K Number
K211135

Validate with FDA (Live)

Device Name
Enlight 2100
Manufacturer
Date Cleared
2022-01-06

(265 days)

Product Code
Regulation Number
868.1505
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ENLIGHT 2100 is a non-invasive, non-radiation medical device that provides information of local impedance variation within a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's lungs.

It is intended for mechanically ventilated adult and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 - 134cm.

ENLIGHT 2100 does not measure regional ventilation of the lungs.

Device Description

ENLIGHT 2100 is a Ventilatory electrical impedance tomograph that uses several electrodes (usually between 16 and 32) placed around the patient's thorax to assess regional impedance variation in a lung slice (tomography). It provides a relative measurement, so it only provides information on variations in local impedance.

ENLIGHT 2100 estimates Local Impedance Variation, occurring in a cross section of the thorax during a respiratory cycles, and which are linearly related to Variations in Regional Air Content within the lung.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) clearance letter for the Timpel S.A. Enlight 2100 device:

Important Note: The provided document is a 510(k) clearance letter from the FDA. 510(k) clearance is based on demonstrating substantial equivalence to a previously legally marketed device (predicate device), not necessarily on rigorous clinical effectiveness studies or extensive AI model validation metrics typically seen for novel AI/ML devices. Therefore, the details around specific acceptance criteria, test sets, ground truth establishment, and MRMC studies for AI performance, as might be expected for an AI/ML software as a medical device (SaMD), are not explicitly detailed in this type of regulatory document. The focus here is on demonstrating that the new device (Enlight 2100) is sufficiently similar to the predicate (Enlight 1810) in terms of safety and performance.

Analysis of Acceptance Criteria and Device Performance

The provided document describes bench testing to demonstrate substantial equivalence, rather than clinical studies with human readers or AI-specific performance metrics like sensitivity/specificity for a diagnostic task. The acceptance criteria are implicit in the comparison table (Table 5.1) between the subject device (Enlight 2100) and the predicate device (Enlight 1810). The "Explanation of Differences" column directly indicates whether the performance attributes are similar or if one is slightly better, implying the new device meets or exceeds the predicate's performance.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied: Performance comparable or better than Predicate Device)

AttributeSubject (ENLIGHT 2100) PerformancePredicate (ENLIGHT 1810) PerformanceExplanation of Differences (Meeting Criteria)
Performance Characteristics – Bench Test with Electrode Belt size S
Signal to Noise Ratio (SNR)(50dB - 95dB)(50dB - 85dB)Met/Exceeded: ENLIGHT 2100 has slightly better performance considering the Signal to Noise Ratio.
Voltage Accuracy(80% - 100%)(85% - 100%)Similar: Voltage Accuracy and Reciprocity Accuracy of ENLIGHT 2100 are similar to ENLIGHT 1810. The document notes imaging quality is not reflected by these absolute parameters due to the normalized difference voltage calculation.
DriftAllan Variance converges to zero (below 100pV2)Allan Variance converges to zero (below 100pV2)Similar: The drift of both devices is similar, as Allan Variance of both devices converge to less than 100pV2, which is negligible.
Reciprocity Accuracy(95% - 100%)(96% - 100%)Similar: Voltage Accuracy and Reciprocity Accuracy of ENLIGHT 2100 are similar to ENLIGHT 1810.
Amplitude response(90% - 104%)(94% - 106%)Similar: ENLIGHT 2100 US Infant and ENLIGHT 1810 have similar performance in all parameters related to imaging quality with no significant differences. The context implies that these ranges are acceptable and comparable.
Position errorSmaller than 4% of the radiusSmaller than 4% of the radiusSimilar: ENLIGHT 2100 US Infant and ENLIGHT 1810 have similar performance in all parameters related to imaging quality with no significant differences. Performance is within the acceptable range defined by the predicate.
RingingSmaller than 0.6Smaller than 0.6Similar: ENLIGHT 2100 US Infant and ENLIGHT 1810 have similar performance in all parameters related to imaging quality with no significant differences. Performance is within the acceptable range defined by the predicate.
ResolutionSmaller than 0.42Smaller than 0.42Similar: ENLIGHT 2100 US Infant and ENLIGHT 1810 have similar performance in all parameters related to imaging quality with no significant differences. Performance is within the acceptable range defined by the predicate.
Percentage error of PlethysmogramBelow 5%Below 5%Similar: ENLIGHT 2100 and ENLIGHT 1810 have similar performance with error below 5%. Performance is within the acceptable range defined by the predicate.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify a "test set" in terms of patient data or a specific number of unique tests for each performance characteristic. Instead, it refers to "bench testing" which implies testing on a controlled system or phantom, not human subjects, for these specific performance metrics. The comparison table focuses on instrument performance rather than AI model performance on a clinical dataset.
  • Data Provenance: Not specified, as these are bench test results of the device itself rather than clinical data from a specific region or patient cohort. It is implicitly "prospective" bench testing of the new device.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Not applicable for this type of submission. The "ground truth" for the performance characteristics listed in the table comes from verifiable engineering measurements on the device, often using calibrated equipment or phantoms, not human expert consensus. This is a technical performance evaluation, not a clinical diagnostic assessment.

4. Adjudication Method for the Test Set

  • Not applicable. As noted above, the "test set" refers to technical performance characteristics measured through bench testing, not a dataset requiring human adjudication of clinical findings.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done as described for AI assistance. This submission is for an Electrical Impedance Tomograph (EIT) device, which provides "information of local impedance variation" as an "adjunctive tool." It does not involve AI for interpretation or diagnosis, nor does it directly assist human readers in image interpretation in the way an AI-powered diagnostic tool would. It provides raw data/images for clinicians to interpret.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, in spirit, the bench testing results represent a "standalone" evaluation of the device's technical performance. The reported performance characteristics (SNR, Voltage Accuracy, Drift, etc.) are inherent properties of the device and its internal algorithms, independent of human interaction or interpretation beyond setting up the test. However, this is not an "algorithm-only" study in the typical sense of a diagnostic AI algorithm that produces a specific output (e.g., classifying a disease). The device itself is the "algorithm" and hardware working together to produce data.

7. Type of Ground Truth Used

  • The ground truth for the "Performance Characteristics" in Table 5.1 is based on engineering measurements and calibrations using established methods for evaluating electronic medical devices. It is not expert consensus, pathology, or outcomes data. For example, SNR is measured against a known signal input, and position error might be measured using phantoms with known, precisely located impedance changes.

8. Sample Size for the Training Set

  • Not applicable. The Enlight 2100 is an EIT device that provides direct physical measurements and derived images based on those measurements, not a machine learning model that requires a "training set" of data in the typical sense. Its "algorithm" is based on the physics of electrical impedance tomography.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no "training set" for an AI model. The device's operation is based on established physical principles and engineering design.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

January 6, 2022

Timpel S.A. % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE Saint Petersburg, Florida 33704

Re: K211135

Trade/Device Name: Enlight 2100 Regulation Number: 21 CFR 868.1505 Regulation Name: Ventilatory Electrical Impedance Tomograph Regulatory Class: Class II Product Code: OEB Dated: December 6, 2021 Received: December 8, 2021

Dear Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K21135

Device Name

Enlight 2100

Indications for Use (Describe)

ENLIGHT 2100 is a non-invasive, non-radiation medical device that provides information of local impedance variation within a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's lungs.

It is intended for mechanically ventilated adult and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 - 134cm.

ENLIGHT 2100 does not measure regional ventilation of the lungs.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "

FORM FDA 3881 (6/20)

Page 1 of 1

{3}------------------------------------------------

Date Prepared:6-Jan-22
Sponsor:TIMPEL S.A.Rua Simão Álvares, 356 Cj. 41/42Pinheiros, São Paulo/SP - BrazilT - +55-113088-0305E - Rafael.holzhacker@timpelmedical.com
Submission Correspondent:Paul DrydenProMedic, LLC131 Bay Point Dr NESt. Petersburg, FL 33704
Proprietary or Trade Name:ENLIGHT 2100 US
Regulation Number:Regulation Name:Product code:21 CFR 868.1505Ventilatory electrical impedance tomographQEB
Predicate Device:Timpel – ENLIGHT 1810 – DEN170072

Device Description:

ENLIGHT 2100 is a Ventilatory electrical impedance tomograph that uses several electrodes (usually between 16 and 32) placed around the patient's thorax to assess regional impedance variation in a lung slice (tomography). It provides a relative measurement, so it only provides information on variations in local impedance.

ENLIGHT 2100 estimates Local Impedance Variation, occurring in a cross section of the thorax during a respiratory cycles, and which are linearly related to Variations in Regional Air Content within the lung.

Principle of Operation:

Electrical impedance tomography (EIT) is a technique in which the electrical properties of tissues are estimated from surface electrode voltage measurements and used to provide information on Local Impedance Variation (LIV) within a cross section of a patient's thorax. Considerable electrical impedivity variations are imposed on lung tissue by variations in the lung's air content.

There is a linear relationship between the variations in air content and the percentage change in lung tissue impedance. This is a characteristic of lung tissue due to the properties of alveoli walls. This linear relationship is explored in Electrical Impedance Tomography, supporting its use as a tool to support the user's assessment of the variations in regional air content within a cross section of a patient's lungs.

The electrode voltage measurement can be made at high rates as 50 times / second, making it possible to display Local Impedance Variations with a high temporal resolution, almost continuously.

Indications for Use:

ENLIGHT 2100 is a non-invasive, non-radiation medical device that provides information of local impedance variation within a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to

{4}------------------------------------------------

support the user's assessment of variations in regional air content within a cross section of a patient's lungs.

It is intended for mechanically ventilated adult and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 - 134 cm.

ENLIGHT 2100 does not measure regional ventilation of the lungs.

Patient Population:

Mechanically ventilated adult and pediatric patients whose thorax perimeter is within the range of 37.5 -134 cm.

Environments of use:

Hospital setting.

Substantial Equivalence Discussion

Indications - There are no differences in the indication for use, besides the patient population.

Patient Population - The intended Patient Population for ENLIGHT 2100 are adult and pediatric patients, whose thorax perimeter is within the range of 37.5 - 134 cm. ENLIGHT 2100 is indicated for mechanically ventilated patients, so the patient is usually sedated, and the application of the electrode belt is performed in uncooperative patients, regardless being adult or pediatric patients. There are no differences in the algorithm or software between different sizes of electrode belt, only the reconstruction meshes are different, but with the same performance.

Environment of Use - There is no difference in the environment of use.

Technological Characteristics - The Operating Principle is the same for ENLIGHT 2100 and ENLIGHT 1810. The hardware system was redesigned, reducing the module volume with improved hardware lavout. The digital part of the hardware was redesigned and the analog hardware (driven shields and common mode circuits of acquisition channels) have the same wellproven topology and only layout was improved.

The enclosure also includes the pneumotachograph OEM Module (the same part number used in ENLIGHT 1810) and DC-DC power supplies, that were previously presented within other mechanical enclosures - Hub and power supply enclosure, respectively.

Improvements on power dissipation were implemented, reducing static consumption of operational amplifiers by reducing the voltage of power supplies without compromising performance. ENLIGHT 1810 OEM AC-DC power supply was replaced by a lower power version OEM module of same manufacturer, as the power consumption of ENLIGHT 2100 US is lower than its predecessor. There is no battery in ENLIGHT 2100 US.

{5}------------------------------------------------

Non-clinical testing

Biocompatibility – There is no difference in the patient contact between the proposed device and the predicate.

Bench testing - We have performed the following performance tests, with similar results between the 2 devices.

Substantial Equivalence Conclusion

Data presented indicated that there is substantial equivalence between ENLIGHT 2100 and ENLIGHT 1810.

{6}------------------------------------------------

Explanation of Differences
AttributesSubjectENLIGHT 2100PredicateENLIGHT 1810
510(k)DEN170072
Product ClassificationCFR868.1505 - QEB868.1505 - QEBSimilar
Indications for UseENLIGHT 1810 is a non-invasive, non-radiation medical device that providesinformation of local impedance variationwithin a cross-section of a patient's thorax.This information is presented to the clinicianuser as an adjunctive tool to other clinicalinformation in order to support the user'sassessment of variations in regional air contentwithin a cross section of a patient's lungs.It is intended for mechanically ventilated adultand pediatric patients in a hospital setting,whose thorax perimeter is within the range of37.5 -134 cm.ENLIGHT 2100 does not measure regionalventilation of the lungs.ENLIGHT 2100 is a non-invasive, non-radiation medical device that providesinformation of local impedance variationwithin a cross-section of a patient's thorax.This information is presented to the clinicianuser as an adjunctive tool to other clinicalinformation in order to support the user'sassessment of variations in regional air contentwithin a cross section of a patient's lungs.It is intended for mechanically ventilated adultpatients in a hospital setting, whose thoraxperimeter is within the range of 78-122 cm.Inclusion of new sizes of accessories andsoftware modifications to expand the patientrange.
Patient PopulationPediatric patients, whose thorax perimeter iswithin the range of 37.5 -134 cm.Adult patients, whose thorax perimeter iswithin the range of 78-122 cm.Inclusion of pediatric patients.
Patient typeMechanically ventilated patientsMechanically ventilated patientsNo differences.
PrescriptiveYesYesNo differences.
Principle of OperationElectrical Impedance Tomography based onvoltage measures to estimate local impedancevariation within a cross-section of a patient'sthorax.Electrical Impedance Tomography based onvoltage measures to estimate local impedancevariation within a cross-section of a patient'sthorax.No differences.
ContraindicationsDeclared on Instructions for useDeclared on Instructions for useNo differences.
Environment of UseHospital SettingHospital SettingNo differences.
Duration of UseUp to 30 days, with Addere Change each 48hours.Up to 30 days, with Addere Change each 48hours.No differences.
AttributesSubjectENLIGHT 2100PredicateENLIGHT 1810Explanation of Differences
Useful lifeENLIGHT 2100 has 7 years of useful life.ENLIGHT 1810 has 7 years of useful life.No differences.
Shelf lifeAddere has 2 year of shelf life.Electrode Belt has 1 year of shelf life.Addere has 1 year of shelf life.Electrode Belt has 1 year of shelf life.There is no change of shelf life for theElectrode Belts, but new validation test datawith real-time aged samples increased Addereshelf life from 1 to 2 years.
Non-sterileThere are no sterile components or accessories.There are no sterile components or accessories.No differences.
Cleaning methodsCleaning and disinfection prescribed only forthe device.Accessories are single patient use.Cleaning and disinfection prescribed only forthe device.Accessories are single patient use.No differences.
Features
Available sizesElectrode Belt sizes P0, P1, P2, 4S, 5S, XXS,XS, S, M, L, XLAddere sizes P0, P1, P2, 4S, 5S, XXS, XS, S,M, L, XLElectrode Belt sizes XS, S, M, LAddere sizes XS, S, M, LNew Electrode Belts and Addere sizesincluded to expand the patient range.
Shaper No 2Reference Cable (Single size)Shaper No 2Reference Cable (Single size)
ShapeMonitor-like device, transportableMonitor-like device, mobile, with incorporatedcartDevice size is smaller to facilitate the shippinglogistics and device handling inside thehospital.
Patient Contact perISO 10993-1Surface contactIntact skinProlonged duration (Up to 30 days)Surface contactIntact skinProlonged duration (Up to 30 days)No differences.
AttributesSubjectENLIGHT 2100PredicateENLIGHT 1810Explanation of Differences
Performance Characteristics – Bench Test with Electrode Belt size S
Signal to Noise Ratio(SNR)(50dB - 95dB)(50dB - 85dB)SNR and Drift are both parameters that reflectthe precision of data acquisition. SNRindicates the system noise in data acquisition.
Voltage Accuracy(80% - 100%)(85% - 100%)indicates the system noise in data acquisition.
DriftAllan Variance converges to zero (below100pV2)Allan Variance converges to zero (below100pV2)Drift indicates a long-term stability in dataacquisition.ENLIGHT 2100 has slight better performanceconsidering the Signal to Noise Ratio whencompared to ENLIGHT 1810. The drift ofboth devices is similar, as Allan Variance ofboth devices converge to less than 100pV2,which is a negligible value when consideringthe background noise measured by SNR.ENLIGHT computes normalized differencevoltages from the acquired voltages togenerate local impedance variation images andnot on the absolute voltages itself. Thesubtraction of a reference voltage reduces theeffect of hardware imperfections related tobias and the normalization by a referencevoltage reduces the effect of imperfectionsrelated to gain. Voltage Accuracy andReciprocity Accuracy reflect thoseimperfections even though attenuated by theimaging algorithm.Voltage Accuracy and Reciprocity Accuracyof ENLIGHT 2100 are similar to ENLIGHT1810, but as indicated above, the imaging
Reciprocity Accuracy(95% - 100%)(96% - 100%)quality is not reflected by these parameters
AttributesSubjectENLIGHT 2100PredicateENLIGHT 1810Explanation of Differences
Amplitude response(90% - 104%)(94% - 106%)ENLIGHT 2100 US Infant and ENLIGHT1810 have similar performance in all
Position errorSmaller than 4% of the radiusSmaller than 4% of the radiusparameters related to imaging quality with no
RingingSmaller than 0.6Smaller than 0.6significant differences
ResolutionSmaller than 0.42Smaller than 0.42
Percentage error ofPlethysmogramBelow 5%Below 5%ENLIGHT 2100 and ENLIGHT 1810 havesimilar performance with error below 5%.

Table 5.1 – Comparison – Subject vs. Predicate

{7}------------------------------------------------

{8}------------------------------------------------

510(k) Summary Page 6 of 7

{9}------------------------------------------------

510(k) Summary Page 7 of 7

§ 868.1505 Ventilatory electrical impedance tomograph.

(a)
Identification. A ventilatory electrical impedance tomograph is a prescription non-invasive, non-radiological ventilatory device that provides an assessment of local impedance variation within a cross-section of a patient's thorax.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following:
(i) Characterization of device parameters, including signal-to-noise ratio, voltage accuracy, drift, reciprocity accuracy, amplitude response, position error, and ringing;
(ii) Real time evaluation of local impedance variation;
(iii) Plethysmogram accuracy testing; and
(iv) Use life testing of reusable components.
(3) Performance data must validate reprocessing instructions for any reusable components of the device.
(4) Performance data must demonstrate the electrical, thermal, and mechanical safety and the electromagnetic compatibility of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include the following:
(i) Guidance for interpretation of the images generated;
(ii) A warning that the device should be removed before use of a defibrillator, or defibrillator interaction information based on defibrillator performance testing with the device;
(iii) A use life for any reusable components; and
(iv) Instructions for reprocessing any reusable components.