K Number
K211513

Validate with FDA (Live)

Date Cleared
2021-08-18

(96 days)

Product Code
Regulation Number
890.5300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The sam X I Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation

Device Description

The sam X1 Long Duration Ultrasound Device consists of:

  • Rechargeable Power Controller and Timer
  • Ultrasound Generating Applicators
  • Single use disposable accessories
  • Ultrasonic Coupling Patch
    The Power Controller can be used to power the Applicator to generate ultrasonic energy at one frequency (3 MHz) and one power setting (0.65 W) per Applicator. In single Applicator mode, the patient can receive 0.65 W at 3 MHz. In dual Applicator mode, the patient can receive 1.3 W at 3 MHz. The Applicators are applied to the skin with onetime use Ultrasonic Coupling Patches.
    The sam X1 Long Duration Ultrasound Device contains firmware that controls the timing display and inputs from the power button to activate/deactivate the device. There is no control function of this firmware.
    The system is intended for prescription home use to apply ultrasonic energy for a long duration (1 hours) to generate deep heat within body tissues and treat selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
AI/ML Overview

This document is a 510(k) Premarket Notification from ZetrOZ Systems, LLC for their sam X1 Long Duration Ultrasound Device. It seeks to demonstrate substantial equivalence to a predicate device (sam 2.0 Long Duration Ultrasound System, K191568).

The document does not describe an AI/ML-based device or a study involving human readers and AI assistance for diagnostic purposes. Instead, it details a physical medical device (an ultrasonic diathermy device) used for therapeutic purposes (deep heat generation for pain relief, muscle spasms, joint contractures, and circulation increase).

Therefore, I cannot provide the information requested in the prompt, as the prompt's questions pertain to acceptance criteria and studies for AI/ML diagnostic devices, such as those that might involve:

  1. A table of acceptance criteria and reported device performance (for AI metrics like sensitivity, specificity, AUC).
  2. Sample sizes and data provenance for test sets (relevant to AI model validation).
  3. Number and qualifications of experts for ground truth establishment (critical for AI imaging diagnostics).
  4. Adjudication methods for test sets (common in AI imaging studies).
  5. MRMC comparative effectiveness studies (specifically for AI assistance to human readers).
  6. Standalone algorithm performance.
  7. Types of ground truth (e.g., pathology, outcomes data - often used in AI validation).
  8. Training set sample size and ground truth establishment (fundamental to AI development).

The provided text focuses on the physical and functional equivalence of the sam X1 device to its predecessor, particularly noting a smaller battery and a maximum 1-hour treatment duration compared to the predicate's 4 hours. The testing mentioned (power output, treatment time, battery conformance, electrical safety, usability, software, cybersecurity) are standard for medical device hardware and firmware safety and performance, not for AI model validation.

In summary, as the provided document is for a non-AI medical device, the requested information regarding AI acceptance criteria and study design is not applicable and cannot be extracted from the text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

August 18, 2021

ZetrOZ Systems, LLC Sabrina Lewis QAR Director 56 Quarry Road Trumbull, Connecticut 06611

Re: K211513

Trade/Device Name: sam X1 Long Duration Ultrasound Device Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic Diathermy Regulatory Class: Class II Product Code: PFW Dated: August 5, 2021 Received: August 6, 2021

Dear Sabrina Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K211513

Device Name sam X1 Long Duration Ultrasound Device

Indications for Use (Describe)

The sam X I Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

Device Trade Name:sam X1 Long Duration Ultrasound Device
Manufacturer:ZetrOZ Systems, LLC56 Quarry RoadTrumbull, CT 06611
Contact:Ms. Sabrina LewisQAR DirectorPhone: 888-202-9831Email: sabrina@zetroz.com
Date Prepared:5/13/2021
Regulation Number:21 CFR 890.5300
Regulation Name:Ultrasonic diathermy
Regulation Class:Class II
Product Code:PFW
Predicate Device:sam 2.0 Long Duration Ultrasound System (K191568)

Indications for Use:

The sam X1 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.

Device Description:

The sam X1 Long Duration Ultrasound Device consists of:

  • Rechargeable Power Controller and Timer
  • Ultrasound Generating Applicators ●

Single use disposable accessories

  • . Ultrasonic Coupling Patch
    The Power Controller can be used to power the Applicator to generate ultrasonic energy at one frequency (3 MHz) and one power setting (0.65 W) per Applicator. In single Applicator mode, the

{4}------------------------------------------------

patient can receive 0.65 W at 3 MHz. In dual Applicator mode, the patient can receive 1.3 W at 3 MHz. The Applicators are applied to the skin with onetime use Ultrasonic Coupling Patches.

The sam X1 Long Duration Ultrasound Device contains firmware that controls the timing display and inputs from the power button to activate/deactivate the device. There is no control function of this firmware.

The system is intended for prescription home use to apply ultrasonic energy for a long duration (1 hours) to generate deep heat within body tissues and treat selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.

Substantial Equivalence:

The sam X1 Long Duration Ultrasound Device is substantially equivalent to the predicate devices cited on the previous page with respect to indications, design, function, and performance. The purpose of this special 510(k) is to allow a smaller battery back to be used. The output and strength of the signal are identical to the predicate. The only difference in the max duration of the subject device is 1 hour where the predicate device can be used for 1 to 4 hours. However, this change does not create a new intended use or different questions of safety or effectiveness as the timeframe of use is within the use of the predicate device.

InformationZetrOZ sam X1 LongDuration UltrasoundDevice(Subject)ZetrOZsam 2.0 Long DurationUltrasound Device(K191568)Comparison(to sam 2.0)
Classification NameUltrasonic DiathermyDeviceUltrasonic DiathermyDeviceIdentical
Service TypePhysical MedicinePhysical MedicineIdentical
Classification21 CFR 890.530021 CFR 890.5300Identical
ClassIIIIIdentical
IndicationsThe sam X1 LongDuration UltrasoundDevice is intended forhome use to applyultrasonic energy togenerate deep heat withinbody tissues for thetreatment of selectedmedical conditions such asthe relief of pain, the reliefof muscle spasms, thetreatment of jointcontractures, and the localincrease in circulation.The sam 2.0 LongDuration UltrasoundDevice is intended forhome use to applyultrasonic energy togenerate deep heat withinbody tissues for thetreatment of selectedmedical conditions such asthe relief of pain, the reliefof muscle spasms, thetreatment of jointcontractures, and the localincrease in circulation.Identical
ManufacturerZetrOZZetrOZIdentical
Console/GeneratorDimensions (L x W x Hcm)3.4 cm x 3.4 cm x 1.4 cm6.10 cm L x 7.09 cm H x1.88cm WSimilarConsole/generatorsized to accommodatesmallerrechargeablebattery.
InformationZetrOZ sam X1 LongDuration UltrasoundDevice(Subject)ZetrOZsam 2.0 Long DurationUltrasound Device(K191568)Comparison(to sam 2.0)
Treatment HeadDimensions (L x W x Hcm)3.81 cm L x 3.30 cm W x1.14 cm H3.81 cm L x 3.30 cm W x1.14 cm HIdentical
Console/Generator Weight(kg)0.02 kg0.10 kgSimilarLesser weightprimarily driven byuse of smallerrechargeable battery.
Treatment Head Weight(kg)0.01 kg0.01 kgIdentical
Power Supply120/240 VAC with 5V DCInput Power Jack andLithium Battery Powered120/240 VAC with 5V DCInput Power Jack andLithium Battery PoweredIdentical
Leakage Current0.3 mA0.3 mAIdentical
Crystal MaterialLead Zirconate-TitanateLead Zirconate-TitanateIdentical
Technology of ultrasoundgeneration (e.g.,piezoelectric.magnetoconstructive)PiezoelectricPiezoelectricIdentical
Treatment Mode(s)Two discrete settings ofpower at same FrequencyTwo discrete settings ofpower at same FrequencyIdentical
Beam Type (collimated ordivergent)DivergentDivergentIdentical
Transducer Diameter (cm)5 cm5 cmIdentical
Acoustic WorkingFrequency and Accuracy(MHz)3MHz ± 20%3MHz ± 20%Identical
Effective Radiating Areaand Accuracy (cm2)One: 6 cm2Two:12 cm2± 20%One: 6 cm2Two:12 cm2± 20%Identical
Beam Nonuniformity Ratioand AccuracyBNR: <5:1 ± 20%BNR: <5:1 ± 20%Identical
Output Mode: (ContinuousWave/Amplitude -Modulated Wave)Continuous Wave - 100%duty cycleContinuous Wave - 100%duty cycleIdentical
Maximum Timer Settingand Accuracy1 Hour +/- 1 minute4 Hours +/- 1 minuteSimilarPredicate devicepermits treatmentdurations of 1, 2, 3, or4 hours.
Beam Maximum Intensityand Accuracy (W/cm2)0.132 W/cm² ± 20%0.132 W/cm² ± 20%Identical
Maximum Value of theOutput Power (RatedOutput Power) andAccuracy (W)Single Applicator:0.65W ± 20%Dual Applicator:1.3W ± 20%Single Applicator:0.65W ± 20%Dual Applicator:1.3W ± 20%Identical
Maximum Value of theEffective Intensity andAccuracy (Not to exceed 3W/cm2 *)0.264 W/cm² ± 20%0.264 W/cm² ± 20%Identical
InformationZetrOZ sam X1 LongDuration UltrasoundDevice(Subject)ZetrOZsam 2.0 Long DurationUltrasound Device(K191568)Comparison(to sam 2.0)
For Amplitude ModulatedWavesNot Amplitude ModulatedNot Amplitude ModulatedIdentical
Peak Temperature Rise vs.Time and Tissue Depth toMaximum Treatment Time(for fixed Treatment HeadPlacement) (deg C)7°C at 1 cm4°C at 3 cm2°C at 5 cmTreatment time: 1 hour7°C at 1 cm4°C at 3 cm2°C at 5 cmTreatment time: 1 hourIdentical
Maximum Patient ContactSurface Temperature ofTreatment Head underSimulated or Actual UseConditions for allOperating Conditions(Continually operated formaximum treatment time)(deg C)44 °C44 °CIdentical
Therapeutically AppliedUltrasound Coupling PatchUltrasound Coupling PatchIdentical
Applicator TypeUp to two circularApplicators with 3 MHzoutputUp to two circularApplicators with 3 MHzoutputIdentical
Applicator TypeApplicator EmittingSurface Areas (cm2)Up to two circularApplicatorsOne Applicator : 5 cm2Two Applicators :10 cm2Up to two circularApplicatorsOne Applicator : 5 cm2Two Applicators :10 cm2Identical
Coupling BandageABS Plastic withintegrated couplingmediumABS Plastic withintegrated couplingmediumIdentical
Applicator Lens MaterialTPXTPXIdentical
Environmental - OperatingTemperature Range0°C to +50°C (32°F to+122°F)0°C to +50°C (32°F to+122°F)Identical
Performance Standards21 CFR 1050.1021 CFR 1050.10Identical
SterilityNon SterileNon SterileIdentical
Designed to meet ElectricalSafety StandardsIEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-1IEC 60601-1-2IEC 60601-1-11Identical
BiocompatibilityYesYesIdentical
Mechanical safetyYesYesIdentical
Radiation safety (if notradioactive state as such)Not RadioactiveNot RadioactiveIdentical
Software/FirmwareYes - For measurement ofbattery levelYes - For measurement ofbattery levelIdentical
Output ChannelsTwo Independent PowerChannelsTwo Independent PowerChannelsIdentical

{5}------------------------------------------------

{6}------------------------------------------------

Preclinical Testing:

The sam X1 Long Duration Ultrasound Device was evaluated to ensure the smaller power control can still deliver the appropriate treatment for the duration of treatment (1 hour). Testing included:

  • Power Output Evaluation .
  • Evaluation of Total Treatment Time .

{7}------------------------------------------------

  • Battery Conformance Evaluation .
  • Electrical Safety and EMC Evaluation
  • Device Operation Validation ●
  • Powering On/Off .
  • Evaluation of Usability ●
  • Software Evaluation
  • Cybersecurity Evaluation .

Clinical Testing:

Clinical data was not necessary to support substantial equivalence.

Conclusion:

The sam X1 Long Duration Ultrasound Device possesses the same intended use and similar technological characteristics as the predicate devices. The sam X1 Long Duration Ultrasound Device is substantially equivalent to the predicate device.

N/A