K Number
K211483

Validate with FDA (Live)

Device Name
SofWave System
Date Cleared
2021-11-15

(187 days)

Product Code
Regulation Number
878.4590
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SofWave System is indicated for use as a non-invasive dermatological aesthetic treatment to improve facial lines and wrinkles, lift the eyebrow, and lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions for subjects aged 22 and older.

Device Description

The SofWave System is an ultrasound system intended for aesthetic purposes. The system generates high frequency ultrasonic pulses that elevate the temperature in the dermis layer and cause controlled isolated areas of thermal damage. The SofWave System consists of two main functional components: 1) the console and 2) the applicator. The console includes the power sources, cooling unit, electrical components and the user interface. The applicator is comprised of an array of ultrasonic transducers that emit continuous acoustic waves and an active cooling element that is used to cool the skin area in contact with the applicator. The applicator is connected by a flexible cable to the console.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Indication/EndpointAcceptance Criteria (Implicit)Reported Device Performance
Eyebrow Lift
% Blinded Reviewer Identification of Pre- vs. Post-Treatment (Eyebrow)Implicitly, significant identification rate demonstrating improvement79% (53/67) identified correctly
Objective Measurement of Maximal Eyebrow Height (MEH)Implicitly, measurable average lift0.78 mm average lift
Objective Measurement of Average Eyebrow Height (AEH)Implicitly, measurable average lift0.69 mm average lift
PGAIS Assessment (% Improved)Implicitly, significant improvement rate80% (51/64) improved at 3-month follow-up
Patient Satisfaction (Eyebrow Lift)Not explicitly defined as an acceptance criterion, but reported for context55% reported improvement and satisfaction
Lax Submental and Neck Tissue Lift
% Blinded Reviewer Identification of Pre- vs. Post-Treatment (Submental/Neck)Implicitly, significant identification rate demonstrating improvement80% (60/75) identified correctly
Objective Measurement of Lax Submental and Neck Tissue Lift (Area)Implicitly, measurable average lift38 mm² average lift
PGAIS Assessment (% Improved)Implicitly, significant improvement rate85% (61/72) improved at 3-month follow-up
Patient Satisfaction (Submental/Neck Lift)Not explicitly defined as an acceptance criterion, but reported for context55% reported improvement, 50% reported satisfaction

Note: The document does not explicitly state numerical acceptance criteria for each endpoint. Instead, it reports the performance and states that "both co-primary endpoints are met," implying that the observed performance was considered sufficient for approval.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Study (Test Set):
    • 92 subjects enrolled.
    • 80 subjects treated.
    • 467 facial areas treated across 3 zones.
    • Specific numbers for each endpoint:
      • Eyebrow Lift (Zone 1): 67 subjects evaluated for blinded review, 64 for PGAIS and patient satisfaction.
      • Submental/Neck Lift (Zone 3): 75 subjects evaluated for blinded review, 72 for PGAIS and patient satisfaction.
  • Data Provenance: Prospective, conducted at 5 investigational sites in the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The text mentions "independent masked evaluators" and "blinded reviewers" for both eyebrow lift and submental/neck lift.
  • Number of Experts: Two blinded reviewers are explicitly mentioned for identifying pre- vs. post-treatment images for both eyebrow and submental/neck lift (e.g., "based on the agreement of two blinded reviewers").
  • Qualifications of Experts: Not specified in the provided text.

4. Adjudication Method for the Test Set

  • For the co-primary effectiveness endpoints related to image identification, the results are "based on the agreement of two blinded reviewers." This implies a form of consensus or 2-reader agreement was used. The specific rule (e.g., unanimous, majority) is not detailed.
  • For other assessments like PGAIS, it's "Physician Global Aesthetic Improvement Scale (PGAIS) assessment," suggesting physician evaluation, but no explicit adjudication method is stated for individual physician assessments.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted to compare human readers with and without AI assistance for this device.
  • The study was designed to evaluate the standalone performance of the SofWave System for aesthetic improvements.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • No, this device is a medical system that requires human operation. The performance described is the device's effectiveness when used on patients, not an AI algorithm's standalone performance. The "device performance" mentioned in the non-clinical section refers to functional bench testing of the system itself, not its output as an algorithm.

7. Type of Ground Truth Used

  • Clinical Ground Truth: The ground truth was established through a combination of:
    • Expert Clinical Assessment: "Independent masked evaluators'" correction identification of pre-post image sequence and "Physician Global Aesthetic Improvement Scale (PGAIS) assessment."
    • Objective Quantitative Measurement: Objective measurements of maximal eyebrow height (MEH), average eyebrow height (AEH), and lax submental and neck tissue lift (area) using "quantitative 2D imaging."
    • Patient Reported Outcomes: Patient satisfaction questionnaires.

8. Sample Size for the Training Set

  • The document does not specify a training set size. The clinical study described appears to be for "expansion of indications" and serves as the primary effectiveness and safety evaluation reported here, consistent with a pivotal study for premarket notification. This device is a physical system, not an AI algorithm requiring a specific data "training set."

9. How the Ground Truth for the Training Set Was Established

  • This question is not applicable as the document does not describe an AI algorithm that would typically require a "training set" with established ground truth in the conventional sense. The "ground truth" discussed in the performance data section refers to the clinical outcomes observed in the human subjects treated with the device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

November 15, 2021

Sofwave Medical Ltd. % Mrs. Janice Hogan, JD Partner Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103

Re: K211483

Trade/Device Name: Sof Wave System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System For Aesthetic Use Regulatory Class: Class II Product Code: OHV Dated: October 14, 2021 Received: October 14,2021

Dear Mrs. Hogan:

We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

OMB No. 0910-0120 Form Approved: Expiration Date: June 30, 2023 See PRA Statement on last page

510(k) Number (if known) K211483

Device Name

SofWave System Indications for Use (Describe)

The SofWave System is indicated for use as a non-invasive dermatological aesthetic treatment to improve facial lines and wrinkles, lift the eyebrow, and lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions for subjects aged 22 and older.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

510(k) SUMMARY Sofwave Medical's SofWave System

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Sofwave Medical Ltd. Beit Tavor 2 Yokneam Ilit, Israel 2069202

Submission Correspondent:

Mrs. Janice Hogan, JD Hogan Lovells US LLP janice.hogan@hoqanlovells.com (267) 675-4611

Date Prepared: November 9, 2021

Name of Device (Trade Name)

SofWave System

Common or Usual Name

Focused Ultrasound Stimulator System for Aesthetic Use

Classification Name

21 CFR 878.4590 (Ultrasound for Tissue Heat or Mechanical Cellular Disruption), Class II, product code OHV

Predicate Devices

Ulthera, Inc.'s Ulthera® System (K180623) (Predicate device)

Sofwave Medical's Sofacia System (K191421) (Reference device)

Indications for Use

The SofMave System is indicated for use as a non-invasive dermatological aesthetic treatment to improve facial lines and wrinkles, lift the eyebrow, and lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions for subjects aged 22 and older.

Technological Characteristics

The SofWave System is an ultrasound system intended for aesthetic purposes. The system generates high frequency ultrasonic pulses that elevate the temperature in the dermis layer and cause controlled isolated areas of thermal damage.

The SofWave System consists of two main functional components: 1) the console and 2) the applicator. The console includes the power sources, cooling unit, electrical components and

{4}------------------------------------------------

the user interface. The applicator is comprised of an array of ultrasonic transducers that emit continuous acoustic waves and an active cooling element that is used to cool the skin area in contact with the applicator. The applicator is connected by a flexible cable to the console.

Comparison of Technological Characteristics with the Predicate Device

The SofWave System has similar technological characteristics compared to its predicate device, the Ulthera System (K180623). Both the subject device and the predicate device consist of a console that includes the power sources, electrical components, user interface (touchscreen), a cart for storage of system components, and a handpiece that is connected to a transducer.

The technological characteristics of the console of the subject device and the predicate device are similar. The console of both the subject and predicate device are assembled in a mobile cart that comprises of a control unit (or personal computer (PC)) with a touchscreen monitor and Graphical User Interface (GUI). The control unit of both the subject and predicate device allow the operator to adjust the treatment settings, view the system status and receive maintenance notices, fault and safety alerts. Further, the main processor of both the subject and predicate device control and monitor all system components. The computers for both systems receive input from the user via the user interface where treatment parameters can be adjusted accordingly. During the treatment, relevant information is displayed on the screen for both the subject and predicate device such as the operating conditions, equipment activation status, treatment parameters, system messages and prompts, and ultrasound images. Further, both the subject and predicate device consist of a main board that includes all electronic circuits required for operating the system.

Both the subject device and predicate device consist of software that enables the continuous monitoring of the overall applicator and console to ensure safe usage. The software for both devices monitors various parameters including operating conditions, equipment activation status, treatment parameters, system messages and prompts.

The technological characteristics of the applicator of both the SoftWave System and the predicate device are also similar. The applicator of both the subject and predicate device are comprised of ultrasonic transducers that emit continuous acoustic waves and is connected by a flexible cable (umbilical cord) connecting the handle to the control unit (console). The thermal coagulation point of both the subject device and predicate device are similar (confined to a focal zone which is <3mm for the subject device as compared to 5mm for the predicate device).

In addition, the treatment depth of the subject device is similar to the predicate device. Although the subject device consists of a thermoelectric cooler (TEC) that maintains the epidermis at a cool temperature whereas the predicate does not have this technology, this difference does not raise new questions of safety or effectiveness as the cooling of the subject device provides additional protection to the epidermis from the spread of heat from the dermis.

The SoftVave System also has the same technological characteristics compared to its reference device, the previously cleared Sofacia System (K191421). Although the subject device has some minor modifications to the technological characteristics, these modifications do not raise new questions of safety or effectiveness because they do not change the device's principles of operation.

In sum, the subject SofWave System has the same or very similar technological characteristics compared to its predicate and reference devices, and the minor differences do not raise new types of safety and effectiveness questions.

{5}------------------------------------------------

Subject DevicePredicate DeviceReference Device
Sofwave Medical'sSofWave SystemUlthera, Inc.'s Ulthera®System(K180623)Sofwave Medical'sSofacia System(K191421)
RegulatoryClassIIIIII
CFRRegulation21 CFR 878.459021 CFR 878.459021 CFR 878.4590
ProductCodeOHVOHV, IYOOHV
Indicationsfor UseThe SofWave System isindicated for use as a non-invasive dermatologicalaesthetic treatment toimprove facial lines andwrinkles, lift the eyebrow,and lift lax submental(beneath the chin) and necktissue; which can also affectthe appearance of lax tissuein the submental and neckregions for subjects aged 22and older.The Ulthera System isindicated for use as a non-invasive dermatologicalaesthetic treatment to:• lift the eyebrow• lift lax submental(beneath the chin)and neck tissue;which can alsoaffect theappearance of laxtissue in thesubmental andneck regionsimprove lines andwrinkles of thedécolletéThe Ulthera System inconjunction with theUlthera DeepSEEtransducer allows forultrasonic visualization ofdepths up to 8 mm belowthe surface of the skin. Theindicated use of theimaging is to visualize thedermal and subdermallayers of tissue to:• ensure propercoupling of thetransducer to theskin (currentclearedindication)• confirmappropriate depthof treatment suchas to avoid boneThe Sofacia System isindicated for use as a non-invasive dermatologicalaesthetic treatment toimprove facial lines andwrinkles for subjects aged22 and older.
TreatmentDepth1-2 mm4.5 mm3 mm1-2 mm
Subject DevicePredicate DeviceReference Device
Sofwave Medical'sSofWave SystemUlthera, Inc.'s Ulthera®System(K180623)Sofwave Medical'sSofacia System(K191421)
Tissue atFocal PointTemperature60°C -70°C65°C60°C -70°C
Type ofEnergyThermal < 5J per channel orhigh intensity therapeuticultrasoundThermal < 2JThermal < 5J per channel orhigh intensity therapeuticultrasound
Systemcomponents- Console thatincludes the powersources, electricalcomponents anduser interface(touchscreen).- Handpiece thatincludes thetransducer- Cart for storage ofsystem components- Console thatincludes the powersources, electricalcomponents anduser interface(touchscreen).- Handpiece- Three differenttransducers- Cart for storage ofsystemcomponents- Console thatincludes the powersources, electricalcomponents anduser interface(touchscreen).- Handpiece thatincludes thetransducer- Cart for storage ofsystem components
ThermalCoagulationPointConfined to focal zone;shallow (&lt3 mm); no thermalcoagulation below focalzoneConfined to focal zone;shallow (&lt5 mm); nothermal coagulation belowfocal zoneConfined to focal zone;shallow (&lt3 mm); nothermal coagulation belowfocal zone
EpidermalImpactNon-invasive; CoolingrequiredNon-invasive; Cooling notrequiredNon-invasive; Coolingrequired
TransducerAcousticCoreEnergizer comprises:- Array ofpiezoelectricceramic plates (7 x5 mm²)- Temperature controlunit (thermistors,Thermoelectriccooler (TEC), HeatExchanger)Energizer comprises:- Focused ultrasonictransducer with adome-shapeddesign, moving ona mechanical axisEnergizer comprises:- Array ofpiezoelectricceramic plates (7 x5 mm²)- Temperature controlunit (thermistors,Thermoelectriccooler (TEC), HeatExchanger)
Energy Type- High Intensity- Ultrasound- Thermal- High Intensity- Ultrasound- Thermal- High Intensity- Ultrasound- Thermal
Frequency10 MHz – 12 MHz4 MHz-10 MHz10 MHz – 12 MHz
Electricalrequirements100-240 VAC; 10 A; 50-60Hz; single phase100-240 VAC; 2.5 A; 50-60Hz; single phase100-240 VAC; 10 A; 50-60Hz; single phase
TreatmentArea Width35mm25mm35mm

{6}------------------------------------------------

Performance Data

The following nonclinical performance testing has been conducted to support the substantial equivalence of the SofWave System to its predicate device, consistent with FDA's "Class II

{7}------------------------------------------------

Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use" (2011). In all instances, the SofWave System functioned as intended.

  • Biocompatibility of the patient-contacting components of the device was . established in accordance with ISO 10993
  • Software verification and validation was performed, and demonstrated that the software performs as intended
  • Electrical Safety and Electromaqnetic Compatibility was established in accordance ● with IEC 60601-1-2: 2014, IEC 60601-1: 2005 (R) 2012 , IEC 60601-1-6: 2010 (R) 2013, and IEC 60601-2-62: 2013
  • . Ultrasound energy safety was established in accordance with IEC 60601-2-62: 2013
  • Functional bench testing was conducted to verify the device performance (acoustic . mapping and parameter measurement testing, applicator performance testing, handle acoustic emission measurements, transducers linearity experiment, power control and output, cooling power, etc.)
  • In vivo testing in an animal model was performed to evaluate and establish the ● safety and effectiveness of the subject device

To support the expansion of indications, the Company conducted a clinical study that evaluated the safety and effectiveness of the device for the non-invasive dermatological aesthetic treatment to lift the eyebrow and lift lax submental and neck tissue. A total of 92 subjects were enrolled and 80 subjects were treated at 5 investigational sites in the United States. Each subject was treated on 3 face and neck zones including the forehead and temples for eyebrow lifting (Zone 1), the cheeks (Zone 2), and the submental and the neck (Zone 3). Overall, 467 facial areas were treated during the study course.

For Zone 1, the co-primary effectiveness endpoints were the independent masked evaluators' correction identification of pre-post image sequence for eyebrow lift, and the objective measurement of maximal evebrow lift according to quantitative 2D imaging. The results demonstrate that the blinded reviewers identified correctly the pre- and post-treatment photographs for 79% (53/67) of the treated subjects (based on the agreement of two blinded reviewers). Objective measurement of eyebrow lift by quantitative 2D imaging demonstrated an average lift of 0.78 mm for maximal eyebrow height (MEH) and 0.69 mm for average eyebrow height (AEH). Therefore, both co-primary endpoints are met. With regard to the secondary endpoints, progressive improvement was noticed along the study course and 80% (51/64) of subjects were improved based on Physician Global Aesthetic Improvement Scale (PGAIS) assessment at the 3-month follow-up visit.

For Zone 3, the co-primary effectiveness endpoints were the independent masked evaluators' correction identification of pre-post image sequence for submental lift, and the objective measurement of lax submental and neck tissue lift according to quantitative 2D imaging. The results demonstrate that the blinded reviewers identified correctly the pre- and post-treatment photographs for 80% (60/75) of the treated subjects (based on the agreement of two blinded reviewers). Objective measurement of lax submental and neck tissue lift by quantitative 2D imaging demonstrated an average lift of 38 mm². Therefore, both co-primary endpoints are met. With regard to the secondary endpoints, progressive improvement was noticed along the study course and 85% (61/72) of subjects were improved based on Physician Global Aesthetic Improvement Scale (PGAIS) assessment at the 3-month follow-up visit.

{8}------------------------------------------------

The patient satisfaction questionnaire demonstrated that 55% of subjects reported improvement and satisfaction for the evebrow lift. For the question, would you undergo further treatments, 37/64 (58%) said yes. 14/64 (17%) replied no and 13/64 (20%) replied not sure. 55% of subjects reported improvement for submental and neck lift. 50% of subjects reported satisfaction. For the question, would you undergo further treatments, 43/72 (60%) of subjects said yes. 14/72 (19%) subjects said no and 15/72 (21%) subjects said not sure.

The clinical study also demonstrated a favorable safety profile for the SoftWave for the proposed use in evebrow lift and lax submental and neck tissue lift. Throughout the study, only one device related adverse events were reported, which was moderate, transient, and resolved completely with topical cream only. No serious or unanticipated adverse event was reported during the study. The mean pain level during treatment was 6.4 out of 10 for Zone 1 and 5.3 out of 10 for Zone 3. Based on a literature review. the study results were consistent with the predicate device.

Conclusions

The SoftWave System has the same intended use and similar indications for use as its predicate device, Ulthera System. Further, the SofWave System has very similar technological characteristics and principles of operation as its predicate device, Ulthera System. The minor technological differences between the subject and the predicate devices do not raise different questions of safety or effectiveness. Performance testing of the device has demonstrated that the device performs as intended and thus, is substantially equivalent.

§ 878.4590 Focused ultrasound stimulator system for aesthetic use.

(a)
Identification. A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.