K Number
K211613

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-01-19

(238 days)

Product Code
Regulation Number
888.3010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STERN FIX Sternal Stabilization System is intended for closure and stabilization of the sternum following sternotomy, through the intercostal spaces, in order to promote its fusion.

Device Description

The STERN FIX is a long-term implantable postoperative sternal stabilization system that closes and stabilizes the sternum after a sternotomy. The STERN FIX is a clamping device consisting of two parts, male and female, which match telescopically with one another to form the sternal stabilization system. Both male and female parts have a curved arm that catches one of the two halves of the sternum laterally and approximates them until the sternum is securely fixed. Five sizes of the STERN FIX are available for use with different sternum thicknesses. The STERN FIX can be cut and removed for emergent, and long-term, re-entry through the sternum.

AI/ML Overview

Here's an analysis of the provided text regarding the Stern Fix Sternal Stabilization System, focusing on acceptance criteria and study details.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally framed as the device meeting "specifications" and demonstrating "adequate performance." The document does not provide a specific table of quantitative acceptance criteria with numerical targets. Instead, it offers qualitative conclusions based on performance testing.

TestAcceptance Criteria (Qualitative)Reported Device Performance
Functional Testing
Breaking strengthMeet specifications; Functionality demonstratedAll tested samples meet specifications. Functionality of the devices is demonstrated.
Fatigue testMeet specifications; Functionality demonstratedAll tested samples meet specifications. Functionality of the devices is demonstrated.
Biomechanical Testing
Breaking strengthMeet specifications; Adequate biomechanical behavior when subjected to relevant effortsAll tested samples meet specifications. The STERN FIX system has an adequate biomechanical behavior when subject to relevant efforts.
Lateral distraction testMeet specifications; Adequate biomechanical behavior when subjected to relevant effortsAll tested samples meet specifications. The STERN FIX system has an adequate biomechanical behavior when subject to relevant efforts.
Cadaver testingCorrect device implantation verified; Adequate performance and safety demonstratedCorrect devices implantation is verified in a simulated real-life situation. The devices show adequate performance and safety.

2. Sample Sizes and Data Provenance

The document does not specify the exact sample sizes used for the functional, biomechanical, or cadaveric testing. It consistently refers to "All tested samples," but the number of samples is not provided.

  • Test Set Sample Size: Not specified.
  • Data Provenance: The studies appear to be in-house or contracted laboratory testing based on the context. No country of origin is mentioned, nor is it specified if the data is retrospective or prospective. Given the nature of the testing (functional, biomechanical, cadaveric), it's inherently prospective data generation for this submission.

3. Number of Experts and Qualifications for Ground Truth of Test Set

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified.

For the cadaver testing, it mentions "Correct devices implantation is verified in a simulated real-life situation," which implies assessment by individuals with expertise in surgical procedures or device implantation, but their specific qualifications or number are not detailed. For the other tests (breaking strength, fatigue, lateral distraction), "ground truth" refers to the measured physical properties against established engineering specifications, rather than expert interpretation of medical images or observations.

4. Adjudication Method for the Test Set

Not applicable. The tests described are objective measurements of physical properties (e.g., breaking strength, fatigue resistance) or direct observation of device implantation in cadavers, not subjective interpretations requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical testing was performed to support this submission." The studies were limited to performance testing (functional, biomechanical, cadaveric) in a non-clinical setting.

6. Standalone (Algorithm Only) Performance Study

Not applicable. The Stern Fix Sternal Stabilization System is a physical medical device (an implant for sternal closure), not an AI algorithm. Therefore, "standalone (algorithm only)" performance is not relevant to this device.

7. Type of Ground Truth Used

  • Functional and Biomechanical Testing: The ground truth is based on engineering specifications and physical measurements (e.g., in a material testing machine) which determine if the device meets required strength, durability, and resistance parameters.
  • Cadaver Testing: The ground truth is the successful and correct implantation of the device in a simulated real-life situation, likely assessed by observing the implementation and stability of the device during and after placement. This would be observed performance against an expected surgical/mechanical outcome.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set was used.

{0}------------------------------------------------

February 10, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Neos Surgery S.L. % Cherita James Regulatory Consultant M Squared Associates, Inc 127 West 30th Street, 9th Floor New York, New York 10001

Re: K211613

Trade/Device Name: Stern Fix Sternal Stabilization System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: JDQ

Dear Cherita James:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 19, 2022. Specifically, FDA is updating this SE Letter to correct a typographical error in the sponsor name as an administrative correction and to include the full indications for use form.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Colin O'Neill, M.B.E., OHT6: Office of Orthopedic Devices, (301) 796-6428, colin.oneill(@fda.hhs.gov.

Sincerely,

Anne D. Talley -S
for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

January 19, 2022

Image /page/1/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

Neoss Surgery SL % Cherita James Regulatory Consultant M Squared Associates, Inc 127 West 30th Street, 9th Floor New York, New York 10001

Re: K211613/S001

Trade/Device Name: Stern Fix Sternal Stabilization System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ Dated: December 16, 2021 Received: December 20, 2021

Dear Cherita James:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{2}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anne D. Talley -S for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K211613

Device Name

STERN FIX, Sternal Stabilization System

Indications for Use (Describe)

The STERN FIX Sternal Stabilization System is intended for closure and stabilization of the sternum following sternotomy, through the intercostal spaces, in order to promote its fusion.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(K) SUMMARY

The following information is provided as required by 21 CFR § 807.87 for the STERN FIX Sternal Stabilization System 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.

  • Neos Surgery S.L. Sponsor: Parc Tecnològic del Vallès C/Ceramistes. 2 08290 Cerdanyola del Vallès Barcelona, Spain
  • Contact: Cherita James M Squared Associates. Inc. 127 West 30th Street, 9th Floor New York, New York 10001 Ph: 347-954-0624 Fax: 703-562-9797 Email: Cjames@MSquaredAssociates.com

Date of Submission: December 16, 2021

Proprietary Name: STERN FIX Sternal Stabilization System

Common Name: Cerclage, fixation

Regulatory Class: II

Regulation: 888.3010 Bone fixation cerclage

Product Codes: JDQ

Predicate Device(s): SYNTHES STERNAL ZIPFIX SYSTEM - K110789 KLS MARTIN STERNAL TALON- K051165 Reference Device:

Device Description

The STERN FIX is a long-term implantable postoperative sternal stabilization system that closes and stabilizes the sternum after a sternotomy. The STERN FIX is a clamping device consisting of two parts, male and female, which match telescopically with one another to form the sternal stabilization system. Both male and female parts have a curved arm that catches one of the two

{6}------------------------------------------------

halves of the sternum laterally and approximates them until the sternum is securely fixed. Five sizes of the STERN FIX are available for use with different sternum thicknesses. The STERN FIX can be cut and removed for emergent, and long-term, re-entry through the sternum.

Indications for Use

The STERN FIX Sternal Stabilization System is intended for closure and stabilization of the sternum following sternotomy, through the intercostal spaces, in order to promote its fusion.

Technological characteristics, comparison to predicate and reference devices.
DeviceSTERN FIXZIPFIX(Predicate device)Sternal Talon(Reference device)
CompanyNEOS Surgery S.L.SynthesKLS Martin, L.P.
510(k) No.Present submissionK110789K051165
Product CodeJDQ- Cerclage, FixationJDQ- Cerclage, FixationHRS- Plate, Fixation, Bone
Indications for UseThe STERN FIX Sternal Stabilization System is intended for closure and stabilization of the sternum following sternotomy, through the intercostal spaces, in order to promote its fusion.The Synthes Sternal ZIPFIXTM system is intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.(Description of the “Intended use” in the 510(k) Summary)The KLS-Martin Sternal Talon is intended for use in stabilization and fixation of anterior chest wall fractures including Sternal Fixation subsequent to Sternotomy and Sternal reconstructive procedures.(Description of the “Intended use” in the 510(k) Summary)
Method of Fixation to SternumPeristernally, through the intercostal spaces. Generally, 5 STERN FIX devices are recommended per median sternotomy closure, but they can be used in combination with other sternal closure systems such as wires or plates.Similar methodSimilar method
DeviceDesignThe STERN FIX deviceconsists of two parts, maleand female, which matchtelescopically with oneanother to form the fixationsystem. Both male andfemale parts have a curvedarm that catches laterally oneof the two halves of thesternum and approximatesthem until the sternum issecurely fixed. Both partspresent a set of teeth that,together, form a ratchetmechanism that allows theirmovement in the closingdirection and, at the sametime, impedes their backwardmovement. The tightening ofboth implant parts (male andfemale) is carried out usingthe specific instrumentsupplied with the product.The STERN FIX can be cutand removed for emergent,and long-term, re-entrythrough the sternum.Similar designSimilar design
Range ofsternumthicknessesand widthscoveredThickness: From 9.5 to 17mmWidth: From 17 to 38 mmSimilarsimilar
AccessoriesFour instruments are used toimplant the device:• A depth gauge (SCI0000)to select the appropriateSTERN FIX size.• Forceps (SCI0501) totighten and adjust theSTERN FIX during itsimplantation.• A cutter (SCI0200) to cutthe excess segment of themale part or cut the devicesafely in case of need ofdevice removal.• A retractor (SCI0300) toease the introduction ofthe STERN FIX in theintercostal spaces.similar toolssimilar tools
MaterialCompositionPEEK- OPTIMA™ LT3,according to ASTM F2026similar materialsSimilar materials
SizesFive sizes that allowimplantation in sternal bonesof all thicknesses considered(from 9.5 mm to 17 mm). Itslength can be adjusted to fiteach sternal bone width bycutting the excess segment ofthe male part.Similar sizesSimilar sizes
Implant lifeLong-term implantLong-term implantLong-term implant
SterilityProvided sterile. E-beamsimilarsimilar

{7}------------------------------------------------

{8}------------------------------------------------

Technological Characteristics and Substantial Equivalence

The STERN FIX is substantially equivalent to its predicate device with regards to indications for use, technological characteristics, principles of operation and performance. The technical differences found between the STERN FIX and its predicate do not raise different questions of safety and effectiveness. These differences have been discussed and justified with the use of the reference device.

Performance testing

Performance testing was conducted to characterize the capacity of the STERN FIX, as a sternal stabilization system, to withstand the forces that act upon the sternum and the whole fixation system once the products have been implanted. The results of the tests confirm the adequate performance of the device, and that it is able to achieve its intended use.

Performance testing has demonstrated that the technological characteristics of STERN FIX do not raise any new safety or effectiveness issues and that the STERN FIX will perform as intended in the clinical setting. A summary of the tests performed follows:

TestConclusions
Functional testing
Breaking strengthAll tested samples meet the specifications. Functionalityof the devices is demonstrated.
Fatigue testAll tested samples meet the specifications. Functionalityof the devices is demonstrated.
Biomechanical testing

{9}------------------------------------------------

Breaking strengthAll tested samples meet the specifications. The STERNFIX system has an adequate biomechanical behaviorwhen subject to relevant efforts.
Lateral distraction testAll tested samples meet the specifications. The STERNFIX system has an adequate biomechanical behaviorwhen subject to relevant efforts.
Cadaver testingCorrect devices implantation is verified in a simulatedreal-life situation. The devices show adequateperformance and safety.

Conformance to Standards

The STERN FIX Sternal Stabilization System is in conformance to the following recognized consensus standards: and the comments of the comments of the comments of the comments of the comments of the comments of the contraction of the contraction of the contribution of the contribution

StandardRecognitionNumber
ISO 14971 Second edition 2007-03-01: Medical devices - Application ofrisk management to medical devices5-40
ISO 14630 Fourth edition 2012-12-01: Non-active surgical implants --General requirements11-254
ISO 11607-1 Second edition 2019-02: Packaging for terminally sterilizedmedical devices - Part 1: Requirements for materials, sterile barriersystems and packaging systems14-530
ISO 11607-2 Second edition 2019-02: Packaging for terminally sterilizedmedical devices - Part 2: Validation requirements for forming, sealingand assembly processes14-531
IEC 62366-1 Edition 1.0 2015-02: Medical devices - Part 1: Applicationof usability engineering to medical devices [Including CORRIGENDUM1 (2016)]5-114
ISO 11137-1 First edition 2006-04-15: Sterilization of health careproducts - Radiation - Part 1: Requirements for development, validationand routine control of a sterilization process for medical devices[Including: Amendment 1 (2013) and Amendment 2 (2018)]14-528
StandardRecognitionNumber
ISO 11137-2 Third edition 2013-06-01: Sterilization of health careproducts - Radiation - Part 2: Establishing the sterilization dose14-409
ISO 11737-1 Third edition 2018-01: Sterilization of health care products- Microbiological methods - Part 1: Determination of a population ofmicroorganisms on product14-514
ISO 11737-2 Third edition 2019-12: Sterilization of medical devices -Microbiological methods - Part 2: Tests of sterility performed in thedefinition, validation and maintenance of a sterilization process14-540
ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medicaldevices - Part 1: Evaluation and testing within a risk management process2-258
ISO 10993-5 Third edition 2009-06-01: Biological evaluation of medicaldevices - Part 5: Tests for in vitro cytotoxicity2-245
ISO 15223-1 Third Edition 2016-11-01: Medical devices - Symbols to beused with medical device labels, labelling, and information to be supplied- Part 1: General requirements5-117
ISO 17665-1 First edition 2006-08-15: Sterilization of health careproducts - Moist heat - Part 1: Requirements for the development,validation and routine control of a sterilization process for medicaldevices14-333
ISO 17664 Second edition 2017-10: Processing of health care products -Information to be provided by the medical device manufacturer for theprocessing of medical devices14-515
ASTM F2026-17: Standard Specification for Polyetheretherketone(PEEK) Polymers for Surgical Implant Applications8-475
ASTM F2503-20: Standard Practice for Marking Medical Devices andOther Items for Safety in the Magnetic Resonance Environment8-528

{10}------------------------------------------------

{11}------------------------------------------------

Discussion of clinical testing

No clinical testing was performed to support this submission.

Conclusion

Based on the design features, the use of an established well-known biocompatible material, the intended use, the technological characteristics and the principles of operation, the subject device has demonstrated substantial equivalence to the identified legally marketed predicate device. The technical differences do not raise new issues of safety or effectiveness, based on the comparison with the identified legally marketed reference device.

The results of the performance testing demonstrate that the subject device meets its specifications and fulfills its intended use.

§ 888.3010 Bone fixation cerclage.

(a)
Identification. A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.(b)
Classification. Class II.