K Number
K212303

Validate with FDA (Live)

Device Name
T2
Date Cleared
2021-09-21

(60 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

T2 is digital X-Ray imaging equipment for dental professionals that convert X-Ray signals into digital signals to acquire 2D images and reconstruct them into 3D images, using Panoramic (Pano), Cephalometric (Ceph), and Computed Tomography (CT) technology for the diagnosis of the anatomical structure of the oral and maxillofacial area.

Device Description

T2 is a digital X-ray CT, panoramic, and Cephalo imaging system device composed of X-ray generator, X-ray controller, X-ray supporter, image processing unit (sensor), PC, and software. The apparatus attached to the equipment column is a structure that can be rotated 360 by the system control unit. This system control unit actuates the motor control, X-ray generator, and image processing unit (sensor). The height controlling unit controls the column and adjusts the height of the equipment. The X-ray generator and image processing unit (sensor) are attached to the rotating apparatus. When the rotating apparatus starts the rotation, X-ray is irradiated from the X-ray generator (generating unit). This X-ray irradiation penetrates the subject and reaches the image processing unit (sensor), and then is converted into electric signals to secure imagery information. Inside the imaging section of the image processing unit (sensor), real time X-ray input is converted into electric signals and consecutively combined, resulting in imagery information. The combined panoramic imagery information is then sent to the PC and saved in patient management software.

AI/ML Overview

The provided document is a 510(k) summary for the T2 imaging device. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical study results with acceptance criteria.

Therefore, many of the requested details regarding acceptance criteria for clinical performance, sample sizes, expert qualifications, and ground truth establishment from a clinical trial perspective are not available in this document. The document explicitly states: "No clinical studies are submitted."

Here's a breakdown of the available information based on the prompt's questions:

1. A table of acceptance criteria and the reported device performance

The document does not provide specific acceptance criteria or reported device performance in terms of clinical accuracy (e.g., sensitivity, specificity) against a clinical ground truth. Instead, the substantial equivalence hinges on non-clinical performance tests demonstrating compliance with relevant standards and technical specifications being similar to the predicate device.

The closest to "acceptance criteria" are the performance specifications listed in the "Substantial Equivalence Matrix" (pages 4-5) which compare the proposed device (T2) to the predicate device (T1-CS, T1-C). These are primarily technical specifications rather than diagnostic performance metrics.

FeatureAcceptance Criteria (Predicate's Performance)Reported Device Performance (T2)Outcome
Device NameT1-CS, T1-CT2Different
510(k) No.K183475ProposedDifferent
ManufacturerOsstem Implant Co., Ltd.Osstem Implant Co., Ltd.Same
Indications for UseSimilar description for imaging oral anatomySimilar description for imaging oral anatomySame
Performance SpecificationPanoramic, cephalometric and computed tomographyPanoramic, cephalometric and computed tomographySame
Input Voltage100-240 VAC100-240VACSame
Tube Voltage60-100 kV60-95KVSimilar (included within range of predicated product)
Tube Current5-16 mA5-11mASimilar (included within range of predicated product)
Focal Spot Size0.5 mm0.5mmSame
Exposure TimeMax. 22 sMax. 22sSame
2D Image Viewing ProgramModel Name: 2D01, 2D02Model Name: 2D02Similar (included within predicated product)
3D Image Viewing ProgramModel Name: 3D01, 3D02Model Name: 3D02Similar (included within predicated product)
Anatomical SitesMaxillofacialMaxillofacialSame
Image Sensor (Computed Tomography)1515DXT, SEN11515DXT, SEN1Same
Image Sensor (Panoramic)1515DXT, SEN11515DXT, SEN1Same
Image Sensor (Cephalometric)XID-C24DCXID-C24DCSame
Size of Imaging Volume (CT)1515DXT: Max. 15x9, SEN1: Max 15x10 (Stitching Mode: Max 15 X 15)1515DXT: Max. 15x9, SEN1: Max 15x10Similar (included within range of predicated product)
Pixel Resolution (CT)1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mm1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mm (-)Same (The listed values for T2 are the high-resolution modes of predicate)
Pixel Resolution (Panoramic)1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mm1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mmSame
Pixel Resolution (Cephalometric)XID-C24DC: 5 lp/mmXID-C24DC: 5 lp/mmSame
Pixel Size (CT)1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μm1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μmSimilar (included within range of predicated product)
Pixel Size (Panoramic)1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μm1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μmSame
Pixel Size (Cephalometric)XID-C24DC: 100 μm * 100 μmXID-C24DC: 100 μm * 100 μmSame
Active Area (CT)1515DXT: 146 x 146 mm, SEN1: 160 x 128 mm1515DXT: 146 x 146 mm, SEN1: 160 x 128 mmSame
Active Area (Pano)1515DXT: 146 x 8.13 mm, SEN1: 160 x 8 mm1515DXT: 146 x 8.13 mm, SEN1: 160 x 8 mmSame
Active Area (Ceph)1.8 x 150 mm1.8 x 150 mmSame

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable, as no clinical studies were submitted. Non-clinical performance tests were conducted, but details on the "test set" for those are not provided in terms of origin or type (e.g., phantom data).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical studies were submitted.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical studies were submitted.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The T2 device is an X-ray imaging system, not an AI-assisted diagnostic tool. No clinical studies, let alone MRMC studies, were reported.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. The T2 device is an X-ray imaging system, not an AI algorithm. Its performance is related to image acquisition and reconstruction, not autonomous diagnostic interpretation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no clinical studies were submitted. For the non-clinical performance, the "ground truth" would be established by physical measurements and adherence to technical standards.

8. The sample size for the training set

Not applicable. This device is a hardware imaging system, not a machine learning algorithm that requires a training set in the conventional sense.

9. How the ground truth for the training set was established

Not applicable. This device is a hardware imaging system, not a machine learning algorithm.

{0}------------------------------------------------

September 21, 2021

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is an emblem representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font, and the words "U.S. FOOD & DRUG" are stacked above the word "ADMINISTRATION".

Osstem Implant Co., Ltd. % Peter Lee QA/RA Manager Hiossen Inc. 85 Ben Fairless Dr. FAIRLESS HILLS PA 19030

Re: K212303

Trade/Device Name: T2 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS, Dated: July 20, 2021 Received: July 23, 2021

Dear Peter Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel M. Digitally signed by
Laurel M. Burk -S
Burk -S
Date: 2021.09.21
09:01:35 -04'00'
, for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212303

Device Name T2

Indications for Use (Describe)

T2 is digital X-Ray imaging equipment for dental professionals that convert X-Ray signals into digital signals to acquire 2D images and reconstruct them into 3D images, using Panoramic (Pano), Cephalometric (Ceph), and Computed Tomography (CT) technology for the diagnosis of the anatomical structure of the oral and maxillofacial area.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

m Implant Co., Ltd ong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea 82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem.com

K212303 - 510(k) Summary

Date: July 20, 2021

1. Company and Correspondent making the submission

- Submitter's Name: Osstem Implant Co., Ltd.
- Address: 2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-siGyeonggi-do, 15079, Republic of Korea
- Contact: Mr. Jinwoo Bae
- Phone: +82-70-4626-0701
- Correspondent's Name: Hiossen Inc.
- Address: 85 Ben Fairless Dr. Fairless Hills, PA 19030
- Contact: Mr. Peter Lee
- Phone: +1-267-759-7031

2. Proposed Device

Trade or (Proprietary) Name: T2
Classification Name: X-Ray, Tomography, Computed, Dental
Regulation Number: 21CFR892.1750
Device Classification: Class II
Classification Product Code: OAS

3. Predicated Device

Primary Predicate -T1-CS, T1-C, Osstem Implant Co., Ltd. (K183475)

4. Description

  • T2 is a digital X-ray CT, panoramic, and Cephalo imaging system device composed of X-ray generator, X-ray controller, X-ray supporter, image processing unit (sensor), PC, and software. The apparatus attached to the equipment column is a structure that can be rotated 360 by the system control unit. This system control unit actuates the motor control, X-ray generator, and image processing unit (sensor). The height controlling unit controls the column and adjusts the height of the equipment. The X-ray generator and image processing unit (sensor) are attached to the rotating apparatus. When the rotating apparatus starts the rotation, X-ray is irradiated from the X-ray generator (generating unit). This X-ray irradiation penetrates the subject and reaches the image processing unit (sensor), and then is converted into electric signals to secure imagery information. Inside the imaging section of the image processing unit (sensor), real time X-ray input is

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Osstem Implant. The logo is in orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. There is a small circle above the right side of the word "OSSTEM".

Osstem Implant Co., Ltd.

2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea Tel: +82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem.com

converted into electric signals and consecutively combined, resulting in imagery information. The combined panoramic imagery information is then sent to the PC and saved in patient management software

Proposed DevicesPredicated DevicesRemark
Device NameT2T1-CS, T1-CDifferent
510(k) No.ProposedK183475Different
ManufacturerOsstem Implant Co., Ltd.Osstem Implant Co., Ltd.Same
Indications for UseT2 is digital X-Ray imagingequipment for dentalprofessionals that convert X-Ray signals into digitalsignals to acquire 2D imagesand reconstruct them into 3Dimages, using Panoramic(Pano), Cephalometric(Ceph), and ComputedTomography (CT)technology for the diagnosisof the anatomical structure ofthe oral and maxillofacialarea.T1-C, T1-CS are are acomputed tomography x-raysystem intended to producepanoramic, cephalometric orcrossectional images of theoral anatomy on a real timebasis by computerreconstruction of X-rayimage data from the sameaxial plane taken at differentangleds. It providesdiagnostic details of theanatomic structures byacquiring 360° rotationalimage sequences of oral andmaxillofacial area from aprecise treatment planninginadult and pediatric dentistry.The device is operated andused by physicians, dentists,and X-ray technicians.Same
Perfomance SpecificationPanoramic, cephalometricand computed tomographyPanoramic, cephalometricand computed tomographySame
Input Voltage100-240VAC100-240 VACSame
Tube Voltage60-95KV60-100 kVSimilar(includedwithinrange ofpredicatedproduct)
Tube Current5-11mA5-16 mASimilar(includedwithinrange ofpredicatedproduct)
Focal Spot Size0.5mm0.5 mmSame
Exposure TimeMax. 22sMax. 22 sSame
2D Image Viewing ProgramModel Name: 2D02Model Name: 2D01, 2D02Similar
(includedwithinpredicatedproduct)
3D Image Viewing ProgramModel Name: 3D02Model Name: 3D01, 3D02Similar(includedwithinpredicatedproduct)
Anatomical SitesMaxillofacialMaxillofacialSame
ImageSensorComputedTomography1515DXTSEN11515DXTSEN1Same
Panoramic1515DXTSEN11515DXTSEN1Same
CephalometricXID-C24DCXID-C24DCSame
1515DXT : Max. 15x9SEN1 : Max 15x10(Stiching Mode : Max 15 X 15)1515DXT : Max. 15x9SEN1 : Max 15x10Similar(includedwithinrange ofpredicatedproduct)
Size of Imaging Volume(cm)
PixelResolutionComputedTomography1515DXT :- 3.94 lp/mmSEN1 :- 4.0 lp/mm1515DXT :- 3.94 lp/mm- 1.97 lp/mmSEN1 :- 4.0 lp/mm- 2.0 lp/mmSame
Panoramic1515DXT : 3.94 lp/mmSEN1 : 4.0 lp/mm1515DXT : 3.94 lp/mmSEN1 : 4.0 lp/mmSame
CephalometricXID-C24DC : 5 lp/mmXID-C24DC : 5 lp/mmSame
Pixel SizeComputedTomography1515DXT :- 127 μm * 127 μmSEN1 :- 125 μm * 125 μm1515DXT :- 127 μm * 127 μm- 254 μm * 254 μmSEN1 :- 125 μm * 125 μm- 250 μm * 250 μmSimilar(includedwithinrange ofpredicatedproduct)
Panoramic1515DXT :- 127 μm * 127 μmSEN1 :- 125 μm * 125 μm1515DXT :- 127 μm * 127 μmSEN1 :- 125 μm * 125 μmSame
CephalometricXID-C24DC- 100 μm * 100 μmXID-C24DC- 100 μm * 100 μmSame
Active Area[CT]- 1515DXT : 146 x 146 mm- SEN1 : 160 x 128 mm[Pano]- 1515DXT : 146 x 8.13 mm- SEN1 : 160 x 8 mm[CT]- 1515DXT : 146 x 146 mm- SEN1 : 160 x 128 mm[Pano]- 1515DXT : 146 x 8.13 mm- SEN1 : 160 x 8 mmSame

5. Substantial Equivalence Matrix

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font above the word "IMPLANT" in a smaller, lighter font. The logo is simple and modern.

Osstem Implant Co., Ltd.

OSSTERM IMPLANT CO., Ltd.
2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Gibentg-si, Gyeonegi-do, 15079, Republic of Korea
Tel: +82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. The logo is simple and modern.

n Implant Co., Ltd

on-gil. Siheung-si. Gyeonggi-do. 15079. Republic of Korea Fax: +82 70 4394 0404 www.osstem.com

[Ceph]- 1.8 x 150 mm[Ceph]- 1.8 x 150 mm
------------------------------------------------------

6. Indications for Use

  • T2 is digital X-Ray imaging equipment for dental professionals that converts X-Ray signals into digital signals to acquire 2D images and reconstruct them into 3D images, using Panoramic (Pano), Cephalometric (Ceph), and Computed Tomography (CT) technology for the diagnosis of the anatomical structure of the oral and maxillofacial area.

7. Summary of Non-clinical Performance Testing

Software validation and verification test, EMC/Electrical Safety tests, and performance tests were conducted and the test results support that the subject device is substantially equivalent to the predicate device

Non-Clinical Performance Data

We performed the applicable non-clinical tests in the FDA guidance document for solid state x-ray detectors.

Software Verification and Validation Testing

Software was designed and developed according to a software development process. Also, Software verification and validation test were conducted and documented in accordance with FDA guidance: "The content of premarket submissions for software contained in medical devices, on May 11, 2005". The software for this device contains "Moderate" level of concern software.

Safety, EMC and Performance Data

T2 complies with the electrical safety and electromagnetic compatibility requirements established by the standards IEC60601-1, IEC60601-1-2.

8. Summary of Clnical Testing

No clinical studies are submitted.

9. Conclusion

The indication for use and the technical characteristics of the subject device is substantially equivalent to the predicate device. It is substantially equivalent to predicate device in output characteristics and operation mode (CT, Cephalo, Panorama). The major difference is size of image volume however, the non-clinical and clinical study results show that the subject device is substantially equivalent to the predicate device. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and Based

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a stylized font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.

Osstem Implant Co., Ltd.

2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea
Tel: +82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem.com

on the information provided in this submission, we conclude that the subject device is safe and effective and substantially equivalent to the predicate device.

N/A