K Number
K091899

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-08-14

(50 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VITEK® 2 Gram Negative Meropenem is designed for antimicrobial susceptibility testing of gramnegative microorganisms and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Meropenem is a quantitative test. Meropenem has been shown to be active against most isolates of the following microorganisms listed below according to the FDA label for the antimicrobial.

Active in vitro and in clinical infections:

Escherichia coli Klebsiella pneumoniae, Pseudomonas aeruginosa, Proteus mirabilis, Acinetobacter species Aeromonas hydrophila

Citrobacter diversus Citrobacter freundii Enterobacter cloacae Hafnia alvei Klebsiella oxytoca Morganella morganii

Pasteurella multocida Proteus vulgaris Salmonella species Serratia marcescens Shigella species

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

Device Description

The principle of the VITEK 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh " and Gerlach (2). The VITEK 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (3)

Each VITEK 2 AST card contains 64 wells. A control well which contains only microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial isolate to be tested is diluted to a standardized concentration with 0.45 -0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the VITEK® 2 Gram Negative Meropenem device, based on the provided text:

Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly refers to performance requirements established by a Guidance Document, which the device is stated to meet. The specific quantitative acceptance criteria are not explicitly detailed in the provided text as ranges or thresholds, but rather the performance metrics achieved by the device are presented. The predicate device's performance is also included for comparative purposes, indicating what would be considered acceptable.

MetricAcceptance Criteria (Implicit from Guidance/Predicate)Reported Device Performance (VITEK 2 GN Meropenem)Predicate Device Performance (VITEK 2 GN Doripenem)
Essential Agreement (EA)Expected to be high (e.g., in the range of the predicate)97.6%97.4%
Category Agreement (CA)Expected to be high (e.g., in the range of the predicate)96.8%97.9%
Reproducibility (Best-case)Expected to be high99.6%99.2%
Reproducibility (Worst-case)Expected to be high99.6%90.6%
Quality Control> 20 tests in range> 20 tests in range> 20 tests in range
Meets Guidance Document Performance RequirementsYesYesYes

2. Sample Size Used for the Test Set and Data Provenance:

The document states "Clinical studies were performed and demonstrate acceptable performance," but it does not provide the specific sample size for the test set used in the "Clinical & Challenge Performance Data."

The data provenance (e.g., country of origin of the data, retrospective or prospective) is not explicitly stated in the provided text. It can be inferred that these were controlled laboratory studies (clinical and challenge) rather than broad population studies.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document does not provide information on the number of experts used or their qualifications to establish the ground truth for the test set.

4. Adjudication Method for the Test Set:

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed as this device is an automated antimicrobial susceptibility test (AST) system. It determines in vitro susceptibility directly and does not involve human readers interpreting results for a "case" in the traditional sense. The "comparison" is against a standard reference method rather than human interpretation.

6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance):

Yes, a standalone performance study was done. The VITEK® 2 system is an automated device designed to determine antimicrobial susceptibility directly from bacterial isolates. The reported performance metrics (EA, CA, Reproducibility) are for the device (algorithm) itself against a reference method, without human intervention as part of the primary diagnostic output. The VITEK 2 Compact has a manual filling, sealing, and loading operation, but the monitoring and result generation are automated.

7. Type of Ground Truth Used:

The ground truth for antimicrobial susceptibility testing is typically established using a standard reference method, such as broth microdilution or agar dilution as defined by CLSI (Clinical and Laboratory Standards Institute) guidelines. The document mentions that "The principle of the VITEK 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh ' and Gerlach (2)." It also refers to "Equivalent Standard Method Concentration." This strongly indicates that a standard microbiological reference method was used to establish the ground truth.

8. Sample Size for the Training Set:

The sample size for the training set is not explicitly stated in the provided text.

9. How the Ground Truth for the Training Set Was Established:

The document does not explicitly describe how the ground truth for the training set was established. However, given the nature of AST devices, it would invariably involve testing bacterial isolates against a standard reference method (as described in point 7) to generate the MIC values and interpretive categories, which would then be used to train and validate the device's algorithms.

{0}------------------------------------------------

K09/899

chromID ™ MRSA Agar

Traditional 510(k) Submission

SECTION 2: EXECUTIVE SUMMARY

AUG 1 4 2009 -

Intended Use:

VITEK® 2 Gram Negative Meropenem is designed for antimicrobial susceptibility testing of gramnegative microorqanisms and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Meropenem is a quantitative test. Meropenem has been shown to be active against most isolates of the following microorganisms listed below according to the FDA label for the antimicrobial.

Active in vitro and in clinical infections:

Escherichia coli Klebsiella pneumoniae, Pseudomonas aeruginosa, Proteus mirabilis, Acinetobacter species Aeromonas hydrophila

Citrobacter diversus Citrobacter freundii Enterobacter cloacae Hafnia alvei Klebsiella oxytoca Morganella morganii

Pasteurella multocida Proteus vulgaris Salmonella species Serratia marcescens Shigella species

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

Device Description:

The principle of the VITEK 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh " and Gerlach (2). The VITEK 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (3)

Each VITEK 2 AST card contains 64 wells. A control well which contains only microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial isolate to be tested is diluted to a standardized concentration with 0.45 -0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

The MIC ranges, interpretive criteria and equivalent concentrations are as follows:

VITEK 2 AST-GNMeropenemEquivalent StandardMethod Concentrationby Efficacy in µg/mlMIC Ranges and FDA InterpretiveCriteria.MIC in µg/ml *
SIR
Enterobacteriaceae,Acinetobacter spp.,Pseudomonasaeruginosa0.5, 2, 6, 12≤ 0.25 - 48≥ 16
  • FDA category interpretation in bold-faced type.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the word "BIOMÉRIEUX" in a bold, sans-serif font. The letters are capitalized and evenly spaced. Above the text, there is a black circle with some lines coming out of it, possibly part of a logo.

Traditional 510(k) Submission

  • S = Susceptible: Attainable levels in blood or tissue on usual usage, including oral administration when applicable.
  • I = Intermediate: "The "Intermediate" category implies clinical efficacy in body sites where the drugs are physiologically concentrated (e.g., quinolones and B-lactams in urine), or when a higher than normal dosage of drug can be used (e.g., B-lactams). The "Intermediate" category also includes a "buffer zone" which should prevent small, uncontrolled, technical factors from causing maior discrepancies in interpretations, especially for drugs with narrow pharmacotoxicity margins."(4)
  • R = Resistant to usually achievable systemic concentrations.

Device Comparison Table:

The similarities and differences of the VITEK 2 Gram Negative Meropenem when compared to the predicate device, VITEK 2 Gram Negative Doripenem (K082346), are described in the following table.

ItemDevice:VITEK 2 Gram NegativeMeropenemPredicate:VITEK 2 Gram Negative Doripenem(K082346)
Similarities
Intended UseVITEK® 2 Gram Negative Meropenem is designed for antimicrobial susceptibility testing of Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa, Proteus mirabilis, Acinetobacter spp., Aeromonas hydrophila, Citrobacter diversus, Citrobacter freundii, Enterobacter cloacae, Hafnia alvei, Klebsiella oxytoca, Morganella morganii, Pasteurella multocida, Proteus vulgaris, Salmonella species, Serratia marcescens, and Shigella species. VITEK 2 Gram Negative Meropenem is a quantitative test. It is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.VITEK® 2 Gram Negative Doripenem is designed for antimicrobial susceptibility testing of Acinetobacter baumannii, Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa, Citrobacter freundii, Enterobacter cloacae, Enterobacter aerogenes, Klebsiella oxytoca, Serratia marcescens. VITEK 2 Gram Negative Doripenem is a quantitative test. It is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
Test methodAutomated Gram negative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems (VITEK 2 Systems) to determine the in vitro susceptibility of gram-negative bacilli.Same
InoculumSaline suspension of bacteriaSame

{2}------------------------------------------------

chromID™ MRSA Agar

Traditional 510(k) Submission

ItemDevice:VITEK 2 Gram NegativeMeropenemPredicate:VITEK 2 Gram Negative Doripenem(K082346)
Differences
AntimicrobialConcentrations of theantimicrobial in the test wells ofthe VITEK 2 AST card and theanalysis algorithms are uniquefor each specific antimicrobial.Same

Discussion:

віом É

RIE U X

Both devices have similar Intended Use statements. The technological characteristics are similar but not identical. Both devices are designed for susceptibility testing of gram-negative bacilli on the VITEK 2 Systems and their technological characteristics are similar. The tests use the same base broth and are intended for use on the same VITEK 2 gram negative AST cards. The difference between the new device and the predicate device is related to the antibiotic. Concentrations of the antibiotic in the test wells and the analysis algorithms are unique for each specific antibiotic. In all other respects the two devices are identical.

The safety and effectiveness of VITEK 2 GN Meropenem is not impacted by the technology differences. Clinical studies were performed and demonstrate acceptable performance of VITEK 2 GN Meropenem. Performance is described in the following table.

ItemDevice:VITEK 2 Gram NegativeMeropenemPredicate:VITEK 2 Gram NegativeDoripenem (K082346)
Performance
Clinical & ChallengePerformance DataEA = 97.6%CA = 96.8%EA = 97.4%CA = 97.9%
ReproducibilityBest-case 99.6%Worst-case 99.6%Best-case 99.2%Worst-case 90.6%
Quality Control> 20 tests in range> 20 tests in range
Meets Guidance DocumentPerformance RequirementsYesYes

Conclusion:

The analytical and clinical performance data presented in this submission support a substantial equivalence decision. The VITEK 2 Gram Negative Meropenem is substantially equivalent to the VITEK 2 Gram Negative Doripenem (K082346).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures, possibly representing health, well-being, and human services.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

Ms. Nancy Weaver Associate Director, Regulatory Affairs bioMérieux 595 Anglum Road Hazelwood, MO 63042

AUG 1 4 2009

Re: K091899

Trade/Device Name: VITEK® 2 Gram Negative Meropenem (≤ 0.25 - ≥ 16 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Short-Term Antimicrobial Susceptibility Test System Regulatory Class: Class II Product Code: LON Dated: June 22, 2009 Received: June 25, 2009

Dear Ms Weaver:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Ms. Weaver

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally a Hynson

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):_ KDGI 899

Device Name: VITEK® 2 Gram Negative Meropenem· (≤ 0.25 - ≥ 16 µg/ml)

Indications For Use:

VITEK® 2 Gram Negative Meropenem is designed for antimicrobial susceptibility testing of gramnegative microorganisms and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Meropenem is a quantitative test. Meropenem has been shown to be active against most isolates of the following microorganisms listed below according to the FDA label for the antimicrobial.

Active in vitro and in clinical infections:

Escherichia coli Klebsiella pneumoniae. Pseudomonas aeruginosa, Proteus mirabilis, Acinetobacter species Aeromonas hydrophila

Citrobacter diversus Citrobacter freundii Enterobacter cloacae Hafnia alvei Klebsiella oxytoca Morganella morganii

Pasteurella multocida Proteus vulgaris Salmonella species Serratia marcescens Shigella species

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2
and VITEK® 2 Compact Systems for the automated quantitative or qualitative susce of isolated colonies for the most clinically significant aerobic gram-negative bacilli. Staphylococus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Fueddi hi. Pedra

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) /K091899

Page 1 of 1

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”