VITEK 2 AST-Gram Negative Piperacillin / Tazobactam (<=4 - =>128 µg/mL)
K211630 · bioMerieux, Inc. · LON · Aug 30, 2021 · Microbiology
Device Facts
| Record ID | K211630 |
| Device Name | VITEK 2 AST-Gram Negative Piperacillin / Tazobactam (<=4 - =>128 µg/mL) |
| Applicant | bioMerieux, Inc. |
| Product Code | LON · Microbiology |
| Decision Date | Aug 30, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1645 |
| Device Class | Class 2 |
Intended Use
VITEK®2 Gram Negative Piperacillin/Tazobactam is designed for antimicrobial susceptibility testing of Acinetobacter baumanii. Escherichia coli. Klebsiella pneumoniae. Pseudomonas aeruginosa, Citrobacter koseri, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, and Salmonella enterica. It is intended for use with the VITEK® 2 and VITEK® 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/mL. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
Device Story
VITEK 2 AST-GN Piperacillin/Tazobactam card; miniaturized, automated doubling dilution technique for MIC determination. Input: pure culture organism suspension (McFarland 0.5). Operation: VITEK 2/Compact systems automatically fill/seal cards; incubate; monitor growth via optical scanner (light transmittance) over 4-18 hours. Output: MIC value and interpretive category. Used in clinical microbiology labs by technicians. Output informs clinicians of in vitro susceptibility to guide antibiotic therapy. Benefits: standardized, automated susceptibility testing for Gram-negative bacilli.
Clinical Evidence
Bench testing only. Re-analysis of 187 clinical/challenge P. aeruginosa isolates from K113200 using updated breakpoints. Combined Essential Agreement (EA) >90%. Category Agreement (CA) was 86.1% (auto-dilution) and 82.4% (manual-dilution). Minor errors increased, leading to a specific testing limitation for MIC 32-64 µg/mL. Trending analysis identified higher MIC trends for P. aeruginosa, M. morganii, and P. rettgeri, and lower trends for S. enterica, addressed via labeling footnotes.
Technological Characteristics
Miniaturized broth microdilution card; automated incubation/reading system. VITEK 2 (automated fill/seal) and VITEK 2 COMPACT (manual fill/seal). Measures bacterial growth via optical monitoring over 18 hours. Software-based MIC determination.
Indications for Use
Indicated for antimicrobial susceptibility testing of specific Gram-negative bacteria (A. baumanii, E. coli, K. pneumoniae, P. aeruginosa, C. koseri, M. morganii, P. mirabilis, P. vulgaris, P. rettgeri, P. stuartii, S. enterica) to determine in vitro susceptibility to antimicrobial agents using VITEK 2 and VITEK 2 COMPACT systems.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
- Vitek 2 Ast-gn Piperacillin/tazobactam (K113200)
Related Devices
- K161217 — VITEK 2 Gram Negative Ceftriaxone (<=0.25->=64 ug/mL) · bioMerieux, Inc. · Jan 12, 2017
- K091899 — VITEK 2 GRAM NEGATIVE MEROPENEM · bioMerieux, Inc. · Aug 14, 2009
- K172944 — VITEK 2 AST-GN Ceftazidime/Avibactam <=0.12 >=16 µg/mL · bioMerieux, Inc. · Dec 19, 2017
- K073406 — VITEK 2 GRAM NEGATIVE TRIMETHOPRIM/SULFAMETHOXAZOLE · bioMerieux, Inc. · Jan 23, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a bold, sans-serif font and the full name in a smaller, sans-serif font below it. The overall design is clean and professional, reflecting the agency's role in regulating food and drugs in the United States.
August 30, 2021
bioMérieux, Inc Mary Beth Anheuser Regulatory Affairs Specialist 595 Anglum Rd. Hazelwood, Missouri 63042 USA
Re: K211630
Trade/Device Name: VITEK 2 AST-Gram Negative Piperacillin / Tazobactam (<4 ≥128 ug/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON, LTW, LTT Dated: May 25, 2021 Received: May 27, 2021
Dear Mary Beth Anheuser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a blue circle on top of a yellow-green circle. The word "BIOMÉRIEUX" is written in white letters inside the blue circle.
# Special 510(k) SUMMARY
## VITEK ®2 Gram Negative Piperacillin/Tazobactam Special 510(k): Device Modification
### 510(k) Submission Information:
| Submitter's Name: | bioMérieux, Inc. |
|----------------------|-----------------------------------------------------------------------------------------------------|
| Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| Contact Person: | Mary Beth Anheuser<br>Staff, Regulatory Affairs Specialist |
| Phone Number: | 618-667-3179 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | May 25, 2021 |
| Formal/Trade Name: | VITEK® 2 Gram Negative Piperacillin/Tazobactam<br>(≤4 - ≥128 µg/mL) |
| Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device, 21 CFR 866.1645 |
| Common Name: | VITEK 2 AST-GN Piperacillin/Tazobactam |
| Predicate Device: | VITEK 2 AST-GN Piperacillin/Tazobactam (K113200) |
### D. 510(k) Summary:
B.
C.
VITEK®2 Gram Negative Piperacillin/Tazobactam is designed for antimicrobial susceptibility testing of Acinetobacter baumanii. Escherichia coli. Klebsiella pneumoniae. Pseudomonas aeruginosa, Citrobacter koseri, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, and Salmonella enterica. It is intended for use with the VITEK® 2 and VITEK® 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/mL. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 COMPACT has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of
### bioMérieux, Inc.
595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700
{4}------------------------------------------------
the incubation cvcle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK 2 Gram Neqative Piperacillin/Tazobactam demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003.
This submission is the result of a FDA 2019 breakpoint change, as published on the FDA STIC website, for Pseudomonas aeruginosa tested with Piperacillin/Tazobactam (≤16, 32-64, 128 ≥). The intended use of the VITEK 2 Gram Negative Piperacillin/Tazobactam, as described in the labeling (package insert), has not changed as a result of this modification to the breakpoints specific to Pseudomonas aeruginosa. The overall performance of the device has been updated to reflect the performance of Pseudomonas aeruginosa when the updated breakpoints are applied. This modification does not affect the fundamental scientific technology of this device.
This premarket Notification (510(k)) presents the data in support of the VITEK 2 Gram Negative Piperacillin/Tazobactam test when the updated breakpoints for Pseudomonas aeruginosa (≤16, 32-64, 128 ≥) are applied. The data are representative of performance on both the VITEK 2 and VITEK 2 COMPACT instrument platforms. VITEK 2 Gram Negative Piperacillin/Tazobactam demonstrated an overall acceptable performance of 92.3% Category Agreement. Quality control and reproducibility are acceptable and remain unchanged from the original clearance (K113200).
The VITEK 2 Gram Neqative Piperacillin/Tazobactam test demonstrated acceptable performance during verification of the breakpoint change specific to Pseudomonas aeruginosa. No new risks were introduced through this modification and the performance equivalency has been established through verification testing. This allows for a substantial equivalence decision when compared to the current device cleared under K113200.