K Number
K211516

Validate with FDA (Live)

Date Cleared
2021-08-24

(99 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device, "Disposable Vinyl Nitrile Synthetic Gloves Powder Free" (K211516). This document details the device's characteristics and its equivalence to a predicate device, as submitted to the FDA.

Based on the nature of the device (disposable patient examination gloves) and the content of the summary, the "device" in question is not an AI/ML-driven diagnostic or assistive technology. Instead, it is a physical medical device. Therefore, the questions related to AI/ML device performance metrics (such as MRMC studies, standalone algorithm performance, number of experts for ground truth, training set size, etc.) are not applicable to this document.

The acceptance criteria and performance are related to the physical properties, safety, and regulatory compliance of the gloves.

Here's the information that can be extracted from the provided text, related to the device's acceptance criteria and how it meets them:

1. A table of acceptance criteria and the reported device performance

Test MethodologyPurposeAcceptance CriteriaReported Device PerformanceResult
ASTM D5250Physical DimensionsLength: ≥230 mm Width (S): 85±5 mm Width (M): 95±5 mm Width (L): 105±5 mm Width (XL): 115±5 mm Thickness (Finger): ≥0.08 mm Thickness (Palm): ≥0.08 mmLength: >230 mm Width (S): 88-90 mm Width (M): 95-97 mm Width (L): 106-108 mm Width (XL): 113-116 mm Finger Thickness: 0.08-0.10 mm Palm Thickness: 0.09 mmPass
ASTM D5151WatertightnessMeet the requirements of ASTM D5151 AQL 2.50/125 leaksPass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0 mg0.11 mg/glovePass
ASTM D412Physical properties (Before Aging)Tensile Strength: ≥11 MPa Ultimate Elongation: ≥300%Tensile Strength: 15.2-17.8 MPa Ultimate Elongation: 417-606%Pass
ASTM D412Physical properties (After Aging)Tensile Strength: ≥11 MPa Ultimate Elongation: ≥300%Tensile Strength: 12.4-16.9 MPa Ultimate Elongation: 370-568%Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells.Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant.Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer.Pass

2. Sample size used for the test set and the data provenance

  • Sample Size for Physical/Performance Tests: The specific sample sizes for each physical test (e.g., number of gloves tested for dimensions, powder content, or tensile strength) are not explicitly stated for all tests. For the watertightness test (ASTM D5151), the result "0/125 leaks" implies a sample size of 125 units were tested.
  • Data Provenance: The document does not specify the country of origin for the test data for the subject device beyond the manufacturer being in China. The tests are "non-clinical" (laboratory-based physical and biocompatibility testing), not human clinical studies. The data would be prospective as it's generated specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is not an AI/ML device requiring expert adjudication/ground truth for image interpretation or diagnosis. The "ground truth" for these physical and biocompatibility tests is based on established laboratory testing methodologies and international standards (ASTM, ISO).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically for human interpretation of data in studies, not for standardized physical/chemical/biocompatibility testing of a physical product.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device. Performance is of the physical glove itself, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance is based on established international standards and laboratory testing protocols (ASTM D5250, ASTM D6124, ASTM D5151, ASTM D412, ISO 10993-5, ISO 10993-10). These standards define the methodologies and acceptance criteria for physical, chemical, and biological properties of examination gloves.

8. The sample size for the training set

Not applicable. There is no "training set" as this is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 24, 2021

Tangshan Hongyun Plastic Products Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161,Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K211516

Trade/Device Name: Disposable Vinyl Nitrile Synthetic Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: May 13, 2021 Received: May 17, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray III, PhD. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211516

Device Name

Disposable Vinyl Nitrile Synthetic Gloves Powder Free

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K211516)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Tangshan Hongyun Plastic Products Co., Ltd. Name: Address: South Shenggezhuang Village,Pachigang Town, Luannan,Tangshan, Hebei 063500, China. Phone Number: +86-13933365259 Contact: Suying Le Date of Preparation: 08/11/2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Disposable Vinyl Nitrile Synthetic Gloves Powder Free Common name: Vinyl Patient Examination Glove Classification name: Non-powdered Patient Examination Glove Model(s): S, M, L, XL

3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Zibo Huiying Medical Products, Co. Ltd. Device: Synmax Synthetic Patient Examination Vinyl Gloves,Powder Free,Blue 510(k) number: K153028

{4}------------------------------------------------

5.0 Indication for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

ItemSubject devicePredicate deviceComparison
510(k) numberK211516K153028/
Product CodeLYZLYZSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseA patient examinationglove is a disposabledevice intended formedical purposes that isworn upon theexaminer's hands toprevent contaminationbetween patient andexaminer.A patient examinationglove is a disposabledevice intended formedical purposes thatis worn upon theexaminer's hands orfingers to preventcontamination betweenpatient and examiner.Same
Powdered or Powered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
Labeling InformationSingle use, powder free,device color, devicename, glove size andquantity, product name,Non-SterileSingle use, powderfree, device color,device name, glovesize and quantity,product name, Non-SterileSimilar

Table1-General Comparison

Table2 Device Dimensions Comparison

PredicateDevice(K153028)DesignationSizeTolerance
Length, mmAverage over 234 on M size-

{5}------------------------------------------------

Width, mmAverage over 96 on M size
Thickness, mm:
FingerAverage 0.98-
PalmAverage 0.096-
Subject Device(K211516)DesignationSizeTolerance
SMLXL
Length, mm230230230230min
Width, mm8595105115±5
Thickness, mm:
Finger0.08min
Palm0.08min
RemarkSimilar

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250.

ItemSubject devicePredicate deviceComparison
ColorantBlueBlueSame
PhysicalPropertiesBeforeAgingTensileStrength11MPa, minAverage 16.9MPaDifferent
UltimateElongation300%minAverage 550%Different
AfterAgingTensileStrength11MPa, minAverage 14.4MPa, minDifferent
UltimateElongation300%minAverage 500%Different
Comply with ASTM D5250Comply with ASTM D5250Comply with ASTM D5250Same
Freedom from HolesBe free from holeswhen tested inaccordance withASTM D5151AQL=2.5Be free from holes whentested in accordance withASTM D5151 AQL=2.5Same
Powder Content0.11 mg per glove.Meet therequirements ofASTM D6124Meet the requirements ofASTM D6124Similar

Table3 Performance Comparison

Analysis: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250.

Table4 Safety Comparison

ItemSubject devicePredicate deviceComparison
MaterialPoly Vinyl ChloridePolyurethanePoly Vinyl ChloridePolyurethaneSimilar

{6}------------------------------------------------

BiocompatibilityIrritation
NitrileDi-(2-ethylhexyl)Terephthalate(DOTP)Diisononyl Phthalate(DINP)
IrritationUnder the conditions of thestudy, not an irritantComply withISO10993-10Same
SensitizationUnder conditions of thestudy, not a sensitizer.
CytotoxicityUnder conditions of thestudy, did not show potentialtoxicity to L-929 cells./Similar
Label and LabelingMeet FDA's RequirementMeetFDA's RequirementSame

Analysis: The materials of the subject device are little different with that of the predicate, but they all meet the performance requirements of ASTM D5250,also biocompatibility test has been performed on subject device and the test result can meet the requirements of ISO 10993 standards.

8.0 Summary of Non-clinical Testing

The subject device were evaluated according to the following standards:

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTMD5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

Biocompatibility Testing:

The biocompatibility evaluation for the subject device were evaluated

according to the following standard:

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

Biocompatibility testing including cytotoxicity test, sensitization test and irritation test according to ISO 10993-1 standards, have been conducted on the Disposable Vinyl Nitrile Synthetic Gloves Powder Free. Under the parameters

{7}------------------------------------------------

of the tests it is concluded that they are biocompatible, and that there are no new issues of biocompatibility regarding their use as intended.

Test MethodologyPurposeAcceptance CriteriaResults
ASTM D5250Physical Dimensions TestLength(mm):≥230;Length: >230
Width(mm):Width:
S: 85±5; M: 95±5; L: 105±5; XL: 115±5S: 88-90M: 95-97L: 106-108XL: 113-116Pass
Thickness (mm) :Finger: 0.08-0.10
Finger: ≥0.08Palm: ≥0.08Palm: 0.09Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 leaksPass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.11 mg/Pass
ASTM D412Physical propertiesBefore Aging Tensile Strength15.2-17.8Pass
Ultimate Elongation ≥300%417-606Pass
After Aging Tensile Strength ≥11MPaUltimate Elongation ≥300%12.4-16.9Pass370-568Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells.Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer.Pass
study, not an irritant.Pass

Table 5: Summary of Non-clinical Testing Table

{8}------------------------------------------------

9.0 Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device in K153028.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.