K Number
K211333

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-01-20

(262 days)

Product Code
Regulation Number
876.5880
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KIDNEY ASSIST-transport is intended to be used for the pulsatile hypothermic oxygenated machine perfusion of kidneys for the preservation, transport and eventual transplantation into a recipient.

Device Description

The KIDNEY ASSIST-transport system of Organ Assist Products B.V. is a portable pump system that continuously allows hypothermic pulsatile perfusion of donor kidneys with oxygenated preservation solution during transport from donor to recipient in transplantation procedures. The system consists of the reusable KIDNEY ASSISTtransport device and a disposable KIDNEY ASSIST-transport perfusion set. The characteristics of the KIDNEY ASSIST-transport are:

  • Pulsatile oxygenated hypothermic machine perfusion of donor kidneys ●
  • Transportable hypothermic machine perfusion technique
  • Hypothermic preservation and reconditioning device
  • Improved preservation compared to cold storage in DBD, ECD and DCD
  • Easy to install disposable perfusion set

The KIDNEY ASSIST-transport device is a thermo isolated enclosure wherein the kidneys are cooled passively by ice and a separate compartment holding the electronics, batteries and a dedicated medical oxygen cylinder. The device has sufficient battery power, holds enough oxygen and ice for an application period of 24 hours of hypothermic oxygenated perfusion. Pulsatile perfusion is generated by a rotary pump driven by an electromotor and is pressure controlled. User-friendly firmware allows the user to change perfusion parameters. Settings and results of the perfusion measurements are numerically displayed on the top of the enclosure. The single-use disposable KIDNEY ASSIST-transport perfusion set contains an easy to install preassembled perfusion cartridge for use in combination with the KIDNEY ASSIST-transport device. The purpose of the KIDNEY ASSIST-transport Perfusion Set is to perfuse human organs to be transplanted with an approved pump perfusion solution. Its set contains a reservoir, kidney holder, cannula, oxygenator, pump head, pressure sensor and compatible tubing. Pulsatile perfusion is maintained by the centrifugal pump head, pulsating the perfusion solution from the reservoir through the oxygenator to the kidney holder in the reservoir. Oxygenation is performed by the hollow fiber membrane oxygenator which facilitates the gas exchange with the perfusion solution. All kidneys will be perfused with University of Wisconsin Machine Perfusion Solution (UW-MP).

KIDNEY ASSIST-transport allows transportable machine perfusion to bridge the timespan between procurement and transplantation of kidneys.

AI/ML Overview

The information provided describes a medical device, the KIDNEY ASSIST-transport, and its substantial equivalence to a predicate device. However, it does not explicitly state "acceptance criteria" for the device's technical performance in a table, nor does it detail a study designed to "prove the device meets the acceptance criteria" in terms of specific performance metrics (like sensitivity, specificity, accuracy, etc.) you would typically find for diagnostic or AI-driven devices.

Instead, the document focuses on demonstrating substantial equivalence through various types of testing and an investigator-initiated clinical study primarily comparing oxygenated HMP (Hypothermic Machine Perfusion) with non-oxygenated HMP, where the KIDNEY ASSIST-transport was used in the oxygenated arm.

However, I can extract information related to the device's characteristics and performance to construct a response that aligns with your request as much as possible given the provided text.

Here's an attempt to answer your questions based on the provided document:


1. A table of acceptance criteria and the reported device performance

The document doesn't provide a formal "acceptance criteria" table for specific performance metrics in the way one might see for an AI algorithm. Instead, it compares the KIDNEY ASSIST-transport to a predicate device based on various characteristics and reports the findings of a clinical study that used the device.

Based on the "Substantial equivalence summary" table (page 5) and "Performance Data" section (page 6-7), here's a synthesis:

Characteristic/Criterion (Implicit)KIDNEY ASSIST-transport Performance / Compliance
Intended UsePulsatile hypothermic oxygenated machine perfusion of kidneys for preservation, transport, and eventual transplantation into a recipient.
Perfusion ParametersPressure: 0-50 mmHg; Flow: 0-250 mL/min; Temp: 2°C -10°C.
Duration of PerfusionUp to 24 hours.
OxygenationMedical oxygen, 100 mL/min gas flow, Hollow fiber oxygenator.
Electrical Safety & EMCComplies with IEC 60601-1:2012 Ed 3.1, ANSI/AAMI ES60601-1:2005/(R)2012 for safety and IEC 60601-1-2:2007 + AC:2010 for EMC.
Software ComplianceComplies with IEC 62304:2006 (Class B software, "major" level of concern).
BiocompatibilityComplies with ISO 10993-1.
Human Factors & UsabilityComplies with IEC 62366:2008 and ANSI/AAMI HE75.
SterilityPerfusion set complies with ISO 11135:2014 (SAL 10⁻⁶) and ISO 10993-7:2008.
Packaging & Shelf LifePackaging complies with ISO 11607:2019; Shelf life complies with ASTM F 1980-16:2016.
Clinical Performance (Oxygenated HMP vs. Non-oxygenated HMP)Improved outcomes: Significantly reduces severe post-operative complications and leads to 44% fewer rejections of the kidney after transplantation, along with improved kidney function and graft survival at 1 year. (Note: These are clinical outcomes using the device, not direct device performance metrics like accuracy).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The relevant study is an "Investigator Initiated Clinical Data" study.

  • Sample Size: Kidney pairs from donors were randomized. There were 106 recipients in each study arm (oxygenated HMP and non-oxygenated HMP), totaling 212 recipients. Kidney pairs imply 2 kidneys per donor, so likely 106 donor kidneys per arm = 212 donor kidneys total.
  • Data Provenance: The document does not specify the country of origin.
  • Study Design: It was an "investigator-initiated clinical study" described as a "paired design" where kidney pairs were randomly assigned. This indicates a prospective study design.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable to the provided document. The device is a kidney perfusion and transport system, not a diagnostic device that requires expert ground truth annotation for images or other data. The "ground truth" for the clinical study would be objective clinical outcomes (e.g., rejection rates, kidney function, graft survival, complications) observed in patients after transplantation. These are physiological endpoints, not expert interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. The clinical study described involves objective clinical outcomes (rejection rates, complications, graft survival, kidney function) in humans. Adjudication methods like "2+1" are typically used for establishing ground truth in image interpretation studies where expert consensus is needed. Clinical outcomes are typically measured directly or through established medical criteria, not by expert adjudication of subjective assessments. The study itself is an effectiveness comparison for a device used in treatment.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a medical apparatus for organ preservation, not an AI-based diagnostic tool. Therefore, an MRMC study comparing human reader performance with and without AI assistance is irrelevant to this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The KIDNEY ASSIST-transport is a physical device that performs machine perfusion; it is not an algorithm, and its operation inherently involves human interaction for setup, monitoring, and clinical decisions.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the "Investigator Initiated Clinical Data" study, the ground truth was based on clinical outcomes data after transplantation. This includes metrics such as:

  • Kidney function (not detailed how measured, but implied clinical performance)
  • Graft survival at 1 year
  • Number of severe post-operative complications (including (S)AE – serious adverse events)
  • Rejections of the kidney after transplantation

8. The sample size for the training set

This is not applicable. The document describes a physical medical device and a clinical study validating its use. It is not an AI/ML model for which a "training set" would be used. The "Investigator Initiated Clinical Data" study is an evaluation of the device's clinical performance, not a dataset for training an algorithm.

9. How the ground truth for the training set was established

This is not applicable for the same reasons as point 8. No "training set" for an algorithm is described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 20, 2022

Organ Assist Products B.V. % Kathleen Johnson President (Regulatory Affairs Consultant) Medical Device Approvals, Inc. 104 E. Harrison Ave. Fairfield, Iowa 52556

Re: K211333

Trade/Device Name: KIDNEY ASSIST-transport Regulation Number: 21 CFR 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: Class II Product Code: KDN Dated: December 20, 2021 Received: December 20, 2021

Dear Kathleen Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-

mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gema Gonzalez Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211333

Device Name KIDNEY ASSIST-transport

Indications for Use (Describe)

The KIDNEY ASSIST-transport is intended to be used for the pulsatile hypothermic oxygenated machine perfusion of kidneys for the preservation, transport and eventual transplantation into a recipient.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter of 510(k)

Owner's name:Address:Organ Assist Products B.V. (an affiliation of XVIVO Perfusion AB)Bornholmstraat 84,9723AZ GroningenThe Netherlands
Contact Person:Arjan van der Plaats
Title:Chief Technology Officer
Phone:+31 682 982 106
E-mail:a.vanderplaats@xvivogroup.com
Official Correspondent:Kathleen Johnson
Title:Regulatory Affairs Consultant
Phone:610-527-0585

kathleen@mdapprovals.com E-mail:

April 30, 2021

Date Prepared:

Device Name

Trade/proprietary name:KIDNEY ASSIST-transport
Common/usual name:Transportable kidney perfusion system
Classification name:Isolated kidney perfusion and transport system and accessories (21 CFR §876.5880)
Product code:KDN
Regulatory Class:Class II

Device Description

The KIDNEY ASSIST-transport system of Organ Assist Products B.V. is a portable pump system that continuously allows hypothermic pulsatile perfusion of donor kidneys with oxygenated preservation solution during transport from donor to recipient in transplantation procedures. The system consists of the reusable KIDNEY ASSISTtransport device and a disposable KIDNEY ASSIST-transport perfusion set. The characteristics of the KIDNEY ASSIST-transport are:

  • Pulsatile oxygenated hypothermic machine perfusion of donor kidneys ●
  • Transportable hypothermic machine perfusion technique
  • Hypothermic preservation and reconditioning device
  • Improved preservation compared to cold storage in DBD, ECD and DCD
  • Easy to install disposable perfusion set

The KIDNEY ASSIST-transport device is a thermo isolated enclosure wherein the kidneys are cooled passively by ice and a separate compartment holding the electronics, batteries and a dedicated medical oxygen cylinder. The device has sufficient battery power, holds enough oxygen and ice for an application period of 24 hours of hypothermic oxygenated perfusion. Pulsatile perfusion is generated by a rotary pump driven by an electromotor and is pressure controlled. User-friendly firmware allows the user to change perfusion parameters. Settings and results of the perfusion measurements are numerically displayed on the top of the enclosure. The single-use disposable KIDNEY ASSIST-transport perfusion set contains an easy to install preassembled perfusion cartridge for use in combination with the KIDNEY ASSIST-transport device. The purpose of the KIDNEY ASSIST-transport Perfusion Set is to perfuse human organs to be transplanted with an approved pump perfusion solution. Its set contains a reservoir, kidney holder, cannula, oxygenator, pump head, pressure sensor and compatible tubing. Pulsatile perfusion is maintained by the centrifugal pump head, pulsating the perfusion

Image /page/3/Picture/20 description: The image shows a gray letter A with a pink circle in the center. The letter A is in a simple, sans-serif font and appears to be a logo or icon. The pink circle is positioned in the middle of the A, creating a visually balanced and modern design. The overall impression is clean and minimalist.

{4}------------------------------------------------

solution from the reservoir through the oxygenator to the kidney holder in the reservoir. Oxygenation is performed by the hollow fiber membrane oxygenator which facilitates the gas exchange with the perfusion solution. All kidneys will be perfused with University of Wisconsin Machine Perfusion Solution (UW-MP).

KIDNEY ASSIST-transport allows transportable machine perfusion to bridge the timespan between procurement and transplantation of kidneys.

Application period

The 'KIDNEY ASSIST-transport' is a portable pump system that continuously allows hypothermic pulsatile perfusion of donor kidneys with oxygenated preservation solution during transport from donor to recipient in transplantation procedures for a period up to 24 hours.

Indications for Use

The KIDNEY ASSIST-transport is intended to be used for the pulsatile hypothermic oxygenated machine perfusion of kidneys for the preservation, transport, and eventual transplantation into a recipient.

Legally Marketed Predicate Device (K111521)

The KIDNEY ASSIST-transport (KAt) has been shown to be substantially equivalent to the legally marketed predicate device:

Trade/proprietary name: Waves (Waters Medical Systems LLC)

Common name: Renal Preservation Systems Classification name: Isolated kidney perfusion and transport system and accessories (21 CFR §876.5880) Product code: KDN Regulatory Class: Class II 510(k) number: K111521

Image /page/4/Picture/10 description: The image shows a gray letter A with rounded edges. Inside the letter A, there is a white circle with a magenta dot in the center. The dot is positioned in the middle of the white circle, creating a simple and modern design.

Title

K211333 Page 2/5

{5}------------------------------------------------

Substantial equivalence summary

The KIDNEY ASSIST-transport can be considered substantial equivalent to the legally marketed predicate device (K111521, Waves system manufactured by Waters Medical Systems LLC).

Subject device: KIDNEY ASSIST-transportImage: KIDNEY ASSIST-transportPredicate device: WavesImage: Waves
ManufacturerOrgan Assist Products b.v.Waters Medical Systems LLC
RegistrationFDA 510(k): pending as per this submissionFDA 510(k): K111521
Intended useThe KIDNEY ASSIST-transport is intended to be used forthe pulsatile hypothermic oxygenated machine perfusionof kidneys for the preservation, transport and eventualtransplantation into a recipient.The Waves device is intended to be used for thepulsatile hypothermic machine perfusion of kidneys forthe preservation, transport and eventualtransplantation into a recipient.
ConfigurationDevice in combination with sterile disposable set.Device in combination with sterile disposable set.
SterilizationEthylene Oxide (ISO 11135)Ethylene Oxide (ISO 11135)
MaterialsValidated against ISO 10993Validated against ISO 10993
Perfusion pumpCentrifugal pumpDisplacement pump
Perfusion modePulsatilePulsatile
PerfusioncontrolPressure control, automated control featuresPressure control, automated control features
SoftwareValidated Embedded software and user interface (watchdog safety)Validated Embedded software and user interface(watch dog safety)
CoolantCrushed ice (6 L)Crushed ice and water (5 L)
Power sourceBattery 4x 11.1 V Li-ionor auxiliary powerBattery 4x 11.1 V Li-ionor auxiliary power
Perfusion ratesPressure: 0-50 mmHgFlow: 0-250 mL/minTemp: 2°C -10°CPressure: 0-100 mmHgFlow: 0-250 mL/minTemp: 3°C-10°C
DurationUp to 24 hoursUp to 24 hours with extra ice filling
PerfusateHypothermic machine perfusion solutionHypothermic machine perfusion solution
OxygenationMedical oxygen100 mL/min gas flowHollow fibre oxygenatorAir or other gas mixturesUp to 1.5-2 L/min gas flowSilicone membrane oxygenator
Dimensions23.6″x15.3″x13.4″600mm x 390mm x 340mm25.5″x16.25″x13.25″(648 mm x 413 mm x 337 mm)
Weight55.8 lbs (25.3 kg)57 lbs (26 kg)
StorageconditionsT: 5 - 40°CH: 5 - 85 %RHT: 0°C tot 40°CH: 0-95%

A detailed comparison overview is presented below

Intended use

Both KIDNEY ASSIST-transport and the predicate device (K111521) are intended to be used for the pulsatile hypothermic machine perfusion of kidneys for the preservation, transport, and eventual transplantation into a recipient. Both systems oxygenate the cold perfusion that is perfused through the kidney; however, this is not included in the Intended Use statement of WAVES. For KIDNEY ASSIST-transport, the oxygenation intention is included for clarification.

Image /page/5/Picture/7 description: The image shows a gray icon in the shape of the letter A. The letter A is rounded and has a circle in the center. The circle is pink. The icon is simple and modern.

{6}------------------------------------------------

Principle of Operation

Principle of operation is equivalent. Both KIDNEY ASSIST-transport and the predicate device (K111521) consist of a pump unit and a single-use disposable perfusion set/cassette. UW-MP solution is pumped through the kidney while being cooled passively by ice and oxygenated through an oxygenator. Perfusion parameters (pressure, flow, resistance and temperature) are continuously monitored, displayed and stored, safeguarded by an alarm system. The systems incorporate embedded software for user interface and safety and control mechanisms.

Confiquration characteristics

Both KIDNEY ASSIST-transport and the predicate device (K111521) are comprised of reusable control/pump unit in combination with a single-use sterile perfusion set/cassette.

Perfusion Characteristics

Both KIDNEY ASSIST-transport and the predicate device (K111521) provide pressure-controlled pulsatile perfusion to the kidney. Perfusion characteristics on pressure are similar. The predicate device (K111521) WAVES allows a higher user-set pressure than the KIDNEY ASSIST-transport (100 mmHg vs 50 mmHg). In most clinical trials a mean pressure of 25 mmHg is used and demonstrated safety and efficacy of the technology. Therefore, the 50-mmHg limit in the KIDNEY ASSIST-transport does not induce new safety and effectiveness questions.

In addition, the predicate device (K111521) uses a displacement pump, where the KIDNEY ASSIST-transport incorporates a centrifugal pump to provide perfusion to the kidney. Since the perfusion parameters (pulsatile pressure and flow) are similar, this has no influence on the perfusion reaching the kidney. In principle, the centrifugal pump is considered a safer alternative for the displacement pump with respect to pressure build as it is an open pump system. Application period of both devices is 24 hours.

Cooling characteristics

Both KIDNEY ASSIST-transport and the predicate device (K111521) provides cooling by melting ice that is added to the insulated device.

Power source characteristics

Power source is identical in both systems: 4 Li-ion batteries, or auxiliary power supply can be used. The WAVES batteries are automatically charged while the device is on auxiliary power.

For the KIDNEY ASSIST-transport, the batteries must be charged using an external charger. As this means that less electrical power is needed inside the device, this situation is considered safer.

Oxygenation characteristics

Both KIDNEY ASSIST-transport and the predicate device (K111521) incorporate oxygenation of the cold perfusion solution that is pumped through the kidney. Both systems actively oxygenated the solution by means of an oxygenator that is part of the sterile single-use perfusion set/cassette. The oxygenator is part of the preassembled single-use sterile perfusion set.

Both systems use oxygen to oxygenate, where the WAVES also allows other gas mixtures.

Performance Data

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on KIDNEY ASSIST-transport. The system complies with the IEC 60601-1:2012 Edition 3.1, ANSI/AAMI ES60601-1:2005/(R)2012 standards for safety and the IEC 60601-1-2:2007 + AC:2010 standard for EMC.

Software

The KIDNEY ASSIST-transport contains software that shows compliance with IEC 62304:2006. The software for this device is classified as IEC 62304 Class B software and is considered as a "major" level of concern in the US.

Image /page/6/Picture/21 description: The image shows a gray letter A with a pink circle in the center. The letter A is in a sans-serif font and has rounded corners. The pink circle is slightly smaller than the opening in the letter A. The background is white.

{7}------------------------------------------------

Biocompatibility testing

The biocompatibility evaluation for the KIDNEY ASSIST-transport was conducted and shows compliance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process."

Human factors and usability

The usability evaluation is conducted on the KIDNEY ASSIST-transport. The system complies with the IEC 62366:2008 application of usability engineering to medical devices and ANSI/AAMI HE75 human factors engineering - design of medical devices.

Sterility

The sterility evaluation is conducted on the KIDNEY ASSIST-transport Perfusion set. The system complies with the ISO 11135:2014 for Ethylene Oxide sterilization of health-care products (SAL 10°) and ISO 10993-7:2008 for Ethylene oxide sterilization residuals.

Packaging and shelf life

Packaging for the sterile components complies with ISO 11607:2019 packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.

Shelf life for the sterile components complies with ASTM F 1980-16:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.

Bench testing

Performance testing -Bench include tests on functional performance, transport and storage validation and usability validation for the complete system.

Investigator Initiated Clinical Data

The published data of an investigator-initiated clinical study supports the claimed Performance and benefit of the KIDNEY ASSIST-transport. The KIDNEY ASSIST-transport provides safe hypothermic machine perfusion of kidneys, while adding beneficial oxygen during perfusion.

In detail, the objective of the investigator-initiated clinical study was to investigate the effects of oxygenation in HMP (hypothermic machine perfusion). Following standard organ procurement, kidney pairs (donors aged 50 or older, 38 % Female and 62 % Male) were cannulated and connected to the KIDNEY ASSIST-transport and perfused with UW-MP solution. Kidney pairs were randomly assigned to oxygenated (HMPQ2) or nonoxygenated (HMP) perfusion in a paired design. In total 106 recipients were present in each study arm and followed one year post transplantation. Baseline characteristics were similar in both groups (HMPO) : Age 60 years on average, 35 % Female, 65 % Male; HMP: Age 61 years on average, 37 % Female, 63 % Male). No data on ethnicity or race were reported.

The study results show that the addition of oxygen to HMP improves the outcomes compared to standard nonoxygenated HMP on kidney function and graft survival at 1 year after transplantation. HMPO2 significantly reduces the number of severe post-operative complications (including (S)AE) and even leads to a drastic 44% fewer rejections of the kidney after transplantation.

Conclusion

Based on the intended use, indications for use, principle of operall technological characteristics the KIDNEY ASSIST-transport has been shown to be substantially equivalent to the predicate device (K111521). Results of validation (including performance data) did not raise any questions on safety and effectiveness. It is concluded that the KIDNEY ASSIST-transport has been shown to be substantially equivalent to the legally marketed predicate device (K111521).

Image /page/7/Picture/17 description: The image shows a gray letter A with rounded edges. Inside the letter A, there is a pink circle. The letter A is simple and modern in design. The pink circle adds a pop of color to the otherwise neutral gray.

Title

§ 876.5880 Isolated kidney perfusion and transport system and accessories.

(a)
Identification. An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.(b)
Classification. Class II (performance standards).