K Number
K210830

Validate with FDA (Live)

Device Name
VUZE System
Manufacturer
Date Cleared
2022-01-03

(290 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VUZE System is intended to enable users to load pre-operative 3D images and planning data and register and overlay this data in real time with intra-operative 2D radiographic images of the same anatomy to support device guidance during interventional spinal procedures. The system also offers pre-operative surgical planning including implant sizing, entry location, and trajectory determination along with intra-operative guidance and tool trajectory / position confirmation.

Device Description

The VUZE System (the "System") enables users to load pre-operative 3D images and planning data and register and overlay this data in real time with intra-operative 2D radiographic images of the same anatomy. The System supports device quidance during minimally invasive spinal surgery, including the stabilization of the spine by means of fixation, fixation coupled with fusion, vertebroplasty or kyphoplasty. Applicable vertebrae are within the range of S1 through T7.

The System offers pre-operative surgical planning including implant sizing, entry location and trajectory determination, along with intra-operative quidance and tool trajectory/position confirmation by displaying a graphical representation of a tool tracked by intraoperative 2D images onto a patient's pre-operative 3D images.

The main system components include:

  • Workstation running the VUZE Planning and Procedure software (pre-installed)
  • Housing for the workstation, with a front door for user access as well as a back service door
  • 32" touchscreen
  • Isolation transformer
  • Internal video acquisition device (frame grabber)
  • Wheeled cart on which the above-listed items are placed.
AI/ML Overview

The VUZE System is a medical imaging system intended to enable users to load pre-operative 3D images and planning data, and then register and overlay this data in real time with intra-operative 2D radiographic images of the same anatomy. This supports device guidance during interventional spinal procedures, including pre-operative surgical planning and intra-operative guidance/tool trajectory confirmation.

Here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly present a table of numerical acceptance criteria for specific performance metrics (e.g., accuracy values with defined thresholds). Instead, it states that the device successfully completed all evaluations and testing to demonstrate substantial equivalence and that technological differences do not raise new or different questions of safety and effectiveness.

The "Performance Data" section indicates:

  • The VUZE System was verified and validated in accordance with 21 CFR 820.30.
  • The device successfully completed all of the evaluations and testing shown below (referring to a list of tests).
  • The results of testing provide assurance that the device is as safe and effective as the predicate.

Given this, the acceptance criteria are implicitly tied to the successful completion of the listed tests and adherence to recognized consensus standards, ensuring the device is "as safe and effective as the predicate." The reported performance is that the device met these implicit criteria by successfully completing all tests.

Implicit Acceptance Criteria and Reported Performance Summary:

Acceptance Criterion (Implicit)Reported Device Performance
Compliance with 21 CFR 820.30 for verification and validationDevice was verified and validated in accordance with 21 CFR 820.30.
Demonstration of substantial equivalence to predicate device K092639 (Innova Vision Applications)Device successfully completed all evaluations and testing, demonstrating substantial equivalence and that technological differences raise no new questions of safety/effectiveness.
Successful completion of hardware component and functional unit verification and validationCompleted successfully.
Successful completion of packaging / transportation validationCompleted successfully.
Successful completion of software verification / validation at unit, integration, and system levelsCompleted successfully.
Adherence to IEC 60601-1 for Basic Safety and Essential PerformanceCompleted successfully, adhering to IEC 60601-1.
Adherence to IEC 60601-1-2 for Electromagnetic Compatibility (EMC)Completed successfully, adhering to IEC 60601-1-2.
Successful Simulated Use / Quantitative Accuracy testingCompleted successfully.
Successful Qualitative Image Output ValidationCompleted successfully.
Successful Summative Usability ValidationCompleted successfully.
Conformance to various voluntary recognized consensus standards (e.g., IEC 62304, IEC 62366-1, ISO 14971, NEMA PS 3.1 3.20, ASTM D4169-16)Device designed and tested in conformance to these standards.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify a sample size for a test set in terms of patient cases or images.
It states: "No animal or human clinical data were needed to demonstrate substantial equivalency."
The testing appears to be primarily based on engineering verification and validation activities, including "Simulated Use / Quantitative Accuracy" and "Qualitative Image Output Validation," rather than a clinical test set with patient data. Therefore, details about data provenance (country of origin, retrospective/prospective) are not applicable or provided.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Since no human clinical data or explicit "test set" in a clinical sense with ground truth established by experts is mentioned, this information is not provided in the document. The validation relied on engineering and simulated testing.

4. Adjudication Method for the Test Set:

As no clinical test set requiring expert ground truth or adjudication is described, an adjudication method is not applicable and not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC comparative effectiveness study was not done.
  • The document explicitly states: "No animal or human clinical data were needed to demonstrate substantial equivalency."
  • Consequently, there is no effect size reported for human readers improving with or without AI assistance, as this type of study was not performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

The device itself is a "Medical Image Management and Processing System" that provides "device guidance" and "tool trajectory/position confirmation" by displaying data for a human user (surgeon). It is not described as a standalone diagnostic or automated detection algorithm in the typical sense.
The listed tests include "Software Verification / Validation at the unit, integration and system levels" and "Simulated Use / Quantitative Accuracy," which would involve evaluating the algorithm's performance. The "Simulated Use" likely evaluates the system's performance as intended with a human-in-the-loop, even if using simulated data.
The prompt asks "algorithm only without human-in-the-loop performance". The information suggests that the system's function is to aid humans, so a purely "standalone" evaluation that completely excludes the human role might not be fully representative of its intended use. However, the software validation and quantitative accuracy testing would assess the algorithm's underlying technical performance. The document does not explicitly describe a "standalone" study in contrast to a human-in-the-loop study, but rather a comprehensive set of engineering tests for the full system.

7. The Type of Ground Truth Used:

For tests like "Simulated Use / Quantitative Accuracy" and "Qualitative Image Output Validation," the ground truth would likely be established through:

  • Known simulated scenarios/phantoms: For "Simulated Use / Quantitative Accuracy," the "ground truth" would be the precisely known positions or trajectories in a simulated environment or phantom model, against which the system's output is measured.
  • Pre-defined specifications/expected outputs: For "Qualitative Image Output Validation" and other verification steps, the ground truth would be the expected display properties, co-registration accuracy, or functional behavior as defined in the product requirements.
  • Predicate device performance: Since substantial equivalence is claimed, the performance of the predicate device (Innova Vision Applications K092639) implicitly serves as a benchmark for safety and effectiveness.

There is no mention of expert consensus, pathology, or outcomes data being used as ground truth, as no human clinical data was used.

8. The Sample Size for the Training Set:

The document does not mention a training set or its sample size. The VUZE System is described as using "3D imaging data of the patient's target anatomy and pre-operative surgical planning information registered against intraoperative 2D x-ray images." This suggests it's an image processing and guidance system rather than a machine learning model that requires a distinct "training set" in the context of deep learning, although underlying algorithms might have been developed using some data. However, for regulatory purposes, specific training set data is not provided.

9. How the Ground Truth for the Training Set Was Established:

Since no training set is described, information on how its ground truth was established is not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

VUZE Medical Ltd. % Clay Anselmo Principal Quality and Regulatory Consultant Shriner & Associates Inc. 429 Whitepine Creek Road Trout Creek, Montana 59874

Re: K210830

February 8, 2022

Trade/Device Name: VUZE System Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ

Dear Clay Anselmo:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 3, 2022. Specifically, FDA is updating this SE Letter due to a typo in the clearance date, which was incorrectly dated as January 3, 2021, as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jessica Lamb, OHT7: Office of In Vitro Diagnostics and Radiological Health, 301-796-6167, jessica.lamb(@fda.hhs.gov.

Sincerely.

Jessica Lamb

Jessica Lamb Assistant Director Mammography Ultrasound and Imaging Software Branch Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

January 3, 2021

Image /page/1/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

VUZE Medical Ltd. % Clay Anselmo Principal Quality and Regulatory Consultant Shriner & Associates Inc. 429 Whitepine Creek Road TROUT CREEK MT 59874

Re: K210830

Trade/Device Name: VUZE System Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: November 18, 2021 Received: November 23, 2021

Dear Clay Anselmo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{2}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K210830

Device Name Vuze System

Indications for Use (Describe)

The VUZE System is intended to enable users to load pre-operative 3D images and planning data and register and overlay this data in real time with intra-operative 2D radiographic images of the same anatomy to support device guidance during interventional spinal procedures. The system also offers pre-operative surgical planning including implant sizing, entry location, and trajectory determination along with intra-operative guidance and tool trajectory / position confirmation.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary — K210830

Introduction:

This document contains the 510(k) Summary for the VUZE Medical - VUZE System. The content of this summary is based on the requirements set forth in 21 CFR 807.92(c).

Submitter Information:

Applicant / ManufacturerVUZE Medical Ltd.
Name and Address13 Zarhin St.
Ra'Anana 4366241, Israel
510(k) contact personClay Anselmo
Principal Quality and Regulatory Con- sultant
Shriner & Associates
429 Whitepine Creek Road
Trout Creek, MT 59874
clay.anselmo@shrinerandassociates.com
(303) 907-2955
Date prepared16-March-2021

Device Identification

Trade namesVUZE System
Common nameVUZE Medical Imaging System for spinalinterventions
Classification nameMedical Image Management and Pro-cessing System
Regulation Number21 CFR Part 892.2050
ClassificationClass II
Product CodePrimary: LLZ
Predicate Device
Trade names:Innova Vision Applications
510(k) number:K092639

{5}------------------------------------------------

Reference Device

Trade names:Renaissance
510(k) number:K140167

Device Description:

The VUZE System (the "System") enables users to load pre-operative 3D images and planning data and register and overlay this data in real time with intra-operative 2D radiographic images of the same anatomy. The System supports device quidance during minimally invasive spinal surgery, including the stabilization of the spine by means of fixation, fixation coupled with fusion, vertebroplasty or kyphoplasty. Applicable vertebrae are within the range of S1 through T7.

The System offers pre-operative surgical planning including implant sizing, entry location and trajectory determination, along with intra-operative quidance and tool trajectory/position confirmation by displaying a graphical representation of a tool tracked by intraoperative 2D images onto a patient's pre-operative 3D images.

The main system components include:

  • Workstation running the VUZE Planning and Procedure software (pre-installed) ●
  • Housing for the workstation, with a front door for user access as well as a back ser-● vice door
  • 32" touchscreen
  • Isolation transformer ●
  • Internal video acquisition device (frame grabber) ●
  • Wheeled cart on which the above-listed items are placed. ●

Indications for Use:

The VUZE System is intended to enable users to load pre-operative 3D images and planning data and register and overlay this data in real time with intra-operative 2D radiographic images of the same anatomy to support device guidance during interventional spinal procedures. The system also offers pre-operative surgical planning including implant sizing, entry location, and trajectory determination along with intra-operative quidance and tool trajectory / position confirmation.

Technological Characteristics Comparison:

Substantial Equivalence: The VUZE System is substantially equivalent to the Innova Vision Application (a.k.a. TrackVision) K092639. A reference device, (Mazor Robotics Renaissance System K140167) was used to address minor technology differences between the VUZE System and the TrackVision device.

The 510(k) Substantial Equivalence Decision-making Process (detailed) from FDA Guidance - The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] was followed as described below:

{6}------------------------------------------------

  • The VUZE System device has the same intended use and similar indications for use as the Predicate device.
  • The VUZE System device uses the same fundamental technology as the Predicate . device and very similar detailed technological solutions as follows:
    • o The VUZE System utilizes 3D imaging data of the patient's target anatomy and pre-operative surgical planning information registered against intraoperative 2D x-ray images to support the correct placement of interventional devices (trajectory and position).
    • This approach to presenting co-registered 3D and 2D imaging information for о the purposes of aiding tool guidance and tool placement confirmation is the same as the predicated device.
  • The minor differences between the VUZE System device and the Predicate do not ● raise new types of questions of safety or effectiveness.
    • o The differences in technology do not raise different questions of safety or effectiveness and were evaluated through comprehensive bench and usability verification and validation testing. The results of testing provide assurance that the device is as safe and effective as the predicate.

For a more detailed comparison refer to the Substantial Equivalence comparison table included below.

Performance Data:

There are no identified special controls or performance standards for this device.

The VUZE System was verified and validated in accordance with 21 CFR 820.30. The following tests were completed to demonstrate substantial equivalence and that any technological differences do not raise new or different questions of safety and effectiveness. The device successfully completed all of the evaluations and testing shown below. The standards shown in the following section were used, where applicable to conduct testing and evaluate results.

  • Hardware component and functional unit verification and validation ●
  • Packaging / Transportation Validation ●
  • Software Verification / Validation at the unit, integration and system levels ●
  • IEC 60601-1 Basic Safety and Essential Performance ●
  • IEC 60601-1-2 Electromagnetic Compatibility (EMC) ●
  • Simulated Use / Quantitative Accuracy ●
  • Qualitative Image Output Validation ●
  • Summative Usability Validation ●

No animal or human clinical data were needed to demonstrate substantial equivalency.

The device has been designed and tested in conformance to the following voluntary recognized consensus standards:

  • ANSI / AAMI / IEC 60601-1: 2005 /(R)2012 +A1:2012: Medical Electrical Equipment ● - Part 1: General Requirements for Basic Safety and Essential Performance (FDA Recognition #19-4)

{7}------------------------------------------------

  • IEC 60601-1-2: Edition 4.0, 2014-02 Medical Electrical Equipment Part 1-2: Gen-● eral Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests Sections 7, 8 professional healthcare facility environment (FDA Recognition #19-8).
  • IEC 62304:2006/A1:2016, Medical Device Software Software Lifecycle Processes ● (FDA Recognition #13-79)
  • IEC 62366-1:2015: Medical devices Part 1: Application of usability engineering to . medical devices (FDA Recognition #5-114)
  • IEC 62366-1-6:2013-10 Edition 3.1: Medical Electrical Equipment Part 1-6: Collat-. eral Standard - Usability (FDA Recognition #5-89)
  • ISO 14971:2019 Medical Devices: Application of Risk Management to Medical De-. vices (FDA Recognition #5-125)
  • ISO 15223:2016 Medical devices Symbols to be used with medical devices labels, . labeling, and information to be supplied - Part 1: General requirements (FDA Recognition #5-117)
  • NEMA PS 3.1 3.20 (2016) Digital Imaging and Communications in Medicine (DI-● COM) Set (FDA Recognition #12-300)
  • ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers . and Systems (FDA Recognition #14-499)
Substantial Equivalence Comparison:
---------------------------------------
DeviceRenaissance
CharacteristicVUZE(Subject Device)Innova Vision Applications(Predicate, K092639)(ReferenceDevice,K140167)Conclusion
Intended UseIntended as an imaging aidfor interventional proce-dures to allow comparisonof planned tool pathsagainst actual tool place-mentIntended as an imaging aidfor interventional proceduresto allow comparison ofplanned tool paths againstactual tool placementN/AIdentical toPredicate
Indications for UseThe VUZE System is intended toenable users to load pre-operative3D images and planning data andregister and overlay this data inreal time with intra-operative 2Dradiographic images of the sameanatomy to support device guid-ance during interventional spinalprocedures. The system also offerspre-operative surgical planningincluding implant sizing, entry loca-tion, and trajectory determinationalong with intra-operative quidanceand tool trajectory / position con-firmation.Innova Vision Applicationssoftware is intended to ena-ble users to load 3D da-tasets and overlay and regis-ter in real time these 3D da-tasets with radioscopic orradiographic images of thesame anatomy in order tosupport catheter/deviceguidance during interven-tional procedures.N/ASubstantiallyEquivalent
Device ClassIIIIIIIdentical toAll
ProcodesLLZLLZOLO, HAW,LLCIdentical toPredicate
DeviceRenaissance
CharacteristicVUZE(Subject Device)Innova Vision Applications(Predicate, K092639)(ReferenceDevice,K140167)Conclusion
AnatomyThoracic and Lumbar SpineNot specific. Anywhere ra-diographic images of theanatomy are available.SpineSubset ofPredicate,Identical toReferenceDevice;SubstantiallyEquivalent
Surgical Access TypePercutaneousPercutaneousOpen andPerc.Identical toPredicate
Patient PopulationGeneral Spine SurgeryAll tool-based interventions,not limited to specific clinicalindicationGeneralSpine Sur-gerySubset ofPredicate:SubstantiallyEquivalent
Target UserInterventional Spine Sur-geonInterventional Surgeons in-cluding Ortho and SpinesurgeonsInterventionalSpine / Or-thopedicSurgeonSubset ofPredicate,Identical toReferenceDevice;SubstantiallyEquivalent
EnvironmentSurgical SuiteCatheter Lab, Surgical SuiteSurgicalSuiteSubset ofPredicate;SubstantiallyEquivalent
Hardware PlatformDedicated WorkstationDedicated WorkstationProprietaryHW platformIdentical toPredicate
Work-Flow3D imaging, planning, 2Dimaging, 3D to 2D co-registration, intervention,confirmation3D imaging, planning, 3D &2D imaging, 3D to 3D/2D co-registration, intervention,confirmation3D imaging,planning, 2Dimaging, 3Dto 2D co-registration,intervention,confirmationSubset ofPredicate,Identical toReferenceDevice;SubstantiallyEquivalent
3D Image Data SourceCTCone-Beam CTCTSubstantiallyEquivalent
Interoperative ImageCo-Registration3D image to X-ray usingDigitally Reconstructed Ra-diographs (DRR's)3D Image to Fluoroscopyand / or X-ray using X-rayposition data.3D image toX-ray usingDRR'sSubset ofPredicate,Identical toReferenceDevice;SubstantiallyEquivalent
Tool UseFree handFree handRoboticAssistanceIdentical toPredicate

{8}------------------------------------------------

Conclusion:

The VUZE System is substantially equivalent to the Innova Vision Applications (K092639).

N/A