AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To provide percutaneous drainage of abscess fluid, biliary, nephrostomy, urinary, pleural empyemas, lung abscesses, and mediastinal collection.

The vanSonnenberg™ Sump Catheter is designed for use as a percutaneously placed drain for intra-abdominal fluid collections. Such conditions include abscesses, cysts, pseudocysts, bilomas, seromas, urinomas, or any loculated fluid collection.

Device Description

All Purpose Drainage catheters are designed to provide constant, unobstructed fluid drainage from various fluid accumulations to an external and/or internal collection sites and to ensure stability during treatment. These ends are met through the choice of drainage material and by incorporating a secure fixation method in the drainage site (internally by pig-tail or "J" style tips and externally, by means of suture or adhesive). Percutaneous placement is accomplished with radiological guidance.

Select Boston Scientific All Purpose Drainage catheters are coated with a hydrophilic material, Glidex™ , for the reduction of surface friction during placement.

Some of these catheters are also available with a biocompatible, distal tip (Temp-Tip™). The tip's internal diameter allows the guidewire to direct the tip for accurate catheter placement; however, soon after placement, the Temp-Tip™ material dissolves, allowing the full diameter of the catheter lumen to contribute to site drainage.

All Purpose Drainage catheters are manufactured with one of two types of resins, Flexima or Percuflex.

  • Flexima is a biocompatible polyurethane (Tecoflex) known for its flexibility, strength, and kink resistance. Flexima comes in durometers ranging in Firm, Regular and Soft and uses a radiopacifier for visualization under fluoroscopy.
  • Percuflex is an ELVAX™ 460, ethylene vinyl acetate (EVA) copolymer for patient comfort. This material is available in firm (white in color) and regular (blue in color) durometers. The firmer durometers provide more pushability during device placement. A radiopaque additive, bismuth subcarbonate, incorporated throughout the catheter tube also aids radiopacity and percutaneous placement.
AI/ML Overview

The provided text is a 510(k) summary for the Boston Scientific APDL Drainage Catheter System and related products. It outlines the regulatory clearance process for drainage catheters and accessories. However, this document does not contain information about an AI/ML-driven device or a study assessing its performance using acceptance criteria related to AI/ML metrics (e.g., sensitivity, specificity, AUC).

Instead, the document details a traditional medical device submission, focusing on establishing substantial equivalence to existing predicate devices. The "performance testing" described refers to:

  • Engineering and materials testing to support the addition of Magnetic Resonance (MR) compatibility information and to justify cumulative changes in manufacturing materials (e.g., resin types, coatings). This includes tests for radio frequency heating, force measurement, and image artifact in an MR environment, as well as packaging integrity and aging tests.
  • Biological Endotoxin (BET) assessment.

Therefore, I cannot provide the requested information regarding acceptance criteria and studies for an AI device. The document does not describe such a device or study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 10, 2022

Boston Scientific Corporation Liz Johnston Regulatory Affairs Specialist One Scimed Place Maple Grove, MN 55311

Re: K211934

Trade/Device Name: APDL Drainage Catheter System, Flexima™ APDL Drainage Catheter System Kit, Flexima™ APDL Drainage Catheter System Kit with Dissolving Tip, Flexima™ APDL Drainage Catheter System with Dissolving Tip, Flexima™ QuickStickTM Drainage Catheter System, vanSonnenberg™ Sump Catheter System Kit, vanSonnenberg™ Sump Catheter System, vanSonnenberg™ Drainage Catheter System, Flexima™ APD™ Drainage Catheter System, Flexima™ APDTM Drainage Catheter System Kit, Flexima™ APD™ Drainage Catheter System with Dissolving Tip, Flexima™ APD™ Drainage Catheter System Kit with Dissolving Tip

Regulation Number: 21 CFR§ 876.5090 Regulation Name: Suprapubic urological catheter and accessories Regulatory Class: II Product Code: FFA Dated: October 13, 2022 Received: October 14, 2022

Dear Liz Johnston:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding

{1}------------------------------------------------

and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen -S

Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

510(k) Number (if known) K211934

Device Name

APDL Drainage Catheter System, Flexima™ APDL Drainage Catheter System Kit, Flexima™ APDL Drainage Catheter System Kit with Dissolving Tip, Flexima™ APDL Drainage Catheter System with Dissolving Tip

Indications for Use (Describe)

To provide percutaneous drainage of abscess fluid, biliary, nephrostomy, urinary, pleural empyemas, lung abscesses, and mediastinal collection.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known) K211934

Device Name

Flexima™ QuickStick™ Drainage Catheter System

Indications for Use (Describe)

To provide percutaneous drainage of abscess fluid, biliary, nephrostomy, urinary, pleural empyemas, lung abscesses, and mediastinal collection.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known) K211934

Device Name

vanSonnenberg™ Sump Catheter System Kit, vanSonnenberg™ Sump Catheter System

Indications for Use (Describe)

The vanSonnenberg™ Sump Catheter is designed for use as a percutaneously placed drain for intra-abdominal fluid collections. Such conditions include abscesses, cysts, pseudocysts, bilomas, seromas, urinomas, or any loculated fluid collection.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known) K211934

Device Name

vanSonnenberg™ Drainage Catheter System

Indications for Use (Describe)

To provide percutaneous drainage of abscess fluid, biliary, nephrostomy, urinary, pleural empyemas, lung abscesses, and mediastinal collection.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known) K211934

Device Name

Flexima™ APD™ Drainage Catheter System, Flexima™ APD™ Drainage Catheter System Kit, Flexima™ APD™ Drainage Catheter System with Dissolving Tip, Flexima™ APD™ Drainage Catheter System Kit with Dissolving Tip

Indications for Use (Describe)

To provide percutaneous drainage of abscess fluid, biliary, nephrostomy, urinary, pleural empyemas, and mediastinal collection.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{7}------------------------------------------------

510(k) Summary

510(K) Summary Complying with 21 CFR 807.92

I. SUBMITTER INFORMATION

Submitter Name: Boston Scientific Corporation

Submitter Address:

Three Scimed Place Maple Grove, MN 55311-1566 USA

Telephone: 763-494-1676 Fax: 763-494-2222 e-mail: liz.johnston@bsci.com

Contact person name: Liz Johnston Date Prepared: November 10, 2022

II. DEVICE INFORMATION

Tables 1 to 4 below summarize the relevant subject device information.

UPNProduct Description
M001271240Flexima™ APDL Drainage Catheter System
M001271310Flexima™ APDL Drainage Catheter System
M001271320Flexima™ APDL Drainage Catheter System
M001275080Flexima™ APDL Drainage Catheter System with Dissolving Tip
M001271330Flexima™ APDL Drainage Catheter System
M001271340Flexima™ APDL Drainage Catheter System
M001271350Flexima™ APDL Drainage Catheter System
M001271380Flexima™ APDL Drainage Catheter System
M001271390Flexima™ APDL Drainage Catheter System
M001271960Flexima™ APDL Drainage Catheter System Kit
M001271970Flexima™ APDL Drainage Catheter System Kit
M001271980Flexima™ APDL Drainage Catheter System Kit
M001271990Flexima™ APDL Drainage Catheter System Kit
M001275010Flexima™ APDL Drainage Catheter System
M001275030Flexima™ APDL Drainage Catheter System Kit
M001275090Flexima™ APDL Drainage Catheter System Kit with Dissolving Tip
M001281340Flexima™ APDL Drainage Catheter System with Dissolving Tip
M001281350Flexima™ APDL Drainage Catheter System with Dissolving Tip
M001281380Flexima™ APDL Drainage Catheter System with Dissolving Tip
M001281390Flexima™ APDL Drainage Catheter System with Dissolving Tip
M001245061APDL Drainage Catheter System
M001245071APDL Drainage Catheter System
M001245081APDL Drainage Catheter System
M001245060APDL Drainage Catheter System
M001245070APDL Drainage Catheter System

Table 1: APDL Drainage Catheters

{8}------------------------------------------------

Table 2: APD Drainage Catheters

UPNProduct Description
M001271250Flexima™ APDTM™ Drainage Catheter System
M001271260Flexima™ APDTM™ Drainage Catheter System
M001271270Flexima™ APDTM™ Drainage Catheter System
M001275060Flexima™ APDTM™ Drainage Catheter System with Dissolving Tip
M001271280Flexima™ APDTM™ Drainage Catheter System
M001271290Flexima™ APDTM™ Drainage Catheter System
M001271300Flexima™ APDTM™ Drainage Catheter System
M001271360Flexima™ APD™ Drainage Catheter System
M001271370Flexima™ APD™ Drainage Catheter System
M001275070Flexima™ APDTM™ Drainage Catheter System Kit with Dissolving Tip
M001281290Flexima™ APDTM™ Drainage Catheter System with Dissolving Tip
M001281300Flexima™ APDTM™ Drainage Catheter System with Dissolving Tip
M001281360Flexima™ APDTM™ Drainage Catheter System with Dissolving Tip
M001281370Flexima™ APDTM™ Drainage Catheter System with Dissolving Tip

Table 3: Flexima Quickstick Drainage Catheters

UPNProduct Description
M001271460FleximaTM QuickStickTM Drainage Catheter System
M001271470FleximaTM QuickStickTM Drainage Catheter System

Table4: vanSonnenberg Drainage Catheters

UPNProduct Description
M001245130vanSonnenberg™ Drainage Catheter System
M001245140vanSonnenberg™ Drainage Catheter System
UPNProduct Description
M001201000vanSonnenberg™ Sump Catheter System Kit
M001201050vanSonnenberg™ Sump Catheter System Kit
M001201060vanSonnenberg™ Sump Catheter System Kit
M001203000vanSonnenberg™ Sump Catheter System
M001203010vanSonnenberg™ Sump Catheter System
M001203020vanSonnenberg™ Sump Catheter System
M001203030vanSonnenberg™ Sump Catheter System
M001203070vanSonnenberg™ Sump Catheter System
M001213060vanSonnenberg™ Sump Catheter System with Dissolving Tip
M001213070vanSonnenberg™ Sump Catheter System with Dissolving Tip
M001273000vanSonnenberg™ Sump Catheter System
M001273010vanSonnenberg™ Sump Catheter System
M001273030vanSonnenberg™ Sump Catheter System
M001273040vanSonnenberg™ Sump Catheter System
M001273060vanSonnenberg™ Sump Catheter System Kit
M001273070vanSonnenberg™ Sump Catheter System Kit
M001273090vanSonnenberg™ Sump Catheter System Kit
M001273100vanSonnenberg™ Sump Catheter System Kit
M001273130vanSonnenberg™ Sump Catheter System
M001273180vanSonnenberg™ Sump Catheter System Kit
M001273190vanSonnenberg™ Sump Catheter System Kit
M001273210vanSonnenberg™ Sump Catheter System Kit

{9}------------------------------------------------

Common orUsual nameRegulationNumberRegulationNameProduct CodeProduct CodeNameRegulatoryClass
DrainageCatheter21 CFR Part876.5090Suprapubicurologicalcatheter andaccessoriesFFATube, Drainage,SuprapubicII

5: Additional Device Information

III. PREDICATE DEVICE IDENTIFICATION

Name of Predicate Device

K924608 Hydrogel Coated Percuflex Drainage Catheters K944290 Modified (Flexima) Hydrogel Coated Drainage Catheters

For devices that are manufactured with Flexima resin (Tecoflex), K944290 is the predicate. For devices that are manufactured from Percuflex resin. K924608 is the appropriate predicate.

These devices are Class II 21 CFR Part 876.4620, FAD (K924608), and Part 876.5090, FFA (K944290) with subsequent product codes of FAD, FGE, LJE. Predicate devices referenced above have not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

All Purpose Drainage catheters are designed to provide constant, unobstructed fluid drainage from various fluid accumulations to an external and/or internal collection sites and to ensure stability during treatment. These ends are met through the choice of drainage material and by incorporating a secure fixation method in the drainage site (internally by pig-tail or "J" style tips and externally, by means of suture or adhesive). Percutaneous placement is accomplished with radiological guidance.

The following All Purpose Drainage catheters are in scope of this submission:

  • . APDL Drainage Catheter System
  • . Flexima™ APDL Drainage Catheter System Kit
  • . Flexima™ APDL Drainage Catheter System Kit with Dissolving Tip
  • . Flexima™ APDL Drainage Catheter System with Dissolving Tip
  • Flexima™ APD™ Drainage Catheter System .
  • Flexima™ APD™ Drainage Catheter System Kit .
  • . Flexima™ APD™ Drainage Catheter System with Dissolving Tip
  • . Flexima™ APD™ Drainage Catheter System Kit with Dissolving Tip
  • . Flexima™ QuickStick™ Drainage Catheter System
  • . vanSonnenberg™ Drainage Catheter System
  • vanSonnenberg™ Sump Catheter System Kit .
  • . vanSonnenberg™ Sump Catheter System

Select Boston Scientific All Purpose Drainage catheters are coated with a hydrophilic material, Glidex™ , for the reduction of surface friction during placement.

Some of these catheters are also available with a biocompatible, distal tip (Temp-Tip™). The tip's internal diameter allows the guidewire to direct the tip for accurate catheter placement; however, soon after placement, the Temp-Tip™ material dissolves, allowing the full diameter of the catheter lumen to contribute to site drainage.

{10}------------------------------------------------

All Purpose Drainage catheters are manufactured with one of two types of resins, Flexima or Percuflex.

  • Flexima is a biocompatible polyurethane (Tecoflex) known for its flexibility, strength, and . kink resistance. Flexima comes in durometers ranging in Firm, Regular and Soft and uses a radiopacifier for visualization under fluoroscopy.
  • . Percuflex is an ELVAX™ 460, ethylene vinyl acetate (EVA) copolymer for patient comfort. This material is available in firm (white in color) and regular (blue in color) durometers. The firmer durometers provide more pushability during device placement. A radiopaque additive, bismuth subcarbonate, incorporated throughout the catheter tube also aids radiopacity and percutaneous placement.

Intended Use/Indications for Use: to provide percutaneous drainage of abscess fluid, biliary, nephrostomy, urinary, pleural empyemas, lung abscesses, and mediastinal collection.

V. INDICATIONS FOR USE

Predicate and subject device Intended use and Indications for Use are the same.

Drainage DeviceIndications for Use
APDL Drainage Catheter SystemFlexima™ APDL Drainage Catheter System KitFlexima™ APDL Drainage Catheter System Kit with Dissolving TipFlexima™ APDL Drainage Catheter System with Dissolving TipFlexima™ APD™ Drainage Catheter SystemFlexima™ APD™ Drainage Catheter System KitFlexima™ APD™ Drainage Catheter System with Dissolving TipFlexima™ APD™ Drainage Catheter System Kit with Dissolving TipFlexima™ QuickStick™ Drainage Catheter SystemvanSonnenberg™ Drainage Catheter SystemTo provide percutaneousdrainage of abscess fluid,biliary, nephrostomy, urinary,pleural empyemas, lungabscesses, and mediastinalcollection.
vanSonnenberg Sump Catheter SystemvanSonnenberg Sump Catheter System KitFor use as a percutaneouslyplaced drain for intra-abdominal fluid collections.Such conditions includeabscesses, cysts,pseudocysts, bilomas,hematomas, seromas,urinomas, or any loculatedfluid collection.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The All Purpose Drainage Catheters incorporate substantially equivalent design, function, packaging, sterilization process, materials, fundamental technology, indications for use, and operating principles as those featured in the predicate drainage devices (K924608, K944290). The difference from the legally marketed predicate device is the addition of Magnetic Resonance (MR) compatibility information to the Directions for Use.

In addition, cumulative changes were implemented since the time the predicate (K924608 and K944290) 510(k)s were cleared such as a coating surfactant change, supplier color additive compound change, and resin additive supplier change.

{11}------------------------------------------------

VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING

In order to support the addition of Magnetic Resonance Compatibility information to the Directions for Use the following nonclinical performance testing was performed:

  • Radio Frequency Heating
  • . Force Measurement
  • Image Artifact

FDA quidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," issued May 20, 2021, and the following standards were used to guide the generation of the non-clinical data:

Document NameDocument NumberDocument Version
Measurement of Radio Frequency Induced HeatingNear Passive Implants During Magnetic ResonanceImagingASTM F2182-11a
Measurement of Magnetically Induced DisplacementForce on Medical Devices in the Magnetic ResonanceEnvironmentASTM F2052-14
Evaluation of MR Image Artifacts from Passive ImplantsASTM F2119-07

In addition, the following tests were performed to support substantial equivalence to the predicate devices:

  • Simulated climatic conditioning and distribution followed by packaging integrity testing ●
  • Accelerated and real time aging followed by packaging integrity testing .
  • Performance testing after aging, shipping and distribution in scope of K924608 and ● K944290
  • BET assessment as recommended in AAMI ST72:2019, Bacterial endotoxins Test methods, . routine monitoring, and alternatives to batch testing

Cumulative changes made to the devices since the clearance of the predicate (K924608 and K944290) were supported by leveraging prior testing and providing justification to support the changes.

VIII. CONCLUSION

Based on the intended use, technological characteristics, and non-clinical performance data and justifications provided, the All Purpose Drainage Catheters are substantially equivalent to the predicate devices (K924608, K944290). The Directions for Use update with magnetic resonance compatibility information for the subject devices and cumulative changes made to the device do not raise new questions of safety or effectiveness, and the subject devices are as safe and effective as the predicate devices.

§ 876.4620 Ureteral stent.

(a)
Identification. A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.(b)
Classification. Class II (performance standards).