 60601-1, IEC 60601-1-6 and IEC 60601-2-10 standards for safety and performance and the IEC 60601-1-2 standard for EMC-

Software Verification and Validation Testing

The software for this device was considered as a "moderate2 level of concern. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices".

Biocompatibility testing

The biocompatibility evaluation for the Neuromuscular Transmission Monitor TOF3D was conducted in accordance with ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Based on the results of the biocompatibility evaluation, it can be concluded that the virological safety risks of the TOF3D sensors with accessories are acceptable for use as medical device surface skin contact pars.

Animal and Clinical Studies

No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for an organization called MIPM. The logo features a stylized drawing of a human figure inside of a square. The human figure is in the style of Leonardo da Vinci's Vitruvian Man. The text around the figure reads "HOMINUM SALUTI MEDICINAM EXERCENS".

Conclusion 5.8

The subject device has the same indications for use, technological characteristics, and design and operating principles as the predicate device. The identified differences do not raise different questions of safety and effectiveness. The performance of the device is supported by non-clinical testing and risk management activities. The Neuromuscular Transmission Monitor TOF3D is substantially equivalent to the TOF-WATCH SX, cleared under K992598.

§ 868.2775 Electrical peripheral nerve stimulator.

(a)
Identification. An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.(b)
Classification. Class II (performance standards).