K Number
K211552

Validate with FDA (Live)

Device Name
Surgical mask
Date Cleared
2021-08-11

(84 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, and provided non-sterile.

Device Description

The Surgical Mask is a non-sterile, single use, and flat pleated mask with ear loops and nose clip. The product is manufactured with three layers, the inner layer is made of white polypropylene spunbond, the outer layer is made of black polypropylene spunbond (a common colorant named carbon black was added) and the middle layer is made of melt blown polypropylene filter. The elastic ear loops are not made with natural rubber latex. The nose clip in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable polyethylene wire. The mask will be provided in black and white. The colour of outer layer is black. The colour of inner layer is white.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Surgical Mask (K211552):

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicReference for Test MethodAcceptance CriteriaReported Device Performance
Fluid ResistanceASTM F1862/F1862M-1729 out of 32 pass at 160 mmHgPass
Particulate FiltrationASTM F2299-03≥ 98%Pass
Bacterial FiltrationASTM F2101-19≥ 98%Pass
Differential PressureEN 14683:2019< 6 mmH2O/cm²Pass
Flammability16 CFR 1610Class IPass
In vitro CytotoxicityISO 10993-5:2009No potential toxicity to L-929 cellsPassed
IrritationISO 10993-10:2010No skin irritation in rabbit skinPassed
Skin SensitizationISO 10993-10:2010No significant evidence of causing skin sensitization in guinea pigPassed

2. Sample size used for the test set and the data provenance:

  • Fluid Resistance (ASTM F1862/F1862M-17): The acceptance criteria states "29 out of 32 pass", implying a sample size of 32 units were tested for fluid resistance.
  • Other performance tests (Particulate Filtration, Bacterial Filtration, Differential Pressure, Flammability): The document does not explicitly state the sample sizes for these tests, only that the device "Passed" the criteria. Typically, standards like those referenced (ASTM F2299, ASTM F2101, EN 14683, 16 CFR 1610) would specify minimum sample sizes.
  • Biocompatibility (ISO 10993-5, ISO 10993-10): The document does not explicitly state the sample size for the biocompatibility tests (in vitro cytotoxicity, irritation, skin sensitization). However, the results are presented as having been performed on the "test article extract" and "applied sample," which refers to the surgical mask material.
  • Data Provenance: The document does not specify the country of origin of the data or whether the tests were retrospective or prospective. It only states that "Performance testing including biocompatibility evaluation was performed on the final finished device" and refers to the device being manufactured in China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This document describes the testing of a medical device (surgical mask) against established performance standards and biocompatibility guidelines. These tests are typically conducted in accredited laboratories by qualified technicians and scientists according to the specified protocols. There is no mention of "experts" in the sense of medical professionals diagnosing or interpreting results (like radiologists for imaging studies) or establishing a subjective "ground truth" through consensus. The ground truth for these tests is defined by the objective measurement criteria and methods specified in the referenced ASTM, ISO, EN, and CFR standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

Not applicable. This device is a medical product (surgical mask) whose performance is evaluated through objective physical and biological tests against predetermined technical standards, not through subjective interpretation by experts that would require an adjudication method.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This document pertains to the regulatory clearance of a surgical mask, not an AI-powered diagnostic or assistive device. Therefore, no MRMC study or AI-related comparative effectiveness analysis was performed or is relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a medical product (surgical mask), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the device's performance is established by the objective, quantitative measurements and qualitative observations obtained through validated test methods outlined in the referenced standards (e.g., ASTM F1862 for fluid resistance, ASTM F2101 for bacterial filtration, ISO 10993 for biocompatibility). These standards define the acceptable performance characteristics for a surgical mask.

8. The sample size for the training set:

Not applicable. This document describes the testing and regulatory submission for a physical medical device (surgical mask), not a machine learning model. Therefore, there is no "training set."

9. How the ground truth for the training set was established:

Not applicable. As there is no "training set" for this physical medical device, there is no ground truth established for one.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

August 11, 2021

Shandong Shengquan New Materials Co., Ltd. % Daniel Qiu Project Manager Shanghai Qisheng Business Consulting Co., Ltd. Room 1301, Bld 46, Jing Gu Zhong Rd. No.58, Min Hang District Shanghai, Shanghai 200240 China

Re: K211552

Trade/Device Name: Surgical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 19, 2021 Received: May 19, 2021

Dear Daniel Qiu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211552

Device Name Surgical mask

Indications for Use (Describe)

The surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, and provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K211552

I. Submitter

Shandong Shengquan New Materials Co.,Ltd. Diaozhen Industrial Development Zone, Zhangqiu District, Jinan, Shandong Province, China.

Telephone: +86-13864136816 Fax: 0086-400-777-8118

Contact person: Xiumei Zhang Date prepared: August 03,2021

II. Device

Name of Device: Surgical mask Model name: SMDP20608 Classification Name: surgical apparel (21 CFR 878.4040) Regulatory Class: II Product Code: FXX

III. Predicative device

510(k) Number: K201629 Name of Device: Medical Face Mask Classification Name: surgical apparel (21 CFR 878.4040) Regulatory Class: II Product code: FXX

IV. Device description

The Surgical Mask is a non-sterile, single use, and flat pleated mask with ear loops and nose clip. The product is manufactured with three layers, the inner layer is made of white polypropylene spunbond, the outer layer is made of black polypropylene spunbond (a common colorant named carbon black was added) and the middle layer is made of melt blown polypropylene filter. The elastic ear loops are not made with natural rubber latex. The nose clip in the layers of facemask is to allow the user

{4}------------------------------------------------

to fit the facemask around their nose, which is made of malleable polyethylene wire. The mask will be provided in black and white. The colour of outer layer is black. The colour of inner layer is white.

V. Indication for use

The surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, and provided non-sterile.

ItemSubject devicePredicate deviceDiscussion
Product NamewSurgical MaskMedial Face Mask
ManufacturerShandong Shengquan NewMaterials Co., Ltd.Shandong ShengquanNew Material Co., Ltd.N/A
510(k) numberK211552K201629N/A
Product codeFXXFXXSame
Regulation No.21 CFR878.404021 CFR878.4040Same
ClassClass IIClass IISame
Intended useThe surgical masks areintended to be worn toprotect both the patientand healthcare personnelfrom transfer ofmicroorganisms, bodyfluids and particulatematerial. These surgicalmasks are intended for usein infection controlpractices to reduce thepotential exposure toblood and body fluids. Thisis a single use, disposabledevice, and provided non-sterile.The medical face Masksare intended to be wornto protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids and particulatematerial. These facemasks are intended foruse in infection controlpractices to reduce thepotential exposure toblood and body fluids.This is a single use,disposable device, andprovided non-sterile.Same
ModelEar Loops. Flat Pleated. 3Ear Loops. Flat Pleated. 3Same

VI. Comparison of technological characteristics with the predicate devices

{5}------------------------------------------------

layerslayers
MaterialOuter layerPolypropylene spunbondSpunbond polypropyleneSame
MiddlelayerMelt blown polypropylenefilterMelt blown polypropylenefilterSame
Inner layerPolypropylene spunbondSpunbond polypropyleneSame
Nose clipMalleable polyethyleneMalleable polyethyleneSame
Ear loopsSpandex, PolyesterSpandex, PolyesterSame
ColorBlack and WhiteWhiteDifferent
Dimension (Length)$17.5cm \pm 0.88cm$$17.5cm\pm0.88cm$Same
Dimension (Width)$9.5cm\pm0.48cm$$9.5cm\pm0.48cm$Same
OTC useYesYesSame
SterilityNon-sterileNon-sterileSame
Single-useYesYesSame
ASTM F2100 LevelLevel 3Level 2Different
BiocompatibilityIn-vitrocytotoxicityConform to ISO 10993-5:2009Conform to ISO 10993-05:2009Same
IrritationConform to ISO 10993-10:2010Conform to ISO 10993-10:2010Same
SkinsensitizationConform to ISO 10993-10:2010Conform to ISO 10993-10:2010Same

The difference in the color and ASTM F2100 level does not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation was performed on the final finished device.

VII. Summary of non-clinical testing

The test results demonstrated that the proposed device complies with the following standards:

  • . ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.
  • . ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
  • ASTM F2100-19, Standard Specification for Performance of Materials Used In . Medical Face Masks.
  • . ASTM F1862M-17, Standard Test Method For Resistance Of Medical Face Masks To Penetration By Synthetic Blood (Horizontal Projection Of Fixed

{6}------------------------------------------------

Volume At A Known Velocity).

  • EN 14683:2019, Medical Face Mask-Requirements and Test Methods.
  • . ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus.
  • . ASTM F2299-03, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres.
  • 16 CFR 1610, Standard for the Flammability of clothing textiles. .
PerformanceCharacteristicsReferences for TestMethodAcceptance CriteriaResults
Fluid ResistancePerformance(mmHg)ASTMF1862/F1862M-1729 out of 32 pass at160 mmHgPass
Particulate FiltrationEfficiency (%)ASTM F2299-03≥ 98%Pass
Bacterial FiltrationEfficiency (%)ASTM F2101-19≥ 98%Pass
Differential Pressure(Delta-P) Test (mmH2O/cm2)EN 14683:2019< 6Pass
Flammability16 CFR 1610Class IPass

Table.1 Performance testing

Table.2 Biocompatibility Testing

ltemTest methodResults
In vitrocytotoxicityISO 10993-5:2009Under the conditions of this study, the test articleextract did not show potential toxicity to L-929 cells.
IrritationISO 10993-10:2010The test result showed that the extract of appliedsample did not induce skin irritation in rabbit skin.
SkinsensitizationISO 10993-10:2010Under the conditions of this study, the test articleextract showed no significant evidence of causing skinsensitization in the guinea pig.

{7}------------------------------------------------

VIII. Summary of clinical testing

Clinical testing was not required to demonstrate the substantial equivalence of Shengquan Surgical Mask to its predicate device.

IX Conclusion

The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.