K Number
K212202

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-04-13

(273 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

WiScope® Digital Cystoscope System is intended to be used by physicians to access, visualize, and perform procedures in the bladder and lower ureter. The instrument enables delivery and use of accessories such as biopsy forceps, laser fibers, graspers and retrieval baskets at a surgical site.

Device Description

WiScope® Digital Cystoscope System is designed for physicians to access, visualize, and perform procedures in the bladder and lower ureter for diagnosis and treatment. This system includes a single-use digital cystoscope and an image system OTU-A. The single-use cystoscope is comprised of a control body with articulation controls and accessory access ports, and a flexible insertion tube with an on-tip camera module and LED lighting source.

AI/ML Overview

The provided text describes the WiScope® Digital Cystoscope System, which is a medical device. The document is a 510(k) premarket notification summary submitted to the FDA.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists numerous tests conducted, and for all of them, the result is "All tests were passed" or "All evaluation acceptance criteria were met." Specific numerical acceptance criteria are not detailed in the provided text, but the claim of passing suggests they were defined and met.

Test ItemAcceptance Criteria/StandardReported Device Performance
Electrical SafetyIEC 60601-1, IEC 60601-2-18All evaluation acceptance criteria were met.
Electromagnetic CompatibilityIEC 60601-1-2All evaluation acceptance criteria were met.
BiocompatibilityISO 10993-1 (Cytotoxicity, Sensitization, Intracutaneous reactivity test)All tests were passed.
Shelf LifeASTM F1980 (Performance test, Sterile Packaging Integrity)All tests were passed.
Sterile Barrier SystemsISO 11607-1:2019All tests were passed.
Sterilization ProcessISO 11135:2014 (SAL 10⁻⁶)The WiScope® Single-Use Digital Flexible Cystoscope is sterile with a determination of lethal rate of the sterilization process to demonstrate achievement of the required SAL of 10⁻⁶.
Technological Characteristics (Performance Verification)ISO8600, ISO 12233, and performance characteristics relevant to functions as intended (e.g., Appearance, Working length, Bending radius, Channel diameter, Outer diameter, Depth of field, Field of view, Direction of view, Distal Tip Temp, Tensile Force, Resolution, Illumination, Angulation range, Working channel leakage, Waterproof, Flow rate)All tests were passed.
Noise and Dynamic RangeISO 15739All evaluation acceptance criteria were met.
Simulated Shipping Distribution (aged devices)ASTM D4169-16All tests were passed.
Photobiological SafetyIEC 62471:2006All tests were passed.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

The document does not specify the sample sizes used for any of the non-clinical tests. It states that "These tests only used to evaluate non-clinical testing of the subject device." There is no mention of country of origin for the data or whether the studies were retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

This information is not applicable to the provided document. The studies described are non-clinical performance and safety tests, not studies requiring expert interpretation of medical images or diagnostic outcomes. Therefore, there is no mention of experts establishing ground truth in this context.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable. As mentioned above, the studies are non-clinical performance and safety tests, not studies involving human interpretation or adjudication of medical results.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical study is included in this submission." The submission focuses on demonstrating substantial equivalence through non-clinical testing and comparison to a predicate device. The device itself (a cystoscope) is an imaging tool, not an AI-powered diagnostic system that would typically be evaluated for reader improvement.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable. The WiScope® Digital Cystoscope System is described as a physical medical device (an endoscope) for visualization and procedures, not an algorithm or AI system for standalone performance evaluation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The concept of "ground truth" as it relates to expert consensus or pathology is not applicable to the non-clinical tests described. The ground truth for these tests would be the established scientific standards, specifications, and physical properties specified in the referenced ISO and ASTM standards. For example, for "Working length of the shaft," the ground truth would be the defined manufacturing specification for the length, and the test verifies if the actual device meets that specification.

8. The sample size for the training set:

This information is not applicable as there is no mention of an algorithm or AI model requiring a training set in the document.

9. How the ground truth for the training set was established:

This information is not applicable for the reason stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 13, 2022

OTU Medical, Inc. % Mingzi Hussey Principal Regulatory Consultant Zi-medical, Inc. 93 Springs Rd Bedford, MA 01730

Re: K212202

Trade/Device Name: WiScope® Digital Cystoscope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ Dated: June 29, 2021 Received: March 10, 2022

Dear Mingzi Hussey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212202

Device Name WiScope® Digital Cystoscope System

Indications for Use (Describe)

WiScope® Digital Cystoscope System is intended to be used by physicians to access, visualize, and perform procedures in the bladder and lower ureter. The instrument enables delivery and use of accessories such as biopsy forceps, laser fibers, graspers and retrieval baskets at a surgical site.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Sponsor

OTU Medical Inc. 2231A Fortune Drive, San Jose, CA 95131 Phone: (408) 797-7313 Contact: Geping Liu Email: info@@otumed.com

2. Submission Correspondent

Mingzi Hussey Zi-medical, Inc. Address: 93 Springs Rd, Bedford, MA 01730 US Phone: 206-981-0675 Email: mingzi@zi-medical.com

3. Date Prepared

Feb 28th, 2022

4. Device Identification

Trade/Proprietary Name: WiScope® Digital Cystoscope System Common Name: Cystoscope and Accessories, Flexible/rigid Classification Name: Endoscope and Accessories Product Code: FAJ Regulation Number: 21 CFR 876.1500 Regulation Class: Class II Review Panel: Gastroenterology/Urology

5. Predicate Devices

The proposed devices are substantially equivalent to the following predicate

ApplicantDevice name510(k) NumberProduct code
Ambu Inc.Ambu® aScope™ 4 CystoK193095FAJ

6. Device Description

WiScope® Digital Cystoscope System is designed for physicians to access, visualize, and perform procedures in the bladder and lower ureter for diagnosis and treatment. This system includes a single-use digital cystoscope and an image system OTU-A.

{4}------------------------------------------------

  • The Image system OTU-A is currently marketed in the United States by OTU Medical Inc. ● under 510(k) # K181977.
  • The single-use cystoscope is comprised of a control body with articulation controls and accessory access ports, and a flexible insertion tube with an on-tip camera module and LED lighting source.

● WiScope® Single-Use Digital Flexible Cystoscope has the following 4 models: OTU-C300S: Standard Deflection Model for OTU-C300 series OTU-C300R: Reverse Deflection Model for OTU-C300 series OTU-C310S: Standard Deflection Model for OTU-C310 series OTU-C310R: Reverse Deflection Model for OTU-C310 series Note:

  • Standard Deflection Models: Pushing the deflection lever forward articulates the distal tip "up" and pushing the lever back articulates the distal tip "down", as shown in Figure 1.

  • Reverse Deflection Models: Pushing the deflection lever forward articulates the distal tip "down" and pushing the lever back articulates the distal tip "up", as shown in Figure 1.

Image /page/4/Picture/7 description: The image shows a collection of line drawings, arranged in a grid-like format. Each drawing features a tool with a handle and a different type of end. Some ends are curved, while others are straight and have small markings. The drawings appear to be technical illustrations or diagrams of some sort of medical or surgical instrument.

Standard Deflection Models

Reverse Deflection Models

Figure 1. Illustration of standard deflection models vs. reverse deflection models.

The structure design and parts of the standard deflection models are the same as those of the reverse deflection model, i.e., OTU-C300S vs. OTU-C300R, OTU-C310S vs. OTU-C310R.

6.1 Specifications of the WiScope® Single-Use Digital Flexible Cystoscope
ItemOTU-C300S/OTU-C300ROTU-C310S/OTU-C310R
Size of CMOS image sensor1mm×1mm1mm×1mm
Optical format of CMOS image1/18"1/18"

{5}------------------------------------------------

OTU Medical Inc.
2231A Fortune Drive, San Jose, CA 95131
2231A Fortune Drive, San Jose, CA 95131
VideoCamerasensor400(H)×400(V)pixels400(H)×400(V)pixels
Resolution of CMOS image sensor
Optical Resolution4lp/mm4lp/mm
Field of view100°100°
Direction of viewForwardForward
Depth of field2 – 50 mm2 – 50 mm
LightSourcePower rating and componentspecification of the LED light source54 mW54 mW
Ratio of luminous energytransmitted to energy delivered100 lm/W100 lm/W
ShaftInsertion tube typeFlexibleFlexible
Distal tip diameter10Fr9Fr
Maximum outer diameter ofinsertion portion16Fr15Fr
Insertion tube outer diameter15.3Fr13.8Fr
Working length of shaft380 mm380 mm
Angulation range210° Up / 210°Down210° Up / 210°Down
WorkingChannelWorking channel diameter6.6 Fr6.6 Fr
Flow rate≥100 ml/min≥100 ml/min
Compatible fluidsNormal SalineNormal Saline

Patient Contacting Materials:

Image /page/5/Picture/4 description: The image shows two views of a medical instrument, possibly an endoscope. The left image is a close-up of the tip, labeled with 'A', 'B', and 'C', likely indicating different components or features. The right image shows a longer section of the instrument, labeled 'B, C', 'D', and 'E', suggesting the shaft and other parts of the device.

Figure 2. Photos of patient contacting materials

S/NComponentsMaterial
ALens of camera moduleGlass
BDistal tipPolycarbonate (PC)
CRigid distal sleevePolycarbonate (PC)
DSheath of deflection sectionThermoplastic polyurethanes (TPU)
EShaftPEBAX

{6}------------------------------------------------

S/NComponentsMaterial
1Camera systemLED, Sensor, PCB, electrical wires
2Working channel connecterStainless steel 304
3Working channelPU
4Deflection sectionStainless steel 304
5ShaftPEBAX, Stainless steel 304

6.2 Non-patient contacting materials on shaft:

7. Indication for Use Statement

The subject device: WiScope® Digital Cystoscope System is intended to be used by physicians to access, visualize, and perform procedures in the bladder and lower ureter. The instrument enables delivery and use of accessories such as biopsy forceps, laser fibers, graspers and retrieval baskets at a surgical site.

The predicate device: Ambu® aScope™ 4 Cysto is a sterile, single-use, flexible cystoscope intended to be used for endoscopic access to and examination of the lower urinary tract. The Ambu® aScope™ 4 Cysto is intended to provide visualization via Ambu® displaying unit and can be used with endoscopic accessories. Ambu® aScope™ 4 Cysto is intended for use in a hospital environment or medical office environment. Ambu® aScope™ 4 Cysto is designed for use in adults.

WiScope® Digital Cystoscope System and the primary predicate device are similar in terms of indication for use. WiScope® Digital Cystoscope System is used in the bladder and lower ureter, while predicate device is used in the lower urinary tract. They are all for endoscopic examinations/diagnoses and therapeutic procedures with endoscopic accessories.

8. Comparison of Technological Characteristics

The following table compares the proposed device with the predicate devices in terms of intended use, technological characteristics and principles of operation, and it provides detailed information for determining substantial equivalences.

ITEMProposed DevicePrimary Predicate Device
Trade nameWiScope® Digital Cystoscope SystemAmbu® aScope™ 4 Cysto
510(K)SubmitterOTU Medical Inc.Ambu Inc.
510(K) Number/K193095
ClassificationRegulation21CRF 876.150021CRF 876.1500
Classificationand CodeClass II,FAJClass II,FAJ
Common nameCystoscope and Accessories,Cystoscope and Accessories,
Table 5A - General Comparison

{7}------------------------------------------------

OTU Medical Inc. 2231A Fortune Drive, San Jose, CA 95131

Flexible/rigidFlexible/rigid
UreteroscopeSingle-UseSingle-Use
Intended UseWiScope® Digital Cystoscope System isintended to be used by physicians toaccess, visualize, and performprocedures in the bladder and lowerureter. The instrument enables deliveryand use of accessories such as biopsyforceps, laser fibers, graspers andretrieval baskets at a surgical site.Ambu® aScope™ 4 Cysto is asterile, single-use, flexiblecystoscope intended to be usedfor endoscopic access to andexamination of thelower urinary tract. The Ambu®aScope™ 4 Cysto is intended toprovide visualization via Ambu®displaying unit and can be usedwith endoscopic accessories.Ambu® aScope™ 4 Cysto isintended for use in a hospitalenvironment or medical officeenvironment.Ambu® aScope™ 4 Cysto isdesigned for use in adults.
Digital videotechnologyCMOSCMOS
IlluminationLEDLED
Field of View100°120°
WorkingLength (mm)380390
Insertion tubeouter diameterOTU-C300 series: 15.3FrOTU-C310 series: 13.8Fr16.2Fr
WorkingChannelDiameter (Fr)6.6Fr6.6Fr
Up/DownDeflectionUP: 210°DOWN: 210°UP: 210°DOWN: 120°
SterilizationEOSAL: 10-6EOSAL: 10-6
PackagingOTU-C300 series: packaged bypaper-plastic bag.OTU-C310 series: packaged in a traywhich is sealed by sterile barrier.Cystoscope is packaged bypaper-plastic bag.
Label andLabelingMeet FDA's RequirementsMeet FDA's Requirements
CytotoxicityComply with ISO 10993-5, nocytotoxicity effectComply with ISO 10993-5, nocytotoxicity effect
IrritationComply with ISO 10993-10, not anirritantComply with ISO 10993-10, notan irritant
SensitizationComply with ISO 10993-10, not aComply with ISO 10993-10, not a
sensitizer.a sensitizer.

{8}------------------------------------------------

WiScope® Single-Use Digital Flexible Cystoscope has two series, OTU-C300 series and OTU-C310 series.

The differences between WiScope® Single-Use Digital Flexible Cystoscope OTU-C300 series and WiScope® Single-Use Digital Flexible Cystoscope OTU-C310 series have been described in 012 Section 10 Device Description in this 510K submission.

The proposed device shares the similar indications for use, same device operation, and overall technical and functional capabilities as the predicate device. It also has the same standards and requirements as the predicate device.

9. Description of Non-clinical Testing

The non-clinical tests of the subject device and predicate device are in compliance with the following standards and guidances.

ITEMProposed DevicePrimary Predicate Device
Trade nameWiScope® Digital CystoscopeAmbu® aScope™ 4 Cysto
System
Safety Testing*AAMI / ANSI ES60601-1:2012*IEC 60601-2-18:2009
*AAMI / ANSI / IEC 60601-1-2:2014*ISO 10993-5:2009
*IEC 60601-2-18:2009*ISO 10993-10:2010
*IEC IEC62471:2006*ISO 11135:2014
*AAMI / ANSI / ISO 10993-5:2009*ISO 8600-1-2015
*ISO 10993-10:2010*ISO 8600-3-2019
*ISO 11135:2014*ISO 8600-4-2014
*ISO 10993-7:2008
*ISO 11607-1:2019
*ASTM F1980-16
*ISO 11737-2:2019
*ASTM D3078-2013
*ASTM F1929-15
*DIN 58953-6:2016
*ASTM F88/F88M-15
*ISO 8600-1-2015
*ISO 8600-3-2019
*ISO 8600-4-2014
*ISO 8600-6-2005
*ASTM D4169-16

WiScope® Digital Cystoscope System has been verified for its safety and effectivity based on the following performance data. These tests only used to evaluate non-clinical testing of the subject device.

Electrical safety of the system was evaluated in accordance with IEC 60601-1and IEC 60601-2-18. Electromagnetic compatibility was evaluated in accordance with IEC 60601-1-2.

{9}------------------------------------------------

All evaluation acceptance criteria were met.

The biocompatibility evaluation for WiScope® Single-Use Digital Flexible Cystoscope was conducted in accordance with ISO 10993-1.

  • · Cytotoxicity (ISO 10993-5)
  • · Sensitization (ISO 10993-10)
  • · Intracutaneous reactivity test (ISO 10993-10)

Result: All tests were passed.

Shelf life test is conducted based on ASTM F1980:

  • Performance test
  • · Sterile Packaging Integrity

Result: All tests were passed.

Sterile barrier systems were evaluated in accordance with ISO 11607-1:2019.

Sterilization Process has been validated accordance with ISO 11135:2014. Result: The WiScope® Single-Use Digital Flexible Cystoscope is sterile with a determination of lethal rate of the sterilization process to demonstrate achievement of the required SAL of 10® is in accordance to half cycle overkill approach as described in Section 14 Sterilization and Shelf Life.

Technological characteristics have been tested for its functions as intended including verification of performance characteristics per ISO8600, ISO 12233 and performances characteristics relevant to functions as intended:

  • · Appearance
  • Working length of the shaft
  • Minimum bending radius
  • Working channel diameter
  • Outer Shaft Diameter
  • · Depth of field
  • Field of view
  • Direction of view
  • Rigid Distal Tip Temperature
  • · Tensile Force Testing
  • Resolution
  • Illumination
  • Angulation range
  • Working channel free from leakage
  • Waterproof
  • Flow rate of water

Result: All tests were passed.

Noise and dynamic range test were conducted based on ISO 15739. Result: All evaluation acceptance criteria were met.

{10}------------------------------------------------

Simulated shipping distribution on aged devices test were conducted based on ASTM D4169-16. Result: All tests were passed.

Photobiological safety test were conducted based on IEC 62471:2006. Result: All tests were passed.

The results of Non-Clinical Performance testing demonstrate that the WiScope® Digital Cystoscope System is considered safe and effective for its intended use.

10. Performance Data - Clinical

No clinical study is included in this submission.

11. Conclusion

lt has been shown in this 510(k) submission that the differences between the proposed device and the predicate device do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards demonstrate that the proposed are substantially equivalent to the relevant aspects of the predicate device in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.

Therefore, the proposed devices are determined to be substantially equivalent to the referenced predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.