(174 days)
The MOZART® Supra® Specimen Tomosynthesis System is a cabinet X-ray system that is used to provide 2-dimensional and 3-dimensional tomographic digital X-ray images of harvested specimens from various anatomical regions.
The MOZART SUPRA (XPERT84) Specimen Radiography System is a Cabinet X-ray System specifically designed to provide high detail radiographic imaging of surgically excised medical specimens both in two-dimensional and three-dimensional tomosynthesis views.
- It is the only cabinet specimen imaging system to utilize 3-D Tomosynthesis technology.
- Creates images in 1mm digital slices of the specimen, allowing physicians to evaluate the specimen layer by layer.
Tomosynthesis is an advanced radiographic application that produces individual coronal "slice" images through an anatomical region of inters (ROI). To produce these slices multiple projection radiographic images are acquired in rapid succession as the X-ray tube sweeps and rotates across the ROI. Once acquired, these projection images are subject to image processing that registers and reconstructs them into individual tomographic slices.
Tomosynthesis provides visualization of human anatomy by -
- Removing overlying anatomical structures, which could otherwise obscure a structure of interest by superimposition in a two dimensional presentation, and
-
- Producing a number of slice images throughout the entire volume of the anatomy
The exceptionally high magnification capability (up to 5X) from the 0.02 mm focal spot with optimized cabinet geometry and the superior contrast available from the low kV capability provides enhanced film and/or digital imaging performance. This device supports radiographic film sizes up to 30 x 35 cm and can be configured to acquire high resolution, DICOM compliant, digital x-ray images through the use of an integrated camera and Kubtec DIGICOM Specimen Radiography software.
To support the Tomosynthesis application, the predicate device, MOZART SUPRA (XPERT84), K200756, has a low-voltage screw drive linear actuator installed in the enclosed head of the cabinet with the same 50 kVp monoblock X-ray tube as the predicate to allow motion of the X-ray source to capture the multiple projection images and an Amorphous Selenium detector mounted stationary in the bottom of the cabinet xray unit
- Producing a number of slice images throughout the entire volume of the anatomy
The Kubtec MOZART SUPRA (XPERT 84) Radiography System is a Cabinet X-ray System for generating 2D and 3D tomographic digital X-ray images of harvested specimens.
Here's an analysis of the acceptance criteria and study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Criteria | Acceptance Criteria | Reported Device Performance | Study Proving Performance |
|---|---|---|---|
| Safety Standards | Compliance with various IEC, ISO, and CFR standards. | Successfully tested to IEC 61010-1 and EMC Directive 2004/108/EC. Design controls and risk management maintained per ISO 13485:2016, 21 CFR Part 820, and 21 CFR 1020.40. | Internal and external safety testing by INTERTEK |
| Radiation Safety | Less than 0.1 mR/hr at 5 cm from any exterior surface at maximum kV. | Less than 0.1 mR/hr at 5 cm from any exterior surface at maximum kV. | Not explicitly detailed, but stated as a specification. |
| Image Resolution (Detector) | Not explicitly defined as acceptance criteria, but resolution is a key performance metric. | Resolution: 85μm for proposed AXS-1824V2 & AXS-2430V2 detectors. | Verification via imaging of line-pair gauges and phantoms. |
| DQE at 1 Lp/mm (Detector) | Not explicitly defined as acceptance criteria. | >50% for AXS-1824V2, >70% for AXS-2430V2. | Verified through performance testing (presumably by Analogic, the detector manufacturer, and then confirmed by Kubtec). |
| MTF at 1 Lp/mm (Detector) | Not explicitly defined as acceptance criteria. | >90% for AXS-1824V2, >95% for AXS-2430V2. | Verified through performance testing (presumably by Analogic, the detector manufacturer, and then confirmed by Kubtec). |
| Bit Depth (Detector) | Not explicitly defined as acceptance criteria. | 16 bits for AXS-1824V2 & AXS-2430V2. | Verified through performance testing. |
| Software Functionality | DICOM Store, Print, and Modality Worklist functionalities. | DIGICOM Software supports DICOM Store, Print, and Modality Worklist. | Internal validation and verification. |
| Overall Performance Equivalence | As safe and effective as the predicate devices. | Performance verified and cleared via 510(k) K193166 when incorporated into the Siemens MAMMOMAT Revelation. | Claims proven internally and validated/verified via imaging of line-pair gauges and phantoms. Clinically validated through reads by a board-certified radiologist. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document only mentions "excised breast biopsies" for clinical validation and "line-pair gauges and phantoms" for internal validation. No specific number of cases or images for the test set is provided.
- Data Provenance: The data provenance for the clinical validation is "reads of excised breast biopsies by a board certified radiologist." This implies retrospective processing of existing biopsies, but does not specify the country of origin. The data for line-pair gauges and phantoms would be generated during internal testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: "A board certified radiologist" (singular) is mentioned for the clinical validation.
- Qualifications: "board certified radiologist." No specific years of experience are provided.
4. Adjudication Method for the Test Set
- Given that only a single board-certified radiologist is mentioned for clinical validation, there is no adjudication method described beyond the single expert's assessment.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC comparative effectiveness study was not done. The document describes a substantial equivalence submission, which typically focuses on demonstrating that the new device performs as well as, or is equivalent to, legally marketed predicate devices, rather than comparing human reader performance with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- The document describes a radiographic system that produces images, not an AI algorithm for image interpretation. Therefore, the concept of a "standalone" AI algorithm performance is not applicable in this context. The device's performance relates to the quality of the images it produces. The software (DIGICOM) primarily handles image processing and DICOM functionalities, not autonomous diagnostic interpretation.
7. The Type of Ground Truth Used
- For clinical validation reported, the ground truth was established by an expert assessment ("reads of excised breast biopsies by a board certified radiologist").
- For technical performance validation (resolution, DQE, MTF), the ground truth relies on physical measurement standards using "line-pair gauges and phantoms."
8. The Sample Size for the Training Set
- The document does not describe a "training set" in the context of an AI algorithm learning from data. The device is a radiographic system, and its software (DIGICOM) is for image handling and processing, not for AI-driven image analysis that requires a training set.
9. How the Ground Truth for the Training Set was Established
- As there is no mention of a training set for an AI algorithm (see point 8), this question is not applicable.
{0}------------------------------------------------
September 20, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
KUB Technologies, Inc. % Chester Lowe, Ph.D. Chief Technology Officer 111 Research Drive STRATFORD CT 06615
Re: K210955
Trade/Device Name: Kubtec MOZART SUPRA (XPERT 84) Radiography System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MWP Dated: August 27, 2021 Received: August 30, 2021
Dear Dr. Lowe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K210955
Device Name Kubtec MOZART SUPRA (Xpert 84) radiography system
Indications for Use (Describe)
The MOZART® Supra® Specimen Tomosynthesis System is a cabinet X-ray system that is used to provide 2dimensional and 3-dimensional tomographic digital X-ray images of harvested specimens from various anatomical regions.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains two logos for KUBTEC. The first logo is for "KUBTEC MEDICAL IMAGING" and the word "KUBTEC" is in blue. The second logo is for "KUBTEC SCIENTIFIC IMAGING" and the word "KUBTEC" is in orange. The word "MEDICAL" and "SCIENTIFIC" are smaller than the word "KUBTEC".
Kubtec MOZART SUPRA (XPERT 84) Premarket Notification 510(k) Summary K210955
Identification of Devices:
| Proprietary/Trade Name: | Kubtec MOZART SUPRA (XPERT84) Radiography System |
|---|---|
| Classification name: | Cabinet X-ray System |
| Classification: | Class II |
| CFR Section: | 21 CFR 892.1680 |
| Product Codes: | MWP |
| Common Name: | Stationary X-ray System |
Applicant:
KUB Technologies, Inc. 111 Research Drive Stratford, CT 06615 USA
Contact Person:
| Name: | Chester Lowe, Ph.D. |
|---|---|
| Title: | Chief Technology Officer |
| Telephone: | +1.203.364.8544 |
| Facsimile: | +1.203.255.7494 |
| Email: | CLowe@kubtec.com |
Establishment Number: KUB Technologies, Inc. # 3006051164
INDICATIONS FOR USE I.
The MOZART® Supra® Specimen Tomosynthesis System is a cabinet X-ray system that is used to provide 2-dimensional and 3-dimensional tomographic digital X-ray images of harvested specimens from various anatomical regions.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows two logos for KUBTEC. The first logo is for KUBTEC Medical Imaging, with the word "KUBTEC" in blue and the words "MEDICAL IMAGING" in a smaller font size below it. The second logo is for KUBTEC Scientific Imaging, with the word "KUBTEC" in yellow and the words "SCIENTIFIC IMAGING" in a smaller font size below it.
II. DEVICE DESCRIPTION
The MOZART SUPRA (XPERT84) Specimen Radiography System is a Cabinet X-ray System specifically designed to provide high detail radiographic imaging of surgically excised medical specimens both in two-dimensional and three-dimensional tomosynthesis views.
- It is the only cabinet specimen imaging system to utilize 3-D Tomosynthesis technology.
- Creates images in 1mm digital slices of the specimen, allowing physicians to ● evaluate the specimen layer by layer.
Tomosynthesis is an advanced radiographic application that produces individual coronal "slice" images through an anatomical region of inters (ROI). To produce these slices multiple projection radiographic images are acquired in rapid succession as the X-ray tube sweeps and rotates across the ROI. Once acquired, these projection images are subject to image processing that registers and reconstructs them into individual tomographic slices.
Tomosynthesis provides visualization of human anatomy by
-
- Removing overlying anatomical structures, which could otherwise obscure a structure of interest by superimposition in a two dimensional presentation, and
-
- Producing a number of slice images throughout the entire volume of the anatomy
The exceptionally high magnification capability (up to 5X) from the 0.02 mm focal spot with optimized cabinet geometry and the superior contrast available from the low kV capability provides enhanced film and/or digital imaging performance. This device supports radiographic film sizes up to 30 x 35 cm and can be configured to acquire high resolution, DICOM compliant, digital x-ray images through the use of an integrated camera and Kubtec DIGICOM Specimen Radiography software.
To support the Tomosynthesis application, the predicate device, MOZART SUPRA (XPERT84), K200756, has a low-voltage screw drive linear actuator installed in the enclosed head of the cabinet with the same 50 kVp monoblock X-ray tube as the predicate to allow motion of the X-ray source to capture the multiple projection images and an Amorphous Selenium detector mounted stationary in the bottom of the cabinet xray unit
III. SUMMARY OF STUDIES AND SAFETY
Kubtec successfully completed internal and external safety testing requirements to the following standards The software and verification testing was also performed.
The Proposed MOZART SUPRA (XPERT84) with a-Se detector was successfully tested by INTERTEK to IEC 61010-1 Safety requirements for electrical equipment for
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows two logos for KUBTEC. The top logo is for "KUBTEC MEDICAL IMAGING" with "KUBTEC" in a larger, bold, blue font and "MEDICAL IMAGING" in a smaller, lighter blue font underneath. The bottom logo is for "KUBTEC SCIENTIFIC IMAGING" with "KUBTEC" in a larger, bold, yellow font and "SCIENTIFIC IMAGING" in a smaller, lighter yellow font underneath.
measurement, control, and laboratory use - Part 1: General requirements as well to EMC Directive 2004/108/EC
Design controls and design transfer with appropriate risk management and analysis have been maintained per ISO 13485:2016, KUBTEC's Standard Operating Procedures via Quality and Regulatory Assurance (QARA) and Engineering, 21 CFR Part 820, and to 21 CFR 1020.40.
| Agency | Standard | RecognitionNumber | Description |
|---|---|---|---|
| IEC | 60601-1-2 | 19-36 | Medical electrical equipment - Part 1-2: Generalrequirements for basic safety and essentialperformance - Collateral Standard: Electromagneticdisturbances - Requirements and tests |
| IEC | 60601-1-10 | 19-37 | Medical electrical equipment - Part 1-10: Generalrequirements for basic safety and essentialperformance - Collateral Standard: Requirementsfor the development of physiologic closed-loopcontrollers |
| IEC | 60601-1-11 | 19-38 | Medical electrical equipment - Part 1-11: Generalrequirements for basic safety and essentialperformance - Collateral Standard: Requirementsfor medical electrical equipment and medicalelectrical systems used in the home healthcareenvironment |
| IEC | 60601-1-12 | 19-39 | Medical electrical equipment - Part 1-12: Generalrequirements for basic safety and essentialperformance - Collateral Standard: Requirementsfor medical electrical equipment and medicalelectrical systems intended for use in theemergency medical services environment |
| IEC | 61010-1 | 19-34 | Safety requirements for electrical equipment formeasurement, control, and laboratory use - Part 1:General requirements |
| IEC | 62304 1st Ed | 13-79 | Medical device software - Software life cycleprocesses |
| ISO | 14971 2nd Ed | 5-40 | Medical devices - Application of risk management tomedical devices |
| ISO | 15223-1 | 5-117 | Medical devices - Symbols to be used with medicaldevice labels, labelling, and information to besupplied - Part 1: General requirements |
| IEC | 61910-1 1st Ed | 12-290 | Medical electrical equipment - Radiation dosedocumentation - Part 1: Radiation dose structuredreports for radiography and radioscopy |
111 Research Drive, Stratford, CT 06615 USA. Tel: +1(203) 364-8544; Fax:+1 (203) 255-7494 KUBTEC is a registered trademark of KUB Technologies, Inc.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a table with information about NEMA and CFR standards. The NEMA standard is PS 3.1 - 3.20 and 12-300 for Digital Imaging and Communications in Medicine (DICOM) Set. The CFR standard is 21 CFR 1020.40 and N/A for Cabinet X-Ray Systems. The image also contains the logos for KUBTEC MEDICAL IMAGING and KUBTEC SCIENTIFIC IMAGING.
IV. BASIS FOR THE SUBMISSION
The XPERT 84 Specimen Radiography System is a fully shielded Cabinet X-ray System that has been designed to comply with 21 CFR 1020.40. The system allows more than 20 times geometric magnification of excised specimens with minimal geometric distortion through the use of a focal spot size that is less than 20 microns. The x-ray coverage of the device allows the use of radiographic film sizes up to 35 x 43 cm. The device can also be configured to provide high resolution, DICOM compliant, digital images through the use of an integrated digital camera and Kubtec DIGICOM Specimen Radiography software. The Kubtec DIGICOM Software supports the DICOM Store, Print and Modality Worklist functionalities.
This device, MOZART SUPRA (XPERT84) will include a modification/replacement of the detectors in previously approved devices and functionally identical to the predicate devices. (See specifications in Section V) CMOS detector to an Amorphous SILICON detector
There is no change for the indication for use for the previously 510(k) approved device.
| Design and Use of the Device | |||
|---|---|---|---|
| Question | YES | NO | |
| Is the device intended for prescription use (21 CFR 801 Subpart D) ? | X | ||
| Is the device intended for over-the-counter use (21 CFR807 Subpart C)? | X | ||
| Is the device provided sterile? | X | ||
| Is the device intended for single use? | X | ||
| Is the device a reprocessed single use device? | X | ||
| If yes, does this device type require reprocessed validation data? | |||
| Does the device contain a drug? | X | ||
| Does the device contain a biologic? | X | ||
| Does the device use software? | X | ||
| Does the submission include clinical information? | X | ||
| Is the device implanted? | X |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows two logos for KUBTEC. The first logo is for KUBTEC Medical Imaging, with "KUBTEC" in blue and "MEDICAL IMAGING" in a smaller font size below it. The second logo is for KUBTEC Scientific Imaging, with "KUBTEC" in gold and "SCIENTIFIC IMAGING" in a smaller font size below it. The logos are stacked vertically, with the medical imaging logo on top.
V. SUBSTANTIAL EQUIVALENCE
The Kubtec MOZART SUPRA(XPERT84) Specimen Radiography System equipped with the Amorphous Selenium detector is substantially equivalent to the following currently cleared devices:
PRIMARY:
510(k) Number: K200756 Trade Name: Kubtec MOZART SUPRA (XPERT84) Specimen Radiography System
SECONDARY: 510(k) Number: K 111508 Trade Name: Hologic TRIDENT SPECIMEN RADIOGRAPHY 510(k) Number: K 193166 Trade Name: Siemens MAMMOMAT Revelation
Predicate Device:
| Device Name: | Kubtec MOZART SUPRA (XPERT84) |
|---|---|
| 510(k) Number: | K200756 |
| Classification | Class: II |
| Regulation Number: | 21 CFR 892.1680 |
| Product Code: | MWP |
| Regulation Name: | Stationary X-ray System |
| Device Name: | Hologic TRIDENT SPECIMEN RADIOGRAPHYSYSTEM Model:RC |
| 510(k) Number: | K111508 |
| Classification | Class: II |
| Regulation Number: | 21 CFR 892.1680 |
| Product Code: | MWP |
| Regulation Name: | Stationary X-ray System |
| Device Name: | Siemens MAMMOMAT Revelation |
| 510(k) Number: | K193166 |
| Classification | Class: II |
| Regulation Number: | 21 CFR 892.1715 |
| Product Code: | MUE |
| Regulation Name: | Full-field digital mammography system |
The proposed and predicate devices utilize similar technology and materials, comparable safety and effectiveness features, and is similar in design and construction.
The Indications for Use and labeling are virtually the same or similar and our labeling contain the required Cautions, Warnings and Contraindications consistent to those required for similar cleared devices.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows two logos for KUBTEC. The top logo is for "KUBTEC MEDICAL IMAGING", with "KUBTEC" in a bold blue font and "MEDICAL IMAGING" in a smaller, lighter blue font underneath. The bottom logo is for "KUBTEC SCIENTIFIC IMAGING", with "KUBTEC" in a bold yellow font and "SCIENTIFIC IMAGING" in a smaller, lighter yellow font underneath.
Both systems produce digital images which can be sent to hardcopy printers, softcopy diagnostic workstations and/or stored in archive.
The proposed MOZART SUPRA(XPERT84) Specimen Radiography System, utilizes the predicate devices, MOZART SUPRA(XPERT84) Specimen Radiography System, with the same 90 kVp monobloc X-ray tube installed in the enclosed head of the cabinet as the predicate and both the Hologic TRIDENT(Predicate 2) and the proposed XPERT 84 Specimen Radiography System utilizes the Amorphous Selenium detector mounted stationary in the bottom replacing the predicate CMOS detector of the same shielded cabinet x-ray unit with proprietary software installed into an off the shelf personal computer, Microsoft Windows 10 Operating System, and a 2 megapixel or greater portrait type monitor.
The performance of the proposed AXS-2430V2 detector has been verified and cleared via 510k K193166 as incorporated into the Siemens MAMMOMAT Revelation.
Both systems utilize the same DIGICOM software.
The "Level of Concern" of MOZART SUPRA (XPERT 84) Specimen Radiography System software (DIGICOM) is "moderate"
The previous claims have been proven internally and validated and verified via imaging of line-pair gauges and phantoms.
| DETECTORS | |||||
|---|---|---|---|---|---|
| Currently ApprovedKUBTEC K200756 | Proposed | HOLOGIC APPROVEDK111508 | |||
| Manufacturer | Manufacturer | Manufacturer | |||
| Rad-Icon | Analogic | Hologic | |||
| MFRModel # | 6K | 2329 | AXS-1824V2 | AXS-2430V2 | 1214 |
| KUBTECSKU | DV136 | DV266 | DV210 | DV270 | n/a |
| Type | Active PixelCMOS | Active PixelCMOS | AmorphousSelenium | AmorphousSelenium | Amorphous Selenium |
| DetectionMethod | In-direct | In-direct | Direct | Direct | Direct |
| Size (cm) | 11.4 x 14.5 | 22.8 x 29.1 | 17.4 x 23.9 | 23.9 x 30.5 | 12 x 14 |
| Resolution | 49.5μm | 49.5μm | 85μm | 85μm | 70μm |
| Pixel size | 2304 x | 4608 x 5890 | 2048 x 2816 | 2816 x 3584 | 2285 x 2571 |
They have been clinically validated by reads of excised breast biopsies by a board certified radiologist.
111 Research Drive, Stratford, CT 06615 USA. Tel: +1(203) 364-8544; Fax:+1 (203) 255-7494 KUBTEC is a registered trademark of KUB Technologies, Inc.
{9}------------------------------------------------
Image /page/9/Picture/0 description: This image shows a table with the specifications for the 2940 model. The table includes the DQE, MTF, and Bit Depth. The DQE at 1 Lp/mm is >45%, >55%, >50%, >70%, and >55%. The MTF at 1 Lp/mm is >80%, >85%, >90%, >95%, and >90%. The Bit Depth is 16 bits, 16 bits, 16 bits, 16 bits, and 14 bits.
Detector Specifications
Radiographic Magnification: 1.0x to 5x Exposure Control: Automatic or Manual X-ray Duty Cycle: 100%
Radiation Safety - 21CFR1020.40 Radiation shielded cabinet Compartment door equipped with dual safety interlocks
Radiation leakage:
Less than 0.1 mR/hr at 5 cm from any exterior surface at maximum kV.
VI. CONCLUSION
The Kubtec MOZART SUPRA (XPERT84) is as safe and effective as the predicate and previously approved MOZART SUPRA (XPERT84) devices, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate devices and conforms to applicable medical device safety standards.
N/A