K Number
K211909

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-06-09

(353 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cysto-Gastro Sets are intended to be used to electrosurgically cannulate pancreatic pseudo-cysts endoscopically (via the transgastric or transduodenal wall) as an alternative to surgical or percutaneous treatment.

Device Description

Each device is made of an outer flexible Teflon tube equipped with a metal tip at the distal end and an HF (high frequency) connector at the proximal end. Some models included an inner Teflon sheet that slides inside the outer sheet and are equipped with an HF needle at the distal tip and an HF connector at the proximal end.

All devices of this family group are packed in a medical grade paper pouch and sterilized by ETO. The sterile devices are distributed in carton boxes along with an Instruction for Use. There are 5 variations of the Cysto-Gastro Sets with different catheter length and diameter, with or without HF needle. However, all variations are delivered sterile and intended for single use.

AI/ML Overview

The provided text details the 510(k) submission for the Cysto-Gastro Sets device. However, it does not describe acceptance criteria or a study that proves the device meets those criteria in the context of an AI/algorithm-driven device for which the prompt's questions (e.g., number of experts, MRMC studies, training set details) are relevant.

The Cysto-Gastro Sets device is an endoscopic electrosurgical accessory, a physical medical device. The studies described are for non-clinical performance data (bench tests, biocompatibility, sterilization, shelf-life) to demonstrate substantial equivalence to a predicate device, not for evaluating the performance of an AI/algorithm.

Therefore, I cannot provide answers to the requested information as the provided document pertains to a physical device and not an AI/algorithm-driven one. There is no mention of acceptance criteria related to AI performance, ground truth, expert adjudication, or MRMC studies.

In summary, the provided document does not contain the type of AI/algorithm performance data requested.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

June 9, 2022

G-Flex Europe SPRL Thierry Cremer OA & RA Manager 20, Rue de l'industrie Nivelles, Brabant Wallon 1400 BELGIUM

Re: K211909

Trade/Device Name: Cysto-Gastro-Sets Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KNS Dated: May 2, 2022 Received: May 4, 2022

Dear Thierry Cremer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Je Hi An, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211909

Device Name Cysto-Gastro Sets

Indications for Use (Describe)

Cysto-Gastro Sets are intended to be used to electrosurgically cannulate pancreatic pseudo-cysts endoscopically (via the transgastric or transduodenal wall) as an alternative to surgical or percutaneous treatment.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

1. SUBMITTER

Submitter Name:G-Flex Europe SPRL
Submitter Address:20, Rue de l'industrie1400 NivellesBelgium
Phone Number:+32 67 88 36 65
Fax Number:+32 67 88 36 88
Contact Person:Thierry CREMER, QA & RA Manager
Email:thierry@g-flex.com
Date Prepared:June 3, 2021

2. DEVICE

Device Trade Name:Cysto-Gastro Sets
Common Name:Endoscopic electrosurgery device
Classification Name:Endoscopic electrosurgical unit and accessories
Regulation Number:876.4300
Product Code:KNS
Class:2
Classification Panel:Gastroenterology/Urology

3. PREDICATE DEVICE

Primary Predicate Device: Wilson-Cook Cystotome (K022595).

4. DEVICE DESCRIPTION

Cysto-Gastro Sets are intended to be used to electrosurgically cannulate pancreatic pseudocysts endoscopically (via the transgastric or transduodenal wall) as an alternative to surgical or percutaneous treatment.

Each device is made of an outer flexible Teflon tube equipped with a metal tip at the distal end and an HF (high frequency) connector at the proximal end. Some models included an inner Teflon sheet that slides inside the outer sheet and are equipped with an HF needle at the distal tip and an HF connector at the proximal end.

All devices of this family group are packed in a medical grade paper pouch and sterilized by ETO. The sterile devices are distributed in carton boxes along with an Instruction for Use. There are 5 variations of the Cysto-Gastro Sets with different catheter length and diameter, with or without HF needle. However, all variations are delivered sterile and intended for single use.

Variety of Cysto-Gastro Sets are described in the table below. The main difference between models is the presence or not of the HF needle, which gives the Cysto-Gastro Sets the possibility to perforate the tissue with the needle before enlarging the hole with the diathermic bigger tip. The other variation is the diameter which will define the final size of the perforation and ultimately the size of the stent that will be used in the next procedure. All devices of this family group are sterile by ETO and intended for single use:

ReferenceDimensions
Length (cm)Diameter (Fr)Details
CYSTO06U1806Without HF Needle Single Use

{4}------------------------------------------------

CYSTO08UK2108.5With HF NeedleSingle Use
CYSTO10U18010Without HF NeedleSingle Use
CYSTO10UK21010With HF NeedleSingle Use
CYSTO85U1808.5Without HF NeedleSingle Use

5. INDICATIONS FOR USE

Cysto-Gastro Sets are intended to be used to electrosurgically cannulate pancreatic pseudocysts endoscopically (via the transqastric or transduodenal wall) as an alternative to surgical or percutaneous treatment.

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

The comparison chart below provides evidence to facilitate the substantial equivalence determination between the Cysto-Gastro Sets to the predicate device (K022595) with respect to intended use, technological characteristics and principles of operation.

FeatureProposed DevicePrimary Predicate DeviceAssessment ofEquivalence
Device nameCysto-Gastro SetsCystotomeNA
510(k) NumberK211909K022595NA
ManufacturerG-Flex Europe SPRLWilson-Cook Medical IncNA
RegulationNumber876.4300876.4300Fully similar
DeviceClassificationNameEndoscopic electrosurgicalunit and accessoriesEndoscopic electrosurgicalunit and accessoriesFully similar
Product CodeKNSKNSFully similar
ClinicalConditionFor electrosurgicalpuncture of the transgastricor transduodenalwall and into pancreaticpseudocystsFor electrosurgicalpuncture of thetransgastric ortransduodenalwall and into pancreaticpseudocystsFully similar
Intended Use/Indications foruseCysto-Gastro Sets areintended to be used toelectrosurgically cannulatepancreatic pseudocystsendoscopically (via thetransgastric ortransduodenal wall) as analternative to surgical orpercutaneous treatment.This device is designed toelectrosurgicallypuncture a hole in thetransgastric ortransduodenal wall andinto a pancreaticpseudocyst, when it isvisibly bulging into thegastrointestinal tract.Fully similar
Contra-indicationsContraindications includeblood coagulationdiseases, interposingvessels, nickel sensitivity,and contraindicationsrelated to the use ofelectrical scalpel. Cysto-Gastro Sets are notrecommended to be usedon pancreatic pseudocyststhat are less than 4cm indiameter similar.Contraindications includethose specific to bloodcoagulation disease,interposing vesselsbetween the pseudocystwall and that of thestomach or the duodenum.If the pseudocyst is <4 cmin diameter do notproceed.Similar, the safety andeffectiveness are notimpacted.Nickel sensitivity isincluded as a warningin the predicatedevice.
Site of useEndoscopic pancreaticpseudocystEndoscopic pancreaticpseudocystFully similar

{5}------------------------------------------------

FeatureProposed DevicePrimary Predicate DeviceAssessment ofEquivalence
IntendedpatientpopulationPatients who are subject toendoscopic puncture ofpancreatic pseudocystPatients who are subject toendoscopic puncture ofpancreatic pseudocystFully similar
PerformancesEffective puncture of thepseudocystEffective puncture of thepseudocystFully similar
DesignDiameter: 6, 8,5 or 10 FrLength: 180 or 210 cmExternal form: diathermicring + needle (dependingon the model) + catheter +handle with electrodeNeedle design: NeedleknifeDiameter: 10 FrLength: 190 cmExternal form: diathermicring + needle + catheter +electrodeNeedle design: NeedleknifeSimilar, the variations(diameter, length,absence of the needledepending on themodel) do not affectthe safety oreffectiveness of theproposed device.
Conditions ofuseSingle UseSingle UseFully similar
SterilizationmodeETOETOFully similar
Operation andclinicalperformanceIntroduction of the devicein the endoscope andpuncture of the pseudocystfollowing endoscopicprocedure.Performance: Effectivepuncture of the pseudocystIntroduction of the devicein the endoscope andpuncture of the pseudocystfollowing endoscopicprocedure.Performance: Effectivepuncture of the pseudocystFully similar
Recommendedsize guide0,035 inch0,035 inchFully similar
MaterialsStainless Steel (needle) /teflon (catheter) / POM +MABS (handle)Needle: stainless steelFully similar
Tissues orbody fluidsin contactTissues near the cyst incontact with the needleTissues near the cyst incontactwith the needleFully similar
Duration ofcontactLimited exposure (<24h)Limited exposure (<24h)Fully similar
Type ofbody-deviceinteractionInvasiveInvasiveFully similar

The Cysto-Gastro Sets is comparable to predicate device with similar technological characteristics and intended use, specifically to perform electrosurgical procedures through an endoscope. The Cysto-Gastro Sets thus meets the requirements for 510(k) substantial equivalence.

As indicated in the table above, several differences were identified between the Cysto-Gastro Sets and the primary predicate, namely the contraindications, diameter, length and absence of the needle depending on the model. Performance testing was conducted to demonstrate substantial equivalence of the Cysto-Gastro Sets to the predicate device. The test results are summarized below.

7. NON-CLINICAL PERFORMANCE DATA

The Cysto-Gastro Sets was subjected to bench tests including visual Bench testing inspection, dimensions verification, simulated use, leaks, flow rate, corrosion resistance, X-ray visibility and mechanical. Results demonstrate that the device meets the design specifications. Additionally, the Cysto-Gastro Sets meets the design specifications and the requirements of the relevant

{6}------------------------------------------------

standards for safety and performance of HF electrosurgical equipment (IEC60601-2-2).
BiocompatibilityThe Cysto-Gastro Sets was the subject of a range of biocompatibility tests inaccordance with ISO 10993 series. Test results confirmed that the Cysto-Gastro Sets is biocompatible for the stated intended use.
SterilizationThe Cysto-Gastro Sets is provided sterile and is intended for single patientuse only. The Cysto-Gastro Sets is sterilized by ETO to meet a minimumsterility assurance level (SAL) of 10⁻⁶.
Shelf-lifeShelf-life study supports a shelf life of 3 years for The Cysto-Gastro Sets whenstored under the recommended environmental conditions. The shelf-lifestudies confirmed that the packaging maintains the integrity of the device andits sterility throughout the shelf life of the device.

Studies for sterilization, packaging and shelf-life conform to the following standard:

Standard referenceStandard title
ISO 11607-1:2019Packaging for terminally sterilized medical devices - Part 1:Requirements for materials, sterile barrier systems andpackaging systems
ISO 11607-2:2019Packaging for terminally sterilized medical devices - Part 2:Validation requirements for forming, sealing and assemblyprocesses
ISO 11135-1:2014Sterilization of healthcare products - Ethylene oxide - Part 1: Requirements fordevelopment, validation and routine control of a sterilizationprocess for medical devices
ANSI AAMI ISO 11737-1:201814-514Sterilization Of Health Care Products - MicrobiologicalMethods -- Part 1: Determination Of A Population OfMicroorganisms On Products
ASTM D7386-16Standard Practice for Performance Testing of Packages forSingle Parcel Delivery Systems
ASTM F1929-15Standard Test Method for Detecting Seal Leaks in PorousMedical Packaging by Dye Penetration

8. CLINICAL PERFORMANCE DATA

No clinical testing was performed.

9. CONCLUSION

The information discussed above and provided in this 510(k) submission demonstrates that the Cysto-Gastro Sets device is substantially equivalent to the predicate.

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).