K Number
K211951

Validate with FDA (Live)

Device Name
GI Genius
Date Cleared
2021-07-23

(30 days)

Product Code
Regulation Number
876.1520
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.

Device Description

GI Genius is an artificial intelligence-based device that has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps, including those with flat (non-polypoid) morphology.

GI Genius is compatible with the following Video Processors: Fujifilm VP-7000, Olympus CV-180 EXERA II, Olympus CV-190 EXERA III, Fujifilm VP-4450HD, and Pentax EPK-i7000 Video Processor.

GI Genius is connected between the video processor and the endoscopic display monitor. When first switched on, the endoscopic field of view is clearly identified by four corner markers, and a blinking green square indicator appears on the connected endoscopic display monitor to state that the system is ready to function.

During live video streaming of the endoscopic video image, GI Genius generates a video output on the endoscopic display monitor that contains the original live video together with superimposed green square markers that will appear when a polyp or other lesion of interest is detected, accompanied by a short sound. These markers will not be visible when no lesion detection occurs.

The operating principle of the subject device is identical to that of the predicate device, this being a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device includes hardware to support interfacing with video endoscopy systems.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the comparison to the predicate device (GI Genius v1.0.2). The goal of the study for GI Genius v2.0.0 was to demonstrate non-inferiority or improved performance compared to the predicate.

CharacteristicAcceptance Criteria (Predicate Device Performance)Reported Device Performance (GI Genius v2.0.0)Result (vs. Predicate)
Lesion-based sensitivity82.0 %86.5 %Improved
Frame-level True Positive228,929269,223Improved
Frame-level True Negative5,235,6825,239,128Improved
Frame-level False Positive108,115104,669Improved
Frame-level False Negative232,861192,567Improved
True Positive Rate per frame Mean49.57 %58.30 %Improved
True Positive Rate per frame % of polyps99.7 %100 %Improved
False Positive Rate per frame Mean2.02 %1.96 %Improved
Frame-Based TPr/FPr ROC curve, AUC0.7230.796Improved
False positive clusters < 500 msBaseline40 less than baselineImproved
False positive clusters > 500 msBaseline1 more than baselineSimilar
Video delay, signal in to signal out1.52 μs1.52 μsSame

2. Sample Size Used for the Test Set and Data Provenance

The non-clinical testing for standalone performance used 150 videos containing 338 polyps.

The data provenance is not explicitly stated regarding country of origin or whether it was retrospective or prospective. However, given that it's a re-training of a neural network with "additional procedure videos," it suggests a retrospective dataset of previously recorded endoscopy procedures.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the given text.

4. Adjudication Method for the Test Set

This information is not provided in the given text.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned for the GI Genius v2.0.0. The evaluation focuses on standalone performance relative to its predicate.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone performance test was done. The text states: "Standalone Performance Testing has been carried out to assess the performance of the subject device in accordance with the same test protocol as that used for the predicate device, the results of which demonstrate substantial equivalence to the predicate device."

7. The Type of Ground Truth Used

The ground truth used for performance evaluation appears to be based on the identified "polyps and adenomas" within the procedure videos. While not explicitly stated as "expert consensus" or "pathology," the context of "colonic mucosal lesions (such as polyps and adenomas)" generally implies a clinical or pathological reference standard.

8. The Sample Size for the Training Set

The document mentions that the neural network was re-trained with "additional procedure videos" and "improved data augmentation," but it does not specify the sample size of the training set (number of videos or lesions).

9. How the Ground Truth for the Training Set Was Established

The document does not explicitly state how the ground truth for the training set was established. However, given that the device aids in "detecting colonic mucosal lesions (such as polyps and adenomas)," it implies that these lesions were identified and annotated within the training videos, likely by expert endoscopists or pathologists.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 23, 2021

Cosmo Artificial Intelligence - AI Ltd % Roger Gray VP Quality and Regulatory Donawa Lifescience Consulting Srl Piazza Albania 10 Rome, 00153 Italy

Re: K211951

Trade/Device Name: GI Genius Regulation Number: 21 CFR 876.1520 Regulation Name: Gastrointestinal Lesion Software Detection System Regulatory Class: Class II Product Code: QNP Dated: June 18, 2021 Received: June 23, 2021

Dear Roger Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211951

Device Name GI Genius

Indications for Use (Describe)

The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cosmo Pharmaceuticals. The logo consists of a stylized figure in light blue with an orange circle above it, followed by the word "COSMO" in a larger, darker blue font. Below the word "COSMO" is the word "Pharmaceuticals" in a smaller, lighter font.

510(k) Summary

510(k) Reference:K211951
Device Name:GI Genius
Type of 510(k) submission:Special
Date of submission:18 June 2021
510(k) Owner and Submitter:Cosmo Artificial Intelligence - Al LtdRiverside II, Sir John Rogerson's QuayDublin D02 KV60Ireland
FDA Establishment Reg. Number:3018899987
Specification Developer:Linkverse Srlvia Ostiense 131/L00154 RomeItaly
Owner/Operator Reg. Number:3018901422
510(k) Application Correspondent:Roger GrayVP Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania 1000153 Rome, Italy
Phone:Fax:Email:+39 06 578 2665+39 06 574 3786rgray@donawa.com
FDA Product Code:QNP
FDA Regulation Number:21 CFR 876.1520
FDA Classification Name:Gastrointestinal lesion software detection system
Classification Panel:Gastroenterology and Urology
FDA Classification:Class II

Indications for Use:

The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Cosmo Pharmaceuticals. The logo features a stylized human figure in blue with an orange circle above its head. To the right of the figure is the word "COSMO" in blue, with the word "Pharmaceuticals" in a smaller font size underneath.

The indications for use statement is identical to that for the original unmodified (predicate) device, as cleared under DEN200055.

Device Description:

GI Genius is an artificial intelligence-based device that has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps, including those with flat (non-polypoid) morphology.

GI Genius is compatible with the following Video Processors: Fujifilm VP-7000, Olympus CV-180 EXERA II, Olympus CV-190 EXERA III, Fujifilm VP-4450HD, and Pentax EPK-i7000 Video Processor.

GI Genius is connected between the video processor and the endoscopic display monitor. When first switched on, the endoscopic field of view is clearly identified by four corner markers, and a blinking green square indicator appears on the connected endoscopic display monitor to state that the system is ready to function.

During live video streaming of the endoscopic video image, GI Genius generates a video output on the endoscopic display monitor that contains the original live video together with superimposed green square markers that will appear when a polyp or other lesion of interest is detected, accompanied by a short sound. These markers will not be visible when no lesion detection occurs.

The operating principle of the subject device is identical to that of the predicate device, this being a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device includes hardware to support interfacing with video endoscopy systems.

Design changes:

This Special 510(k) describes the design changes incorporated into GI Genius following its FDA clearance under DEN200055. The device software version number as cleared under DEN 2000555 was 1.0.2; the device software version that is the subject of this Special 510(k) is 2.0.0.

The design changes include a single software design change that falls within the FDA guideline for submittal of a new 510(k). A labeling change is also planned, to add non-clinical performance data of the new software version and to include one additional video processor type in the list of compatible units in the User Manual. Other minor design changes have been the subject of internal documentation and are being advised to FDA for information only.

The software design change includes an upgrade to the inference engine to optimize algorithm speed and improve energy efficiency, together with a retraining of the neural network with additional procedure videos, and improved data augmentation, to improve polyp detection capability robustness.

Non-clinical testing:

The following non-clinical verification/ validation activities have been completed:

  • Verification of the revised software at the system level. Each element of the SRS was tested and found to meet specified requirements, testing 14 Units and 9 Items, encompassing 133 software requirements.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Cosmo Pharmaceuticals. The logo consists of a stylized human figure in light blue with an orange circle above its head, followed by the word "COSMO" in teal. Below the word "COSMO" is the word "Pharmaceuticals" in a smaller, lighter font.

  • · Validation of the revised software at the user level, testing 14 Units and 9 Items, encompassing 133 software requirements.
  • · Protective measures identified by risk management have been verified during Installation Qualification and Operational Qualification.
  • Electromagnetic Compatibility (EMC) and Electrical Safety compliance tests have been completed according to IEC 60601-1 and IEC 60601-1-2 requirements.
  • Standalone Performance Testing has been carried out to assess the performance of the subject device in accordance with the same test protocol as that used for the predicate device, the results of which demonstrate substantial equivalence to the predicate device.
  • Non-inferiority of performance of GI Genius with a new video processor has been established by means of a per-frame assessment on 40 pre-recorded procedures.

The results of the above testing aid in demonstration of substantial equivalence of the subject device with the predicate device, as the same test protocols have been used.

Substantial equivalence

The predicate device for the subject device is the pre-modification version of the same device, Gl Genius, FDA-cleared under DEN200055 on 9 April 2021:

Predicate Device:GI Genius
Sponsor:Cosmo Artificial Intelligence - Al Ltd
De Novo Number:DEN200055
Clearance Date:9 April 2021
FDA Product Code:QNP
Classification Name:Gastrointestinal lesion software detection syster
Regulation No:21 CFR 876.1520
Class:II

Predicate device comparison table:

Table 1 provides evidence of substantial equivalence of the subject device with the predicate device.

Table 1: Predicate device comparison table
CharacteristicSubject devicePredicate deviceComparison
Device nameGI Genius v.2.0.0GI Genus v.1.0.2N/A
ManufacturerLinkverse Srl, ItalyLinkverse Srl, ItalySame
FDA clearanceK211951DEN200055N/A
FDA Reg nameGastrointestinal lesion softwaredetection systemGastrointestinal lesion softwaredetection systemSame
FDA Reg #21 CFR 876.152021 CFR 876.1520Same
FDA Product CodeQNPQNPSame
Table 1: Predicate device comparison table
CharacteristicSubject devicePredicate deviceComparison
Indications for UseThe GI Genius System is acomputer-assisted reading tooldesigned to aid endoscopists indetecting colonic mucosallesions (such as polyps andadenomas) in real time duringstandard white-light endoscopyexaminations of patientsundergoing screening andsurveillance endoscopic mucosalevaluations. The GI Geniuscomputer-assisted detectiondevice is limited for use withstandard white-light endoscopyimaging only. This device is notintended to replace clinicaldecision making.The GI Genius System is acomputer-assisted reading tooldesigned to aid endoscopists indetecting colonic mucosallesions (such as polyps andadenomas) in real time duringstandard white-light endoscopyexaminations of patientsundergoing screening andsurveillance endoscopic mucosalevaluations. The Gl Geniuscomputer-assisted detectiondevice is limited for use withstandard white-light endoscopyimaging only. This device is notintended to replace clinicaldecision making.Same
Video delay, signal into signal out1.52 μs1.52 μsSame
Lesion-basedsensitivity86.5 %82.0 %Improvedperformance
Frame levelperformance(150 videos / 338polyps)True positive: 269,223True negative: 5,239,128False positive: 104,669False negative: 192,567True positive: 228,929True negative: 5,235,682False positive: 108,115False negative: 232,861Improvedperformance
True positive rate perframeMean: 58.30 %% of polyps: 100 %Mean: 49.57 %% of polyps: 99.7 %Improvedperformance
False positive rate perframeMean: 1.96 %Mean: 2.02 %Improvedperformance
Frame-Based TPr/FPrROC curve, AOC0.7960.723Improvedperformance
False positive clustersper patient< 500 ms: 40 less than baseline> 500 ms: 1 more than baseline< 500 ms: Baseline> 500 ms: BaselineImprovedSimilar
Additional videoprocessorYes: Fujifilm VP-7000N/AImprovedperformance
Electrical safetyIEC 60601-1IEC 60601-1Same
ElectromagneticcompatibilityIEC 60601-1-2IEC 60601-1-2Same
LAN portYes, non-functional to userNoDifferent

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Cosmo Pharmaceuticals. The logo consists of a stylized human figure in light blue with an orange circle for the head, followed by the word "COSMO" in a larger, teal font. Below "COSMO" is the word "Pharmaceuticals" in a smaller, lighter font.

The subject device and the predicate device have many identical or similar characteristics. None of the identified differences introduce new aspects of safety or effectiveness.

Conclusion

The subject and predicate devices have identical indications for use and fundamental technological characteristics. Any differences in performance between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device, which is already in interstate commerce within the USA. Therefore, the subject device is as safe, as effective, and performs better than the predicate device.

§ 876.1520 Gastrointestinal lesion software detection system.

(a)
Identification. A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.