K Number
K211473

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-06-04

(23 days)

Product Code
Regulation Number
892.1720
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Micro C Medical Imaging System, M01 is a handheld and portable general purpose X-ray system that is indicated for use by qualified/trained clinicians on adult and pediatric patients for taking diagnostic static and serial radiographic exposures of extremities. The device is not intended to replace a radiographic system that has both variable tube current and voltages (kVp) in the range that may be required for full optimization of image quality and radiation exposure for different exam types.

Device Description

The Micro C Medical Imaging System, M01 (subject device) is a handheld X-ray system designed to aid clinicians with point of care visualization through diagnostic X-rays of distal extremities. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The Micro C Medical Imaging System, M01 consists of three major subsystems: The Emitter, Cassette, and Control Unit. The System is intended to interface an external Monitor (touchscreen or non-touchscreen display), keyboard and a mouse, and can provide a remote operator interface over the network to a laptop. The Micro C Medical Imaging System, M01 utilizes a computer vision positioning system to allow the emitter to be positioned above the patient anatomy and aligned to the cassette by the operator. The device is used in a clinical environment.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the OXOS Medical Micro C Medical Imaging System, M01. This submission aims to expand the indications for use to include pediatric patients and surgical settings. The document heavily relies on the substantial equivalence to a previously cleared predicate device (also Micro C Medical Imaging System, M01, K203658).

However, the document does not include detailed acceptance criteria or a study that rigorously proves the device meets specific performance criteria in a way that would be typical for an AI/CADe device. The focus is on demonstrating that the expanded indications (pediatric use and surgical use) do not raise new questions of safety or effectiveness compared to the predicate device. The information provided is primarily a comparison between the subject device and its predicate, highlighting that the core technological characteristics are identical.

Therefore, many of the requested details about acceptance criteria, specific performance metrics, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, and standalone AI performance are not present in the provided text, as this is a submission for an imaging device, not an AI/CADe device.

Based on the provided text, here's the information that can be extracted or inferred:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria or detailed performance metrics in the way one would expect for an AI/CADe device (e.g., sensitivity, specificity, AUC). Instead, the "performance" discussed relates to the device's fundamental imaging capabilities and its intended use, established by its substantial equivalence to the predicate. The crucial aspect from the document is the claim that the updated indication for pediatric use and surgical use is acceptable as demonstrated by "performance data" that does not raise different questions of safety and effectiveness.

Acceptance Criteria (Inferred from equivalence to predicate and non-clinical testing)Reported Device Performance (from text)
Image quality for diagnostic static and serial radiographic exposures of extremities (comparable to predicate K203658)Device is "identical" to predicate device K203658 in technological characteristics like detector, resolution, DQE, MTF.
Safe and effective for adult patients (established by predicate K203658)Established as substantially equivalent.
Safe and effective for pediatric patients (new indication)Supported by "performance data" related to the addition of the pediatric population, implying that the device performs adequately for this cohort without new safety/effectiveness concerns.
Safe and effective for use in surgery (new indication)Supported by "usability testing for use in surgery" and "cleaning and disinfection studies."
Compliance with relevant guidance documents and standardsMention of conformance with guidance for Medical X-ray Imaging Devices, Solid State X-ray Devices, Hand-held X-ray Equipment, and Pediatric Information for X-ray Imaging Device Premarket Notifications.

2. Sample size used for the test set and the data provenance

Not provided in the document. The filing is for an X-ray imaging device, not an AI/CADe system. The "performance data" for pediatric and surgical use is not detailed with sample sizes or data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable/Not provided. This relates to AI/CADe studies, not the 510(k) for an X-ray imaging system.

4. Adjudication method for the test set

Not applicable/Not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI-enabled device.

7. The type of ground truth used

Not applicable. The "performance data" for pediatric and surgical use is not specified in terms of clinical "ground truth" used for evaluation. It likely refers to technical performance evaluations and usability rather than diagnostic accuracy against a clinical gold standard for disease detection.

8. The sample size for the training set

Not applicable. This is not an AI-enabled device.

9. How the ground truth for the training set was established

Not applicable. This is not an AI-enabled device.

{0}------------------------------------------------

June 4, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

OXOS Medical, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k SAINT PAUL MN 55114

Re: K211473

Trade/Device Name: Micro C Medical Imaging System, M01 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: May 11, 2021 Received: May 12, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211473

Device Name Micro C Medical Imaging System, M01

Indications for Use (Describe)

The Micro C Medical Imaging System, M01 is a handheld and portable general purpose X-ray system that is indicated for use by qualified/trained clinicians on adult and pediatric patients for taking diagnostic static and serial radiographic exposures of extremities. The device is not intended to replace a radiographic system that has both variable tube current and voltages (kVp) in the range that may be required for full optimization of image quality and radiation exposure for different exam types.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 GER 801 Subpart D) ☐ Over-The-Counter Use (21 CER 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for OXOS Medical. The logo consists of four symbols: a circle with two gaps, an X, another circle with two gaps, and the letter S. Below the symbols, the word "MEDICAL" is written in a smaller font.

Section 5 510(k) Summary K211473

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Micro C Medical Imaging System, M01 Traditional 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.

Submitter:OXOS Medical, Inc1230 Peachtree St NESuite #300Atlanta, GA 30309Tel: 1-855-SEE-XRAY
Submission Contact:Mo KhosravanipourDirector of Program ManagementOXOS Medical, Inc.Email: Mo@oxos.com
Grace Powers, MS, MBA, RACFounder/Principal ConsultantPowers Regulatory ConsultingEmail: grace@powersregulatory.com
Submission Date:April 16, 2021
Subject Device:Trade Name: Micro C Medical Imaging System, M01Common Name: System, X-Ray, MobileRegulation: 21 CFR § 892.1720Regulatory Classification: 2Product Code: IZLClassification Panel: Radiology
Predicate Device:Legally marketed device to which substantial equivalence is claimed:Micro C Medical Imaging System, M01 (K203658)Common Name: System, X-Ray, MobileRegulation: 21 CFR § 892.1720Regulatory Classification: 2Product Code: IZLClassification Panel: Radiology

{4}------------------------------------------------

Section 5 510(k) Summary

Device Description:

The Micro C Medical Imaging System, M01 (subject device) is a handheld X-ray system designed to aid clinicians with point of care visualization through diagnostic X-rays of distal extremities. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The Micro C Medical Imaging System, M01 consists of three major subsystems: The Emitter, Cassette, and Control Unit. The System is intended to interface an external Monitor (touchscreen or non-touchscreen display), keyboard and a mouse, and can provide a remote operator interface over the network to a laptop. The Micro C Medical Imaging System, M01 utilizes a computer vision positioning system to allow the emitter to be positioned above the patient anatomy and aligned to the cassette by the operator. The device is used in a clinical environment. A description of the three major sub-systems is listed below.

  • Emitter: This component contains the operator control panel, X-ray tube, and computer vision camera. The control panel allows the operator to control the major functions of the device, including the technique factors. This component is controlled and held in the operator's hand.
  • Cassette: This component contains the X-ray detector that collects the X-ray energy and provides a digital representation to the control unit for eventual display. This component also contains status lights and IR lights to assist in X-ray field positioning. The patient anatomy of interest is placed on top of this module.
  • . Control Unit: This component contains the High Voltage generator, computing power, monitor and keyboard inputs, and other electronics required for the functioning of the device. This module is typically placed on a shelf, cart, counter, or other flat surface convenient to the operator and environment.

The system is intended to work in conjunction with a DICOM monitor, keyboard and a mouse and the mains power outlet. These are accessories are not included in the submission.

The Micro C Medical Imaging System, M01 has custom validated software that includes a user interface that allows the operator to view and adjust captured radiographs and transfer radiographs to a PACS server or flash drive. The device software is based on the predicate device and no changes to the software were performed for this submission.

Image /page/4/Figure/9 description: The image shows a white box connected to a white handheld device and a white square device via cables. Two blue cables connect the box to the handheld device and the square device. A gray cable connects the box to the square device.

Figure 5-1: Subject Device - Micro C Medical Imaging System, M01

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows a computer setup with a monitor, keyboard, mouse, and a white box-shaped device. The monitor is connected to the white box, which also has connections for the keyboard, mouse, and an Ethernet cable. The power cords for the monitor and the white box are plugged into wall outlets.

Intended Use:

The M01 System is a hand-held X-ray system designed to aid clinicians with point of care visualization through diagnostic X-rays of extremities.

Indications for Use:

The Micro C Medical Imaging System, M01 is a handheld and portable general purpose X-ray system that is indicated for use by qualified/trained clinicians on adult and pediatric patients for taking diagnostic static and serial radiographic exposures of extremities.

The device is not intended to replace a radiographic system that has both variable tube current and voltages (kVp) in the range that may be required for full optimization of image quality and radiation exposure for different exam types.

Technological Characteristics

The Micro C Medical Imaging System, M01 is identical to the predicate device cleared via K203658. The purpose of this submission is to add the pediatric population to the indication and to allow use in surgery.

The table below compares the subject and predicate device.

Comparison
Subject Device:Micro C Medical ImagingSystem, M01Predicate Device: MicroC Medical ImagingSystem, M01 (K203658)
Product CodeIZL (Mobile X-RaySystem)IZL (Mobile X-RaySystem)Identical
Regulation21 CFR 892.172021 CFR 892.1720Identical
ClassificationNameMobile X-Ray SystemMobile X-Ray SystemIdentical
ClassificationClass 2Class 2Identical
Indication forUseThe Micro C MedicalImaging System, M01 is ahandheld and portablegeneral purpose X-raysystem that is indicatedThe Micro C MedicalImaging System, M01 isa handheld andportable generalpurpose X-ray systemDifferent- The subject devicecan be used in pediatricpopulations.
for use byqualified/trainedclinicians on adultpatients for takingdiagnostic static andserial radiographicexposures of extremities.The device is notintended to replace aradiographic system thathas both variable tubecurrent and voltages(kVp) in the range thatmay be required for fulloptimization of imagequality and radiationexposure for differentexam types.that is indicated for useby qualified/trainedclinicians on adult andpediatric patients fortaking diagnostic staticand serial radiographicexposures ofextremities.The device is notintended to replace aradiographic systemthat has both variabletube current andvoltages (kVp) in therange that may berequired for fulloptimization of imagequality and radiationexposure for differentexam types.
Age of DeviceUseAdults and PediatricAdultsDifferent- The subject devicecan be used in pediatricpopulations.
Principle ofOperationGeneral purposediagnostic X-rayGeneral purposediagnostic X-rayIdentical
Image typeproducedStatic, serial radiographicand photographic imagesfor convenience.Static, serialradiographic andphotographic images forconvenience.Identical
Detector6 x 6" digital detector6 x 6" digital detectorIdentical
CollimatorThe removable fixedcollimators (referred toas pucks)The removable fixedcollimators (referred toas pucks)Identical
WeightEmitter: 2.86kg (6.3lbs)Cassette: 6.5kg (14.3lbs)Control Unit: 8.6kg(19.0lbs)Emitter: 2.86kg (6.3lbs)Cassette: 6.5kg (14.3lbs)Control Unit: 8.6kg(19.0lbs)Identical
Dimension/SizeEmitter: 9.3"H x 3.5"W x8.3"L (excluding SSDCone)Cassette: 15.5"H x15.5"W x 2.8"LControl Unit: 16.4"H x12.9"W x 5.5"LEmitter: 9.3"H x 3.5"W x8.3"L (excluding SSDCone)Cassette: 15.5"H x15.5"W x 2.8"LControl Unit: 16.4"H x12.9"W x 5.5"LIdentical
TriggeringMechanismTwo stage triggeringTwo stage triggeringIdentical
MinimumSource toskin distance(SSD)20 cm SSD Cone ensuresminimum SSD of 20 cm20 cm SSD Cone ensuresminimum SSD of 20 cmIdentical
Source toDetectordistance20 - 45 cm20 - 45 cmIdentical
Light FieldVirtual light field onMonitor UI. No projectedlight field.Virtual light field onMonitor UI. Noprojected light field.Identical
EnergySource120 VAC / 60 Hz (norechargeable battery)120 VAC / 60 Hz (norechargeable battery)Identical
ExposureTime33ms, 66ms and 99ms33ms, 66ms and 99msIdentical
mA1.0 mA fixed1.0 mA fixedIdentical
kVp40kVp, 50kVp, and 60kVp40kVp, 50kVp, and60kVpIdentical
ScintillatorCesium lodide (CsI)Cesium lodide (CsI)Identical
Resolution/Pixel size99 μm99 μmIdentical
DQE @0Lp/mm70%70%Identical
MTF @ 1Lp/mm,RQA560%60%Identical
IngressProtectionRatingIPOOIPOOIdentical
ImageProcessingUser Interface can beused to drag, zoom,rotate and also adjustbrightness, contrast, andsharpness.User Interface can beused to drag, zoom,rotate and also adjustbrightness, contrast,and sharpness.Identical
ConnectivityOptionsWiFi, Ethernet, Four USB2.0 portsWiFi, Ethernet, FourUSB 2.0 portsIdentical
DICOMYes- DICOM 3.0CompliantYes- DICOM 3.0CompliantIdentical
DevicePackageContents● Cassette● Control Unit● Emitter● Collimation Pucks● SSD Cone● Cassette PowerCable● Cassette Data Cable● Control Unit PowerCable● Cassette● Control Unit● Emitter● Collimation Pucks● SSD Cone● Cassette PowerCable● Cassette Data Cable● Control Unit PowerCableIdentical
Connector Covers Instructions for Use CaseConnector Covers Instructions for Use Case

Table 5-2: Device Comparison

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Non-Clinical Performance Data

Testing specific to the use in surgery was performed and includes:

  • Usability testing for use in surgery
  • Cleaning and Disinfection Studies ●

The following specific guidance document was utilized in the device development to ensure the safety of this device for both the operators and patients:

  • Guidance for Medical X-ray Imaging Devices Conformance with IEC Standards ●
  • . Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices
  • Radiation Safety Considerations for X-ray Equipment Designed for Hand-held Use
  • Pediatric Information for X-ray Imaging Device Premarket Notifications

Conclusion

The Micro C Medical Imaging System, M01 is identical in design to the legally marketed predicate device. The updated indication and contraindication are acceptable as demonstrated by the performance data and does not raise different questions of safety and effectiveness.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.