K Number
K212065

Validate with FDA (Live)

Date Cleared
2021-10-07

(97 days)

Product Code
Regulation Number
870.2100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Remington Medical, Inc. VascuChek™ Clinical Device is intended for the non-invasive transcutaneous evaluation of blood flow in Peripheral Vasculature.

Device Description

VascuChek™ is a cardiovascular blood flowmeter comprised of two components: a nonsterile, single-use VascuChek™ Clinical Probe which connects to the reusable, nonsterile VascuChek™ Transceiver.

The VascuChek™ device follows Track 1.

A transmitter in the transceiver drives the ultrasonic transmitting crystal located at the tip of the probe component. The ultrasonic waves generated by the sensor travel through the tissue just under the probe tip in a narrow beam. The reflected ultrasonic waves are received by the transducer and are converted via the piezoelectric effect into a high frequency electronic signal. The received electronic signal is amplified and detected. The result is a base band audio Doppler shifted signal which is filtered and converted to audio via a speaker. During the intervals when the unit is not transmitting, the device passes any reflected signals that it receives to a receiving circuit. This circuit amplifies the returning echoes, compares their frequency to that of the transmitted signal and converts any frequency differences into an audible tone.

AI/ML Overview

The provided document is a 510(k) summary for the Remington Medical, Inc. VascuChek™ Clinical Device. The product is a cardiovascular blood flowmeter intended for non-invasive transcutaneous evaluation of blood flow in peripheral vasculature. The submission demonstrates substantial equivalence to the predicate device, Vascular Technology Incorporated (VTI) Surgical Doppler (K082870).

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state "acceptance criteria" for performance metrics in a quantitative manner (e.g., minimum sensitivity or specificity). Instead, the performance evaluation is based on demonstrating substantial equivalence to a predicate device through various non-clinical and an in vivo simulated use test. The criteria for acceptance are implied by the successful completion of these tests and a conclusion of substantial equivalence.

Acceptance Criteria Category (Implied)Reported Device PerformanceStudy Type
Software/Firmware PerformanceTests performed and met.Non-clinical
Mechanical PerformanceTests performed and met.Non-clinical
BiocompatibilityTests performed and met.Non-clinical
Device LifecycleTests performed and met.Non-clinical
Reprocessing (Cleaning and Low-Level Disinfection)Tests performed and met.Non-clinical
Distribution SimulationTests performed and met.Non-clinical
Electrical SafetyTests performed and met.Non-clinical
EMC (Electromagnetic Compatibility)Tests performed and met.Non-clinical
Environmental Performance TestingTests performed and met.Non-clinical
Audio Quality in Blood Flow Velocity MeasurementAssessed and compared performance of the new device to its predicate at different vessel depths/sizes. The results contributed to the conclusion of substantial equivalence.In vivo Simulated Use

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions an "in vivo Simulated Use" test. However, it does not specify the sample size used for this test (e.g., number of subjects, number of measurements). It also does not provide details about the data provenance beyond stating it was an in vivo test. Therefore, information regarding country of origin of the data or whether it was retrospective or prospective is not available in the provided text.

3. Number of Experts Used to Establish the Ground Truth and Qualifications:

The document does not mention the use of experts to establish ground truth for any test set. The in vivo simulated use test focused on assessing audio quality and comparing performance to the predicate device, implying that the predicate's output or a direct measurement was used as a reference rather than expert consensus on a diagnostic outcome.

4. Adjudication Method for the Test Set:

Since the document does not describe the use of multiple experts or a diagnostic task requiring consensus, there is no mention of an adjudication method (e.g., 2+1, 3+1).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or reported. The study focused on demonstrating substantial equivalence of the device's technical characteristics and performance to a predicate device, not on how human readers' performance might improve with or without AI assistance. The device itself is a blood flowmeter, not an AI-assisted diagnostic tool for interpretation by human readers in the traditional sense of an MRMC study.

6. Standalone Performance:

The primary focus of the performance data section is on the standalone performance of the VascuChek™ Clinical Device, both through non-clinical bench testing and the in vivo simulated use test. The purpose was to show that the device itself performs comparably to the predicate device.

7. Type of Ground Truth Used:

For the in vivo simulated use test, the ground truth appears to be established by comparison to the predicate device in terms of measuring the velocity of blood flow. The phrasing "compare performance of the new device to its predicate" suggests the predicate's output served as the reference for evaluating the subject device's audio quality in measuring blood flow velocity. There is no indication of pathology, outcomes data, or expert consensus being used as ground truth for this device's specific function.

8. Sample Size for the Training Set:

The document does not mention a training set because the VascuChek™ Clinical Device is a cardiovascular blood flowmeter, not a device that employs machine learning or artificial intelligence requiring a training set in the typical sense.

9. How the Ground Truth for the Training Set Was Established:

As there is no mention of a training set, there is no information on how its ground truth might have been established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 7, 2021

Remington Medical, Inc. C. Matt Brown VP of Quality and Regulatory Affairs 6830 Meadowridge Court Alpharetta, Georgia 30005

Re: K212065

Trade/Device Name: VascuChek Kit, VascuChek Transceiver, VascuChek Clinical Probe, VascuChek Charger Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: September 10, 2021 Received: September 13, 2021

Dear C. Matt Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212065

Device Name VascuChek™ Clinical Device

Indications for Use (Describe)

The Remington Medical, Inc. VascuChek™ Clinical Device is intended for the non-invasive transcutaneous evaluation of blood flow in Peripheral Vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Remington Medical. The logo consists of a blue geometric shape on the left and the words "Remington MEDICAL" on the right. The word "Remington" is in a larger, bolder font than the word "MEDICAL", which is stacked below it.

Preparation DateJuly 1, 2021 – revised on September 8, 2021
ApplicantRemington Medical, Inc.6830 Meadowridge Court,Alpharetta, GA, USA 30005Registration Number:Owner/Operator Number:10565539006473
Contact PersonC. Matt Brown, MS, CQE, RACVP of Quality and Regulatory Affairs470-719-1121mattb@remmed.com
Trade ProprietaryName(s)Remington Medical, Inc. VascuChek™ Clinical Device
Common Name (s)Flowmeter, Blood, Cardiovascular
Classification Name21 CFR 870.2100 Cardiovascular blood flowmeterProduct Code: DPW
Device Class:II

Section 5 - 510(k) Summary

Predicate Device:

Vascular Technology Incorporated (VTI) Surgical Doppler (K082870)

Description of the Device:

VascuChek™ is a cardiovascular blood flowmeter comprised of two components: a nonsterile, single-use VascuChek™ Clinical Probe which connects to the reusable, nonsterile VascuChek™ Transceiver.

The VascuChek™ device follows Track 1.

A transmitter in the transceiver drives the ultrasonic transmitting crystal located at the tip of the probe component. The ultrasonic waves generated by the sensor travel through the tissue just under the probe tip in a narrow beam. The reflected ultrasonic waves are received by the transducer and are converted via the piezoelectric effect into a high frequency electronic signal. The received electronic signal is amplified and detected. The result is a base band audio Doppler shifted signal which is filtered and converted to audio via a speaker. During the intervals when the unit is not transmitting, the device passes any reflected signals that it receives to a receiving circuit. This circuit amplifies the returning echoes, compares their frequency to that of the transmitted signal and converts any frequency differences into an audible tone.

Intended Use/Indications for Use

The Remington Medical, Inc. VascuChek™ Clinical Device is intended for the noninvasive transcutaneous evaluation of blood flow in Peripheral Vasculature.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Remington Medical. The logo consists of a blue geometric shape on the left and the words "Remington MEDICAL" on the right. The word "Remington" is in a larger font than the word "MEDICAL". The geometric shape is a stylized cross or plus sign.

Comparison to Predicate Device:

The technological characteristics (design, specifications, materials, and performance) of the subject device and the predicate device are substantially equivalent.

Subject Device:Remington Medical, Inc.VascuChek™ Clinical DevicePredicate Device:Vascular TechnologyIncorporated (VTI) SurgicalDoppler (K082870)
Device ClassClass IIClass II
FDA Product CodeDPWDPW
Regulation21 CFR 870.2100 (Flowmeter, Blood, Cardiovascular)21 CFR 870.2100 (Flowmeter, Blood, Cardiovascular)
Intended UseThe Remington Medical, Inc. VascuChek™ Clinical Device is intended for the non-invasive transcutaneous evaluation of blood flow in Peripheral Vasculature.The Vascular Technology Incorporated (VTI) Intraoperative Doppler Systems are intended for the intraoperative and transcutaneous evaluation of blood flow.
Indications for Use StatementThe Remington Medical, Inc. VascuChek™ Clinical Device is intended for the non-invasive transcutaneous evaluation of blood flow in Peripheral Vasculature.The VTI Intraoperative Doppler Systems are intended for the intraoperative and transcutaneous evaluation of blood flow for the following clinical applications:• Intraoperative (Microvascular and Vascular),• Intraoperative Neurological Transesophageal, Transrectal, Laparoscopic and Peripheral Vascular
Type of UsePrescription UsePrescription Use
Use EnvironmentHospital, Outpatient Surgery CenterHospital, Outpatient Surgery Center
Patient PopulationAdults OnlyVascular Patients
Installation and UseHand-Held (Transceiver and Probe)Hand Held (Probe)
Theory of OperationUse of the Doppler effect to evaluate the flow velocity of blood in peripheral vasculature.Use of the Doppler effect to evaluate the flow velocity of blood in peripheral vasculature.
Frequency9 MHz8 MHz and 20 MHz
Global Maximum Outputs / Worst Case SettingISPTA 3 (mW/cm²)359.7 mW/cm²< 94 mW/cm²
Mode of OperationUltrasonic Doppler/Continuous WaveUltrasonic Doppler/ Pulsed Wave
ReusableProbe No, Single UseProbe No, Single Use
Transceiver Yes, Cleaning/Low Level DisinfectionTransceiver Yes, Cleaning
Sterilization MethodNot Applicable - Transceiver and Probe are nonsterile.Probe ETO

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Remington Medical. The logo consists of a blue geometric shape on the left and the words "Remington MEDICAL" in blue on the right. The word "Remington" is in a larger font than the word "MEDICAL", and it is placed above the word "MEDICAL".

Dimensions221x 33 x 31 mm(8.7 in x 1.29 in x 1.22 in.)6.5 in. D x 10 in. W x 4 in. H
Weight90 grams (0.198 lb)2.6 lb, nom
The degree of protectionProbeProbe
against harmful ingress ofliquid (IPX rating)TransceiverCharger1Transceiver0
Power SupplyRechargeable battery assembly withAC to DC ChargerBatteries or External Power Source(AC to DC)
Battery Operating Voltage6.4 VDC12 VDC
Battery Chemistry1 - LiFePO4 rechargeable batteryassembly8 AA (LR6) alkaline batteries
The degree of protectionagainst electric shockProbeType CFProbeTransceiverType CF
ButtonsThree - Power, Volume Up, andVolume DownFive – Power, Volume Up, VolumeDown, Channel A, Channel B
Status LEDOne: Power and Battery IndicatorSix: Four indicate power on andvolume, one indicates low batteries,one indicates active channel
Calibration RequiredNoNo
MaintenanceNoNo

Performance Data:

To demonstrate that the subject device, Remington Medical Inc. VascuChek™ Clinical Device, is as safe and as effective as the predicate device, Vascular Technology Incorporated (VTI) Surgical Doppler, technological characteristics and performance criteria were evaluated. The following tests were performed on the subject device:

  • Software/Firmware Performance ●
  • Mechanical Performance ●
  • Biocompatibility
  • Device Lifecycle ●
  • Reprocessing Manual Cleaning and Low Level Disinfection ●
  • Distribution Simulation ●
  • Electrical Safety ●
  • EMC
  • Environmental Performance Testing

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Remington Medical. The logo consists of a blue geometric shape on the left and the words "Remington MEDICAL" on the right. The word "Remington" is in a larger font than the word "MEDICAL". The geometric shape is made up of several blue triangles that form a cross.

In addition, the following in vivo tests were performed on the subject device and predicate device:

  • . Simulated Use – assess audio quality in measuring the velocity of blood flow at different vessel depths/sizes and compare performance of the new device to its predicate
    The performance data (design, specifications, materials, and performance) of the Subject Device and the Predicate Device are substantially equivalent.

Clinical testing:

No clinical testing was required.

Conclusion:

The results of the non-clinical testing demonstrated that the subject device, Remington Medical Inc. VascuChek™ Clinical Device, is substantially equivalent to the predicate device, Vascular Technology Incorporated (VTI) Surgical Doppler, with respect to intended use, design, materials, and technological characteristics.

§ 870.2100 Cardiovascular blood flowmeter.

(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).