K Number
K212047

Validate with FDA (Live)

Device Name
Pantheris System
Manufacturer
Date Cleared
2021-11-16

(139 days)

Product Code
Regulation Number
870.4875
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pantheris System is intended to remove plaque (atherectomy) from partially occluded native or restenotic vessels, including in-stent restenosis (ISR), in the peripheral vasculature with a reference diameter of 3.0 mm to 7.0 mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures, and vessel morphologies. The Pantheris system is not intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

Device Description

The Pantheris System consists of a Pantheris atherectomy catheter, a Lightbox Sled with integrated umbilical (referred to as "Sled"), and a Lightbox HS Imaging Console (referred to as "Lightbox"). The Pantheris atherectomy catheter is a 7 French device with a working length of 110 cm. It incorporates an optical fiber that allows real-time diagnosis of vessel condition and morphology as well as OCT-guided atherectomy during the procedure with its connection to the Lightbox via the Sled accessory.

AI/ML Overview

The following provides an analysis of the acceptance criteria and study findings for the Pantheris System, based on the provided FDA 510(k) summary.

1. Table of Acceptance Criteria and Reported Device Performance

EndpointAcceptance Criteria (Performance Goal)Reported Device Performance (Pantheris System)
Primary Safety: Freedom from Major Adverse Events (MAE) at 30 daysMAEs occurring in < 20% of subjects97% freedom from MAEs (3% experienced MAE); 95% one-sided upper confidence bound of 6.5%. Met.
Secondary Safety: Absence of new or worsening stent fractureNot established with a sample size requirement; extremely low incidence expected1% (1 out of 97 procedures) inadvertently made contact with a stent. Met (due to low incidence and subsequent physician retraining efficacy).
Primary Effectiveness: Technical Success (residual stenosis ≤ 50% post-Pantheris alone)> 79%89% (86 out of 97 subjects) had <50% residual stenosis; 95% one-sided upper confidence bound of 95% and lower confidence bound of 82%. Met.
Secondary Effectiveness (Powered): Freedom from Target Lesion Revascularization (TLR) at 6 months> 61%93% (79 out of 85 subjects) freedom from TLR; 95% one-sided upper confidence bound of 98% and lower bound of 87%. Met.
Secondary Effectiveness: Procedural Success (residual stenosis ≤ 30% post-Pantheris + adjunctive therapy)Not explicitly defined as a performance goal with a specific threshold.80% (78 out of 97 subjects) had residual stenosis < 30%; mean stenosis of 15% ± 10.1%.
Secondary Effectiveness: Change in Ankle-Brachial Index (ABI)Not explicitly defined as a performance goal with a specific threshold.ABI improved 39% from baseline at 6 months.
Secondary Effectiveness: Change in Rutherford ClassesNot explicitly defined as a performance goal with a specific threshold.Rutherford Classification improved 71% at 6 months.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: A total of 97 subjects were enrolled in the INSIGHT trial.
  • Data Provenance: The study was a prospective, multi-center trial conducted across 17 investigational sites in the USA (n=15) and EU (n=2). Therefore, the data is prospective and from both the US and Europe.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The document specified that the primary safety endpoint (freedom from MAEs) was adjudicated by an independent Clinical Events Committee (CEC).
  • The primary effectiveness endpoint (technical success) and secondary effectiveness endpoint (procedural success) involving residual diameter stenosis were assessed by an independent angiographic core laboratory.
  • The specific number or qualifications of experts within the CEC or the core laboratory are not detailed in this summary.

4. Adjudication Method for the Test Set

  • For the primary safety endpoint (MAEs), an independent Clinical Events Committee (CEC) was used for adjudication. The specific method (e.g., 2+1, 3+1) is not provided, but the independence suggests a standardized, expert-driven review process.
  • For effectiveness endpoints related to stenosis, an independent angiographic core laboratory assessed the outcomes. This implies multiple expert readers are likely involved in standardized image analysis, but the exact adjudication method is not specified.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was reported. The study focused on the performance of the Pantheris System itself, a medical device for atherectomy, not an AI-assisted diagnostic or interpretative system.

6. Standalone Performance

  • The study was a "single-arm trial" to determine the safety and performance outcomes of the Pantheris System.
  • The primary effectiveness endpoint specifically refers to "residual diameter stenosis ≤ 50% after use of the Pantheris device alone," indicating that standalone performance (algorithm only without human-in-the-loop performance) was assessed for this specific aspect of effectiveness, meaning the intervention of the device itself.
  • The "algorithm" in this context refers to the device's mechanism of action, as it's an atherectomy system, not a software-based AI algorithm.

7. Type of Ground Truth Used

  • Clinical Outcomes: The ground truth relied on direct clinical outcomes assessed during the study, including:
    • Occurrence of Major Adverse Events (MAEs).
    • Angiographic measurements of residual stenosis (assessed by an independent core lab).
    • Target Lesion Revascularization (TLR) rates.
    • Changes in physiological markers (ABI) and clinical classification (Rutherford Classes).
  • Expert Adjudication: For MAEs and angiographic assessments, the "ground truth" was established by expert adjudication (independent CEC and independent angiographic core laboratory).
  • Histopathology/Pathology: For preclinical animal studies, pathology (light microscopy, radiography, and SEM) was used to assess acute treatment effects on arteries and stents.

8. Sample Size for the Training Set

  • The document describes a clinical trial (INSIGHT Trial) which is a clinical validation study for expanded indications, not a training set for an AI algorithm.
  • No training set for an AI algorithm is mentioned as the device is a physical atherectomy system, not an AI diagnostic tool.

9. How the Ground Truth for the Training Set Was Established

  • As the device is an atherectomy system and not an AI algorithm, there is no mention of a "training set" or "ground truth for a training set" in the context of machine learning. The clinical trial data serves as validation for the device's performance against pre-defined clinical endpoints.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 16, 2021

Avinger, Inc. Thomas Lawson VP. Clinical & Regulatory Affairs 400 Chesapeake Drive Redwood City, California 94063

Re: K212047

Trade/Device Name: Pantheris System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW, NQQ Dated: October 6, 2021 Received: October 8, 2021

Dear Thomas Lawson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212047

Device Name Pantheris System

Indications for Use (Describe)

The Pantheris System is intended to remove plaque (atherectomy) from partially occluded native or restenotic vessels, including in-stent restenosis (ISR), in the peripheral vasculature with a reference diameter of 3.0 mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures, and vessel morphologies. The Pantheris System is not in the iliac, coronary, cerebral, renal or carotid vasculature.

Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K212047

510(k) Summary

General Information

SubmitterAvinger, Inc.
Address400 Chesapeake DriveRedwood City, CA 94063
FDA Registration Number3007498664
Correspondence PersonThomas Lawson, PhDDirector, Clinical & Regulatory AffairsAvinger Inc.
Contact InformationEmail: tlawson@avinger.comPhone: 510-206-1794
Date Prepared16 November 2021

Proposed Device

Trade NamePantheris System
Common NamePantheris Catheter
Regulation Number andClassification Name21 CFR§870.4875, Intraluminal Artery Stripper21 CFR§892.1560, Imaging System Optical CoherenceTomography (OCT)
Product CodeMCW, NQQ
Regulatory ClassII

Predicate Device #1—Primary Predicate Device

Trade NameB-Laser Atherectomy System
Common NameB-Laser
Premarket NotificationK181642
Regulation Number andClassification Name21 CFR§870.4875, Intraluminal Artery Stripper
Product CodeMCW
Regulatory ClassII
Note: This predicate device has not been subject to a design-related recall.

{4}------------------------------------------------

Trade NamePantheris System
Common NamePantheris Catheter
Premarket NotificationK173862
Regulation Number andClassification Name21 CFR§870.4875, Intraluminal Artery Stripper21 CFR§892.1560, Imaging System Optical CoherenceTomography (OCT)
Product CodeMCW
Regulatory ClassII
Note: This predicate device has not been subject to a design-related recall.

Predicate Device #2-Reference Predicate Device

Device Description and Proposed Modifications

The Pantheris System consists of a Pantheris atherectomy catheter, a Lightbox Sled with integrated umbilical (referred to as "Sled"), and a Lightbox HS Imaging Console (referred to as "Lightbox"). The Pantheris atherectomy catheter is a 7 French device with a working length of 110 cm. It incorporates an optical fiber that allows real-time diagnosis of vessel condition and morphology as well as OCT-guided atherectomy during the procedure with its connection to the Lightbox via the Sled accessory.

The Pantheris System was reviewed and cleared under K173862. Neither the catheter, the Sled, nor the Lightbox were modified for use in the clinical trial conducted to support this submission and all three elements of the svstem are the same as they were when cleared under K173862.

This Traditional 510(k) is submitted in order to expand the indications for use statement to include in-stent restenosis as a condition for use of the Pantheris catheter.

Proposed Indications for Use Statement

The indication for use for Pantheris atherectomy catheter cleared in K173862 is to be modified to include restenotic vessels, including in-stent restenosis (ISR):

The Pantheris System is intended to remove plaque (atherectomy) from partially occluded native or restenotic vessels, including in-stent restenosis (ISR), in the peripheral vasculature with a reference diameter of 3.0 mm to 7.0 mm, using OCTassisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures, and vessel morphologies. The Pantheris system is not intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

{5}------------------------------------------------

Equivalence of the Subject Device with the Two Predicate Devices

Avinger Inc. has identified the B-Laser Atherectomy System (K181642) as the primary predicate device due to a shared indication to address in-stent restenosis.

The Pantheris catheter is substantially equivalent to the B-Laser Atherectomy System based upon the following similarities:

  • · Both devices are intended to be used to address in-stent restenosis in peripheral vessels by atherectomy, under product code MCW;
  • · Both devices are used in cardiac catheter labs in either a hospital or an officebased lab:
  • · Both devices are advanced to the target in-stent occlusion through an indwelling vascular sheath and over a guide wire;
  • Advancement to the target tissue for both devices is monitored by external fluoroscopy imaging; and
  • · Both devices have equivalent sizes in terms of outer diameter and working length of the catheter.

The Pantheris 7Fr catheter (subject device) and Pantheris 7Fr catheter (reference predicate device) cleared under K173862 are the exact same device. This predicate device covers specifications, materials, biocompatibility, sterilization, and packaging for the device used in the IDE trial to gather clinical data for ISR treatment.

Table 14.1 shows the comparison of the Pantheris 7 Fr catheter to the predicate devices. The equivalences among these three devices satisfy the criteria for a 510(k) submission.

Table 14.1 Comparison of the Pantheris 7Fr catheter (subject device) to the predicate devices, the Laser-B atherectomy catheter (K181642) and the Pantheris 7Fr catheter (K173862).

{6}------------------------------------------------

Subject DevicePrimary Predicate DeviceReference Predicate Device
PantherisAtherectomy System7 French(Avinger, Inc.)(This Submission)B-LaserAtherectomySystem(Eximo MedicalLtd.)K181642PantherisAtherectomy System7 French(Avinger, Inc.)K173862
Overview
ClassIIII
Product CodeMCW NQQMCW
ClassificationSection21 CFR §870.4875Intraluminal ArteryStripperSame
Indication forUseThe Pantheris Systemis intended to removeplaque (atherectomy)from partiallyoccluded native orrestenotic vessels,including in-stentrestenosis (ISR), inthe peripheralvasculature with areference diameter of3.0 mm to 7.0 mm,using OCT-assistedorientation andimaging. The systemis as an adjunct tofluoroscopy byproviding images ofvessel lumen, wallstructures and vesselmorphologies.The Pantheris Systemis NOT intended foruse in the iliac,The B-LaserAtherectomySystem is intendedfor use in thetreatment,includingatherectomy, ofinfra-inguinalstenoses andocclusions,including in-stentrestenosis (ISR).The Pantheris Systemis intended to removeplaque (atherectomy)from partiallyoccluded vessels inthe peripheralvasculature with areference diameter of3.0 mm to 7.0 mm,using OCT-assistedorientation andimaging. The systemis as an adjunct tofluoroscopy byproviding images ofvessel lumen, wallstructures and vesselmorphologies.The Pantheris Systemis NOT intended foruse in the iliac,coronary, cerebral,renal or carotid
MCW NQQ
Same
coronary, cerebral, renal or carotid vasculature.vasculature.
Intended useRemove plaque (atherectomy) from partially occluded native or restenotic vessels, including in-stent restenosis (ISR), in the peripheral vasculatureRemove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature, including in-stent restenosisRemove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature
Treatment MethodAtherectomy using a rotating cutter headAtherectomy with an array of optic fibers surrounded by a circumferential blunt bladeAtherectomy using a rotating cutter head
Technical Characteristics
DesignSystem is comprised of (1) catheter, (2) imaging console, and (3) connector between the catheter and the consoleSystem is comprised of (1) catheter, (2) laser console, and (3) connector between the catheter and the consoleSystem is comprised of (1) catheter, (2) imaging console, and (3) connector between the catheter and the console
Configuration of the CatheterA cannula that contains a rotating cutter head and an OCT imaging fiberA cannula that contains an array of optic fibers surrounded by a circumferential blunt bladeA cannula that contains a rotating cutter head and an OCT imaging fiber
Outer Diameter of the Catheter2.3 mm (7Fr)2.0 and 2.35 mm (6 & X Fr)2.3 mm (7 Fr)
Working Length of the Catheter110 cm125 to 150 cm110 cm
Mechanism of Tissue ExcisionRotating cutter head that cuts the target tissueLaser pulses for photoablation of the target tissueRotating cutter head that cuts the target tissue
Excised Tissue ManagementExcised tissue is collected and contained in an integrated storageExcised tissue is collected and removed using vacuum aspirationExcised tissue is collected and contained in an integrated storage
segment of thecannulasegment of thecannula
Imaging duringthe procedureAngiography forinitial placement withonboard opticalcoherencetomography on thecannulaSameSame
Electrical SafetyClass I, Type CF,defibrillation proofIEC 60601-1SameSame
ElectromagneticCompatibilityIEC 60601-1-2SameSame
LaserClassificationClass IClass IVClass I
Provided SterileYesYesYes
SterilizationMethode-beam irradiationEthylene oxidee-beam irradiation
SterilityAssurance Level10-6SameSame
Single-useCatheterYesYesYes
PackagingPlaced in a traycontained in a TyvekpouchSameSame
ClinicalCharacteristics
Anatomical Siteof Use in theBodyPeripheral VesselsSameSame
ClinicalConditionOccluded vessels andvascular stents in theperipheralvasculatureOccluded vesselsand vascular stentsin the peripheralvasculatureOccluded vessels inthe peripheralvasculature
Population inwhich Device isUsedMale or FemaleAdultsSameSame
Advancement tothe TargetTissueOver an 0.014 inchguide wire through avascular sheathSameSame
SheathCompatibilityfor the Catheter7 Fr4 to 7 Fr7 Fr
MedicalProcedure SiteHospital CardiacCatheter LabSameSame
Office-based Lab
TreatmentMethodAtherectomySameSame
BiologicalCharacteristics
Biocompatibilityof MaterialsUsed in theConstruction ofthe CannulaMeets ISO 10993requirementsSameSame
Length of Timein Contact withthe TargetTissue< 24 hoursSameSame

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Biocompatibility testing

There have been no new materials or coatings to the Pantheris catheter since review and clearance of K173862, so no new tests were warranted to support this submission. The tests completed and reviewed in K173862 were:

    1. Sensitization,
    1. Irritation.
    1. Systemic toxicity,
    1. Materials-mediated pyrogenicity,
    1. Hemocompatibility (dog thrombogenicity),
    1. Hemocompatibility (platelet and leukocyte PLC with predicate device),
    1. Cytotoxicity,
    1. Hemocompatibility (hemolysis direct and indirect),
    1. Hemocompatibility (complement activation), and
    1. Hemocompatibility (partial thromboplastin time, human plasma).

{10}------------------------------------------------

Electrical safety and electromagnetic compatibility (EMC)

The subject and reference predicate devices comply with IEC 60601-1 standard for electrical safety and the IEC 60601-1-2 standard for electromagnetic compatibility, as reviewed and cleared in K173862.

Software Verification and Validation Testing

The catheter does not contain any software. The software of the Lightbox component of the system was reviewed and cleared in K173862. The software is considered a "moderate" level of concern, per the FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

Performance Testing - Bench

Bench testing was conducted to ensure that the Pantheris catheter performs as intended in an ISR model. Following use of the catheter in this model, the stents were evaluated under 20X magnification and there was no evidence of damage to the stents. In addition, the excised tissue was examined and no stent particulate material-was noted to be present when viewed under 20X magnification. No interaction between the Pantheris catheter and the stents during simulated treatment either a single stent or overlapping stent models, which was confirmed by examination of the stents-there was no evidence of cracks, scratches, permanent set or coating delamination.

Performance Testing - Animal

Preclinical (animal) data were collected from the use of the Pantheris catheter in a simulated ISR model in the pig. Pathologically, acute treatment with the Avinger Pantheris Atherectomy Catheter was associated with no evidence of adverse effects on either the arteries (e.g., no thrombosis or dissection) or the implanted stents (e.g., strut exposure, deformation, or discontinuities/fractures caused by interaction with the cutter), as assessed by light microscopy, radiography, and SEM.

{11}------------------------------------------------

Performance Testing - Clinical

The INSIGHT Trial was a prospective, multi-center, single-arm trial to determine whether the Pantheris System is substantially equivalent in safety and performance outcomes in the treatment of in-stent restenosis in peripheral vessels.

The primary safety endpoint was defined as freedom from a composite of major adverse events (MAEs) through 30 days after the procedure, as adjudicated by an independent Clinical Events Committee (CEC). The primary effectiveness endpoint was technical success, defined as the percent of target lesions that have a residual diameter stenosis ≤ 50% after use of the Pantheris device alone, as assessed by an independent angiographic core laboratory. The secondary safety endpoint was absence of new or worsening stent fracture following use of the Pantheris catheter. A secondary powered effectiveness endpoint was freedom from target lesion revascularization (TLR) at 6 months following the index procedure. Additional secondary effectiveness included procedural success, defined as the percent of target lesions that have residual diameter stenosis ≤ 30% post-Pantheris and any other adjunctive therapy, as determined by an independent core lab, and changes in Ankle- Brachial Index (ABI), and Rutherford Classes at 30 days and 6 months after the procedure in relation to the measurements prior to the procedure.

A total of 97 subjects were enrolled in the INSIGHT trial, with the treated population consisting of subjects presenting with documented symptomatic in-stent restenosis (stenosis >70% by visual estimation) and who met all eligibility criteria. The target in-stent restenotic lesion had to be located in vessels with diameters of > 3 mm and < 7 mm and not exceed 30 cm in length. Subjects were followed through 30 days and six months postprocedure. The clinical data for 97 subjects enrolled across 17 investigational sites in the USA (n=15) and EU (n=2) who reported for clinic visits 30 days and 85 subjects who reported for clinic visits 6 months after the index procedure were analyzed.

The primary safety endpoint was defined as freedom from a composite of major adverse events (MAEs) through 30 days after the procedure, as adjudicated by an independent Clinical Events Committee (CEC). Only 3 subjects (3%) experienced a MAE, with 97% of subjects from MAEs within 30 days. With only 3% subjects reporting an MAE, with a 95% one-sided upper confidence bound of 6.5%, the primary safety performance goal of MAEs occurring in < 20% of subjects was met.

The secondary safety endpoint was absence of new or worsening stent fracture following use of the Pantheris catheter. Only one (1) catheter inadvertently made contact with a stent during the 97 procedures, a rate of 1%. This endpoint was not established with a sample size requirement, so this performance goal was met not only due to its extremely low incidence rate but also by the experience that after re-training of the one physician who had this event on use of real-time optical coherence tomography imaging during the procedure, he completed 12 cases subsequently with no further events.

{12}------------------------------------------------

The primary effectiveness endpoint of this study was technical success, defined as the percent of target lesions that have a residual diameter stenosis ≤ 50% after use of the Pantheris device alone, as assessed by an independent angiographic core laboratory. In this analysis, 86 out of 97 (89%) subjects had <50% residual stenosis following use of the Pantheris catheter alone, with a 95% one-sided upper confidence bound of 95% and a lower confidence bound of 82%, which met the adjusted performance goal of > 79%.

A secondary powered effectiveness endpoint was freedom from target lession revascularization (TLR) at 6 months following the index procedure. The freedom from TLR of the 85 subjects that have completed their 6-month follow-up visits after the index procedure was 93% (79/85), with a 95% one-sided upper confidence bound of 98% and a lower bound of 87%, which met the performance goal of > 61%.

Additional secondary effectiveness included procedural success, defined as the percent of target lesions that have residual diameter stenosis ≤ 30% post-Pantheris and any other adjunctive therapy, as determined by an independent core lab, and changes in Ankle-Brachial Index (ABI), and Rutherford Classes at 30 days and 6 months after the procedure in relation to the measurements prior to the procedure.

Procedural success was determined if the residual diameter stenosis was ≤ 30% following adjunctive treatment. In this cohort 78 of the 97 subjects (80%) were determined to have a residual stenosis < 30% following review of angiograms by the core lab, with a mean stenosis of 15% ±10.1%.

The ABI measures improved 39% from baseline by the time of the 6-month visit and the Rutherford Classification measures improved by 71% at the same time.

Adjunctive devices used in the procedure were primarily balloons (83%), with balloon angioplasty followed by placement of a stent occurring in 13% of the cases, and no adjunctive treatment provided in 4% of the procedures.

In summary:

{13}------------------------------------------------

Substantially equivalent safety and performance outcomes in treating in-stent restenosis with the B-Laser (predicate device) and the Pantheris Catheter (study device).

Conditions and ResultsB-Laser System(K181642)Pantheris System(This Submission)
Sample size17 enrolled andcompletingfollow-up visit 6months post-procedure*97 total enrolled85 completingfollow-up visit 6months post-procedure
Length of ISR lesion(mean ± SD)7.6 ± 6.1 cm13.0 ± 9.6 cm
Percent stenosis of theISR occlusion prior totreatment (mean ± SD)85.8 ± 12.8%83.4 ± 13.8%
Percent stenosis of theISR occlusion after useof the catheter (mean ±SD)52.9 ± 11.7%39.1 ± 12.1%
Percent stenosis of theISR occlusion afteradjunctive therapy(mean ± SD)17.1 ± 9.9%15.0 ± 10.1%
Major adverse event(MAE) from procedureto 30 days post-procedure (percent andconfidence intervals)3%0% to 3.6%3%0% to 6.5%
Technical success(percent with residualstenosis < 50% with useof Pantheris catheteralone and confidenceintervals)Not reported89%82% to 95%
Freedom from targetlesion revascularizationat 6 months (percentand confidenceintervals)97%94% to 100%93%87% to 98%
ABI at baseline(mean ± SD)0.70**(SD notreported)0.69 ± 0.21
ABI at 6 months post-procedure(mean ± SD)0.90**(SD notreported)0.96 ± 0.17
Rutherford Class atbaseline (mean ± SD)2.8 ± 0.6**2.8 ± 0.84
Rutherford Class at 6months post-procedure(mean ± SD)0.7 ± 1.1**0.8 ± 1.1
Improvement ofRutherford Class at 6months post-procedurefrom baseline (percentand confidenceintervals)75%**71%
59% to 79%61% to 80%

{14}------------------------------------------------

*Data reported in Rundback ], P Chandra, M Brodmann, B Weinstock, et al. Novel laser-based catheter for peripheral atherectomy: 6-month results from the Eximo Medical B-Laser IDE study. Catheter Cardiovasc Interv 2019: 1-8 for 17 subjects treated for ISR out of 97 subjects total in this laser atherectomy trial.

** Outcomes for all 97 subjects in this laser atherectomy study; data for the 17 subject ISR cohort were not reported separately from the full study data set for these outcomes.

The results from the INSIGHT trial demonstrate that the Pantheris catheter is safe and effective when used to address in-stent restenosis. The study endpoints achieved the effectiveness performance goals while demonstrating a strong safety profile indicating that the Pantheris catheter can be used to safely excise tissue from occluded vascular stents with precision. The study results also demonstrate extremely low, acute device-related adverse events.

Conclusion

The information submitted in this premarket notification confirms that the addition of instent restenosis to the indications for use of the Pantheris System raises no new questions of safety and effectiveness and that the Pantheris catheter is substantially equivalent to the primary predicate device for treatment of in-stent restenosis.

§ 870.4875 Intraluminal artery stripper.

(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).