Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

AI/ML Overview

The provided document is a 510(k) summary for a medical device: "Disposable Vinyl Nitrile Synthetic Gloves Powder Free". This document describes the device's characteristics, its comparison to a predicate device, and the results of non-clinical tests conducted to demonstrate its safety and effectiveness.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodologyPurposeAcceptance CriteriaReported Device Performance
ASTM D5250Physical Dimensions TestLength (mm): ≥230 Width (mm): S: 85±5; M: 95±5; L: 105±5; XL: 115±5 Thickness (mm): Finger: ≥0.05, Palm: ≥0.05Length: >230 Width: S: 84-89; M: 93-97; L: 105-107; XL: 115-116 Finger Thickness: 0.08-0.2 Palm Thickness: 0.07-0.08 Result: Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 leaks Result: Pass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.07 mg Result: Pass
ASTM D412Physical Properties - Before AgingTensile Strength: ≥11MPa Ultimate Elongation: ≥300%Tensile Strength: 11.6-16.2 Ultimate Elongation: 308-391 Result: Pass
ASTM D412Physical Properties - After AgingTensile Strength: ≥11MPa Ultimate Elongation: ≥300%Tensile Strength: 11.0-14.0 Ultimate Elongation: 323-337 Result: Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells. Result: Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant. Result: Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. Result: Pass

2. Sample Size Used for the Test Set and Data Provenance

  • ASTM D5151 (Watertightness): The test result states "0/125 leaks", indicating a sample size of 125 gloves for this particular test.
  • Other tests (ASTM D5250, ASTM D6124, ASTM D412, ISO 10993-5, ISO 10993-10): The specific sample sizes for these tests are not explicitly stated in the provided summary. However, these are generally performed on a sufficient number of samples to meet the requirements of the respective ISO/ASTM standards.
  • Data Provenance: The device manufacturer (Tangshan Hongyun Healthcare Products Co., Ltd.) is located in China. The non-clinical tests were conducted to verify the device met design specifications. The document does not specify if the data is retrospective or prospective, but given the nature of product testing for regulatory submission, it is typically prospective for the specific batch/samples used for testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • This document describes physical, chemical, and biological testing of a medical device (gloves) against established standards. It does not involve interpretation of medical images or patient data that would require "experts" in the sense of clinicians establishing ground truth on a test set. The "ground truth" for these tests is defined by the acceptance criteria within the international and national standards (ASTM, ISO) themselves.

4. Adjudication Method for the Test Set

  • Not applicable. The tests performed are objective measurements against predefined criteria as per established standards (ASTM, ISO), not subjective assessments requiring adjudication by multiple readers or experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically associated with diagnostic imaging or AI algorithms that require human interpretation and comparison with and without AI assistance to measure reader improvement. The device in question is a physical glove, and its performance is evaluated through standardized physical, chemical, and biological tests, not through human reader interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

  • Yes, the testing described is effectively "standalone" performance in the sense that the device's characteristics are measured objectively against standards without human-in-the-loop performance in a clinical scenario being assessed for regulatory purposes. The device itself is not an algorithm, so the term "algorithm only" isn't directly applicable, but the performance evaluation is solely based on the device's intrinsic properties.

7. The Type of Ground Truth Used

  • The "ground truth" is defined by the acceptance criteria established in recognized international and national standards (ASTM D5250, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-5, ISO 10993-10). These standards specify objective physical, chemical, and biological parameters that the gloves must meet to be deemed safe and effective for their intended medical purpose.

8. The Sample Size for the Training Set

  • Not applicable. A "training set" is relevant for machine learning or AI algorithms. This document describes the non-clinical testing of a physical medical device (gloves), which does not involve machine learning models and therefore has no "training set."

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 24, 2021

Tangshan Hongyun Healthcare Products Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K211262

Trade/Device Name: Disposable Vinyl Nitrile Synthetic Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: April 19, 2021 Received: April 26, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211262

Device Name

Disposable Vinyl Nitrile Synthetic Gloves Powder Free

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K211262

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Tangshan Hongyun Healthcare Products Co.,Ltd. Address: South Shenggezhuang Village, Pachigang Town, Luannan, Tangshan, Hebei 063500, China. Phone Number: +86-13933365259 Contact: Suying Le Date of Preparation: 07/23/2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Disposable Vinyl Nitrile Synthetic Gloves Powder Free Common name: Vinyl Patient Examination Glove Classification name: Non-powdered Patient Examination Glove Model(s): S, M, L, XL

3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Zibo Huiying Medical Products, Co. Ltd. Device: Synmax Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue 510(k) number: K153028

5.0 Indication for Use

{4}------------------------------------------------

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

ItemSubject devicePredicated deviceComparison
510(k) numberK211262K153028
Product CodeLYZLYZSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassSame
Intended Use/Indications foru UseA patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.Same
Powdered or Powered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
Labeling InformationSingle use, powder free, device color, device name, glove size and quantity, product name, Non-SterileSingle use, powder free, device color, device name, glove size and quantity, product name, Non-SterileSimilar

Table1-General Comparison

Table2 Device Dimensions Comparison

PredicateDevice(K153028)DesignationSizeTolerance
Length, mmAverage over 234 on M size-
Width, mmAverage over 96 on M size-
Thickness, mm:

{5}------------------------------------------------

FingerAverage 0.98-
PalmAverage 0.096-
Subject DeviceDesignationSizeTolerance
SMLXL
Length, mm230230230230min
Width, mm8595105115+5
Thickness, mm:
Finger0.05min
Palm0.05min
RemarkSimilar

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250.

ItemSubject devicePredicated deviceComparison
ColorantBlueBlueSame
PhysicalPropertiesBeforeAgingTensileStrength11MPa, minAverage 16.9MPaAnalysis
UltimateElongation300%minAverage 550%Analysis
AfterAgingTensileStrength11MPa, minAverage 14.4MPa, minAnalysis
UltimateElongation300%minAverage 500%Analysis
Comply with ASTM D5250Comply with ASTM D5250Comply with ASTM D5250Same
Freedom from HolesBe free from holes when tested in accordance with ASTM D5151 AQL=2.5Be free from holes when tested in accordance with ASTM D5151 AQL=2.5Same
Powder Content0.07 mg per glove.Meet the requirements ofASTM D6124Meet the requirements ofASTM D6124Similar

Table3 Performance Comparison

Analysis: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250

Table4 Safety Comparison

ItemSubject devicePredicated deviceComparison
MaterialPoly Vinyl ChloridePolyurethanePoly Vinyl ChloridePolyurethaneSimilar

{6}------------------------------------------------

NitrileDi-(2-ethylhexyl)Terephthalate(DOTP)Diisononyl Phthalate(DINP)
BiocompatibilityIrritationUnder the conditions of thestudy, not an irritantComply withISO10993-10SAME
SensitizationUnder conditions of thestudy, not a sensitizer.
CytotoxicityUnder conditions of thestudy, did not show potentialtoxicity to L-929 cells./Different
Label and LabelingMeet FDA's RequirementMeetRequirementFDA'sSAME

Analysis: The materials of the subject device are little different with that of the predicate, but they all meet the performance requirements of ASTM D5250,also biocompatibility test has been performed on subject device and the test result can meet the requirements of ISO 10993 standards.

8.0 Discussion of Non-clinical and Clinical Test Performed

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTMD5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

TestMethodologyPurposeAcceptance CriteriaResults
ASTM D5250PhysicalDimensionsTestLength(mm):≥230;Width(mm):S: 85±5;M: 95±5;L: 105±5;XL: 115±5;Length: >230Width:S: 84-89M: 93-97L: 105-107XL: 115-116Pass

{7}------------------------------------------------

Thickness (mm):Finger: ≥0.05Palm: ≥0.05Finger: 0.08-0.2Palm: 0.07-0.08Pass
ASTM D5151WatertightnessTest forDetection ofHolesMeet the requirements ofASTM D5151 AQL 2.50/125 leaksPass
ASTM D6124PowderContentMeet the requirements ofASTM D6124 < 2.0mg0.07 mg;Pass
ASTM D412PhysicalpropertiesBeforeAgingTensileStrength≥11MPa11.6-16.2Pass
UltimateElongation≥300%308-391Pass
AfterAgingTensileStrength≥11MPa11.0-14.0Pass
UltimateElongation≥300%323-337Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditionsof the study, didnot show potentialtoxicity to L-929cells.Pass
ISO 10993-10IrritationNon-irritatingUnder theconditions of thestudy, not anirritant.Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditionsof the study, not asensitizer.Pass

9.0 Clinical Test Conclusion

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device in K153028.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.