K Number
K211183

Validate with FDA (Live)

Date Cleared
2022-03-02

(316 days)

Product Code
Regulation Number
882.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sugita AVM Microclips are designed for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures.

Sugita AVM Microclips are not intended as a replacement for bipolar electrocoagulation of smaller vessels nor do they negate the need for immediate post-operative angiography following arteriovenous malformation surgery.

Device Description

The Sugita AVM Microclips were cleared by the FDA through 510(k) K960037. The Sugita AVM Microclips are manufactured from Elgiloy, a cobalt-chromiummolybdenum alloy. They are designed for the occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures. The Sugita AVM Microclips are available in the following sizes: 2mm, 3mm, 4mm and 5mm. Other than the modifications identified above, there have been no other changes to the Sugita AVM Microclips.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Subject Device K211183)Reported Device Performance (Sugita AVM Microclips)
MRI Safety - Magnetic Field Interaction (Force)Force no greater than gravitational field. (Tested according to ASTM F2052-15 - translational attraction determined to be 14°)
MRI Safety - Magnetic Field Interaction (Torque)Magnetically induced torque is not substantial and requires no further evaluation. (Tested according to ASTM F2213-17 – torque showed no movement or alignment)
MRI Safety - MRI Related HeatMaximum Whole Body Averaged SAR 2-W/Kg for 60 min. continuous RF exposure. (Tested according to ASTM F2182-19 – maximum temperature rise of 1.5°C after 15 min. of continuous scan)
MRI Safety - Imaging ArtifactThe presence of this implant produces an imaging artifact. Therefore, should carefully select pulse sequence parameters if the implant is located in the area of interest. (Tested according to ASTM F2119-2013 – image artifact extends approx. 10mm from implant with 3-Tesla gradient echo pulse sequence)
MRI Safety - Spatial Gradient Magnetic FieldTesting supports using a value of 2,000 gauss/cm (20-T/m) for the MRI related labeling of the Sugita AVM Microclips. (5mm Sugita AVM Microclips were attached to a porcine blood vessel in order to conduct a digital force gauge-based Pull-Test.)
Steam Sterilization ValidationValidated steam sterilization parameters are provided in the product labeling. (Tested according to ISO 17665-1 to validate steam sterilization parameters for the Sugita AVM Microclips using an FDA cleared nonwoven wrap to a sterility assurance level of (SAL) of 10-6.)
MR Conditional ConditionsMR Conditional for Use in Both 1.5T and 3T MR Systems. (Demonstrated through the above MRI safety testing, which included tests for 1.5T and 3T MR systems where applicable.) The predicate device was only cleared for 1.5T MR Systems, making this an improvement and a new acceptance criterion for the modified device.
Maintenance of Device Performance (Shelf-life)Performance data are not needed to establish maintenance of device performance over the shelf life of the Sugita AVM Microclips because the materials of construction of these products are highly stable metal alloys with a long history of use in surgical implants. (The device is provided non-sterile and constructed of inert materials, implying no degradation over time for its core functionality before use, as the device itself doesn't possess a shelf-life.)

2. Sample Size for the Test Set and Data Provenance

The provided text does not specify exact sample sizes for each MRI compatibility test beyond implying "Sugita AVM Microclips" were used for testing. For the Spatial Gradient Magnetic Field testing, it mentions "5mm Sugita AVM Microclips" were attached to "a porcine blood vessel." The provenance of this data is not explicitly stated as retrospective or prospective, nor is the country of origin. It can be inferred that these are prospective in-vitro tests conducted specifically for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

Not applicable. The tests performed are engineering-based performance tests (MRI compatibility, sterilization validation) and do not involve human expert interpretation or ground truth establishment in the traditional clinical sense. These tests rely on established scientific and engineering standards (e.g., ASTM, ISO).

4. Adjudication Method for the Test Set

Not applicable. As the tests are objective engineering measurements against predefined standards, there is no need for an adjudication method by human experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical medical implant (microclips), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study is irrelevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device.

7. The Type of Ground Truth Used

The "ground truth" for the performance tests relies on established engineering and scientific standards and measurements as defined by:

  • ASTM F2503-13 (Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment)
  • Guidance for Industry and Food and Drug Administration Staff (Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, document issued on May 20, 2021)
  • ASTM F2052-15 (Standard Test Method for Measurement of Magnetically Induced Translational Force on Medical Devices in the Magnetic Resonance Environment)
  • ASTM F2213-17 (Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment)
  • ASTM F2182-19 (Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging)
  • ASTM F2119-2013 (Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants)
  • ISO 17665-1 (Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices)

8. The Sample Size for the Training Set

Not applicable. This device undergoes physical performance testing, not machine learning model training.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

March 2, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Mizuho America, Inc. Richard Chadwick Senior Manager, Operations and QA 30057 Ahern Avenue Union City, California 94587

Re: K211183

Trade/Device Name: Sugita AVM Microclips Regulation Number: 21 CFR 882.5200 Regulation Name: Aneurysm Clip Regulatory Class: Class II Product Code: HCH Dated: January 25, 2022 Received: January 31, 2022

Dear Richard Chadwick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradvan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K211183

Device Name Sugita AVM Microclips

Indications for Use (Describe)

The Sugita AVM Microclips are designed for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures.

Sugita AVM Microclips are not intended as a replacement for bipolar electrocoagulation of smaller vessels nor do they negate the need for immediate post-operative angiography following arteriovenous malformation surgery.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

[X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Mizuho America, Inc. The logo features a blue circular graphic on the left, followed by the word "MIZUHO" in large, bold, blue letters. Below "MIZUHO" is the text "Medical Innovation" in a smaller font. Underneath that is the text "Mizuho America, Inc." in a similar font size.

K211183

510(k) SUMMARY

Applicant Information:

Owner Name:Mizuho America, Inc.
Address:30057 Ahern AvenueUnion City, CA 94587
Phone number:+1(510) 324-4500
Establishment Registration Number:1223656
Contact Person:Richard Chadwick
Date Prepared:February 28, 2022
Device Information:
Device Classification:Class II
Trade Name:Sugita AVM Microclips
Common name:Aneurysm Clip
Classification name:Aneurysm Clip
Regulation number:882.5200
Product Code:HCH

Predicate Device:

Sugita AVM Microclips (K960037).

Device Modification:

The following changes have been made to the Sugita AVM Microclips:

  • . The device has been demonstrated to be safe for imaging in both 1.5- Tesla and 3-Tesla MR systems.
  • New parameters for steam sterilization are provided for end users.

Device Description:

The Sugita AVM Microclips were cleared by the FDA through 510(k) K960037. The Sugita AVM Microclips are manufactured from Elgiloy, a cobalt-chromiummolybdenum alloy. They are designed for the occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures. The Sugita AVM Microclips are available in the following sizes: 2mm, 3mm, 4mm and 5mm. Other than the modifications identified above, there have been no other changes to the Sugita AVM Microclips.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Mizuho Medical Innovation. The logo features a blue circular graphic with white lines emanating from the center, resembling a stylized globe or sun. To the right of the graphic is the word "MIZUHO" in a bold, blue sans-serif font. Below "MIZUHO" are the words "Medical Innovation" in a smaller, lighter font, followed by "Mizuho America, Inc." in a similar font.

Indications for Use:

The Sugita AVM Microclips have the same Indications for Use as the predicate device.

The Sugita AVM Microclips are designed for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures.

Sugita AVM Microclips are not intended as a replacement for bipolar electrocoagulation of smaller vessels nor do they negate the need for immediate post-operative angiography following arteriovenous malformation surgery.

Comparison of Technological Characteristics to the Predicate Device:

Both the Sugita AVM Microclips and the predicate device have the same intended use. Both are intended for occlusion of small vessels in Arteriovenous Malformations (AVM's) and other similar structures.

The materials, design, technological characteristics and operating principles of the current device are unchanged from the predicate cleared through K960037. See the table below.

Substantial equivalence in materials, technological characteristics, and performance of the Sugita AVM Microclips to the predicate device is outlined in the table below:

ProductSugita AVM MicroclipsSugita AVM MicroclipsConclusion
510(k) numberSubject Device – K211183K960037
ManufacturerMizuho America, Inc.Mizuho America, Inc.
MaterialsElgiloy (cobalt-chromium-molybdenum alloy – ASTMF1058)Elgiloy (cobalt-chromium-molybdenum alloy – ASTMF1058)SAME
TechnologicalCharacteristics
Designed with apredeterminedholding forceHolding force of 50 to 70gHolding force of 50 to 70gSAME
Various Sizes forspecificrequirementsFour Sizes Available –2mm, 3mm, 4mm and 5mmFour Sizes Available –2mm, 3mm, 4mm and 5mmSAME
Delivery toSurgical SiteSugita Microclip Applierdesigned specifically forSugita AVM Microclips toaid in delivery to surgicalsiteMicroclip Applier designedspecifically for Sugita AVMMicroclips to aid in deliveryto surgical siteSAME

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Mizuho Medical Innovation. The logo consists of a circular graphic on the left and the word "MIZUHO" in blue, bold letters on the right. Below the word "MIZUHO" is the phrase "Medical Innovation" in smaller, lighter-colored letters. The circular graphic appears to be a stylized representation of a medical device or technology.

ProductSugita AVM MicroclipsSugita AVM Microclips
510(k) numberSubject Device – K211183K960037
ManufacturerMizuho America, Inc.Mizuho America, Inc.Conclusion
Steam Sterilizedby end userCycle Type: Pre-vacuum• Temperature: 132°C• Exposure Time: 4 min.• Drying Time: 20 min.• Temperature: 134°C• Exposure Time: 3 min.• Drying Time: 20 min.Cycle Type: Pre-vacuum• Temperature: 252°F• Exposure Time: 20 minutesSubstantiallyEquivalent
PerformanceTesting
MRI SafetyMR Conditional for Use inBoth 1.5T and 3T MRSystems Based onTesting:-Magnetic FieldInteractions-MRI Related Heating-MRI Artifact TestMR Conditional for Use in1.5T MR SystemsSubstantiallyEquivalent

Non-clinical Performance Testing

Performance testing was conducted to demonstrate the safety of the Sugita AVM Microclips for imaging in both 1.5-Tesla and 3-Tesla MR Systems and to provide labeling for MRI safety of the device conforming to ASTM F2503-13 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment and Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, Guidance for Industry and Food and Drug Administration Staff, document issued on May 20, 2021.

Additionally, sterilization validation testing was conducted to provide recommended steam sterilization parameters for the device.

TestTest SummaryConclusions
MRI Compatibility Testing –Magnetic Field Interaction (force)Testing according to ASTMF2052-15 -translationalattraction determined to be 14°.Force no greater thangravitational field.
MRI Compatibility Testing –Magnetic Field Interaction(torque)Testing according to ASTMF2213-17 – torque showed nomovement or alignment.Magnetically induced torque isnot substantial and requires nofurther evaluation.
MRI Compatibility Testing – MRIRelated Heat TestingTesting according to ASTMF2182-19 – maximumtemperature rise of 1.5°C after 15min. of continuous scan.Maximum Whole Body AveragedSAR 2-W/Kg for 60 min.continuous RF exposure.

Non-clinical Performance Testing Summary

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Mizuho Medical Innovation. The logo consists of a blue circular graphic on the left, followed by the word "MIZUHO" in large, bold, blue letters. Below the word "MIZUHO" are the words "Medical Innovation" in smaller, thinner, black letters. The circular graphic on the left appears to be a stylized representation of a globe or sphere with curved lines.

MIZUNO America, Inc.MRI Compatibility Testing -Imaging Artifact TestingTesting according to ASTMF2119-2013 – image artifactextends approx. 10mm fromimplant with 3-Tesla gradientecho pulse sequence.The presence of this implantproduces an imaging artifact.Therefore, should carefully selectpulse sequence parameters if theimplant is located in the area ofinterest.
MRI Compatibility Testing -Spatial Gradient Magnetic FieldTesting5mm Sugita AVM Microclips wereattached to a porcine blood vesselin order to conduct a digital forcegauge-based Pull-Test.Testing supports using a value of2,000 gauss/cm (20-T/m) for theMRI related labeling of theSugita AVM Microclips.
Steam Sterilization ValidationTestingTesting according to ISO 17665-1 was performed to validatesteam sterilization parameters forthe Sugita AVM Microclips usingan FDA cleared nonwoven wrapto a sterility assurance level of(SAL) of 10-6.Validated steam sterilizationparameters are provided in theproduct labeling.

Shelf-life

The Sugita AVM Microclips do not have a shelf life because they are provided non-sterile and are constructed of inert materials. Performance data are not needed to establish maintenance of device performance over the shelf life of the Sugita AVM Microclips because the materials of construction of these products are highly stable metal allovs with a long history of use in surgical implants.

Biocompatibility

The proposed changes do not impact the contact duration or biocompatibility profile of the Sugita AVM Microclips.

Conclusion

The modified Sugita AVM Microclips subject to this submission are substantially equivalent to the predicate. Sugita AVM Microclips, in the following ways: they have the same intended use and Indications for Use, the same technological characteristics and operating principles, and incorporate the same design and materials.

Performance testing has demonstrated that the Sugita AVM Microclips are substantially equivalent to the predicate and are safe for imaging in both 1.5- Tesla and 3-Tesla MR systems. In addition, validated parameters for steam sterilization and cleaning are provided to update the instructions for end users of the device.

Page 4 of 4

§ 882.5200 Aneurysm clip.

(a)
Identification. An aneurysm clip is a device used to occlude an intracranial aneurysm (a balloonlike sac formed on a blood vessel) to prevent it from bleeding or bursting.(b)
Classification. Class II (performance standards).