K Number
K212002

Validate with FDA (Live)

Date Cleared
2021-09-23

(87 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

AI/ML Overview

This document is a 510(k) Summary for Synthetic Vinyl Examination Gloves, a Class I medical device. It details the device's characteristics and its equivalence to a predicate device, supported by non-clinical testing.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

Test/CharacteristicAcceptance CriteriaReported Device Performance
Physical Dimensions (ASTM D5250)
Length (mm)≥ 230≥ 230 / Pass (for all sizes)
Width (mm)XS: 75 ± 5S: 85 ± 5M: 95 ± 5L: 105 ± 5XL: 115 ± 5XS: 78-79 / PassS: 87-88 / PassM: 95-98 / PassL: 103-107 / PassXL: 116-117 / Pass
Thickness (mm) - Finger≥ 0.080.08-0.09 / Pass
Thickness (mm) - Palm≥ 0.080.10 / Pass
Watertightness Test for Detection of Holes (ASTM D5151)Meet the requirements of ASTM D5151 AQL 2.50/125 leaks / Pass
Powder Content (ASTM D6124)Meet the requirements of ASTM D6124 < 2.0 mg0.12 mg / Pass
Physical Properties (ASTM D412)
Before Aging:
Tensile Strength≥ 11MPa13-17 / Pass
Ultimate Elongation≥ 300%370-495 / Pass
After Aging:
Tensile Strength≥ 11MPa12-16 / Pass
Ultimate Elongation≥ 300%370-445 / Pass
Cytotoxicity (ISO 10993-5)Non-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells. / Pass
Irritation (ISO 10993-10)Non-irritatingUnder the conditions of the study, not an irritant / Pass
Sensitization (ISO 10993-10)Non-sensitizingUnder conditions of the study, not a sensitizer. / Pass

2. Sample Size Used for the Test Set and Data Provenance

  • Physical Dimensions (ASTM D5250): The sample sizes are implied by the ranges provided for width and thickness (e.g., "XS: 78-79/ Pass" suggests testing across multiple samples within each size). No explicit total sample size or country of origin for the data is provided, but the manufacturer is based in China. The testing is non-clinical.
  • Watertightness Test for Detection of Holes (ASTM D5151): 125 leaks were tested from a sample size that is not explicitly stated. The method is "0/125 leaks / Pass", suggesting 125 units were tested (or 125 opportunities for leaks in a larger sample).
  • Powder Content (ASTM D6124): "Medium glove is the representative sample of the product." The exact number of medium gloves tested is not specified.
  • Physical Properties (ASTM D412): "Medium glove is the representative sample of the product." The exact number of medium gloves tested is not specified.
  • Biocompatibility (ISO 10993-5, ISO 10993-10): The studies were conducted on the subject device. No explicit sample size for these biological tests is provided, nor the country of origin of the data.

All data provenance is implied to be from non-clinical laboratory testing performed by or for Yunnan Huazhiyuan Medical Technology Co., Ltd. in China, as indicated by the submitter's information. All studies are retrospective in the sense that they are evaluations of a fully developed product.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This document describes non-clinical performance and biocompatibility testing for a medical glove. No human-derived ground truth or expert consensus was used to establish the "ground truth" for the test set. The acceptance criteria are based on established international standards (ASTM and ISO).

4. Adjudication Method for the Test Set

There was no adjudication method as this involved quantitative laboratory testing against pre-defined scientific standards rather than subjective human assessment requiring consensus.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No MRMC comparative effectiveness study was done. This is a non-clinical submission for a Class I general hospital device (examination gloves). Such studies are not applicable or required for this type of device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not Applicable. This device is a physical medical glove, not an algorithm or AI-driven system. The testing is standalone in the sense that the device's performance was evaluated intrinsically against standards, without human intervention affecting its core function during testing.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance assessment is based on established international standards and specifications:

  • ASTM D5250: Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.
  • ASTM D5151: Standard Test Method for Detection of Holes in Medical Gloves.
  • ASTM D6124: Standard Test Method for Residual Powder on Medical Gloves.
  • ASTM D412: Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension (implied basis for physical properties).
  • ISO 10993-5: Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.
  • ISO 10993-10: Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

8. The Sample Size for the Training Set

Not Applicable. This is a physical medical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not Applicable. As no training set was used, no ground truth needed to be established for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. Food & Drug Administration" in blue text.

September 23, 2021

Yunnan Huazhiyuan Medical Technology Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China

Re: K212002

Trade/Device Name: Synthetic Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: June 21, 2021 Received: June 28, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Liqun Zha

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212002

Device Name Synthetic Vinyl Examination Gloves

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K212002)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Yunnan Huazhiyuan Medical Technology Co., Ltd. Address: Yangjie Industrial Park,Jianshui County,Honghe Hani and Yi Autonomous Prefecture, Yunnan Province, 654300 China Phone Number: +86-18252909158 Fax Number: +86-873-7921166 Contact: Yun Gao Date of Preparation: Aug.31,2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Synthetic Vinyl Examination Gloves Common name: Vinyl Patient Examination Glove Classification name: Non-powdered Patient Examination Glove Model(s): XS、S、M、L、XL

3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Zibo Huiying Medical Products, Co. Ltd. Trade/Device Name: Synthetic Patient Examination Vinyl Gloves,Powder

{4}------------------------------------------------

Free,Blue

510(k) number: K153028

5.0 Indication for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

Table1-General Comparison
ItemSubject devicePredicate deviceComparison
510(k) numberK212002K153028Different
Product CodeLYZLYZSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseA patient examination gloveis a disposable deviceintended for medicalpurposes that is worn uponthe examiner's hands orfingers to preventcontamination betweenpatient and examiner.A patient examination gloveis a disposable deviceintended for medicalpurposes that is worn uponthe examiner's hands orfingers to preventcontamination betweenpatient and examiner.Same
Powdered orPowered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
LabelingInformationSingle use, powder free,device color, device name,glove size and quantity,product name. Non-SterileSingle use, powder free,device color, device name,glove size and quantity,product name. Non-SterileSimilar

Table1-General Comparison

Table2 Device Dimensions Comparison

PredicateDevice(K153028)DesignationSizeTolerance
Length, mmAverage over 234 on M size-
Width, mmAverage over 96 on M size-

{5}------------------------------------------------

Thickness, mm:
FingerAverage 0.096-
PalmAverage 0.098-
Subject Device(K212002)DesignationSizeTolerance
XSSMLXL
Length, mm230230230230230min
Width, mm758595105115± 5
Thickness, mm:
Finger0.08min
Palm0.08min
RemarkSIMILAR

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250, so the differences do not raise any new safety or performance questions.

ItemSubject devicePredicate deviceComparison
ColorantBlueBlueSame
PhysicalPropertiesBefore AgingTensile Strength11MPa, minAverage 16.9MPaDifferent
UltimateElongation300%minAverage 550%Different
After AgingTensile Strength11MPa, minAverage 14.4MPa, minDifferent
UltimateElongation300%minAverage 550%Different
Comply with ASTM D5250Comply with ASTM D5250Same
Freedom from HolesBe free from holes when tested in accordance with ASTM D5151 AQL=2.5Be free from holes when tested in accordance with ASTM D5151 AQL=2.5Same
Powder Content0.12 mg per glove. Meet the requirements of ASTM D6124Meet the requirements of ASTM D6124Similar
Table3 Performance Comparison

Analysis: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250, so the differences do not raise any new safety or performance questions.

Table4 Safety Comparison

{6}------------------------------------------------

ItemSubject devicePredicate deviceComparison
MaterialPoly Vinyl ChloridePolyurethaneNitrileDi-(2-ethylhexyl)Terephthalate(DOTP)Poly Vinyl ChloridePolyurethaneDiisononylPhthalate(DINP)Similar
BiocompatibilityIrritationUnder the conditions of the study, not an irritantComply with ISO10993-10Same
SensitizationUnder conditions of the study, not a sensitizer.
CytotoxicityUnder conditions of the study, did not show potential toxicity to L-929 cells.Not providedDifferent
Label and LabelingMeet FDA'sRequirementMeet FDA's RequirementSame

Analysis: The materials of the subject device are little different with that of the predicate, but they all meet the performance requirements of ASTM D5250, also biocompatibility test has been performed on subject device and the test result can meet the requirements of ISO 10993 standards. Therefore, the differences will not raise any safety and effectiveness issues.

8.0 Discussion of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

TestPurposeAcceptance CriteriaResults
---------------------------------------------

{7}------------------------------------------------

Methodology
ASTM D5250PhysicalDimensionsTestLength(mm): $\geq$ 230;Length: $\geq$ 230/ Pass
Width(mm):Width:
XS:75±5;XS: 78-79/ Pass
S: 85±5;S: 87-88/Pass
M: 95±5;M: 95-98/Pass
L: 105±5;L: 103-107/Pass
XL: 115±5;XL: 116-117/Pass
Thickness (mm):
Finger: $\geq$ 0.08Finger: 0.08-0.09/Pass
Palm: $\geq$ 0.08Palm: 0.10/Pass
ASTM D5151WatertightnessTest forDetection ofHolesMeet the requirements of ASTM D5151AQL 2.50/125 leaks / Pass
ASTM D6124PowderContent(Medium gloveis therepresentativesample of theproduct)Meet the requirements of ASTM D6124$<$ 2.0mg0.12 mg/Pass;
ASTM D412Physicalproperties(Medium gloveis therepresentativesample of theproduct)BeforeAgingTensileStrength$\geq$ 11MPa13-17/Pass
UltimateElongation$\geq$ 300%370-495/Pass
After AgingTensileStrength$\geq$ 11MPa12-16/Pass
UltimateElongation$\geq$ 300%370-445/Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of thestudy, did not show potentialtoxicity to L-929 cells./ Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of thestudy, not an irritant/ Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of thestudy

9.0 Clinical Test Conclusion

{8}------------------------------------------------

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device in K153028.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.