K Number
K211971

Validate with FDA (Live)

Device Name
Drug Relief v1
Manufacturer
Date Cleared
2021-11-05

(134 days)

Product Code
Regulation Number
882.5896
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Drug Relief v1 is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial nerves V, VII, IX and X, and the occipital nerves identified by transillumination.

Device Description

The Drug Relief v1 is designed to be used as an aid to reduce opioid withdrawal symptoms by the method of cranial electrical stimulation at the auricular stimulation points. The Drug Relief v1 is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over five days / 120 hours from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of cranial Nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 120 hours. There atimulation electrode and one ground electrode which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin. The stimulation pattern consists of rectangular pulses with differing inter-pulse intervals and a duty cycle of 2 hours ON/1 minute OFF. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

AI/ML Overview

The provided text describes the Drug Relief v1 device and its substantial equivalence to a predicate device, Drug Relief (K173861). The document primarily focuses on demonstrating that the new device has similar performance specifications and safety characteristics to the previously cleared predicate.

Here's an analysis of the provided information concerning acceptance criteria and studies:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of predetermined acceptance criteria with corresponding performance results. Instead, it relies on demonstrating that the "Drug Relief v1 has equivalent performance specifications when compared to the predicate device." This is a common approach for 510(k) submissions, where the new device's performance is compared to a legally marketed predicate rather than against absolute, pre-defined acceptance metrics for efficacy in a clinical setting.

However, the core of the performance evaluation for this type of device, as indicated, revolves around:

  • Electrical Safety Parameters: Max Charge Density (µC/cm²) and Max Average Power Density (W/cm²).
  • Functional Performance: Pulse width, duty cycle, amplitude, and current values over the intended operating time (120 hours).
  • Biocompatibility: Ensuring materials are safe for patient contact.
  • Software Verification and Validation: Confirming the software functions as intended.
  • Sterilization: Achieving a defined sterility assurance level (SAL).

The document states that the reported device performance for the Drug Relief v1 shows these characteristics are "identical or near identical" to the predicate device. Specifically, the electrical output parameters are presented in "Table 13.3.2 Comparison of Technical System Characteristics" (pages 6-7).

System Characteristic / ParameterPredicate Device (Drug Relief K173861) PerformanceSubject Device (Drug Relief v1) PerformanceJustification / Outcome
Max Charge Density (µC / cm²) per needle65.67 @ 1 kΩ; 7.96 @ 10 kΩ67.16 @ 1 kΩ; 8.01 @ 10 kΩNear identical; minor technological differences do not impact safety/effectiveness.
Max Average Power Density (W / cm²)0.346 @ 1 kΩ; 0.0509 @ 10 kΩ0.363 @ 1 kΩ; 0.0516 @ 10 kΩNear identical; minor technological differences do not impact safety/effectiveness.
Pulse ShapeRectangleRectangleIdentical
Frequency (HZ)1 - 10 (Pulse with modulating frequency)1 - 10 (Pulse with modulating frequency)Identical
WaveformRectangular PulseRectangle pulseIdentical
Pulse Width (ms)0.9800.980Identical
Duty Cycle2 hours ON / 1 min OFF2 hours ON / 1 min OFFIdentical
Battery Operating Time (hours)120120Identical
BiocompatibilityPassed (Implied, as K211971 passed)PassedDemonstrated via nonclinical testing.
Electrical Safety (EMC and Safety)Passed (Implied, as K211971 passed)PassedDemonstrated via nonclinical testing, shown as "near identical" to predicate.
Software Verification and ValidationPerformedPerformedDemonstrated via nonclinical testing.
Sterility Assurance Level (SAL)Achieved (Implied, as K211971 passed)AchievedEvidenced by bio-burden, EtO sterilization, and sterility tests per ISO standards.

2. Sample Size Used for the Test Set and Data Provenance

The document describes nonclinical testing, including "performance bench testing" and "electrical safety (electromagnetic compatibility and safety)" for the Drug Relief v1 device (page 7). It also mentions "software verification and validation" and "biocompatibility testing."

  • Sample Size: The specific sample sizes for these tests (e.g., how many devices were tested for electrical safety or how many batches for sterilization) are not explicitly stated in the provided text.
  • Data Provenance: The studies are described as nonclinical (bench testing) and likely conducted by the manufacturer (DyAnsys, Inc.). There is no mention of clinical data, human subjects, country of origin related to human data, or retrospective/prospective study design, as this submission is based on demonstrating substantial equivalence through technical and performance comparisons to a predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable and not provided because the evaluation presented is based on nonclinical testing (bench tests, electrical safety, etc.) and direct comparison to a predicate device's technical specifications. There is no "ground truth" derived from expert interpretation of clinical data in this document.

4. Adjudication Method for the Test Set

This is not applicable and not provided as the submission does not involve clinical studies with human readers or diagnostic interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable and not provided. The Drug Relief v1 is a percutaneous nerve stimulator, not an AI-based diagnostic imaging or analysis device. There is no mention of human readers, AI assistance, or MRMC studies.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable and not provided. The device is a nerve stimulator, not an algorithm, and the submission does not refer to AI or algorithmic performance evaluation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

As this is a nonclinical performance and electrical safety submission, the "ground truth" refers to established engineering standards, electrical measurement protocols, and material science specifications. For example:

  • Electrical Safety: Standards like those for current leakage, output parameters, and EMC.
  • Biocompatibility: ISO standards for biological evaluation of medical devices.
  • Sterilization: ISO standards for sterilization processes and sterility assurance.
  • Functional Performance: The established specifications of the predicate device (Drug Relief K173861) serve as the benchmark for "equivalent performance."

8. The Sample Size for the Training Set

This is not applicable and not provided, as there is no mention of an algorithm or AI model requiring a training set in this submission.

9. How the Ground Truth for the Training Set was Established

This is not applicable and not provided, for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 5, 2021

DyAnsys, Inc. Srini Nageshwar CEO 300, North Bayshore Boulevard San Mateo, CA 94401

Re: K211971

Trade/Device Name: Drug Relief v1 Regulation Number: 21 CFR 882.5896 Regulation Name: Percutaneous Nerve Stimulator For Substance Use Disorders Regulatory Class: Class II Product Code: PZR Dated: October 5, 2021 Received: October 6, 2021

Dear Srini Nageshwar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name Drug Relief V 1

Indications for Use (Describe)

The Drug Relief v1 is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial nerves V, VII, IX and X, and the occipital nerves identified by transillumination.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling an abstract letter 'A' or a stylized checkmark, followed by the company name "DyAnsys" in bold, black font. A registered trademark symbol is placed to the upper right of the company name.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.87 and 21 CFR 807.92.

510(k) Number: K211971

1. SUBMITTER

Name:DyAnsys, Inc
Address:300, North Bayshore Boulevard,San Mateo, CA 94401, USA
Contact Person:Srini Nageshwar
Phone Number:408.480.4700
Facsimile Number:(650)-556-1621

II. DEVICE INFORMATION

Trade Name: Drug Relief v1
Classification Name: Percutaneous nerve stimulator for substance use disorders(21 CFR 882.5896)
Classification: Class II
Product Code: PZR

III. PREDICATE DEVICE INFORMATION

Model Name: Drug Relief
Manufacturer: DyAnsys Inc.,
K Number: K173861
Classification Name: Percutaneous nerve stimulator for substance use disorder(21 CFR 882.5896)
Classification: Class II
Product Code: PZR

IV. DEVICE DESCRIPTION

The Drug Relief v1 is designed to be used as an aid to reduce opioid withdrawal symptoms by the method of cranial electrical stimulation at the auricular stimulation points. The Drug Relief v1 is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over five days / 120 hours from the time of activation of the device.

The electrical pulse from the device will be delivered to the branches of cranial Nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 120 hours. There atimulation electrode and one ground electrode which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling an abstract letter 'A' or an arrow pointing to the right. To the right of the symbol is the company name "DyAnsys" in a bold, sans-serif font, with a registered trademark symbol '®' positioned at the upper right corner of the name.

This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin.

The stimulation pattern consists of rectangular pulses with differing inter-pulse intervals and a duty cycle of 2 hours ON/1 minute OFF.

A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance

V. INDICATIONS FOR USE

The Drug Relief v1 is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination.

The Indications For Use statement of Drug Relief v1 is identical to that of Drug Relief, the cited predicate device.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The DyAnsys, Inc Drug Relief v1 device is substantially equivalent to its own legally marketed predicate device Drug Relief (K173861). It was evaluated through performance and non-clinical testing.

MODEL NAMEDrug Relief(K173861)Drug Relief v1(Subject)JUSTIFICATIONON SAFETY &EFFECTIVENESS VARIATIONS
MANUFACTURERDyAnsys IncDyAnsys IncNA
INDICATIONSFOR USEThe Drug Relief is apercutaneous nerve fieldstimulator (PNFS) system,that can be used as an aid toreduce the symptoms ofopioid withdrawal, throughapplication to branches ofCranial Nerves V, VII, IXand X, and the occipitalnerves identified bytransillumination.The Drug Relief v1 is apercutaneous nerve fieldstimulator (PNFS) system,that can be used as an aid toreduce the symptoms ofopioid withdrawal, throughapplication to branches ofCranial Nerves V, VII, IXand X, and the occipitalnerves identified bytransillumination.NA
PATIENTPOPULATIONAdultsAdultsNA
TECHNOLOGICAL CHARACTERISTICS

Table 6.1: Comparison of Technological Characteristics

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a red abstract symbol resembling an arrow or a stylized letter 'A', followed by the word 'DyAnsys' in black, sans-serif font. A registered trademark symbol is placed to the upper right of the word 'DyAnsys'.

PULSE GENERATOR
SHAPERectangleRectangleNA
PRODUCTDIMENSION(mm)$50 * 23 * 7$$38 * 21 * 10$NA
MASS (g)8 (including battery)6 (including battery)NA
PRODUCTSHAPERectangleRectangleNA
POWER
FREQUENCY(HZ)1 - 10 (Pulse withmodulating frequency)1 - 10 (Pulse withmodulating frequency)NA
WAVEFORMRectangular PulseRectangle pulseNA
ENERGYSOURCEBattery OperatedBattery OperatedNA
BATTERY TYPEP10 Zinc Air batteriesP13 Zinc Air batteriesImprovedspecification
BATTERYCAPACITY(mAh)100310validated byelectrical safetyand bench testing.
NO. * VOLTAGE$3 * 1.4V$$3 * 1.4 V$NA
PULSEWIDTH(ms)0.9800.980NA
DUTY CYCLE2 hours ON / 1 min OFF2 hours ON/ 1 min OFFNA
BATTERYOPERATINGTIME (hours)120120NA
ENVIRONMENTAL
OPERATINGTEMPERATURE5°C to 45°C5°C to 45°CNA
OPERATINGHUMIDITY(NONCONDENSING)40% to 80%40% to 80%NA
ENVIRONMENTOF USEClinics, Hospital and HomeenvironmentsClinics, Hospital and HomeenvironmentsNA
STERILIZATIONEtO SterilizationEtO SterilizationNA
RE-USESingle use DeviceSingle use DeviceNA
SHELF LIFE6 months6 monthsNA

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow or checkmark, followed by the company name "DyAnsys" in black text. A registered trademark symbol is present to the upper right of the company name.

SystemCharacteristicDrug Relief(K173861)Drug Relief v1(Subject)
Pulse Generator
Battery TypeZinc Air Batteries, P10Zinc Air Batteries, P13
SoftwareControlledYesYes
Mass (g)68
Dimensions(mm)50 * 23 * 738 * 21 * 10
Housing MaterialABS Plastic MaterialABS Plastic Material
Electrode Needle Arrays
IntroductionMethodPercutaneousPercutaneous
Needle MaterialTitaniumTitanium
Number of leads4 (3 signal and 1 ground)4 (3 signal and 1 ground)
ElectrodeConfigurationSingle needle in all the 4 leadsSingle needle in all the 4 leads
NeedleDimensions –Diameter (mm) xLength (mm)0.4 x 20.4 x 2
Surface area ofneedle (cm²)0.02010.0201
System Characteristics (Output Specs)
Max ChargeDensity( $\mu$ C / cm²)perneedle65.67 @ 1 kΩ7.96 @ 10 kΩ67.16 @ 1 kΩ8.01 @ 10 kΩ
Max AveragePower Density(W/ cm²)0.346 @ 1 kΩ0.0509 @ 10 kΩ0.363 @ 1 kΩ0.0516 @ 10 kΩ
Where and how itis usedLow levels of electrical current aredelivered by the external stimulatorthrough electrodes placedpercutaneously on the ear to targetthe associated cranial and occipitalnerves.Low levels of electrical current aredelivered by the external stimulator throughelectrodes placed percutaneously on the earto target the associated cranial and occipitalnerves.
Anodes /Cathodes pleaseidentify theThe ground electrode of Drug Reliefis a single isolated wire separatedfrom the assembly of signal wires,The ground electrode of Drug Relief vl is asingle isolated wire separated from theassembly of signal wires, which creates a
groundwhich creates a closed circuit forsafe and proper functioning of thestimulation device.closed circuit for safe and properfunctioning of the stimulation device.
Wire Assembly4 units of wire with snap-fit ringwhere 3 nos constitute a singleassembly for stimulation and theother one separate wire will act as aground electrode.4 units of wire with snap-fit ring where 3nos constitute a single assembly forstimulation and the other one separate wirewill act as a ground electrode.
VoltageMeasurementFeasibilityA 3-pin connector is provided tomeasure the output voltage of thedevice once it is activated.A 3-pin connector is provided to measurethe output voltage of the device once it isactivated.

Table 13.3.2 Comparison of Technical System Characteristics

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a red symbol resembling a stylized "greater than" sign with a curved element, followed by the text "DyAnsys" in black. A circled "R" symbol, indicating a registered trademark, is positioned to the upper right of the word "DyAnsys".

Similar to Drug Relief, the Drug Relief v1 device is intended to be a prescription (Rx) device for use by or on the order of a licensed healthcare practitioner. Both devices are body-worn, have identical indications for use and deliver electrical stimulation therapy as an aid in the reduction of opioid withdrawal symptoms.

The two minor technological difference are 1) the inclusion of an advanced micro-controller to replace the older version and 2) upgrading the battery to a P13 zinc air battery. To accommodate the changes, the design of the PCB has been modified. These differences do not impact substantial equivalence in terms of safety and effectiveness of the Drug Relief v1 as shown by the electrical safety and non-clinical performance test reports.

The technical system characteristics and essential electrical safety parameters are identical or near identical as noted in Table 13.3.2. The minor technological differences do not impact the electrical safety parameters in a significant manner. Note that electrical safety parameters includes the actual electrical output delivered by the subject and predicate devices. The actual electrical output is near identical as mentioned in the technical sections of this submission hence the subsequent effectiveness remains equivalent to that of the predicate device.

VII. PERFORMANCE DATA

The Drug Relief v1 device and its components are subjected to performance testing to validate the effectiveness of each unit. The final product testing is perfy and compare the effectual output along with that of the predicate device. The functional test is performed for 120 hours to monitor the continuous performance. The pulse width, duty cycle, amplitude and current values are captured for the Drug Relief v1 device. The Drug Relief v1 has equivalent performance specifications when compared to the predicate device.

The nonclinical testing of Drug Relief v1 device included biocompatibility testing, electrical safety (electromagnetic compatibility and safety), performance bench testing and software verification and validation.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow or a stylized letter 'A', followed by the text "DyAnsys" in black. A registered trademark symbol is present to the upper right of the text.

VIII. STERILIZATION TESTING SUMMARY

The needle package was subjected to Bio-burden test. The needle packs are exposed to EtO Sterilization to curtail the presence of microorganisms and to achieve the defined sterility assurance level (SAL). During the sterilization validation process the biological indicators are used to ensure the desired sterility assurance level. These Bls were placed at the appropriate location, where the sterilizing conditions are the most difficult to achieve. These needle packages carry a chemical indicator on the rear side which indicates the exposure to EtO. The sterility test performed on the needles indicates that there is no turbidity. The residual risk report carried out on the sterilized needle packs evidenced that the results are inline with standards requirement. All sterilization testing was performed in accordance with ISO 11135:2014-Sterilization of healthcare products- ethylene oxide, ISO 11140-1:2005/(R) 2010 Sterilization of healthcare products- chemical indicators, ISO 10993-7:2008/(R) 2012 Biological evaluation of medical devices- ethylene oxide sterilization residuals, ISO 11737-1:2006/(R) 2011 Sterilization of medical devices- Microbiological methods-Part 1: Determination of a population of microorganisms on products, ISO 11737-2: 2009 Sterilization of medical devices- Microbiological methods- Part2: Tests of Sterility performed in the definition, validation and maintenance of a Sterilization process and ISO 11138-2:2006/(R) 2010-Sterilization Of Healthcare Products-Biological Indicators- Part2: Biological Indicators for ethylene oxide Sterilization Processes.

IX. CONCLUSION

Drug Relief v1 has been shown to be substantially equivalent to the identified predicate device based on identical device classification, intended use, indications for use statement, basic operating principles and near identical electrical safety parameters. The minor technological differences do not impact the electrical safety parameters and subsequent effectiveness of the device.

Hence it is concluded that by demonstrating the performance testing and with the indications for use. environment for use, compliance with the appropriate safety standards, the product Drug Relief v1 is substantially equivalent to the predicate device that was cleared by FDA - Drug Relief (K173861).

§ 882.5896 Percutaneous nerve stimulator for substance use disorders.

(a)
Identification. A percutaneous nerve stimulator for substance use disorders is a device that stimulates nerves percutaneously to aid in the reduction of withdrawal symptoms associated with substance use disorders.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the specified shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.