K Number
K210968

Validate with FDA (Live)

Date Cleared
2021-12-21

(265 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The reVive Light Therapy® Essentials is an Over the Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.

Device Description

The reVive Light Therapy® Essentials is an over-the counter light emitting diode (LED) device, that emits energy for use in dermatology for the treatment of mild to moderate inflammatory acne. The device uses two types of LEDs: 630nm red and 415nm blue. The reVive Light Therapy® Essentials components include the device containing the LED module. There are no user settings or adjustments required. The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components.

AI/ML Overview

The provided text describes the regulatory clearance of the "reVive Light Therapy® Essentials" device. While it details performance testing conducted for safety and functionality, it does not include information about clinical studies proving the device's efficacy in treating mild to moderate inflammatory acne against specific acceptance criteria.

The document primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics, intended use, and safety testing (electrical, EMC, biocompatibility, software validation, and usability). It explicitly states: "The conclusions drawn from nonclinical tests demonstrate that the device is safe, as effective, and performs as well as the legally marketed devices." This implies that clinical efficacy, such as a reduction in acne severity, was not a primary part of this submission for demonstrating substantial equivalence, but rather inferred from the predicate devices.

Therefore, for the aspects of your request related to clinical efficacy and performance against specific acceptance criteria for treatment outcomes (e.g., reduction in inflammatory lesions), the provided text does not contain the necessary information. I will fill in what is available and state when information is not present.

Here's a breakdown of the requested information based on the provided text:


Acceptance Criteria and Device Performance Study (as described in the document):

The document does not detail acceptance criteria related to a specific clinical efficacy outcome (e.g., percentage reduction in acne lesions) nor does it present results from a clinical study proving such efficacy for the new device. Instead, the "Performance Testing and Standards" section focuses on safety and functionality.

1. A table of acceptance criteria and the reported device performance

Since no clinical efficacy acceptance criteria are presented, this table will focus on the non-clinical performance and safety criteria mentioned.

Acceptance Criterion (Type: Functional/Safety)Reported Device Performance
Electrical Safety (compliance with ANSI/AAMI ES 60601-1:2005 + A1:2012, IEC 60601-1:2005)Device conforms to standard. Requires further information for specifics.
Electromagnetic Compatibility (EMC) (compliance with ANSI/AAMI/IEC 60601-1-2 (2014))Device conforms to standard. Requires further information for specifics.
Biocompatibility - Cytotoxicity (compliance with ISO 10993-5:2009)Device conforms to standard. Requires further information for specifics.
Biocompatibility - Intracutaneous Reactivity (compliance with ISO 10993-10:2010)Device conforms to standard. Requires further information for specifics.
Biocompatibility - Skin Sensitization (compliance with ISO 10993-10:2010)Device conforms to standard. Requires further information for specifics.
Biocompatibility - Acute Systemic Toxicity (compliance with ISO 10993-11:2017)Device conforms to standard. Requires further information for specifics.
Biocompatibility - Material Medicated Pyrogenicity (compliance with USP General Chapter 151)Device conforms to standard. Requires further information for specifics.
Software Validation (compliance with FDA Guidance for Content of Premarket Submissions for Software)Software was tested and validated. Requires further information for specifics.
Usability - Correct preparation of device for use100% of participants were able to demonstrate correct preparation.
Usability - Correct device usage100% of participants were able to demonstrate correct device usage.

2. Sample size used for the test set and the data provenance

  • Sample Size for Usability Study: 16 participants.
  • Data Provenance: Not specified in terms of country of origin or whether it was retrospective/prospective for the usability study. The document is an FDA submission for a US market, suggesting US relevance, but this is not explicitly stated for the data collection.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Applicability: This question is typically relevant for studies involving expert interpretation of data (e.g., medical images).
  • In this context: For the usability study, there wouldn't be "ground truth" established by experts in the same way. The performance was measured by participant actions. For the functional/safety testing, compliance was assessed against recognized standards, implying expert evaluation by the testing body, but specific numbers and qualifications are not provided here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Applicability: This is typically for studies where multiple annotators establish ground truth for complex data.
  • In this context: Not applicable to the usability study or the stated safety/functional testing.

5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Applicability: This is for AI-assisted diagnostic devices.
  • In this context: Not applicable. The "reVive Light Therapy® Essentials" is a light therapy device for acne, not an AI diagnostic tool. No MRMC study was mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Applicability: This is for AI/algorithm performance directly.
  • In this context: Not applicable. This device is a direct treatment device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For Usability Study: The "ground truth" was predefined correct procedures for device preparation and usage, against which participants' actions were evaluated.
  • For Safety/Functional Testing: Compliance with specified international and national standards (e.g., IEC 60601 series, ISO 10993 series, USP chapters). The ground truth is the performance specification defined by these standards.

8. The sample size for the training set

  • Applicability: This is for AI/machine learning models.
  • In this context: Not applicable. No AI/ML model for which "training data" would be used is described for this device.

9. How the ground truth for the training set was established

  • Applicability: This is for AI/machine learning models.
  • In this context: Not applicable. No AI/ML model for which "training data" would be used is described for this device.

Summary of what the document focuses on for "proving the device meets acceptance criteria":

The core of the submission, as per the text, relies on substantial equivalence to legally marketed predicate devices, combined with comprehensive non-clinical safety and functional testing. The rationale is that since the new device has "similar technical characteristics," "same control mechanism, same operating principle, and energy type," and the "same method of operation" as already cleared predicate devices for the same intended use (mild to moderate inflammatory acne), and it passes all safety and functional tests, it is deemed safe and effective. The minor changes in size were also assessed and found not to introduce new risks.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2021

LED Technologies, Inc. Jelena Barbaric Compliance Manager 12821 Starkey Rd., Suite 4900 Largo, Florida 33773

Re: K210968

Trade/Device Name: reVive Light Therapy Essentials Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: November 24, 2021 Received: November 29, 2021

Dear Jelena Barbaric:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210968

Device Name reVive Light Therapy® Essentials

Indications for Use (Describe)

The reVive Light Therapy® Essentials is an Over the Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) Summary – K210968

This summary of 510 (k) information is being submitted in accordance with the requirements of 21 CFR § 807.92.

Submission Date: December 16th, 2021

1. Submitter Information:LED Technologies, Inc. - Jelena Barbaric
12821 Starkey Rd., Suite 4900
Largo, FL 33773
Tel: 727-285-8300 Ext 203
Email: jbarbaric@ledtechnologies.com

For Specification Developer: LED Technologies, Inc. Attn: Lloyd Nelson 12821 Starkey Rd., Suite 4900 Largo, FL 33773 Tel.: 727-285-8300. Ext 201 Email: Inelson@ledtechnologies.com

2. General Information

  • 2.1 Classification Name: Over the Counter Powered Light Based Laser for Acne
  • 2.2 Common/usual name: reVive Light Therapy® Essentials
  • 2.3 Proprietary Names: reVive Light Therapy® Essentials
  • Classification: Class II 2.4
  • Classification Number: 878.4810 2.5
  • 2.6 Product Code: OLP
  • 2.7 Review Panel: General & Plastic Surgery

3. Device Description

The reVive Light Therapy® Essentials is an over-the counter light emitting diode (LED) device, that emits energy for use in dermatology for the treatment of mild to moderate inflammatory acne. The device uses two types of LEDs: 630nm red and 415nm blue.

{4}------------------------------------------------

The reVive Light Therapy® Essentials components include the device containing the LED module.

There are no user settings or adjustments required.

The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components.

Indications/Intended Use:

The reVive Light Therapy® Essentials is an Over the Counter (OTC) LED device intended for use in treating mild to moderate inflammatory acne.

4. Predicate Devices:

Primary Predicate - K180447 – reVive Light Therapy® LED Cleansing System (LED Technologies, Inc.)

K180320 – dpl® IIa Panel (LED Technologies, Inc.)

Comparison Chart

DevicereVive LightTherapy® EssentialsK210968dpl® Ila PanelLED Technologies,Inc.K180320Primary PredicatereVive Light Therapy® LEDCleansing SystemLED Technologies, Inc.K180447
Wavelengths630nm, 415nm605nm, 630nm,660nm, 880nm415nm630nm, 415nm
IrradiancesourceLEDLEDLED
Treatment Area(cm²)741518.86
Treatment Time3 minutes pertreatment3 minutes pertreatment3 minutes per treatment
Type/ClassOTCOTCOTC
IFUFor treatment ofmild to moderateinflammatory acneFor treatment ofwrinkles & mild tomoderateinflammatory acneFor treatment of mild tomoderate inflammatoryacne

{5}------------------------------------------------

Summary of the technological characteristics of the device compared to predicate device:

    1. Has the same intended use as the predicate devices (i.e., treatment of mild to moderate inflammatory acne).
    1. Has similar technical characteristics.

The device has same control mechanism, same operating principle, and energy type. It uses the same method of operation as described in K180447 and K180320.

The subject device is smaller in size. The changes do not have significant effect on the use of the device. The risk assessment did not identify any new or significantly modified risks. Moreover, there were no unexpected issues from validation and verification activities.

Therefore, the minor changes in the device appearance do not affect the substantial equivalence of the subject device with the predicates.

    1. Has the same method of operation.
    1. Utilizes the same treatment duration as the predicate devices.

The reVive Light Therapy® Essentials and the above referenced predicate devices are Over the Counter Devices used to treat mild to moderate inflammatory acne as defined in 21 CFR § 878.4810. These devices utilize red and blue diodes with 630 nm and 415 nm to provide narrow bands of light energy to treat acne. The performance achieved by these devices is same with similar power output.

The reVive Light Therapy® Essentials is powered by Batteries) or via Universal USB power source.

The devices are intended to be placed directly on the skin. They are manufactured out of similar materials. Based upon comparison to the predicate devices, the reVive Light Therapy® Essentials has the same intended uses, with similar technological characteristics as predicate devices. The system performs as intended and does not raise any new safety or effectiveness issues.

5. Performance Testing and Standards:

Testing of the reVive Light Therapy® Essentials, included functional performance testing, software validation, testing, and user safety testing.

Safety and functionality testing demonstrate that the reVive Light Therapy® Essentials conforms to various international consensus standards.

ANSI/AAMI ES 60601-1:2005® + 2012 and A1:2012 IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012: Medical Electrical Equipment part 1: General Requirements for Basic Safety and Essential Performance.

ANSI/AAMI/IEC 60601-1-2 (2014): Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance: Collateral Standard: Electromagnetic Compatibility.

{6}------------------------------------------------

Biocompatibility

ISO 10993-5:2009 – Cytotoxicity Test ISO 10993-10:2010 - Intracutaneous reactivity test ISO 10993-10:2010 – Skin Sensitization Test ISO 10993-11:2017 – Acute Systemic Toxicity USP General Chapter 151 – Material Medicated Pyrogenicity

Additionally, the subject device tissue contacting components are the ABS plastic and the PC. The materials data from previous submissions is identical in materials selection, manufacturing process, chemical composition of the materials, as well as nature of patient contact, proving the ABS and PC of the reVive Light Therapy® Essentials medical device in its final finished form is identical to the ABS and PC material of the predicate devices.

The reVive Light Therapy® Essentials software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices".

A Usability Study was conducted with 16 participants.

The results of the study found that:

100% of the participants were able to demonstrate the correct preparation of device for use.

100% of the participants were able to demonstrate correct device usage.

The conclusions drawn from nonclinical tests demonstrate that the device is safe, as effective, and performs as well as the legally marketed devices.

6. Conclusion

After analysis of safety, indications, intended uses, dose rates, performance, features, design materials, power output, technological properties, treatment regiment regimens and methods of operation, the manufacturer asserts that no significant differences exist between the subject device and predicate, and no different questions of safety and effectiveness arise. Therefore, the subject device is substantial equivalence to the predicate.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.