K Number
K211897

Validate with FDA (Live)

Date Cleared
2021-08-03

(43 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Device Description

The Surgical Face Masks are blue color, single use, three-layer, flat-folded masks with nose piece and ear loops. The blue colorant is polypropylene (PP) master batch.

The Surgical Face Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter.

The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex or fiberglass.

The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Iron core coated with polypropylene resin.

The surgical face masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

This document describes the premarket notification (510(k)) for a Surgical Face Mask (K211897), comparing it to a predicate device (K182515). The acceptance criteria and the study proving the device meets them are detailed in the "Summary of Non-Clinical Test" section.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance:

Test Method (Purpose)Acceptance CriteriaReported Device PerformanceResult
Bacterial Filtration Efficiency (BFE) (%) (ASTM F2101-19)≥98%94/96 samples passed at ≥99%Pass
Differential Pressure (mmH2O/cm²) (EN 14683: 2019, Annex C)<6.0 mmH2O/cm²92/96 samples passed at <4 mmH2O/cm²Pass
Sub-micron Particulate Filtration Efficiency (PFE) at 0.1 micron (%) (ASTM F2299-03)≥98%96/96 samples passed at ≥99%Pass
Resistance to Penetration by Synthetic Blood (mmHg) (ASTM F1862M-17)Fluid resistant claimed at 160 mm Hg95 of 96 test articles passed at 160mmHgPass
Flammability (ASTM F2100:2019, 16 CFR Part 1610) (Flame spread)Class 1 (≥3 seconds burn Time)96/96 samples passed at ≥3 seconds burn time - Class 1Pass

Biocompatibility Testing (According to ISO 10993-1:2009 for surface device, skin contact, ≤24h duration):

Test Method (Purpose)Acceptance CriteriaReported Device PerformanceResult
In vitro Cytotoxicity Test (ISO 10993-5:2009)Device extract determined to be non-cytotoxicSubject device extract was determined to be non-cytotoxicSame (Pass)
Skin Sensitization Tests (ISO 10993-10:2010)Device extract determined to be non-sensitizerSubject device extract was determined to be non-sensitizerSame (Pass)
Skin Irritation Tests (ISO 10993-10:2010)Device extract determined to be non-irritantSubject device extract was determined to be non-irritantSame (Pass)

2. Sample size used for the test set and the data provenance:

  • Sample Size for performance tests (BFE, Delta-P, PFE, Synthetic Blood Penetration, Flammability): 3 lots were tested, with a total of 96 samples (e.g., "3 lots tested with total 96 samples"). This implies 32 samples per lot.
  • Sample Size for Biocompatibility tests: The document doesn't explicitly state the sample size for biocompatibility tests (e.g., number of test articles or subjects used in in vitro or in vivo studies for cytotoxicity, irritation, and sensitization). It only states that the tests "were conducted."
  • Data Provenance: Not explicitly stated regarding country of origin for the test data. The manufacturer is "ZHEJIANG HONGYU MEDICALI COMMODITY CO., LTD" from Yiwu City, Zhejiang Province, China, suggesting the testing likely occurred in China.
  • Retrospective or Prospective: The testing described is non-clinical performance testing, which is generally conducted prospectively for regulatory submission purposes.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. For this type of device (surgical face mask) and the non-clinical performance tests conducted, "ground truth" is established by direct measurement against established international standards (ASTM, EN, ISO). There is no "expert consensus" on image interpretation or diagnosis involved.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. As mentioned above, this is non-clinical performance testing based on direct measurements and adherence to specified pass/fail criteria within the standards. There is no need for expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This submission is for a surgical face mask, not an AI-assisted diagnostic device. Therefore, no MRMC study or AI assistance evaluation was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a physical medical device, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Standardized Test Methods and Criteria: The "ground truth" for the device's performance is defined by the acceptance criteria specified in the relevant international standards (ASTM, EN, ISO, CFR). For example, the "truth" for bacterial filtration efficiency is a measured percentage, compared to a required percentage (e.g., ≥98%).

8. The sample size for the training set:

  • Not Applicable. This device is a physical product, not a machine learning model; therefore, there is no "training set" in the context of data used to train an algorithm.

9. How the ground truth for the training set was established:

  • Not Applicable. (As above, no training set for a physical product.)

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 3, 2021

Zhejiang Hongyu Medicali Commodity CO., LTD % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K211897

Trade/Device Name: Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: June 21, 2021 Received: June 21, 2021

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K211897

Device Name Surgical Face Mask

Indications for Use (Describe)

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(K) Summary

K211897

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

Date prepared: August 3, 2021

A. Applicant Information:

ZHEJIANG HONGYU MEDICALI COMMODITY CO., LTD Address: No. 668 ChanHua Road, Fotang Town Industrial Functional Area, 322002, Yiwu City, Zhejiang Province, China Contact person: Sophie Title: Business Manager Tel: +86- 0579-85980762 Fax: +86- 0579-85070207 Email: 2853571191@qq.com

Submission Correspondent:

Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Subject Device Information:

Trade Name: Surgical Face Mask Common Name: SURGICAL MASK Model: Ear Loops

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

{4}------------------------------------------------

C. Predicate device Information: K182515 Surgical Face Mask Wuhan Dymex Healthcare Co., Ltd.

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

D. Indications for use of the device:

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

E. Device Description:

The Surgical Face Masks are blue color, single use, three-layer, flat-folded masks with nose piece and ear loops. The blue colorant is polypropylene (PP) master batch.

The Surgical Face Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter.

The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex or fiberglass.

The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Iron core coated with polypropylene resin.

The surgical face masks are sold non-sterile and are intended to be single use, disposable devices.

F. Comparison with predicate device

DeviceProposed DevicePredicate DeviceResult
510K #K211897K182515-
ManufacturerZHEJIANG HONGYU MEDICALICOMMODITY CO., LTDWuhan Dymex Healthcare Co., Ltd.-
Model NameSURGICAL FACE MASKEar loopsSURGICAL FACE MASKEar loopsSame
ClassificationClass II Device, FXX (21CFR878.4040)Class II Device, FXX (21 CFR878.4040)Same
Indication for useThe Surgical Face Mask isintended to be worn to protectboth the patient andhealthcare personnel fromThe Surgical Face Masks areintended to be worn to protect boththe patient and healthcarepersonnel from transfer ofSame
transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
Design FeaturesEar Loops, Flat-pleated, 3 layersEar Loops, Flat-pleated, 3 layersSame
MaterialsOuter layerSpunbond PolypropyleneSpunbond PolypropyleneSame
Inner layerSpunbond PolypropyleneSpunbond PolypropyleneSame
Filter layerMelt-blown PolypropyleneMelt-blown PolypropyleneSame
Nose wireIron core coated with polypropylene resinMalleable polyethylene wireDifferent
Ear loopsPolyester + spandexSpandexDifferent
ColorBlueYellowDifferent
Dimension (Length)17.5cm±0.2cm17.5cm±0.2cmSame
Dimension (Width)9.5cm±0.2cm9.5cm±0.2cmSame
Dimension (Ear Loops)17cm±0.2cm16.5cm±0.5cmSimilar
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTM F2100 LevelLevel 3Level 2Different
BiocompatibilityCytotoxicityUnder the conditions of the study, the subject device extract was determined to be non-cytotoxicUnder the conditions of the study, the subject device extract was determined to be non-cytotoxicSame
IrritationUnder the conditions of the study, the subject device extract was determined to be non-irritantUnder the conditions of the study, the subject device extract was determined to be non-irritantSame
SensitizationUnder the conditions of the study, the subject device extract was determined to be non-sensitizerUnder the conditions of the study, the subject device extract was determined to be non-sensitizerSame

{5}------------------------------------------------

The proposed device has different material of nose piece, ear loops and different color to the predicate device, while the proposed device has been tested and the test results shown that the material and color differences do not affect the safety of the proposed device.

The proposed device is similar in design, intended use, technological characteristics, and is composed of the same or similar components as the predicate device. The product proposed under this premarket notification

{6}------------------------------------------------

submission has the same or similar performance characteristics and conform to the same or similar standards. Differences of the penetration level between the Surgical Face Masks and predicate devices did not raise any new concerns regarding safety and effectiveness.

G. Summary of Non-Clinical Test

A

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was same to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004:

Test MethodPurposePass CriteriaResults
ASTM F2101-19 StandardTestMethod for Evaluatingthe Bacterial FiltrationThe purpose of the test is toevaluate the Bacterial filtrationefficiency (BFE) (%)≥98%3 lots tested with total96 samples,94/96 Passed at≥99% /Pass
EN 14683: 2019, Annex CMedical face masks -Requirements andtest methods according toASTM F2100:2019The purpose of the test is toevaluate the Different pressure(Delta-P)<6.0mmH20/cm²3 lots tested with total96 samples,92/96 Passed<4 mmH2O/cm²/ Pass
ASTM F2299-03 StandardTest Method forDetermining the InitialEfficiency of Materials Usedin Medical Face Masks toPenetration by ParticulatesUsing Latex Spheresaccording toASTMF2100:2019The purpose of the test is toevaluate the Sub-micronparticulate filtration efficiency at0.1 micron, % (PFE)≥98%3 lots tested with total96 samples,96/96 Passed at≥99% / Pass
ASTM F1862M-17 StandardTest Method for Resistanceof Medical Face Masks toPenetration by SyntheticBlood (Horizontal Projectionof Fixed Volume at a KnownVelocity) according toASTMF2100:2019for passresultThe purpose of the test is toevaluate the Resistance topenetration by synthetic blood,Minimum pressure in mmHgFluidresistantclaimed at160 mm Hg3 lots tested with total96 samples,95 of 96 test articlespassed at 160mmHg/Pass
16 CFR Part 1610 Standard forthe Flammability of Clothingaccording to ASTMF2100:2019The purpose of the test is toevaluate the Flame spread3 lots tested with total96 samples,96/96 Passed $\geq$ 3seconds burn Time-Class 1 /Pass
Class 1

{7}------------------------------------------------

  • Biocompatibility Testing According to ISO 10993-1:2009, the nature of body contact ● for the subject device is Surface Device category, Skin Contact and duration of contact is A-Limited (≤24h). The following tests for the subject device were conducted to demonstrate that the subject device is biocompatible and safe for its intended use:
  1. In vitro Cytotoxicity Test per ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity,

  2. Skin Sensitization Tests per ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization,

  3. Skin Irritation Tests per ISO 10993-10:2010 Biological evaluation of medical devices— Part 10: Tests for irritation and skin sensitization.

H. Summary of Clinical Test Conclusion

No clinical study is included in this submission.

l. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K211897, the Surgical Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K182515.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.