K Number
K211681

Validate with FDA (Live)

Device Name
EON
Date Cleared
2022-03-30

(302 days)

Product Code
Regulation Number
878.5400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EON (1064nm laser) is intended for non-invasive lipolysis of the abdomen and flanks to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic effect. It is intended for individuals with a Body Mass Index of 30 or less.

Device Description

EON aesthetic laser system is 1064nm diode laser device that is intended for non-invasive lipolysis of the abdomen and flanks in individuals with a Body Mass Index (BMI) of 30 or less. This wavelength is used to affect the appearance of visible fat bulges in the abdomen and flanks.

Both laser energy and cooling air are delivered concurrently through the same treatment head that is positioned by a robotic arm. The treatment head is never in contact with the skin.

During treatment, the desired area will be defined by the physician and will be treated with the 1064nm laser over a 20-minute timeframe. The mechanism of action is to preferentially heat the adipose tissue to temperatures that will induce apoptosis (42°C to 50°C). The cooling system will maintain the skin at comfortable temperature (<43°C).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the EON device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated Objectives)Reported Device Performance and Study Success
Primary Objective: Confirm safety of the 1064-nm laser device for non-invasive subcutaneous fat reduction in the flank area.Low incidence of adverse effects (only 1 subject developed palpable thickening, resolved by Week 12). Mean pain score during procedure was 1.95 (0-10 scale), decreasing to 0.9 at 30 minutes post-procedure.
Primary Efficacy Endpoint: Blinded evaluation of pre- and post-treatment photos. Study success if at least two of three independent readers correctly identify at least 9 of 11 photo sets as pre-treatment.27 (81.8%) of 33 image sets were correctly scored overall. Two of the three readers correctly identified pre- and post-treatment photos in at least 9 of 11 subjects, thus achieving study success.
Secondary Efficacy Endpoint: Mean fat reduction by ultrasound.At Week 12 after one treatment, the mean reduction in subcutaneous adipose thickness was 6.1 mm per flank and 12.1 mm per patient (-15%; p<0.01, both measures).
Secondary Objective: Assess extent of subject discomfort during treatment.Mean pain score during procedure was 1.95 (0-10 scale), decreasing to 0.9 at 30 minutes post-procedure.
Secondary Objective: Assess overall subject satisfaction with the results of the procedure.Subject satisfaction was "Excellent" for all subjects (100%).

2. Sample Size Used for the Test Set and Data Provenance

The text doesn't explicitly state the total number of subjects enrolled in the clinical study, but it implies a minimum of 11 subjects for the blinded photo evaluation ("at least 9 of the 11 photos as pre-treatment"). For the "27 (81.8%) of 33 image sets" for blinded evaluation, this implies 33 image sets were evaluated, which likely corresponds to 33 subjects or 33 treatment areas. Considering the 12.1mm reduction "per patient," it refers to patient-level data.

The data provenance is prospective clinical study. The country of origin of the data is not specified in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

Three independent readers were used for the blinded evaluation of pre- and post-treatment photos. Their specific qualifications (e.g., radiologist, years of experience) are not specified in the provided text.

4. Adjudication Method for the Test Set

The adjudication method for the blinded photo evaluation was 2 out of 3 concensus. The study was declared successful if "at least two of the three independent readers correctly identified at least 9 of the 11 photos as pre-treatment."

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not explicitly described in the provided text. The study involved a blinded evaluation by human readers, but it was for assessing the visual outcomes of the EON device itself, not to compare human reader performance with and without AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

The EON device is a laser system for lipolysis, not an AI-powered diagnostic tool. Therefore, a standalone (algorithm only) performance study in the context of diagnostic AI is not applicable and was not performed. The performance evaluation focuses on the efficacy and safety of the laser device itself.

7. The Type of Ground Truth Used

The ground truth used in the study included:

  • Expert Consensus (implied): For the blinded photo evaluation, the consensus of at least two out of three independent readers served as the ground truth for identifying pre- and post-treatment images.
  • Objective Measurements: Ultrasound measurements were used to objectively quantify subcutaneous fat reduction.
  • Subjective Outcomes: Patient satisfaction surveys and pain scores were also collected.

8. The Sample Size for the Training Set

The provided text does not mention a training set specific to the EON device's clinical study. This is expected as the EON is a physical device (laser system) and not a software algorithm that requires a training set in the typical AI/machine learning sense.

9. How the Ground Truth for the Training Set was Established

As there is no mention of a training set for the EON device in the context of the provided clinical study description, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.

Dominion Aesthetic Technologies, Inc. Ahmed Mohammed VP, Product Development 2431 Aloma Avenue Suite 225 Winter Park, Florida 32792

August 11, 2022

Re: K211681 Trade/Device Name: EON Regulation Number: 21 CFR 878.5400 Regulation Name: Low level laser system for aesthetic use Regulatory Class: Class II Product Code: PKT

Dear Ahmed Mohammed:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 30, 2022. Specifically, FDA is updating this SE Letter to remove OOK and GEX as an administrative correction. because OOK (Dermal Cooling Pack/Vacuum/Massager) and GEX (Powered Laser Surgical Instrument) do not apply to your device.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Long Chen, Ph.D., OHT4: Office of Surgical and Infection Control Devices, 301-796-6389, Long.Chen@fda.hhs.gov.

Sincerely,

Long H. Chen -S Digitally signed by Long H. Chen -S
Date: 2022.08.11 13:57:54 -04'00'

Long Chen, Ph.D. Director (Acting) DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. On the left is the Department of Health and Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 30, 2022

Dominion Aesthetic Technologies, Inc. Ahmed Mohammed VP, Product Development 2431 Aloma Avenue Suite 225 Winter Park, Florida 32792

Re: K211681

Trade/Device Name: EON Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, OOK, PKT Dated: February 28, 2022 Received: February 28, 2022

Dear Ahmed Mohammed:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Purva U. Pandya -S

Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K211681

Device Name EON

Indications for Use (Describe)

The EON (1064nm laser) is intended for non-invasive lipolysis of the abdomen and flanks to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic effect. It is intended for individuals with a Body Mass Index of 30 or less.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Section 5: 510(k) Summary

Table Of Contents

D

Section 5: 510(k) Summary5-1
5-1 Submitter5-2
5-2 Device Information.5-2
5-3 Predicate Device.5-3
5-4 Device Description...5-3
5-5 Indications for Use...5-3
5-6 Comparison of Technological Characteristics with the Predicate Device5-4
5-7 Performance Data5-7
5.7.1 Electrical Safety and Electromagnetic Compatibility (EMC) Testing Conducted5-7
5.7.2 Software Validation & Verification5-7
5.7.3 Other Non-Clinical Testing Conducted.5-7
5.7.4 Summary of the Clinical Study Conducted5-8
5-8 Conclusions5-10

{5}------------------------------------------------

5-1 Submitter

Dominion Aesthetic Technologies, Inc. 2431 Aloma Avenue Suite # 300 Winter Park, FL 32792

Contact Person: Ahmed Mohammed Phone: 763-439-4602 Email: amohammed@dominionaethetic.com

Date Prepared: March 15, 2022

5-2 Device Information

Name of Device: EON. Common or Usual Name: Aesthetic Laser System Classification Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology (21 CFR 878.4810) Regulatory Class: II Product Code: GEX, OOK and PKT

{6}------------------------------------------------

5-3 Predicate Device

Cynosure, Inc., Sculpsure, K160470

5-4 Device Description

EON aesthetic laser system is 1064nm diode laser device that is intended for non-invasive lipolysis of the abdomen and flanks in individuals with a Body Mass Index (BMI) of 30 or less. This wavelength is used to affect the appearance of visible fat bulges in the abdomen and flanks.

Both laser energy and cooling air are delivered concurrently through the same treatment head that is positioned by a robotic arm. The treatment head is never in contact with the skin.

During treatment, the desired area will be defined by the physician and will be treated with the 1064nm laser over a 20-minute timeframe. The mechanism of action is to preferentially heat the adipose tissue to temperatures that will induce apoptosis (42°C to 50°C). The cooling system will maintain the skin at comfortable temperature (<43°C).

5-5 Indications for Use

The EON (1064nm laser) is intended for non-invasive lipolysis of the abdomen and flanks to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic effect. It is intended for individuals with a Body Mass Index of 30 or less.

{7}------------------------------------------------

K211681

5-6 Comparison of Technological Characteristics with the Predicate Device

ParameterDominionEONCynosure, Inc.SculpSureComparison
Indications for Use (IFU)The EON (1064nm laser) isintended for non-invasivelipolysis of the abdomen andflanks to achieve disruptionof adipocyte cells intendedfor non-invasive aestheticuse to achieve a desiredaesthetic effect. It isintended for individuals witha Body Mass Index of 30 orless.Cynosure SculpSure is adiode laser system intendedfor non-invasive lipolysis ofthe abdomen and flanks inindividuals with a Body MassIndex (BMI) of 30 or less. Thedevice is intended to affectthe appearance of visible fatbulges in the abdomen andflanks.Similar
K NumberK211681K160470N/A
Lipolysis MethodHeat-assistedHeat-assistedSame
Laser TypeDiodeDiodeSame
Wavelength1064 nm1064 nmSame
Power ModeContinuous Wave (CW)Continuous Wave (CW)Same
Pulse Length1 to 20s1 to 20sSame
ParameterDominionEONCynosure, Inc.SculpSureComparison
Applicator Size75 cm², 110 cm², 150 cm²24 cm² (4X)Similar
Application MethodArticulated Scanning Arm, Non-contactingBeltDifferent: EON uses an articulated scanning arm to continuously move the laser beam over the treatment area and, simultaneously, provide skin cooling. EON is similar to SculpSure in principal of operation where both devices deliver laser energy and cool down the skin in the treated area. To ensure safety, EON monitors skin temperature and turn off the laser if the skin temperature exceeds 43°C. EON uses jet impingement cooling (which is more effective than contact cooling), this ensures that the skin in the treated area is safe and comfortable while the laser energy is heating the subcutaneous adipose tissue. EON is as safe, and as effective as SculpSure.
Treatment AreaAbdomen and FlanksAbdomen and FlanksSame
Maximum Power DensityUp to 1.4 W/cm²Up to 1.4 W/cm²Same
ParameterDominionEONCynosure, Inc.SculpSureComparison
Supply Voltage110 V; Single Phase220 V; Single PhaseDifferent: EON uses 110V supply voltage toimprove usability since 110V is the standardsupply voltage the United States. Supply voltagedoesn't affect efficacy. EON design meets allapplicable electrical safety standards.
Supply Current20A20ASame
Laser CoolingClosed cycle, internalClosed cycle, internalSame

{8}------------------------------------------------

K211681

{9}------------------------------------------------

I

DN Premarket Notification 510(k)

{10}------------------------------------------------

5-7 Performance Data

5.7.1 Electrical Safety and Electromagnetic Compatibility (EMC) Testing Conducted

  • . IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance
  • IEC 60601-1 :2005 (Third Edition) + CORR. 1 :2006 + CORR. 2:2007 +A1 :2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1: 2012 reprint) including compliance with US National Differences demonstrating conformance to the FDA-recognized standard for electrical safety (AAMI/ANSI ES60601-1:2005(R) 2012).

5.7.2 Software Validation & Verification

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or operator.

5.7.3 Other Non-Clinical Testing Conducted

  • IEC 60601-2-22 Edition 3.1 2012-10 Medical Electrical Equipment Part 2-22: Particular ● Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment
  • IEC 60825-1 :2014 (Third Edition) Safety of Laser Products Part 1: Equipment . Classification and Requirements

{11}------------------------------------------------

5.7.4 Summary of the Clinical Study Conducted

5.7.4.1 Objectives

The primary objective of this study was to confirm the safety of a1064-nm laser device with a novel robotic arm for noninvasive subcutaneous fat reduction in the flank area. Secondary objectives included assessing the extent of subject discomfort during treatment, overall subject satisfaction with the results of the procedure, and a determination of subcutaneous fat reduction in the treated area.

5.7.4.2 Efficacy Endpoints

Primary efficacy endpoint: blinded evaluation of pre- and post-treatment photos Secondary efficacy endpoints: mean fat reduction by ultrasound and subject satisfaction.

5.7.4.3 Materials and Methods

A 110-cm² area on both flanks of enrolled subjects was treated for 20 minutes with an FDAcleared robotic non-contact 1064-nm laser system (EON®; Dominion Aesthetic Technologies, Inc.). Patients were followed for 12 weeks, and examined routinely at 2 weeks, 12 weeks and additionally as needed, post-treatment. Pre- and post-treatment photos were taken by a professional photographer using a standardized DSLR and lighting set-up. The ability of three independent readers to correctly identify blinded pre-treatment vs. post-treatment photos was assessed. The study was declared to be a success if at least two of the three independent readers correctly identified at least 9 of the 11 photos as pre-treatment. Ad hoc surveys were administered to assess patient satisfaction. A 2-week post-treatment ultrasound scan was used to check for changes in the treated area. The study was powered to assess patient-level and flank-level subcutaneous adipose tissue thickness using ultrasound measurements taken at a center of each treatment zone prior to treatment and at 12 weeks post-treatment for efficacy determination, with mean thicknesses calculated per subject and per flank.

5.7.4.4 Results

The treatment had a low incidence of adverse effects, with only one subject developing a palpable thickening in the subcutaneous tissue following treatment. This was noted at the two-week time period and had resolved by the 12-week post-treatment exam. No other predefined adverse effects were noted. On a scale of 0 to 10, the mean pain score during the procedure was 1.95, decreasing to 0.9 at 30 minutes post-procedure. Subject satisfaction was "Excellent" for all subjects (100%). For blinded evaluation of pre- and post-treatment photos, 27 (81.8%) of 33 image sets were correctly scored overall, with two of the three readers correctly identifying pre- and post-treatment photos in at least 9 of the 11 subjects, thus achieving study success. At Week 12 after one treatment, the mean reduction in

{12}------------------------------------------------

subcutaneous adipose thickness on the treated flanks was 6.1 mm per flank and 12.1 mm per patient (-15%; p<0.01, both measures).

Similar to the prior abdominal study with the same robotic laser device, this study confirms the safety of this 1064-nm non-contact laser device for treating subcutaneous fat on the flanks. The study provided objective evidence of EON safety and efficacy in treating the flank area.

{13}------------------------------------------------

5-8 Conclusions

D

The clinical and non-clinical testing conducted on EON support the safety of the device and the hardware and software verification and validation demonstrate that EON should perform as intended in the specified use conditions. Dominion believes that EON is as safe, as effective, and performs as well as the predicate device. The clinical performance of EON demonstrates that EON has a safety and effectiveness profile that is similar to its predicate device. It has been concluded that the EON device is safe and effective for its intended use.

§ 878.5400 Low level laser system for aesthetic use

(a)
Identification. A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.