K Number
K211349

Validate with FDA (Live)

Date Cleared
2022-01-07

(249 days)

Product Code
Regulation Number
870.5100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Selebrek PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

Device Description

The Selebrek PTCA Balloon Dilatation Catheter is a rapid exchange (RX) PTCA Balloon Dilatation Catheter used for the purpose of dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

The Selebrek is a sterile, single-use, intravascular medical device with a working length of 142cm. The proximal shaft is PTFE coated stainless steel tube, which allows for exceptional push ability and a smooth transition to the distal shaft, which is composed of an outer tube, an inner tube, and a balloon. A hydrophilic coating is applied to the distal section. The semicompliant balloons are available in diameters of 1.5mm and lengths from 6-30mm, and have a rated burst pressure of 14 atm.

The proximal shaft of the catheter has two marker sections of 5mm length that aid in gauging dilatation catheter position relative to the guiding catheter located closest to the dilatation catheter adaptor is for femoral guiding catheters and the other marker is for brachial guiding catheters).

The distal shaft of the catheter has an integrated shaft system. The shaft has a combination of single lumen and dual lumen tubing. One lumen is used for inflation of the balloon with contrast medium. The other lumen, in the distal shaft, permits the use of a quide wire to facilitate advancement of the dilatation catheter to and through the stenosis to be dilated.

The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial quidance and rapid exchange of the catheter with a single standard length guidewire.

One radiopaque Platinum/lridium marker band is located within the balloon segment to aid in positioning the balloon in the stenosis, and are designed to provide an expandable segment of known diameter and length at a specific pressure. The design of this dilatation catheter does not incorporate a lumen for distal dye injections and distal pressure measurements.

AI/ML Overview

The provided document is a 510(k) Summary for a medical device (Selebrek PTCA Balloon Dilatation Catheter). It details the device's comparison to a predicate device and non-clinical testing performed to establish substantial equivalence.

However, the document does NOT contain information about AI/ML device performance or acceptance criteria related to AI/ML. There is no mention of an algorithm, AI assistance, human readers, ground truth establishment for AI, or any of the metrics typically associated with AI/ML model evaluation (e.g., sensitivity, specificity, AUC).

Therefore, I cannot provide the requested information regarding AI/ML acceptance criteria and studies. The document only describes a traditional medical device (balloon catheter) and its non-clinical bench testing for mechanical, physical, and biocompatibility properties.

If you are looking for an example of a 510(k) submission that does include AI/ML performance data, you would need to refer to a different type of device, such as an AI-powered diagnostic imaging tool.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

January 7, 2022

Kossel Medtech (Suzhou) Co., Ltd. Zane Wang Quality Director F2-F3, BLDG 6, No. 8, Jinfeng Road Suzhou, Jiangsu Province 215163 China

Re: K211349

Trade/Device Name: Selebrek PTCA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: December 2, 2021 Received: December 8, 2021

Dear Zane Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211349

Device Name Selebrek PTCA Balloon Dilatation Catheter

Indications for Use (Describe)

The Selebrek PTCA Balloon Dilatation Catheted for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a logo for KOSSEL. The logo consists of a circular graphic on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular graphic is made up of several curved lines that create a swirling effect. The text is in a sans-serif font and is a teal color.

510(k) Summary – K211349

This 510(k) Summary is submitted in accordance with 21 CFR 807.92(c).

Submitter:Kossel Medtech (Suzhou) Co., Ltd.F2-F3, BLDG 6, No. 8 Jinfeng RoadSuzhou New District, Jiangsu Province, China
Contact Person:Zane WangF3, BLDG 6, No. 8 Jinfeng RoadSuzhou New District, Jiangsu Province, ChinaPhone: +86 512 8717 4080-6003Mail: wangzhigao@kosselmed.com
Date Prepared:December 30, 2021
Trade Name:Selebrek PTCA Balloon Dilatation Catheter
Common Name:Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter
Classification Name:Catheters, transluminal coronary angioplasty, percutaneous (21 CFR 870.5100), Class II (special controls)
Product Code:LOX
Predicate Device:Sapphire II PRO Balloon Dilatation Catheter (K180921; cleared June 28, 2018)

Intended use/Indication for use:

The Selebrek PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

Device Description:

The Selebrek PTCA Balloon Dilatation Catheter is a rapid exchange (RX) PTCA Balloon Dilatation Catheter used for the purpose of dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

The Selebrek is a sterile, single-use, intravascular medical device with a working length of 142cm. The proximal shaft is PTFE coated stainless steel tube, which allows for exceptional push ability and a smooth transition to the distal shaft, which is composed of an outer tube, an inner tube, and a balloon. A hydrophilic coating is applied to the distal section. The semicompliant balloons are available in diameters of 1.5mm and lengths from 6-30mm, and have a rated burst pressure of 14 atm.

The proximal shaft of the catheter has two marker sections of 5mm length that aid in gauging dilatation catheter position relative to the guiding catheter located closest to the dilatation catheter adaptor is for femoral guiding catheters and the other marker is for brachial guiding catheters).

The distal shaft of the catheter has an integrated shaft system. The shaft has a combination of single lumen and dual lumen tubing. One lumen is used for inflation of the balloon with contrast medium. The other lumen, in the distal shaft, permits the use of a quide wire to facilitate advancement of the dilatation catheter to and through the stenosis to be dilated.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for KOSSEL. The logo consists of a circular graphic on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular graphic is made up of several curved lines that form a spiral shape. The text is in Chinese and English, with the English text "KOSSEL" appearing below the Chinese text.

The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial quidance and rapid exchange of the catheter with a single standard length guidewire.

One radiopaque Platinum/lridium marker band is located within the balloon segment to aid in positioning the balloon in the stenosis, and are designed to provide an expandable segment of known diameter and length at a specific pressure. The design of this dilatation catheter does not incorporate a lumen for distal dye injections and distal pressure measurements.

Comparison with predicate device:

Percutaneous Transluminal Coronary Angioplasty (PTCA) is the technological principle for both Selebrek PTCA Balloon Dilatation Catheter and the predicate device. PTCA is based on the use of Percutaneous Coronary Intervention (PCI) devices for the purpose of myocardial perfusion.

The Selebrek PTCA Balloon Dilatation Catheter and predicate device have following same technological elements:

  • Operating principle - balloon dilatation of stenotic portion by pressurization of inflation medium
  • Intended User – for patients evidencing coronary ischemia for the purpose of improving myocardial perfusion
  • Fundamental catheter design – balloon, shaft, radiopaque marker, hub, coating
  • Shaft type – Rapid exchange
  • Sterilization - Ethylene oxide
  • Pressure - Nominal Pressure and Rated Burst Pressure

There are following minor technological differences between Selebrek PTCA Balloon Dilatation Catheter and predicate device:

  • Combination of balloon diameter and balloon length
  • Distal shaft diameters
  • Catheter effective length
  • Crossing profile
  • Marker bands
  • Miniwrap folds

Any differences that exist between the Selebrek PTCA Balloon Dilatation Catheter and the predicate device did not raise any new questions of safety or effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for KOSSEL, a medical company. The logo features a circular design with concentric rings on the left. To the right of the circular design, the company name "KOSSEL" is written in both Chinese and English.

Subject DevicePredicate Device
Comparison itemsSelebrek PTCA Balloon CatheterOrbus Neich Medical, Sapphire II PROBalloon Dilatation Catheter K180921Comments
Intended Use/Indications foruseThe Selebrek PTCA Balloon DilatationCatheter is indicated for balloondilatation of the stenotic portion of acoronary artery or bypass graft stenosisfor the purpose of improvingmyocardial perfusion.The Sapphire® II PRO Balloon DilatationCatheter (Ø1.0-1.25mm configurations) isindicated for:● Balloon pre-dilatation of a stenotic portionof a coronary artery or bypass graft stenosis(≥70% stenosis) for the purpose of improvingmyocardial perfusion.The Sapphire® II PRO Balloon DilatationCatheter (Ø1.5-4.0mm configurations) isindicated for:● Balloon dilatation of the stenotic portionof a coronary artery or bypass graft stenosis inpatients evidencing coronary ischemia for thepurpose of improving myocardial perfusion.● Balloon dilatation of a coronary arteryocclusion for the treatment of acute myocardialinfarction.The Sapphire® II PRO Balloon DilatationCatheter is also indicated for:● percutaneous transluminal angioplasty in theperipheral vasculature, including renal, femoral,popliteal, infra-popliteal, tibial, and peronealarteries.● The intended use of predicatedevice is larger than that of the subjectdevice, which not only for coronary usebut also support the peripheral use;● The intended use of subject deviceis identical to that of the predicatedevice balloon size 1.5~4.0mm, whilethe description of predicate device ismore detailed and specific.● The Subject Device Selebrek PTCABalloon Catheter is intended use inpatients evidencing coronary ischemiafor improving myocardial perfusion.The differences in description are notcritical to the intended therapeutic, orsurgical use of the device, so thedifferences do not raise any newquestions of safety or effectiveness
Subject DevicePredicate Device
Comparison itemsSelebrek PTCA Balloon CatheterOrbus Neich Medical, Sapphire II PROBalloon Dilatation Catheter K180921Comments
Classification namePercutaneous Transluminal CoronaryAngioplasty (PTCA) Catheter21 CFR 870.5100 (class II, special control)Percutaneous Transluminal CoronaryAngioplasty (PTCA) Catheter21 CFR 870.5100 (class II, special control)Identical
Product codeLOXLOXIdentical
ChemicalcompositionBalloonmaterialPebaxPebaxIdentical
CoatingPTFE / HydrophilicHydrophilic (distal tip to guidewire exitmarker); Hydrophobic (guidewire lumen)Equivalent
Physicalcharacteristics/DevicedesignBalloonDiameters(mm)1.51.0 - 4.0Equivalent
Balloonlengths (mm)6-305-30Equivalent
BallooncomplianceSemi-CompliantSemi-CompliantIdentical
Catheter TypeRxRxIdentical
Marker bands1 marker1.0-1.5mm: 1 marker1.5-4.0mm: 2 markersEquivalent
Comparison itemsSubject DeviceSelebrek PTCA Balloon CatheterPredicate DeviceOrbus Neich Medical, Sapphire II PROBalloon Dilatation Catheter K180921Comments
Guide wire size(inch)0.014"0.014"Identical
Proximal shaftdiameter2.0 F2.0 FIdentical
Distal shaftdiameter2.36 F1.0-1.5mm: 2.36F1.75-4.0mm: 2.36F – 2.7FEquivalent
CrossingProfile (inch)0.026"~0.043"3.0mm: 0.031"Equivalent
Miniwrap folds3 folds1.0mm: 2 folds;1.25-4.0mm: 3 foldsEquivalent
Catheterlength (cm)142140Equivalent
RatedburstPressure (atm)1414Identical
Nominalpressure (atm)66Identical
Sterilization MethodEOEOIdentical

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains a logo for KOSSEL, a medical company. The logo features a circular design with swirling lines on the left. To the right of the circular design, the company's name is written in both Chinese characters and English.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains a logo with a circular design on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular design consists of multiple curved lines arranged in a spiral pattern. The text is in Chinese and English, with "KOSSEL" written in capital letters below the Chinese characters. The logo appears to represent a medical or healthcare-related organization.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for KOSSEL. The logo consists of a circular graphic on the left and the text "科塞尔医疗 KOSSEL" on the right. The circular graphic is made up of several concentric circles. The text is in a sans-serif font, with the Chinese characters "科塞尔医疗" appearing above the English word "KOSSEL".

Non-clinical testing/Performance data:

Non-clinical bench testing was performed on the subject device to determine substantial equivalence. The testing performed is as follows:

  • Dimensional verification
  • Balloon rated burst pressure
  • Balloon fatigue
  • Balloon compliance
  • Balloon inflation and deflation time
  • Catheter bond strength
  • Tip pull strength
  • Flexibility and kinkinq
  • Torque strength
  • Radiopacity
  • Coating integrity
  • Particulate evaluation
  • Shelf life testing

The results met all acceptance criteria and ensure that the Balloon Dilatation Catheter design and construction are suitable for its intended use as recommended by the Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters (FDA; September 8, 2010).

Biocompatibility:

To demonstrate the biological safety of the body-contacting materials and substantial equivalence of the Selebrek PTCA Balloon Dilatation Catheter to the compared device Selethru PTCA Balloon Dilatation Catheter which had been cleared by FDA (K182699), the following biocompatibility testing, which was performed on the compared device (Selethru PTCA Balloon Dilatation Catheter) in accordance with "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"; Guidance for Industry and Food and Drug Administration Staff" were leveraged:

Cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility (hemolysis, complement activation, and in vivo thromboresistance), pyrogenicity were conducted.

The results of the testing show that the subject device in this submission met all acceptance criteria and the subject device is biocompatible.

Conclusion:

The Selebrek PTCA Balloon Dilatation Catheter included in this notification is equivalent to the previously cleared predicate device, the Sapphire® II PRO PTCA Balloon Dilatation Catheter (K180921) in terms of intended use and technological characteristics. Any differences that exist between the Selebrek PTCA Balloon Dilatation Catheter and the predicate device did not raise any new questions of safety or effectiveness.

§ 870.5100 Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter.

(a)
Standard PTCA Catheter —(1)Identification. A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion.(2)
Classification. Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
Cutting/scoring PTCA Catheter —(1)Identification. A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
Classification. Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.