K Number
K211487

Validate with FDA (Live)

Device Name
MAF Guard
Date Cleared
2021-12-08

(209 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MAF Guard is a device intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of microorganisms, body fluid and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

Device Description

The surgical face mask is a non-sterile, single use, three-layer, flat, pleated style with ear loops or tie-on strings and a nose piece. The mask is 175 ± 5 mm by 95 ± 3 mm. The outer, blue layer and the inner, white layer are spunbond polypropylene, and the middle layer consists of melt-blown polypropylene filter. One model of the mask is secured to the face with two ear loops made of polyester Spandex that are 150 ± 10 mm long. The other model is secured with four spin bond polypropylene tie-on strings that are 450 ± 10 mm long. A nosepiece of steel core wire coated with polyethylene resin that is 110 ± 5 mm long conforms the mask to the shape of the nose.

AI/ML Overview

The provided document is a 510(k) Summary for a surgical face mask (MAF Guard). It describes the non-clinical testing performed to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets acceptance criteria for an AI/algorithm-based medical device.

Therefore, most of the requested information regarding AI/algorithm performance, such as sample sizes for test/training sets, expert consensus, MRMC studies, or standalone algorithm performance, is not applicable to this document.

However, I can extract the acceptance criteria and performance data for the physical characteristics and filtration efficacy of the surgical mask based on the provided tables.

Here's the information that can be extracted from the document, focusing on the appropriate acceptance criteria and "study" (non-clinical testing) for a surgical mask:


Device: MAF Guard Surgical Face Mask (Ear Loop model SM3P-MB25-2511; Tie-on model SM3P-MB25-2521)
Device Type: Surgical Face Mask
Regulatory Class: Class II
Product Code: FXX
Predicate Device: Non Woven Face Mask (K173062)

The study described is non-clinical performance testing to demonstrate that the MAF Guard surgical face mask meets established consensus standards for surgical masks, aligning with the "Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004."

1. Table of Acceptance Criteria and the Reported Device Performance:

Test Method (Standard)PurposeAcceptance Criteria (for Level 3 Surgical Mask)Reported Device PerformanceResult
Fluid Resistance Performance (ASTM F1862)Evaluate resistance to penetration by synthetic blood, minimum pressure in mm HgLevel 3 requirement: 29 out of 32 passed at 160 mm Hg32 out of 32 passed at 160 mm Hg for three non-consecutive lotsPass
Particulate Filtration Efficiency (ASTM F2299)Evaluate Sub-micron particulate filtration efficiency at 0.1 micron (%)Level 3 requirement: ≥ 98%Mean for three non-consecutive lots = 99.88%, 99.89%, and 98.50%Pass
Bacterial Filtration Efficiency (ASTM F2101)Evaluate BFE using a biological aerosol of Staphylococcus aureus (%)Level 3 requirement: ≥ 98%Mean for three non-consecutive lots = 98.9%, 98.5%, and 98.4%Pass
Differential Pressure (Delta P) (EN 14683)Evaluate Differential pressure (mm H2O/cm²)Level 3 requirement: < 6 mm H2O/cm²Mean for three non-consecutive lots = 2.11, 2.57, and 4.60 mm H2O/cm²Pass
Flammability (16 CFR 1610)Evaluate flammabilityClass 1Class 1 for 3 non-consecutive lots of each mask modelPass
Biocompatibility (ISO 10993-5)Evaluate for CytotoxicityNon-CytotoxicNon-CytotoxicPass
Biocompatibility (ISO 10993-10)Evaluate for SensitizationNon-SensitizingNon-SensitizingPass
Biocompatibility (ISO 10993-10)Evaluate for IrritationNon-IrritatingNon-IrritatingPass

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for performance tests: The performance tests (Fluid Resistance, Particulate Filtration, Bacterial Filtration, Differential Pressure, and Flammability) were conducted on three non-consecutive lots of the device. For Fluid Resistance, 32 samples were tested per lot. The specific number of samples for other tests (e.g., for efficiency percentages or differential pressure) within each lot is not explicitly stated beyond "mean for three non-consecutive lots."
  • Data Provenance: The document does not specify the country of origin where the testing was performed. It is implicitly retrospective as it's a 510(k) submission summarizing completed testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Not applicable. This submission pertains to physical and performance specifications of a medical device (surgical face mask), not an AI/algorithm-based device requiring expert clinical ground truth evaluation. The ground truth is established by laboratory test methods and international/national standards (e.g., ASTM, EN, ISO).

4. Adjudication Method for the Test Set:

  • Not applicable. This is not a clinical study involving human readers or subjective interpretations that would require an adjudication method. The testing relies on objective measurement and adherence to defined technical standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI/algorithm-based device, and no MRMC study was performed. The device is a physical surgical face mask.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI/algorithm-based device. "Standalone performance" in this context refers to the intrinsic performance of the physical mask itself as measured by the specified tests.

7. The Type of Ground Truth Used:

  • The ground truth is established by objective laboratory testing methods defined by recognized consensus standards (e.g., ASTM F1862 for fluid resistance, ASTM F2299 for particulate filtration, ASTM F2101 for bacterial filtration, EN 14683 for differential pressure, 16 CFR 1610 for flammability, and ISO 10993 for biocompatibility). The acceptance criteria are derived from these standards, specifically aiming for Level 3 performance for surgical masks.

8. The Sample Size for the Training Set:

  • Not applicable. This is not an AI/algorithm-based device that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. This is not an AI/algorithm-based device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 8, 2021

MAF Clothing Pvt Ltd Unit-3 Thomas Knott Senior Regulatory Advisor Benjamin L. England and Associates, LLC 810 Landmark Dr. Suite 126 Glen Burnie, Maryland 21061

Re: K211487

Trade/Device Name: MAF Guard (Ear Loop model SM3P-MB25-2511; Tie-on model SM3P-MB25-2521) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 11, 2021 Received: November 15, 2021

Dear Thomas Knott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211487

Device Name

MAF Guard (Ear Loop model SM3P-MB25-2511; Tie-on model SM3P-MB25-2521)

Indications for Use (Describe)

MAF Guard is a device intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of microorganisms, body fluid and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Date of Summary Prepared: November 30, 2021 510(k) Number: K211487

Submitted By:Benjamin L. England And Associates, LLC
Address:810 Landmark DriveSuite 126Glen Burnie, MD 21061
Phone:(410) 220-2800
Fax:(443) 583-1464
E-mail:tcknott@englandlawgroup.com
Contact:Thomas C. Knott, Senior Regulatory Consultant
On Behalf of:MAF CLOTHING PVT LTD UNIT-3
AddressSy. No59/3, 59/4. Rajathadripura, Near K B CrossTiptur Taluk, Tumkur District - 572114.Karnataka, India.
Phone:+91- 9513696166
Fax:None
E-Mail:faizal@mafclothing.com / kirit@mafclothing.com
Contact:Mr. Kirit Jagani (Business Head)

1. Regulatory Information

Proprietary Name:MAF Guard (Ear Loop model SM3P-MB25-2511; Tie-onmodel SM3P-MB25-2521)
Classification Name:Surgical Face Mask, Apparel
Classification:Class II
Product Code:FXX
PanelGeneral Surgery Devices
Regulation:21 CFR 878.4040

2. Predicate Device

510(k) Number:K173062
Device Name:Non Woven Face Mask (Models: VQN0185W (earloop) andVQN0185B (ties))
Recalls:A search of the recall database revealed no design-related recalls.

{4}------------------------------------------------

3. Device Description

The surgical face mask is a non-sterile, single use, three-layer, flat, pleated style with ear loops or tie-on strings and a nose piece. The mask is 175 ± 5 mm by 95 ± 3 mm. The outer, blue layer and the inner, white layer are spunbond polypropylene, and the middle layer consists of melt-blown polypropylene filter. One model of the mask is secured to the face with two ear loops made of polyester Spandex that are 150 ± 10 mm long. The other model is secured with four spin bond polypropylene tie-on strings that are 450 ± 10 mm long. A nosepiece of steel core wire coated with polyethylene resin that is 110 ± 5 mm long conforms the mask to the shape of the nose.

4. Indications for Use

MAF Guard is a device intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of microorganisms, body fluid and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

5. Summary of Comparison and Technological Characteristics

CLASSIFICATION INFORMATION
510(k) NumberK211487K173062N/A
ManufacturerMAF Clothing Pvt. Ltd.V&Q Manufacturing CorporationN/A
ProprietaryNameMAF Guard (Models Ear Loop –SM3P-MB25-2511, Tie-on –SM3P-MB25-2521.Non Woven Face Mask (Models:VQN0185W (earloop) andVQN0185B (ties))N/A
CommonNameSurgical Face MaskSurgical Face MaskSame
Product CodeFXXFXXSame
ClassificationClass IIClass IISame
Regulation21 CFR 878.404021 CFR 878.4040Same
Indications forUseMAF Guard is a deviceintended to be worn byoperating room personnelduring surgical procedures toprotect both the surgical patientand the operation roompersonnel from transfer ofmicroorganisms, body fluid andparticulate material. These facemasks are intended for use ininfection control practices toreduce potential exposure toblood and body fluids. The facemask is single use, disposabledevice, provided non-sterile.Non Woven Face Mask (Models:VQN0185W (ear loop) andVQN0185B (ties)) is intended forsingle use by operating roompersonnel and other generalhealthcare workers to protectboth patients and healthcareworkers against transfer ofmicroorganisms, blood and bodyfluids, and particulate materials.Similar
ModelThree ply, flat pleated with earloops and ties3 Ply, Flat-Pleated Style with earloops and tiesSame

Table 1 - General Comparison

{5}------------------------------------------------

MATERIALS
Outer LayerSpunbond PolypropyleneSame
Middle LayerMelt Blown Polypropylene FilterSame
Inner LayerSpunbond PolypropyleneSame
Nose PieceSteel core wire coated with Polyethylene resinWhite aluminum strip covered by PP coveringDifferent
Ear LoopsPolyester SpandexUrethane elastic fiberDifferent
Tie strips:Spunbond polypropyleneSpunbond polypropyleneSame
ColorBlue outer layerBlueSame
Width95 ± 3 mm95 mmSame
Length175 ± 5 mm175 mmSame
OTC Use?YesYesSame
SterilityNon-SterileNon-SterileSame
Single Use?YesYesSame
ASTM F2100Level 3Level 2Different
ASTM F1862 Fluid Resistance Performance32 out of 32 pass at 160 mm Hg for each of three lotsPass at 120 mm HgDifferent
ASTM F2299 Particulate Filtration EfficiencyMean for three lots = 99.88%, 99.89%, and 98.50%, respectivelyAverage 99.74% at 0.1µmSimilar
ASTM F2101 Bacterial Filtration EfficiencyMean for three lots = 98.9%, 98.5%, and 98.4%Average 99.4%Similar
Differential Pressure (Delta P)Mean for three lots = 2.11, 2.57, and 4.60 mm H2O/cm²Average 2.7 mm H2O/cm²Similar
Flammability 16 CFR 1610Class 1Class 1Same
Biocompatibility TestingNon-Cytotoxic, Non-Sensitizing, Non-IrritatingNon-Cytotoxic, Non-Sensitizing, Non-IrritatingSame

Differences in technological characteristics do not raise different questions of safety and effectiveness.

6. Non-clinical Tests Performed on the Proposed Device

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004.

{6}------------------------------------------------

Table 2 - Performance Testing

TestMethodPurposeAcceptance CriteriaResult
ASTM 2100Meetsrequirements forLevel 3
Fluid ResistancePerformanceASTM F1862To evaluate the resistance topenetration by syntheticblood, minimum pressure inmm HgLevel 3 requirement: 29 outof 32 passed at 160 mm HgPass32 out of 32passed at 160mm Hgfor three non-consecutive lots
ParticulateFiltrationEfficiency ASTMF2299To evaluate the Sub-micronparticulate filtrationefficiency at 0.1 micron (%)Level 3 requirement: ≥ 98%PassMean for threenon-consecutivelots = 99.88%,99.89%, and98.50%
BacterialFiltrationEfficiency ASTMF2101To evaluate the BFE of thetest article using a biologicalaerosol of Staphylococcusaureus (%)Level 3 requirement: ≥ 98%PassMean for threenon-consecutivelots = 98.9%,98.5%, and98.4%
DifferentialPressure (DeltaP) EN 14683To evaluate the Differentpressure (mm H2O/cm²)Level 3 requirement: <6 mm H2O/cm²PassMean for threenon-consecutivelots = 2.11, 2.57,and 4.60 mmH2O/cm²
Flammability16 CFR 1610To evaluate the flammabilityClass 1Pass3 non-consecutive lotsof each maskmodel

{7}------------------------------------------------

Table 3 - Biocompatibility Testing

ltemProposed DeviceAcceptance CriteriaResult
ISO 10993-5Non-CytotoxicNon-CytotoxicPass
ISO 10993-10Non-SensitizingNon- SensitizingPass
ISO 10993-10Non-IrritatingNon-IrritatingPass

7. Conclusion

There is no clinical study included in this submission. The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.

END OF 510(k) SUMMARY**

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.