K Number
K211905

Validate with FDA (Live)

Device Name
UltraCal XS
Date Cleared
2021-10-13

(114 days)

Product Code
Regulation Number
872.3820
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

UltraCal XS calcium hydroxide paste is indicated to be used for:

  • temporary dressing for root canals
  • apexification and/or perforation
  • vital pulpotomy
  • direct pulp capping
  • indirect pulp capping
  • root resorption
  • root canal filling for primary teeth
Device Description

UltraCal XS calcium hydroxide paste is an aqueous, 12.0 to 12.5 pH, radiopaque, premixed paste for direct application.

AI/ML Overview

This document is a 510(k) submission for a medical device called UltraCal XS, a calcium hydroxide paste used in dentistry. It's a "substantial equivalence" determination, not a study proving the device meets specific performance acceptance criteria in the way an AI/ML device would. Therefore, most of the requested information for an AI/ML study (like sample size for test sets, expert ground truth, MRMC studies, standalone performance, training sets, etc.) is not applicable here.

However, I can extract and infer some related information regarding the device's technical characteristics and how its equivalence to predicate devices was established.

Here's a breakdown of the applicable information:

1. A table of acceptance criteria and the reported device performance

The document does not specify "acceptance criteria" in the typical sense of numerical thresholds for clinical performance metrics (e.g., sensitivity, specificity, AUC) that an AI/ML study would define. Instead, it demonstrates "substantial equivalence" to legally marketed predicate devices based on technological characteristics, intended use, and safety. Its performance is implicitly shown through comparison to these predicates and successful completion of verification and validation testing.

Characteristic / "Acceptance Criteria" (Implicit)Reported Device Performance (UltraCal XS)Comparison to Predicate (DiaPaste K210333)
Product CodeKIF, EJK (Similar to predicates, within accepted codes for this type of device)Similar
Intended Use•Temporary dressing for root canals•Apexification and/or perforation•Vital Pulpotomy•Direct Pulp Capping•Indirect Pulp Capping•Root resorption•Root canal filling for primary teeth (Within the scope of predicate devices)Similar: The Indications for Use of the subject device is within that of the predicate and reference device(s).
Intended UserLicensed Dentist or Dental ProfessionalIdentical
Characteristics (Description)Aqueous 12.0 to 12.5 pH calcium hydroxide, radiopaque, premixed paste for direct application.Similar: Subject device and predicates are calcium hydroxide paste with relatively high pH.
Composition-Calcium hydroxide-Water-Barium Sulfate-Propylene Glycol-Hydroxypropyl MethylcelluloseSimilar: Primary ingredient (calcium hydroxide) is equivalent. Barium sulfate included as radio-opacifier. Differences in other ingredients (thickeners, humectants, etc.) do not raise new safety/effectiveness questions.
Delivery System/DeploymentProduct is provided in a 1.2mL syringe with disposable delivery tips.Similar: Both the subject device and the predicate and reference devices are provided in a pre-filled syringe with or without delivery tips.
Physical Properties-Viscosity-Water Removal-Radio-Opacity (Evaluated through similar testing as predicates, though not directly to ISO 6876 which was deemed not applicable for a temporary dressing as opposed to a sealing material).Similar: Testing performed to evaluate characteristics of the chemistry. (Note: ISO 6876 applicable for root canal sealing material, not dressing, so direct comparison not made).
Calcium Hydroxide Concentration35%Similar: Concentration range is within ranges cleared by FDA and does not raise new questions regarding safety and efficacy.
Patient PopulationIndividuals of all ages and gender and shall be assessed by the administering dental professional.Identical
Biocompatibility and SafetyISO 7405:2018 (Successfully passed all verification and validation testing, including biocompatibility assessment).ISO 7405 was applied as it is a recommended biocompatibility evaluation standard for this product type.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This is a 510(k) submission for a dental filling material, not an AI/ML device relying on a "test set" of clinical data to evaluate algorithmic performance. The submission relies on laboratory testing and comparison to existing devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. The "ground truth" for this device's safety and effectiveness is established through adherence to recognized standards (e.g., ISO 7405 for biocompatibility) and demonstrated substantial equivalence to legally marketed predicate devices, supported by laboratory characterization of its physical and chemical properties.

8. The sample size for the training set

Not applicable.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

October 13, 2021

Ultradent Products, Inc. Adam Black Regulatory Affairs Manager 505 West Ultradent Drive (10200 South) South Jordan, Utah 84095

Re: K211905

Trade/Device Name: UltraCal XS Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF, EJK Dated: July 15, 2021 Received: July 16, 2021

Dear Adam Black:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211905

Device Name UltraCal XS

Indications for Use (Describe)

UltraCal XS calcium hydroxide paste is indicated to be used for:

· temporary dressing for root canals

  • apexification and/or perforation
  • · vital pulpotomy
  • direct pulp capping
  • · indirect pulp capping
  • · root resorption
  • root canal filling for primary teeth

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary-K211905

This summary of the traditional 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92 for UltraCal XS.

l. Applicant's Name and Address

Ultradent Product, Inc. 505 West Ultradent Drive (10200 South) South Jordan, UT 84095

Contact Person:Mr. Adam Black
Title:Regulatory Affairs Manager
Telephone:801-553-4425
Cell Phone:435-459-9302
Fax:801-553-4609

Date Summary Prepared: 22 September 2021

II. Name of the Device

Device:Resin, Root Canal Filling
Trade/Device Name:UltraCal XS
Common Name:Root Canal Filling Resin
Review Panel:Dental
Regulation Number:21 CFR 872.3820
Device Class:Class II
Classification Product Code:KIF
Subsequent Product Code:EJK

III. Device Description

UltraCal XS calcium hydroxide paste is an aqueous, 12.0 to 12.5 pH, radiopaque, premixed paste for direct application.

IV. Statement of Intended Use

UltraCal XS calcium hydroxide paste is indicated to be used for:

  • . Temporary dressing for root canals
  • . Apexification and/or perforation

{4}------------------------------------------------

  • Vital Pulpotomy
  • Direct Pulp Capping
  • Indirect Pulp Capping
  • Root resorption
  • Root canal filling for primary teeth

V. Predicate Device

UltraCal XS identified primary predicate: DiaPaste (K210333); secondary predicates: CleaniCal (K201799), TheraCal DC (K180344), Dia-Root Bio MTA (K200174).

VII. Comparison of Technological Characteristics

Predicate technological comparison:

The technology, delivery, and intended use of UltraCal XS are substantially equivalent to the identified predicate as outlined in Table 5-1:

{5}------------------------------------------------

Ultradent Products, Inc. 505 West Ultradent Drive (10200 South) South Jordan, UT 84095 Est. Reg. No. 1718912

Table 5-1: UltraCal XS substantial equivalence comparison

DescriptiveInformation/characteristicPrimary Predicate:DiaPaste (K210333)Device: UltraCal XSSecondary Predicate:CleaniCal (K201799)Secondary Predicate:TheraCal DC (K180344)Secondary Predicate:Dia-Root Bio MTA(K200174)Differences
Product CodeKIFKIF, EJKEJK, KIFEJKKIFSimilar
Intended UseAqueous ointmentmaterial thattemporarily fills theroot canal for thefollowing indications:-Apexification-Temporary root filling-Root canal filling forprimary teeth-Vital pulpotomy-Temporary pulpcappingUltraCal XS calciumhydroxide paste isindicated to be usedfor:•Temporary dressingfor root canals•Apexification and/orperforation•Vital Pulpotomy•Direct Pulp Capping•Indirect Pulp Capping•Root resorption•Root canal filling forprimary teethCleaniCal is a CalciumHydroxide paste thathas a creamyconsistency and issuitable for severalindications including.-Temporarydisinfectant dressingsin the obturation ofroot canals-Indirect pulp cappingor management ofdeep caries lesions; or-Direct pulp capping1. Pulpotomy2. Temporary FillingMaterial3. Repair of RootPerforations4. Repair of FurcationPerforations5. Repair of PerforatingInternal and ExternalResorptions6. Root-End Filling inEndodontic Surgery7. Pulp Exposures (DirectPulp Capping)8. Protective Liner(Indirect Pulp Capping)and Base for Use Undera Variety of SubstratesDIA-ROOT BIO MTA isused for pulp capping(direct pulp capping orpartial pulpotomy) andrepair of root perforation.Other indications for useinclude: repair of rootresorption, root endfilling, apexification andpulpotomy.SimilarThe Indications for Useof the subject device iswithin that of thepredicate andreference device(s).
Intended UserLicensed Dentist orDental ProfessionalLicensed Dentist orDental ProfessionalDental ProfessionalDental ProfessionalDental ProfessionalIdentical
CharacteristicsCalcium hydroxidepaste with bariumsulfate, used as aAqueous 12.0 to 12.5pH calcium hydroxide,radiopaque, premixedPre-filled syringe withcalcium hydroxidepaste with a pH of 12.3.TheraCal DC is abiocompatible, basic,dual-cured, resin-SimilarThe subject device aswell as the predicate
DescriptiveInformation/characteristicPrimary Predicate:DiaPaste (K210333)Device: UltraCal XSSecondary Predicate:CleaniCal (K201799)Secondary Predicate:TheraCal DC (K180344)Secondary Predicate:Dia-Root Bio MTA(K200174)Differences
temporary root canalfilling material.paste for directapplication.modified calcium silicateprovided in a pre-filledsyringeand reference device(s)are both calciumhydroxide paste withrelatively high pH.
Composition-Calcium Hydroxide-Titanium Dioxide-Zinc Oxide-Barium Sulfate-Water-Polyether Polyol-Polysorbate(Tween80)-Calcium hydroxide-Water-Barium Sulfate-Propylene Glycol-HydroxypropylMethylcellulose-Calcium hydroxide-Zirconium dioxide- Excipients (nMethyl-2-pyrrolidone,Hypromellose)-Calcium hydroxide-Portland Cement-All other componentsare proprietary-Calcium silicate-Zirconium dioxide-Citric Acid-Silanamine, 1,1,1-trimethyl-N-(trimethylsilyl-,hydrolysisproducts with silica-HydroxypropylmethylcelluloseSimilarThe primary ingredientis equivalent in thepredicate and subjectdevice, calciumhydroxide. Bariumsulfate is included inboth as a radio-opacifier. Differences inthe ingredientsincluded varyingchemicals used asthickeners,humectants,emulsifiers, etc.Dia-Root Bio MTA doesnot include calciumhydroxide as acomponent, howeverthe hydration reactionduring setting occursbetween tricalcium
DescriptiveInformation/characteristicPrimary Predicate:DiaPaste (K210333)Device: UltraCal XSSecondary Predicate:CleaniCal (K201799)Secondary Predicate:TheraCal DC (K180344)Secondary Predicate:Dia-Root Bio MTA(K200174)Differences
silicate $(3CaO·SiO2)$ anddicalcium silicate$(2CaO·SiO2)$ to form acalcium hydroxide andcalcium silicate hydrategel, producing analkaline pH.
Delivery Systemor DeploymentMethodsProduct is provided ina 2g syringe withdisposable deliverytipsProduct is provided ina 1.2mL syringe withdisposable deliverytipsPre-loaded syringePre-loaded syringePowder within a cap &body to be mixed withdistilled water on amixing pad and appliedintraorally with a spatulaSimilarBoth the subject deviceand the predicate andreference devices areprovided in a pre-filledsyringe with or withoutdelivery tips.The difference withDia-Root Bio MTAdelivery is based on thenecessary reaction withwater to achieveperformance.
DescriptiveInformation/characteristicPrimary Predicate:DiaPaste (K210333)Device: UltraCal XSSecondary Predicate:CleaniCal (K201799)Secondary Predicate:TheraCal DC (K180344)Secondary Predicate:Dia-Root Bio MTA(K200174)Differences
PhysicalPropertiesConfirmed to ISO 6876-Flowability-Film Thickness-Radio-opacity-Viscosity-Water Removal-Radio-OpacityConfirmed to ISO 6876-Packaging-Extraneous matter-Flowability-Radio-opacityConfirmed to ISO 6876-Radiopaque-Calcium releasing-Basic-Water sorption-SolubilityConfirmed to ISO 6876-Setting time-Solubility-RadiopacitySimilarISO 6876 is applicablefor root canal sealingmaterial, as UltraCal XSis not intended to sealroot canals and wastherefore deemed notapplicable. Similartesting was performedto evaluate thecharacteristics of thechemistry.
CalciumHydroxideConcentration22-26%35%30%Not specifiedNot specified - based onratio of distilled waterand calcium silicate usedSimilarConcentration range iswithin ranges clearedby FDA and does notraise new questionsregarding safety andefficacy.
PatientPopulationIndividuals of all agesand gender and shallbe assessed by theadministering dentalprofessional.Individuals of all agesand gender and shallbe assessed by theadministering dentalprofessional.Individuals of all agesand gender and shall beassessed by theadministering dentalprofessional.Not specifiedNot specifiedIdentical
DescriptiveInformation/characteristicPrimary Predicate:DiaPaste (K210333)Device: UltraCal XSSecondary Predicate:CleaniCal (K201799)Secondary Predicate:TheraCal DC (K180344)Secondary Predicate:Dia-Root Bio MTA(K200174)Differences
Biocompatibilityand SafetyISO 10993ISO 7405:2018ISO 10993ISO 7405:2008ISO 10993-1:2009ISO 7405:2018ISO 7405 was appliedas it is a recommendedbiocompatibilityevaluation standard forthis product type.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

VII Conclusion:

The purpose of this submission was to address the change in indications for use of the previously submitted UltraCal XS (K970114). As outlined in the comparison table above, UltraCal XS is similar to the identified predicate device with respect to its intended use, its Intended User, the Device Design, Types of Material used, Delivery System and or Deployment Method, Physical Properties and Patient Population. The subject device also successfully passed all verification and validation testing, including the biocompatibility assessment.

In summary it can be stated that the development of the subject device, UltraCal XS, is based on a well-established technology in the form of the predicate device, DiaPaste (K210333) and is chemically equivalent to the previously submitted UltraCal XS (K970114). The decision to rely on DiaPaste (K210333) as the predicate rather than the previously submitted UltraCal XS (K970114) is due to the updated indications of the subject device. In order to establish substantial equivalence, DiaPaste (K210333) was selected as the best fit predicate based off of chemistry and cleared indications for use.

Based on the completed testing and the comparisons to the predicate device, we believe that UltraCal XS is substantially equivalent to the predicate device and does not raise any different questions regarding safety and effectiveness.

§ 872.3820 Root canal filling resin.

(a)
Identification. A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.(b)
Classification. (1) Class II if chloroform is not used as an ingredient in the device.(2) Class III if chloroform is used as an ingredient in the device.
(c)
Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to a root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976. Any other root canal filling resin shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.