K Number
K211068

Validate with FDA (Live)

Device Name
PrevisEA Device
Manufacturer
Date Cleared
2021-08-04

(114 days)

Product Code
Regulation Number
870.1875
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PrevisEA is a compact, non-invasive device placed on the abdomen to capture the digestive sounds of patients. PrevisEA displays information which assists physicians in assessing the digestive health of patients. This device is for prescription use only and should be used under the direction of a licensed healthcare practitioner. The PrevisEA device has not been tested for and it is not intended for pediatric use.

Device Description

PrevisEA is a non-invasive, self-contained, medical device with a proprietary algorithm used to assess the digestive health of patients. The device includes a stethoscope diaphragm, sound chamber, embedded microphone and CPU, as well as a display screen and buttons. The PrevisEA device is a convenient, single-use, disposable unit, which attaches to the abdomen of the patient via an adhesive wafer on the back of the device. The PrevisEA is a single-use, disposable unit, which avoids the need for cleaning and disinfection. The PrevisEA requires no cable connections for operation of the device. Integration of the above components allow auscultation and digital capture of digestive sounds, such as MH4. Messages are transmitted to the easy-to-read display screen on the front of the device. The PrevisEA detects the acoustic biomarker MH4 and quantifies the number of MH4 detections over a 4-minute interval, which it compares to a predefined threshold to determine the presence of abnormal digestive health in patients following intestinal surgery, such as gastrointestinal impairment (GII).

AI/ML Overview

Based on the provided text, here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and the Reported Device Performance:

The document doesn't explicitly state a table of quantifiable acceptance criteria with corresponding performance metrics for the clinical effectiveness of the PrevisEA device. However, it does mention that "All testing met acceptance criteria" for electrical safety, electromagnetic compatibility, and battery safety. For clinical performance, the acceptance criteria are implicitly tied to demonstrating the device's capability in "listening/capturing the acoustic biomarker, MH4 while in clinical use" and validating its intended use in assessing digestive health.

CategoryAcceptance Criteria (Implicit/Stated)Reported Device Performance
BiocompatibilityConformance to ISO 10993-1:2018Testing demonstrated PrevisEA Device is biocompatible.
Electrical SafetyConformance to safety standards (e.g., IEC60601 1-1:2005 3rd Ed.)All testing met acceptance criteria and demonstrated safety.
ElectromagneticConformance to EMC standards (e.g., EN60601-1-2, 2007/03)All testing met acceptance criteria and demonstrated EMC.
Battery SafetyConformance to battery safety standardsAll testing met acceptance criteria and demonstrated safety.
Auscultation Performance (Bench Testing)Demonstrated conformityPassed/Conforms
Clinical PerformanceCapability in listening/capturing MH4 acoustic biomarker in vivo; validation of intended use.Demonstrated capability of capturing MH4 while in clinical use; validated intended use.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: 75 subjects
  • Data Provenance: The study involved subjects undergoing intestinal resection surgery, which implies a clinical setting. While the specific country of origin is not explicitly stated, the FDA clearance suggests it was likely conducted in the US or under protocols acceptable to the FDA. The study is prospective as devices were applied at the completion of surgery.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

The document does not specify the number of experts used to establish the ground truth for the test set or their qualifications. The output of the PrevisEA device is "information which assists physicians in assessing the digestive health of patients," and it "quantifies the number of MH4 detections over a 4-minute interval, which it compares to a predefined threshold to determine the presence of abnormal digestive health." The "ground truth" seems to be related to the device's ability to detect MH4 and, implicitly, its correlation with digestive health post-surgery. However, the method for establishing this ground truth (e.g., against a gold standard for "abnormal digestive health") is not detailed.

4. Adjudication Method for the Test Set:

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, the document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The study described focuses on the device's ability to detect a specific biomarker in a clinical setting, not on comparing the performance of human readers with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, the described "Clinical Testing - PrevisEA Optimization Trial" appears to be a standalone (algorithm only) performance study. The device itself "detects the acoustic biomarker MH4 and quantifies the number of MH4 detections over a 4-minute interval, which it compares to a predefined threshold to determine the presence of abnormal digestive health." The study demonstrated the device's capability in "listening/capturing the acoustic biomarker, MH4 while in clinical use." There is no mention of human readers interpreting the raw sounds or the device's direct output except that the device "displays information which assists physicians." The focus is on the device's intrinsic ability to detect and quantify MH4.

7. The Type of Ground Truth Used:

The type of ground truth is indirectly a form of outcome data or physiological measurement (acoustic biomarker detection). The device specifically listens for the "MH4 acoustic biomarker" and "quantifies the number of MH4 detections." The study's objective was to demonstrate the device's capability in "listening/capturing the acoustic biomarker, MH4 while in clinical use." The implication is that the detection of MH4 itself serves as the ground truth for the device's primary function, with the ultimate goal of assisting in the assessment of digestive health. However, the direct clinical ground truth for "abnormal digestive health" against which the MH4 detection was validated is not elaborated.

8. The Sample Size for the Training Set:

The document does not provide any information regarding the sample size for the training set. The clinical trial described (75 subjects) is presented as a validation or optimization trial, not a training set. The term "proprietary algorithm" suggests internal development, but details on its training are absent.

9. How the Ground Truth for the Training Set Was Established:

The document does not provide any information on how the ground truth for the training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

August 4, 2021

Entac Medical Inc. % Adam Harris Director, Regulatory Affairs Target Health LLC 261 Madison Avenue, 24th Floor New York, NY 10016

Re: K211068

Trade/Device Name: PrevisEA Device Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DOD Dated: May 5, 2021 Received: May 7, 2021

Dear Adam Harris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211068

Device Name PrevisEA Device

Indications for Use (Describe)

PrevisEA is a compact, non-invasive device placed on the abdomen to capture the digestive sounds of patients. PrevisEA displays information which assists physicians in assessing the digestive health of patients. This device is for prescription use only and should be used under the direction of a licensed healthcare practitioner. The PrevisEA device has not been tested for and it is not intended for pediatric use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PrevisEA Device K211068 Premarket Notification

5. 510(k) Summary

5.1

Entac Medical Inc. 680 Oakleaf Office Lane, Suite 201 Memphis, TN 38117

5.2 Contact Person

Adam Harris, MM, RAC Director, Regulatory Affairs Target Health LLC 261 Madison Avenue, 24th Floor New York, NY 10016 Telephone: (646) 218-2009 aharris@targethealth.com

Date of Summary: July 8, 2021

Device Proprietary NamePrevisEA™ Device
Common/Usual NameElectronic Stethoscope
Classification Names /Numbers and Code21 CFR870.1875Classification NameElectronic StethoscopeCodeDQD
Regulatory Class2
Prescription StatusPrescription Device
Classification PanelGastroenterology
Predicate DeviceK150782GI Logic AbStats® Gateway
Description of DevicePrevisEA is a non-invasive, self-contained, medical device with aproprietary algorithm used to assess the digestive health of patients. Thedevice includes a stethoscope diaphragm, sound chamber, embeddedmicrophone and CPU, as well as a display screen and buttons. ThePrevisEA device is a convenient, single-use, disposable unit, whichattaches to the abdomen of the patient via an adhesive wafer on the backof the device. The PrevisEA is a single-use, disposable unit, which avoidsthe need for cleaning and disinfection.The PrevisEA requires no cable connections for operation of the device.Integration of the above components allow auscultation and digitalcapture of digestive sounds, such as MH4.Messages are transmitted to the easy-to-read display screen on the frontof the device. The PrevisEA detects the acoustic biomarker MH4 andquantifies the number of MH4 detections over a 4-minute interval, which itcompares to a predefined threshold to determine the presence of abnormal

{4}------------------------------------------------

digestive health in patients following intestinal surgery, such as gastrointestinalimpairment (GII).
Indications for Use/IntendedUsePrevisEA is a compact, non-invasive device placed on the abdomen tocapture the digestive sounds of patients. PrevisEA displays informationwhich assists physicians in assessing the digestive health of patients. Thisdevice is for prescription use only and should be used under the directionof a licensed healthcare practitioner. The PrevisEA device has not beentested for and it is not intended for pediatric use.

{5}------------------------------------------------

Subject DevicePredicate Device
Entac PrevisEA DeviceGI Logic Abstats GatewayComparison
510(k) NumberK211068K150782Same
Product CodeDQDDQDSame
Regulation21 CFR 870.1875 Stethoscope21 CFR 870.1875 StethoscopeSame
Intended UsePrevisEA is a compact, non-invasive device placed on theabdomen to capture the digestivesounds of patients. PrevisEAdisplays information which assistsphysicians in assessing thedigestive health of patients.This device should be used underthe direction of a licensedhealthcare practitioner. ThePrevisEA device has not beentested for and it is not intended forpediatric use.The AbStats Gateway is a compactdevice with integrated sensorinterfaces and embeddedcomputing system (stethoscope)that in conjunction with externalsensors constitutes the AbStatssystem. The AbStats systemexternal sensors are placed on theabdomen and identify the vibratorysignals associated with digestiveprocesses. This device should beused under the direction of alicensed healthcare practitionerwhen it is required to determinethis patient digestive state. Thedevice has not been tested for andit is not intended for pediatric use.Substantially equivalent
Number of Sensors1 - self contained2 - externalDifferent number and location butsubstantially equivalentfunctionality
Sensor Size (mm) Active Area3 inches30 mm x 20 mm400 mm2Different size but substantiallyequivalent functionality
Subject DeviceReference DeviceComparison
Entac PrevisEA DeviceGI Logic Abstats Gateway
AlgorithmThe PrevisEA utilizesalgorithm technology tospecifically evaluategastrointestinal auditorybiomarker MH4The Abstats Gateway utilizes algorithmtechnology to recognize and classifygastrointestinal signalsDifferent but Subject andReference devices have similarfunctionality
Sensor TechnologyStandard ElectricalMicrophoneStandard Electrical MicrophoneSubstantially equivalent
System TechnologySensor data signal processingand computingSensor data signal processingand computingSubstantially equivalent
Interface to PCNone required (no PCInterface)None required (no PC Interface)Substantially equivalent
Power SupplyLithium-Ion Battery5V USB Power SourceDifferent but substantiallyequivalent functionality
Sensor Cable Length (m)N/A – none required2.0 mDifferent but substantiallyequivalent functionality
IEC60601 1-1:2005 3rd EditionMedical Electrical Equipment Part1: General Requirement or SafetyConformsConformsSubstantially equivalent
EN60601-1-2, 2007/03, EMCRequirements for Safety, 2.Collateral Standard –Electromagnetic CompatibilityRequirements and TestsConformsConformsSubstantially Equivalent
Immunity Requirements for MedicalElectrical Equipment Part 1:GeneralConformsConformsSubstantially Equivalent
Subject DeviceReference DeviceComparison
Entac PrevisEA DeviceGI Logic Abstats Gateway
BiocompatibilityTesting demonstrated allpatient contacting materialsfor the PrevisEA device to besafe and biocompatibleNo biocompatibility testing performed as theonly patient contact consists of commerciallyavailable dermal adhesive dressingsSubstantially Equivalent
SoftwareThe PrevisEA softwareprocesses audio signals fordisplay and presentation tothe user.The AbStats software processesaudio signals for displayand presentation to the user.Substantially Equivalent
Auscultation Performance Tests:Bench TestingPassed/ConformsPassed/ConformsSubstantially Equivalent
Clinical performance validationListens for the MH4 acousticbiomarkerListens for acoustic event rates to determinean intestinal motilityDifferent mechanism/parameterof assessment but does not raisenew issues of safety and efficacy.
Abdomen PlacementThe PrevisEA is placed inany quadrant of the patient'sabdomen.Image: PrevisEA device on a human torsoAbStats external sensors placed on theabdomen,Image: AbStats external sensors placed on a human abdomenSubstantially Equivalent

Table 1: Substantial Equivalence Comparison Table

{6}------------------------------------------------

Table 1: Substantial Equivalence Comparison Table (Continued)

{7}------------------------------------------------

Table 1: Substantial Equivalence Comparison Table (Continued)

{8}------------------------------------------------

CharacteristicsPrevisEAAbstats Gateway
Electronic StethoscopeYesYes
Non-invasive systemYesYes
Attach to abdomenYesYes
Auscultation deviceYesYes
Assess digestion specificallyYesYes
Algorithm technologyYesYes
Self-contained sensorsYesNo
Batter poweredYesNo

Table 2: Summary of Technological Characteristics Compared to the Predicate Device

5.3 Biocompatibility Testing

Biocompatibility tests were conducted with the PrevisEA Device according to ISO 10993-1:2018 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. These studies demonstrated the PrevisEA Device is biocompatible.

5.4 Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical product safety, electromagnetic compatibility, and battery safety testing were performed for the PrevisEA Device. All testing met acceptance criteria and demonstrated the PrevisEA Device to be safe and electromagnetically compatible.

ર્સ્ડ Clinical Testing - PrevisEA Optimization Trial

Clinical data was submitted in support of this 510k Premarket Notification. Clinical testing demonstrated the PrevisEA Device accomplishes its intended use. Trial Registration: ClinicalTrials.gov NCT03505476

75 subjects undergoing intestinal resection surgery had PrevisEA applied at the completion of surgery.

This study demonstrated the capability of the PrevisEA Device in listening/capturing the acoustic biomarker, MH4 while in clinical use (attached to patient abdomens).

5.6 Conclusion

Since the comparison of bench testing to clinical outcomes is still not well understood for this type of device, clinical testing was required to support substantial equivalence.

{9}------------------------------------------------

The clinical data submitted validates the intended use of the PrevisEA Device. The nonclinical data support the safety and effectiveness of the device, and the bench demonstrates the device meets its own specifications. The hardware and software verification and validation demonstrate that the PrevisEA device performs as intended in the specified use conditions. Though there are minor differences in technological characteristics, the data provided in this submission and the comparisons demonstrate these differences do not raise new questions of safety and efficacy. It is the conclusion of the sponsor that the subject device is substantially equivalent to the predicate and reference devices.

§ 870.1875 Stethoscope.

(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.