K Number
K211091

Validate with FDA (Live)

Device Name
Deka Tiac II
Date Cleared
2021-10-15

(186 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DEKA TIAC II RF device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The DEKA TIAC II massage device is intended to provide a temporary reduction in the appearance of cellulite.

Device Description

The DEKA TIAC II is a medical device that used high frequency RF currents to produce a localized increase in temperature in subcutaneous tissue, for temporary relief of pain and muscle spasm and increase local circulation. The device has integrated massaging handpieces intended to provide a temporary reduction in the appearance of cellulite.

The DEKA TIAC II consists of the following main components:

  • microprocessor-driven control unit
  • high-frequency electromagnetic energy generator
  • user interface with 10.4" color touch screen
  • 2 RF handpieces for application of radiofrequency
  • 2 integrated massaging balls handpieces

Contact quality monitoring system is present for monitoring of the contact quality under the RF handpiece electrodes.

The electrical specifications are 100-240Vac, 50/60Hz, 1500VA.

AI/ML Overview

This document describes the DEKA TIAC II device, a medical radio frequency and massage device intended for temporary relief of pain, muscle spasms, increased local circulation, and temporary reduction in the appearance of cellulite. The study proving the device meets acceptance criteria is detailed in the Non-Clinical Performance Data section on pages 6 and 7 of the provided document.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Test PerformedAcceptance CriteriaReported Device Performance
Non-clinical - Tissue heating performance- The sample temperature in the range 40-43°C for at least for 10 minutes in the active zone of the handpieces (hot spots);- Skin surface increase in temperature not greater than 3°C;- Temperature outside the hot spot never exceeding 40°C;- Dissipation phase not increasing the temperature at the end of the treatment.- Number of planned tests: 2- Number of tests executed: 2- Number of positive outcome: 2 (100%)- Device able to heat the target tissue maintaining the tissue temperature in the range 40°C - 43°C for at least 10 minutes in the active zone, with no significant temperature increase out of depth.
Non-clinical – Human skin temperature maintaining performance- Skin temperature is maintained in the range 40-45°C for at least 10 minutes;- Skin temperature doesn't exceed 45°C over the entire test period.- Number of planned tests: 36- Number of tests executed: 36- Number of positive outcome: 36 (100%)- Device able to keep the skin surface temperature in the range 40º-45ºC for at least 10 minutes.

2. Sample Size Used for the Test Set and the Data Provenance:

  • Non-clinical Tissue heating performance:

    • Sample Size: Ex-vivo pig muscle sample (20x20x5 cm). Two tests were executed (2 for Deep handpiece and 2 for Shallow handpiece).
    • Data Provenance: Not explicitly stated, but "ex-vivo pig muscle sample" implies a laboratory setting, likely in Italy where the submitter is located. This is retrospective as the tests were performed before submission.
  • Non-clinical Human skin temperature maintaining performance:

    • Sample Size: 3 people of different skin types (Fitzpatrick I, III, IV). 36 tests were executed (e.g., each handpiece on each person, across three body areas at different power settings).
    • Data Provenance: Not explicitly stated, but likely conducted in Italy. This is retrospective as the tests were performed before submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

This information is not provided in the document. The tests performed are objective measurements of temperature, not subjective evaluations requiring expert consensus for ground truth.

4. Adjudication Method for the Test Set:

This information is not applicable as the tests performed are objective measurements with pre-defined acceptance criteria, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The DEKA TIAC II is a medical device for heating tissue and massage, not an AI-powered diagnostic or interpretive tool that would involve human readers. The document states "Clinical Performance Data: None."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. The DEKA TIAC II is a physical medical device. The "tests performed" are evaluations of the device's physical output and effect on tissue/skin, not autonomous algorithmic performance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

  • Non-clinical Tissue heating performance: The ground truth was direct temperature measurement within the tissue and on the skin surface.
  • Non-clinical Human skin temperature maintaining performance: The ground truth was direct skin surface temperature measurement.

8. The Sample Size for the Training Set:

This information is not applicable/not provided. This device is not an AI/machine learning device that requires a training set in the conventional sense. The "training" for the device's design and engineering would be based on general biophysics, engineering principles, and potentially prior device development, rather than a specific labeled dataset.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable/not provided for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 15, 2021

El.En. Electronic Engineering S.p.A. Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, Firenze 50041 Italy

Re: K211091

Trade/Device Name: DEKA TIAC II Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: PBX Dated: September 10, 2021 Received: September 13, 2021

Dear Paolo Peruzzi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211091

Device Name DEKA TIAC II

Indications for Use (Describe)

The DEKA TIAC II RF device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The DEKA TIAC II massage device is intended to provide a temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

DEKA TIAC II

Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

Date Summary Prepared:

October 15, 2021

Device Trade Name:

Deka TIAC II

Common Name:

Medical Radio Frequency and massage device

Classification Name:

Massager, vacuum, radio frequency induced heat (PBX)

Classification Number:

21 CFR 878.4400

Predicate Devices:

DEKA TIAC 1 (K183371)

{4}------------------------------------------------

Device Description:

The DEKA TIAC II is a medical device that used high frequency RF currents to produce a localized increase in temperature in subcutaneous tissue, for temporary relief of pain and muscle spasm and increase local circulation. The device has integrated massaging handpieces intended to provide a temporary reduction in the appearance of cellulite.

The DEKA TIAC II consists of the following main components:

  • microprocessor-driven control unit
  • high-frequency electromagnetic energy generator
  • user interface with 10.4" color touch screen
  • 2 RF handpieces for application of radiofrequency
  • 2 integrated massaging balls handpieces

Contact quality monitoring system is present for monitoring of the contact quality under the RF handpiece electrodes.

The electrical specifications are 100-240Vac, 50/60Hz, 1500VA.

Indication for Use:

The DEKA TIAC II RF device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The DEKA TIAC II massage device is intended to provide a temporary reduction in the appearance of cellulite.

Substantial equivalence discussion:

The DEKA TIAC II is substantially equivalent to the DEKA TIAC 1 (K183371).

FeatureProposed 510(k) DevicePredicate Device
DEKA TIAC IIK183371DEKA TIAC 1
Device Trade NameProposed 510(k) DeviceDEKA TIAC 1Proposed 510(k) DeviceDEKA TIAC 1
FeatureProposed 510(k) DeviceDEKA TIAC IIPredicate DeviceK183371DEKA TIAC 1
Indications for UseThe DEKA TIAC II RF device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.The DEKA TIAC II massage device is intended to provide a temporary reduction in the appearance of cellulite.The DEKA TIAC1 RF device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.The DEKA TIAC 1 massage device is intended to provide a temporary reduction in the appearance of cellulite.
Regulation number21 CFR 878.4400: Electrosurgical cutting and coagulation21 CFR 878.4400: Electrosurgical cutting and coagulation
Product CodePBXPBX
Device Technologies• Application of the heat to the tissue via RF energy.• Mechanical Massaging of body parts• Application of the heat to the tissue via RF energy.• Mechanical Massaging of body parts
RF Maximum output power200W120W
RF mode of operationBipolarBipolar
RF Output Frequency2.45 GHz2.45 GHz
Effective temperature40-43°C40-43°C
Contact quality monitoring systemYESYES
Massage handpiece dimensionsHandpiece 1: Diameter: 68 mmHandpiece 2: Diameter: 78 mmHandpiece 1: Diameter: 89 mmHandpiece 2: Diameter: 99 mm
Massage handpiece materialAISI 316PPSU

{5}------------------------------------------------

K211091

The DEKA TIAC II has the same indications for use as the abovementioned predicate device, with same principle of operation and same performances.

{6}------------------------------------------------

Clinical Performance Data:

None

Non-Clinical Performance Data:

The following verification and validation activities have been performed on the modified device:

। Test according to AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, Cl:2009/(R)2012 and A2:2010/(R)2012 – Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance.

Test according to IEC 60601-1-2 Ed. 4 :2014- Medical electrical equipment – part 1-2: General । Requirements for Basic Safety and Essential Performance– Collateral standard: Electromagnetic Disturbances - Requirements and tests.

Test according to IEC 60601-2-6:2012+A1:2016- Medical electrical equipment - Part 2-6: Particular requirements for basic safety and essential performance of microwave therapy equipment.

Non-clinical performance tests on ex-vivo animal tissue, in order to show that DEKA TIAC II RF device is able to maintain the tissue temperature in the range 40°C - 43°C for at least 5 minutes, at the maximum setting of RF output power (same test method used for the predicate device).

– Non-clinical human tests, in order to demonstrate that DEKA TIAC II RF device is capable of maintaining a skin surface temperature of 409-459C for at least 10 minutes when using the device as intended (same test method used for the predicate device).

Verification of biocompatibility of AISI 316 material -

Non-clinical tissue-heating and human tests conducted on TIAC II are summarized in the table below.

TestsPerformedDevice Description/Sample SizeTest method/Applicable StandardsAcceptance CriteriaUnexpectedResults/SignificantDeviationsResults
Non-clinical -Tissue heatingperformanceBoth Deep handpieceand Shallow handpiecetested on 20x20x5 cmex-vivo pig musclesample. Tissue sampleimmersed in a basinwith waterthermostated at 37 °C.Worst case conditionsfor both handpieces:• Highest RF Power(200W)• Lowest skin coolingeffect• Smallest treatedarea• Highest skinfeedbackTemperature.• The sampletemperature in therange 40-43°C for atleast for 10 minutes inthe active zone of thehandpieces (hot spots);• skin surface increase intemperature notgreater than 3°C.• temperature outsidethe hot spot neverexceeding 40°C.• dissipation phase notincreasing thetemperature at the endof the treatmentNone• Number of plannedtests: 2• Number of testsexecuted: 2• Number of positiveoutcome: 2 (100%)

{7}------------------------------------------------

Non-clinical –human skintemperaturemaintainingperformanceBoth Deep handpieceand Shallow handpiecetested on 3 people ofdifferent skin types(Fitzpatrick I, III, IV), onthree areas of the bodythat are consistent withinstructions andindications for use(abdomen, arm, thigh).The room was air-conditioned at aconstant temperature of22°C.• Both handpiecestested at lowest(20W) and highest(200W) RF power• skin cooling 30°C• 15x15cm treatedarea• Feedbackthreshold 42°C• Smooth andcontinuousmovements on thetreatment area• skin temperature ismaintained in the range40-45°C for at least 10minutes• skin temperaturedoesn't exceed 45°Cover the entire testperiodNone• Number of plannedtests: 36• Number of testsexecuted: 36• Number of positiveoutcome: 36 (100%)

The tests performed on ex-vivo animal tissue show that TIAC II RF device is able to heat the target tissue maintaining the tissue temperature in the range 40°C - 43°C for at least 10 minutes in the active zone of the handpieces, while not showing any significant increase in temperature out of the depth of the handpieces.

Non-clinical human tests show that that DEKA TIAC II RF device is able to keep the skin surface temperature in the range 40º-45ºC for at least 10 minutes when using the device as intended.

The non-clinical tests have been performed with the same test methods and acceptance criteria used for the predicate device.

Conclusion:

Based on the comparison of indications for use and the technological characteristics, and on the outcome of non-clinical performance data provided, we can conclude that DEKA TIAC II is as safe, as effective, and performs as well as the legally marketed predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.