K Number
K211034

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-05-06

(29 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.

The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.

Device Description

CORVUS is a semi-automatic planning system: For forward planning, the system allows the user to design a treatment plan. For IMRT, rather than simply verifying a user-designed plan, the system itself suggests a plan. A clinician then reviews and approves the plan.

CORVUS is designed to generate plans for treatment delivery systems that can create multiple radiation patterns composed of pencil beams on which the intensity can be individually controlled. The treatment beams are weighted so that when they are projected into the treatment space they superimpose to give the desired dose distribution.

Each radiation field is generated using one of several optimization methods provided with the system, including simulated annealing and gradient descent.

The treatment beams are set not only to deliver the prescribed dose to the identified target volume, but also to keep the dose to other sensitive volumes below user-defined limits. Planning is done volumetrically: the beam weights for treating the entire target volume are generated simultaneously. The dose matrix is volumetric. The dose to each point is calculated to be that received from all beams and from all gantry angles. Dosage is calculated using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data. The degree to which a treatment plan is optimized is determined in part by constraints placed on the planning algorithm. The user has direct control over these constraints, which include dose goals to the target structures, dose limits to the sensitive structures, and the specification of arcs or fixed gantry positions in the treatment plan.

CORVUS treatment plans need not have the isocenter located within the target volume. An unlimited number of targets falling within the treatment volume can be planned for at the same time. Dose may be prescribed for up to 32 structures, 29 of them user-selectable, any number of which may be separate targets or radiation-sensitive structures. Each structure can have a separate dose prescription.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the CORVUS device, based on the provided FDA 510(k) summary:

The CORVUS device is a radiation treatment planning system. The supplied document focuses on demonstrating substantial equivalence to a predicate device (CORVUS 14) rather than a direct clinical performance study against specific acceptance criteria for diagnostic accuracy in the typical sense of AI/ML devices.

Therefore, the acceptance criteria and study detailed below are primarily centered around dosimetric accuracy, software functionality, and compliance with relevant standards, which are the key performance indicators for a treatment planning system seeking 510(k) clearance.


1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a 510(k) submission for a treatment planning system and not a diagnostic AI device, the "acceptance criteria" are generally related to dosimetric accuracy and software performance compared to established benchmarks or the predicate device.

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from documentation)Reported Device Performance
Dosimetric Accuracy- Treatment plan calculations (dose distribution) align with medical physics measurements. - Calculations for Cobalt-60 based Forward Planning align with expected accuracy (implied by adding this new feature). - Pencil-beam algorithm and dose calculation options (homogeneous, EPL) maintain accuracy.- "The accuracy of treatment plans was evaluated through comparison with medical physics measurements 2D array and ion chamber measurements." - The submission states that the "verification and validation results demonstrate that the CORVUS 15 system met its design requirements and specifications." - The fundamental scientific technology and intended use/indications are unchanged from the predicate, implying similar dosimetric performance for existing features.
Software Functionality- All device functionalities work as per its intended use. - Risk mitigation is ensured for software. - Conforms to required software standards (IEC 62304, IEC 62366-1, IEC 62083). - New features (Cobalt-60 Forward Planning) function correctly and safely.- "Design verification and validation testing was performed to ensure that the device functionality works as per its intended use, all risks are mitigated, is substantially equivalent, and the product conforms to the required standards." - "The verification activities included system tests, module tests, anomaly verification, code reviews, and run-through integration tests." - "The validation activities included clinical workflow, treatment planning software usability, dosimetric accuracy, and import-export."
Substantial Equivalence- No new issues of safety or effectiveness are raised compared to the predicate device (CORVUS 14). - Intended use and indications for use remain substantially similar or unchanged. - Fundamental technical characteristics are the same.- "The fundamental scientific technology of the CORVUS 15 with respect to its predicate device (CORVUS 14) system has not changed. The intended use and indications for use of the device have not changed. Based upon the performance testing results for CORVUS 15 (as detailed in the submission), the system raises no new issues of safety or effectiveness."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in the sense of a dataset of patient cases used for clinical efficacy as would be seen for a diagnostic AI. Instead, the testing focuses on system verification and dosimetric validation.

  • Sample Size for Test Set: Not explicitly stated in terms of patient cases. The testing involved "medical physics measurements 2D array and ion chamber measurements." This typically involves a set of phantoms or carefully controlled geometric setups, but the number of such measurements or scenarios is not provided.
  • Data Provenance: Not applicable in the context of patient data. The "measurements" would be generated in a lab setting, not from patients.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not explicitly stated. The "accuracy of treatment plans was evaluated through comparison with medical physics measurements." This implies evaluation by medical physicists.
  • Qualifications of Experts: The involvement of "medical physicists" is mentioned. Their specific years of experience or board certifications are not provided in this summary.

4. Adjudication Method for the Test Set

Not applicable. This is not a study requiring adjudication of expert interpretations of images or patient outcomes. The "ground truth" for dosimetric accuracy would be independent physical measurements.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. This type of study is typically performed for diagnostic or AI-assisted interpretation devices to compare human performance with and without AI assistance. CORVUS is a treatment planning system, and its evaluation focuses on the accuracy of its calculations and software functionality, not on improving human reader performance in interpreting images.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in the context of a treatment planning system. The "dosimetric accuracy" testing evaluates the algorithm's output (calculated dose distributions) against physical measurements, which is an algorithm-only evaluation. While the system is used by humans (medical physicists, dosimetrists, radiation oncologists), the core dose calculation component performs its function algorithmically, and its output is verified independently.


7. The Type of Ground Truth Used

The ground truth used for evaluating dosimetric accuracy was medical physics measurements (2D array and ion chamber measurements). This represents a physical, objective standard for verifying dose distribution accuracy.


8. The Sample Size for the Training Set

Not applicable. CORVUS is a rule-based or physics-based treatment planning system, not a machine learning (ML) or artificial intelligence (AI) device that typically requires a large 'training set' of data in the common sense for algorithm development. Its algorithms are based on established physics models and mathematical optimization techniques.


9. How the Ground Truth for the Training Set Was Established

Not applicable, as there isn't a "training set" in the context of machine learning. The algorithms are built upon fundamental physics and mathematical principles, and characterized/calibrated using machine-specific data (e.g., beam data from a linear accelerator). This characterization data is not a "ground truth for training" in the AI/ML sense but rather input parameters that define the physical behavior of the treatment machine.

{0}------------------------------------------------

May 6, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Best Nomos % Vineet Gupta, Ph.D. Director R&D One Best Drive PITTSBURGH PA 15202

Re: K211034

Trade/Device Name: CORVUS Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: April 5, 2021 Received: April 7, 2021

Dear Dr. Gupta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211034

Device Name CORVUS

Indications for Use (Describe)

The CORVUS system is a radiation treatment planning package designed to allow medical physicists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.

The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Best nomos. The word "Best" is written in a cursive, blue font. To the right of "Best", the word "nomos" is written in a green, sans-serif font.

K211034

510(k) Summary

A. SUBMITTERS NAME

Best NOMOS

B. ADDRESS

One Best Drive Pittsburgh, PA 15202

C. CONTACT

Name:Vineet Gupta, Ph.D.
Phone:(412) 312-6819
Fax:(412) 312-6701
Email:vineet.gupta@nomos.com

D. DATE PREPARED

April 5, 2021

E. DEVICE NAME

Device Trade Name:CORVUS
Common Name:Radiation Therapy Treatment Planning System
Classification Name:Medical Charged-Particle Radiation Therapy System(21CFR 892.5050)

F. DEVICE CLASS

Class II Panel: Radiology Product Code: 90-MUJ Regulation Number: 21CFR 892.5050

Image /page/3/Picture/15 description: The image contains the logo for Best healthcare. The logo features the word "Best" in a stylized, blue font. Above the word "Best" is an image of people holding hands in a circle. Below the word "Best" is the phrase "healthcare for everyone" in a smaller, blue font.

One Best Drive (formerly One Conair Drive), Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS fax 412 312 6701 www.nomos.com

{4}------------------------------------------------

G. PREDICATE DEVICES

CORVUS Radiation Therapy Treatment Planning System 14 (K201350)

H. STATEMENT OF INDICATIONS FOR USE

The intended use and indications for use for the CORVUS Radiation Therapy Treatment Planning System have not changed due to this modification to the product. This modification does not change the fundamental principle of using photons for creating conformal treatment plans which is the same as that of its predicate device.

Intended Use

CORVUS is intended for use as a planning tool for conformal radiation therapy. For IMRT, using operator-supplied input and patient scans, it creates a plan for treatment delivery systems and generates a set of beam weights that, when applied to a compatible system, facilitates delivery of an intensity-modulated 3D conformal radiation therapy treatment. For Forward Planning, CORVUS helps the user manually create a plan for treatment delivery based on user supplied inputs and patient scans including dose calculation. Forward Planning is supported only for Cobalt-60 based treatment planning.

CORVUS is intended only to suggest a delivery plan. It is the physician's responsibility to verify that the dose distributions which would result from plan implementation are appropriate for a particular patient.

The CORVUS system is intended to be used as an integrated system with a delivery device for planning and delivery of conformal radiation therapy. CORVUS produces radiation fields to conform to the projected tumor volume plus margins as desired during planning. The system provides statistical calculations to help achieve target goals while sparing sensitive structures.

Indications for Use

The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.

The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.

{5}------------------------------------------------

I. DEVICE DESCRIPTION

CORVUS is a semi-automatic planning system: For forward planning, the system allows the user to design a treatment plan. For IMRT, rather than simply verifying a user-designed plan, the system itself suggests a plan. A clinician then reviews and approves the plan.

CORVUS is designed to generate plans for treatment delivery systems that can create multiple radiation patterns composed of pencil beams on which the intensity can be individually controlled. The treatment beams are weighted so that when they are projected into the treatment space they superimpose to give the desired dose distribution.

Each radiation field is generated using one of several optimization methods provided with the system, including simulated annealing and gradient descent.

The treatment beams are set not only to deliver the prescribed dose to the identified target volume, but also to keep the dose to other sensitive volumes below user-defined limits. Planning is done volumetrically: the beam weights for treating the entire target volume are generated simultaneously. The dose matrix is volumetric. The dose to each point is calculated to be that received from all beams and from all gantry angles. Dosage is calculated using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data. The degree to which a treatment plan is optimized is determined in part by constraints placed on the planning algorithm. The user has direct control over these constraints, which include dose goals to the target structures, dose limits to the sensitive structures, and the specification of arcs or fixed gantry positions in the treatment plan.

CORVUS treatment plans need not have the isocenter located within the target volume. An unlimited number of targets falling within the treatment volume can be planned for at the same time. Dose may be prescribed for up to 32 structures, 29 of them user-selectable, any number of which may be separate targets or radiation-sensitive structures. Each structure can have a separate dose prescription.

J. PREDICATE DEVICE INFORMATION

The CORVUS 15 system is substantially equivalent to its primary predicate device CORVUS 14 (K201350; Decision Date: 20-September-2020). The CORVUS 14 system was determined to be substantially equivalent to its predicate device as of September 2020.

The fundamental scientific technology of the CORVUS 15 with respect to its predicate device (CORVUS 14) system has not changed. The intended use and indications for use of the device have not changed. Based upon the performance testing results for CORVUS 15 (as detailed in the submission), the system raises no new issues of safety or effectiveness.

{6}------------------------------------------------

K. COMPARISON TO THE PREDICATE DEVICE

This section provides the summary of comparison of CORVUS 14 to the predicate device.

Proposed DevicePredicate Device
CORVUS 15CORVUS 14 (K201350)
CORVUS is intended for use as aplanning tool for conformal radiationtherapy. For IMRT, using operator-supplied input and patient scans, itcreates a plan for treatment deliverysystems and generates a set of beamweights that, when applied to acompatible system, facilitates deliveryof an intensity-modulated 3Dconformal radiation therapy treatment.For Forward Planning, CORVUS helpsthe user manually create a plan fortreatment delivery based on usersupplied inputs and patient scansincluding dose calculation. ForwardPlanning is supported only for Cobalt-60 based treatment planningCORVUS is intended for use as aplanning tool for conformal radiationtherapy. For IMRT, using operator-supplied input and patient scans, itcreates a plan for treatment deliverysystems and generates a set of beamweights that, when applied to acompatible system, facilitates deliveryof an intensity-modulated 3Dconformal radiation therapy treatment.For Forward Planning, CORVUS helpsthe user manually create a plan fortreatment delivery based on usersupplied inputs and patient scansincluding dose calculation. ForwardPlanning is supported only for Cobalt-60 based treatment planning
Intended UseCORVUS is intended only to suggest adelivery plan. It is the physician'sresponsibility to verify that the dosedistributions which would result fromplan implementation are appropriatefor a particular patient.CORVUS is intended only to suggest adelivery plan. It is the physician'sresponsibility to verify that the dosedistributions which would result fromplan implementation are appropriatefor a particular patient.
The CORVUS system is intended to beused as an integrated system with adelivery device for planning anddelivery of conformal radiationtherapy. CORVUS produces radiationfields to conform to the projectedtumor volume plus margins as desiredduring planning. The system providesstatistical calculations to help achievetarget goals while sparing sensitivestructures.The CORVUS system is intended to beused as an integrated system with adelivery device for planning anddelivery of conformal radiationtherapy. CORVUS produces radiationfields to conform to the projectedtumor volume plus margins as desiredduring planning. The system providesstatistical calculations to help achievetarget goals while sparing sensitivestructures.
Indications forUseThe CORVUS system is a radiationtreatment planning package designed toThe CORVUS system is a radiationtreatment planning package designed to
allow medical physicists, dosimetrists,and radiation oncologists to createconformal treatment plans usingphoton (x-ray, gamma ray) externalbeam radiation therapy. The treatmentplans generated by CORVUS are basedupon treatment machine-specific dataand are intended to provide a guide todelivering external beam radiationtherapy which conforms to the targetvolume defined by the radiationoncologist.The CORVUS system is valid for useonly with external beam photontherapy; calculations for electrons andintracavity sources (Brachytherapy) areNOT supported.allow medical physicists, dosimetrists,and radiation oncologists to createconformal treatment plans usingphoton (x-ray, gamma ray) externalbeam radiation therapy. The treatmentplans generated by CORVUS are basedupon treatment machine-specific dataand are intended to provide a guide todelivering external beam radiationtherapy which conforms to the targetvolume defined by the radiationoncologist.The CORVUS system is valid for useonly with external beam photontherapy; calculations for electrons andintracavity sources (Brachytherapy) areNOT supported.

Table 1 Indications for Use & Intended Use Comparison

{7}------------------------------------------------

Comparison ItemProposed DeviceCORVUS 15Predicate DeviceCORVUS 14 (K201350)
Product Code90-MUJ90-MUJ
ClassIIII
Regulation Number892.5050892.5050
Operating SystemMac OS XMac OS X
Treatment Planning Supportfor X-RaysMLC-IMRT, Serial Tomotherapy(MIMiC-IMRT)MLC-IMRT, Serial Tomotherapy(MIMiC-IMRT)
Treatment Planning Supportfor Cobalt-60 (Gamma Rays)Serial Tomotherapy (MIMiC-IMRT),MLC-IMRT, Forward PlanningSerial Tomotherapy (MIMiC-IMRT),MLC-IMRT, Forward Planning

Table 2 General Comparison

CORVUS 15 contains all of the features of CORVUS 14 (K201350) and adds the feature of supporting Cobalt-60 based external beam radiation treatment planning based Forward Planning using jaw defined fields and wedges. The similarities and differences are discussed in detail as part of this submission.

In both products, CORVUS 15 and CORVUS 14, the pencil-beam algorithm is used and dose calculation can be performed using either a homogeneous option or an Effective Path Length (EPL) option. The dose calculation based on the Lateral Disequilibrium Inclusive option is only supported for X-Ray based treatment planning in CORVUS 15 as it was in the predicate

{8}------------------------------------------------

device CORVUS 14 (K201350). The optimization algorithms used in both products are the same.

The target population, intended use, and indications for use are similar to that of the predicate devices. In addition, the fundamental technical characteristics are the same as those of the predicate devices and any minor differences in features do not raise any concerns for safety, performance, or effectiveness of the device. The characteristics/features of CORVUS 15 with respect to the predicate device is described in the comparison chart in the submission.

L. SUMMARY OF TESTING

Design verification and validation testing was performed to ensure that the device functionality works as per its intended use, all risks are mitigated, is substantially equivalent, and the product conforms to the required standards. The software for this device is considered as a "Major" level of concern. The details of the design verification and validation activities are explained in the submission as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

The verification activities included system tests, module tests, anomaly verification, code reviews, and run-though integration tests.

The validation activities included clinical workflow, treatment planning software usability, dosimetric accuracy, and import-export. The accuracy of treatment plans was evaluated through comparison with medical physics measurements 2D array and ion chamber measurements.

Conclusion:

Detailed results of these tests are included as part of this submission. The verification and validation results demonstrate that the CORVUS 15 system met its design requirements and specifications, is substantially equivalent to its predicate device, and conforms to the applicable sections of standards that includes:

  • IEC 62304 Edition 1.1 2015-06 Medical device software Software life cycle processes .
  • IEC 62366-1 Edition 1.0 2020-06 Application of usability engineering to medical devices .
  • IEC 62083 Edition 2.0 2009-09 Requirements for the safety of radiotherapy treatment . planning systems

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.