Intended Use

Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.

These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy:

Test Chemotherapy Drug (Concentration)Minimum Breakthrough Detection Time in minute
Fluorouracil (50.0mg/ml)>240
Etoposide (20.0mg/ml)>240
Cyclophosphamide (20.0mg/ml)>240
Carmustine (3.3mg/ml)14.1
Thiotepa (10.0mg/ml)16.9
Paclitaxel (6.0mg/ml)>240
Doxorubicin Hydrochloride (2.0mg/ml)>240
Methotrexate (25.0mg/ml)>240
Vincristine Sulfate (1.0mg/ml)>240

"WARNING: Do not use with Carmustine and Thiotepa"

Device Description

It is the powder-free variation of the class I sterile latex surgical gloves made on-line by surface treatment on the outside and polymer coating on inner surface.

AI/ML Overview

The provided text is related to the 510(k) premarket notification for surgical gloves and does not contain information about an AI/ML powered medical device. Therefore, I cannot extract the requested information regarding acceptance criteria, study details, expert involvement, and ground truth establishment as it pertains to an AI/ML device.

The document details the substantial equivalence of new surgical gloves to a legally marketed predicate device based on various physical, chemical, and biocompatibility tests, as well as their resistance to chemotherapy drugs. It explicitly states that "Clinical testing is not applicable to this device." This confirms that there was no study involving human readers or AI in the way your prompt describes.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 3, 2021

Careplus (M) SDN BHD Lim Shyan CEO/ Managing Director Lot 120 & 121, Jalan Senawang 3, Senawang Industrial Estate Seremban, Negeri Sembilan 70450 Malaysia

Re: K210724

Trade/Device Name: ENCORE Latex Acclaim, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (50 ug/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs, GAMMEX Latex Standard, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (50 ug/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO, LZC Dated: June 22, 2021 Received: June 28, 2021

Dear Lim Shyan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray III -S

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below

510(k) Number (if known) K210724

Device Name

ENCORE® Latex Acclaim, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 µg/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

Indications for Use (Describe)

Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.

These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy:

Test Chemotherapy Drug (Concentration)Minimum Breakthrough Detection Time in minute
Fluorouracil (50.0mg/ml)>240
Etoposide (20.0mg/ml)>240
Cyclophosphamide (20.0mg/ml)>240
Carmustine (3.3mg/ml)14.1
Thiotepa (10.0mg/ml)16.9
Paclitaxel (6.0mg/ml)>240
Doxorubicin Hydrochloride (2.0mg/ml)>240
Methotrexate (25.0mg/ml)>240
Vincristine Sulfate (1.0mg/ml)>240

"WARNING: Do not use with Carmustine and Thiotepa"

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below

510(k) Number (if known) K210724

Device Name

GAMMEX® Latex Standard, Powder free Latex Surgical Gloves with Protein Labeling Claim (< 50 µg/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

Indications for Use (Describe)

Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.

These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy:

Test Chemotherapy Drug (Concentration)Minimum Breakthrough Detection Time in minutes
Fluorouracil (50.0mg/ml)>240
Etoposide (20.0mg/ml)>240
Cyclophosphamide (20.0mg/ml)>240
Carmustine (3.3mg/ml)14.1
Thiotepa (10.0mg/ml)16.9
Paclitaxel (6.0mg/ml)>240
Doxorubicin Hydrochloride (2.0mg/ml)>240
Methotrexate (25.0mg/ml)>240
Vincristine Sulfate (1.0mg/ml)>240

"WARNING: Do not use with Carmustine and Thiotepa"

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K210724

510(K) SUMMARY

  • 1.0 Applicant: CAREPLUS (M) SDN BHD
    • Lot 120 & 121 , Jalan Senawang 3, Address: Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia.

Phone No. 60-6-6772781 Fax No. 60-6-6772780

Contact Person Lim Kwee Shyan

  • 2.0 Date of Summary 31st July, 2021

3.0 Device Information

Device Name: Surgeons Glove, Latex Powder Free with Protein Content Labeling Claim (< 50 µg/dm² per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

Glove Trade Name: ENCORE® Latex Acclaim, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 µg/dm² per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

GAMMEX® Latex Standard. Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 uq/dm² per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

Common Name: Surgical Gloves Product Code: KGO Subsequent Product Code: LZC Classification Name: Surgeon's Gloves Patient Examination Gloves Specialty (21 CFR 880.6250, KGO)

4.0 Device Description

It is the powder-free variation of the class I sterile latex surgical gloves made on-line by surface treatment on the outside and polymer coating on inner surface.

5.0 Predicate Device

Predicate device: Encore Acclaim Sterile Powder Free Latex Surgical Gloves, Tested for use with Chemotherapy Drugs with a Protein Content Label Claim <50 ug/dm2 per Glove of Extractable Protein, 510(k) number K103714, product code KGO.

6.0 Intended Use of Device

Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.

{5}------------------------------------------------

K210724

7.0 Technological Characteristics Comparison
CharacteristicsSubject Glove(ENCORE® Latex Acclaim, Powder freeLatex Surgical Gloves with ProteinContent Labeling Claim (< 50 μg/dm²per glove of total aqueous extractableprotein) and Tested for use withChemotherapy drugs,K210724)Predicate Device(Encore Acclaim Sterile Powder FreeLatex Surgical Gloves, Tested for usewith Chemotherapy Drugs with aProtein Content Label Claim, K103714)Discussion
Product CodeKGOLZCKGOLZCSame
Regulation Number21CFR878.446021CFR880.625021CFR878.446021CFR880.6250Same
Device ClassClass 1Class 1Same
Freedom from holesMeets ASTM D3577Meets ASTM D5151Meets ASTM D3577Meets ASTM D5151Same
DimensionSame
Length (size: 5.5), mmLength (size: 6.0), mmLength (size: 6.5), mmLength (size: 7.0), mmLength (size: 7.5), mmLength (size: 8.0), mmLength (size: 8.5), mmLength (size: 9.0), mmThickness (cuff), mmThickness (palm), mmThickness (finger), mmWidth (size: 5.5), mmWidth (size: 6.0), mmWidth (size: 6.5), mmWidth (size: 7.0), mmWidth (size: 7.5), mmWidth (size: 8.0), mmWidth (size: 8.5), mmWidth (size: 9.0), mmMeet 245mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 0.10mm minMeet 0.10mm minMeet 0.10mm minMeet 70 ± 6 mmMeet 76 ± 6 mmMeet 83 ± 6 mmMeet 89 ± 6 mmMeet 95 ± 6 mmMeet 102 ± 6 mmMeet 108 ± 6 mmMeet 114 ± 6 mmMeet 245mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 265mm minMeet 0.10mm minMeet 0.10mm minMeet 0.10mm minMeet 70 ± 6 mmMeet 76 ± 6 mmMeet 83 ± 6 mmMeet 89 ± 6 mmMeet 95 ± 6 mmMeet 102 ± 6 mmMeet 108 ± 6 mmMeet 114 ± 6 mm
Physical Properties(Before Ageing)i) Tensile Strength(Mpa)ii) Ultimate Elongation(%)iii) Stress at 500%Elongation(After Ageing)i) Tensile Strength(Mpa)ii) Ultimate Elongation(%)Meets ASTM D3577Meets ASTM D3577Same
Powder ContentMeets ASTM D3577Meets ASTM D6124Meets ASTM D3577Meets ASTM D6124Same
Biocompatibility Testi) Primary Skin IrritationTestii) Dermal SensitizationTestPassesi) Primary Skin Irritation Test.Conclusion: Under the conditions of thisstudy, the test material did not causeand irritant response.ii) Dermal Sensitization Test.Conclusion: Under the conditions of thisstudy, the test material did not producea skin sensitization effect.Passesi) Primary Skin Irritation Test.Conclusion: Under the conditions of thisstudy, the test material did not causeand irritant response.ii) Dermal Sensitization Test.Conclusion: Under the conditions of thisstudy, the test material did not producea skin sensitization effect.Same
iii) In Vitro Cytotoxicity Testiii) In vitro Cytotoxicity test.Conclusions: Under the conditions ofthis study, the test material is Cytotoxicat dilutions of 1:2, 1:4, 1:8, 1:6 andNon-cytotoxic, grade 2 at 1:32 and 1:64dilutions.iii) No data availableDifferent
iv) Acute Systemic Toxicityiv) Acute Systemic Toxicity study.Conclusion: Under the conditions of thisstudy, the test material both inner andouter surface did not reveal systemictoxicity.iv) No data availableDifferent
Protein Label ClaimContains 50 µg/dm² or less per glove oftotal aqueous extractable protein oftotal water extractable protein per gram.Contains 50 micrograms or less of totalaqueous extractable protein of totalwater extractable protein per gram.Same
Chemo Drugs ClaimChemo ClaimChemo ClaimSame
ColorNaturalNaturalSame
White PigmentTitanium DioxideTitanium DioxideSame
Indications for usePowder Free Surgical gloves are steriledisposable devices intended to be wornby operating room personnel to protecta surgical wound from contamination.Powder Free Surgical gloves are steriledisposable devices intended to be wornby operating room personnel to protecta surgical wound from contamination.Same
Shelf-life3 years3 yearsSame
Sterile PackagingPacked 1 pair in a wallet. One wallet ina pouch.Packed 1 pair in a wallet. One wallet ina pouch.Same
SterilitySterileSterileSame
Sterilization MethodGamma IrradiationGamma IrradiationSame
CharacteristicsSubject GlovePredicate DeviceDiscussion
(GAMMEX® Latex Standard, Powderfree Latex Surgical Gloves with ProteinContent Labeling Claim (< 50 µg/dm²per glove of total aqueous extractableprotein) and Tested for use withChemotherapy drugs,K210724)(Encore Acclaim Sterile Powder FreeLatex Surgical Gloves, Tested for usewith Chemotherapy Drugs with aProtein Content Label Claim, K103714)
Product CodeKGOLZCKGOLZCSame
Regulation Number21CFR878.446021CFR880.625021CFR878.446021CFR880.6250Same
Device ClassClass 1Class 1Same
Freedom from holesMeets ASTM D3577Meets ASTM D5151Meets ASTM D3577Meets ASTM D5151Same
DimensionSame
Length (size: 5.5), mmMeet 245mm minMeet 245mm min
Length (size: 6.0), mmMeet 265mm minMeet 265mm min
Length (size: 6.5), mmMeet 265mm minMeet 265mm min
Length (size: 7.0), mmMeet 265mm minMeet 265mm min
Length (size: 7.5), mmMeet 265mm minMeet 265mm min
Length (size: 8.0), mmMeet 265mm minMeet 265mm min
Length (size: 8.5), mmMeet 265mm minMeet 265mm min
Length (size: 9.0), mmMeet 265mm minMeet 265mm min
Thickness (cuff), mmMeet 0.10mm minMeet 0.10mm min
Thickness (palm), mmMeet 0.10mm minMeet 0.10mm min
Thickness (finger), mmMeet 0.10mm minMeet 0.10mm min
Width (size: 5.5), mmMeet 70 ± 6 mmMeet 70 ± 6 mm
Width (size: 6.0), mmMeet 76 ± 6 mmMeet 76 ± 6 mm
Width (size: 6.5), mmMeet 83 ± 6 mmMeet 83 ± 6 mm
Width (size: 7.0), mmMeet 89 ± 6 mmMeet 89 ± 6 mm
Width (size: 7.5), mmMeet 95 ± 6 mmMeet 95 ± 6 mm
Width (size: 8.0), mmMeet 102 ± 6 mmMeet 102 ± 6 mm
Width (size: 8.5), mmMeet 108 ± 6 mmMeet 108 ± 6 mm
Width (size: 9.0), mmMeet 114 ± 6 mmMeet 114 ± 6 mm
Physical Properties(Before Ageing)i) Tensile Strength(Mpa)ii) Ultimate Elongation(%)iii) Stress at 500%ElongationMeets ASTM D3577Meets ASTM D3577Same
(After Ageing)i) Tensile Strength(Mpa)ii) Ultimate Elongation(%)Meets ASTM D3577Meets ASTM D3577Same
Powder ContentMeets ASTM D3577Meets ASTM D6124Meets ASTM D3577Meets ASTM D6124Same
Biocompatibility Testi) Primary Skin IrritationTestPassesi) Primary Skin Irritation Test.Conclusion: Under the conditions of thisstudy, the test material did not causeand irritant response.Passesi) Primary Skin Irritation Test.Conclusion: Under the conditions of thisstudy, the test material did not causeand irritant response.Same
ii) Dermal SensitizationTestii) Dermal Sensitization Test.Conclusion: Under the conditions of thisstudy, the test material did not producea skin sensitization effect.ii) Dermal Sensitization Test.Conclusion: Under the conditions of thisstudy, the test material did not producea skin sensitization effect.Same
iii) In Vitro Cytotoxicity Testiii) In vitro Cytotoxicity test.Conclusions: Under the conditions ofthis study, the test material is Cytotoxicat dilutions of 1:2, 1:4, 1:8, 1:6 andNon-cytotoxic, grade 2 at 1:32 and 1:64dilutions.iii) No data availableDifferent
iv) Acute Systemic Toxicityiv) Acute Systemic Toxicity study.Conclusion: Under the conditions of thisstudy, the test material both inner andouter surface did not reveal systemictoxicity.iv) No data availableDifferent
Protein Label ClaimContains 50 µg/dm² or less per glove oftotal aqueous extractable protein oftotal water extractable protein per gram.Contains 50 micrograms or less of totalaqueous extractable protein of totalwater extractable protein per gram.Same
Chemo Drugs ClaimChemo ClaimChemo ClaimSame
ColorNaturalNaturalSame
White PigmentTitanium DioxideTitanium DioxideSame
Indications for usePowder Free Surgical gloves are steriledisposable devices intended to be wornby operating room personnel to protecta surgical wound from contamination.Powder Free Surgical gloves are steriledisposable devices intended to be wornby operating room personnel to protecta surgical wound from contamination.Same
Shelf-life3 years3 yearsSame
Sterile PackagingPacked 1 pair in a wallet. One wallet ina pouch.Packed 1 pair in a wallet. One wallet ina pouch.Same
Sterility StatusSterilitySterileSame
Sterilization MethodGamma IrradiationGamma IrradiationSame

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

K210724

ChemotherapyDrugASTMD6978-05Subject Glove, K210724(ENCORE® Latex Acclaim, Powder free Latex Surgical Gloves withProtein Content Labeling Claim (< 50 µg/dm² per glove of totalaqueous extractable protein) and Tested for use withChemotherapy drugs andGAMMEX® Latex Standard, Powder free Latex Surgical Gloves withProtein Content Labeling Claim (< 50 µg/dm² per glove of totalaqueous extractable protein) and Tested for use withChemotherapy drugs)
Test ChemotherapyDrugConcentrationMinimum Breakthrough Detection Time (min)
Fluorouracil50.0mg/ml> 240
Etoposide20.0mg/ml> 240
Cyclophosphamide(Cytoxan)20.0mg/ml> 240
*Carmustine (BCNU)3.3mg/ml14.1
*Thiotepa10.0mg/ml16.9
Paclitaxel6.0mg/ml> 240
DoxorubicinHydrochloride2.0mg/ml> 240
Methotrexate25.0mg/ml> 240
Vincristine Sulfate1.0mg/ml> 240
Warning Statement*Warning: Do not use with Carmustine and ThioTepa"

Summary of Non-Clinical Testing 8.0

The non-clinical performance test data demonstrate that the subject devices met the acceptance criteria or the specification for the test methodology or standard shown below.

ENCORE® Latex Acclaim, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (<50 µg/dm² per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

Standard/Test MethodPurposeAcceptance CriteriaResults
ASTM D 3577ASTM D 5151Freedom From HolesMeet requirement inspection level G-I, AQL 1.5Pass
ASTM D 3577DimensionSize 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 Length, min. mm 245 Thickness, min. mm 0.10 Width, ± 6 mm 70 76 83 89 95 102 108 114Pass
ASTM D 3577,Type IPhysical PropertiesBefore Aging After Accelerated Aging TensileStrength UltimateElongation Stress at 500%Elongation TensileStrength UltimateElongation 24 MPa min 750% min 5.5 MPa max 18 MPa min 560% minPass
ASTM D 3577ASTM D 6124Powder ContentNot more than 2mg per glovePass
ASTM D 3577ASTM D 5712Protein ContentThis glove contains 50 µg/dm² or less per glove of total aqueousextractable protein of total water extractable protein per gramPass
ISO10993-10BiocompatibilityPrimary Skin IrritationTestDermal SensitizationTestNot a skin irritant under the conditions of test.Not a skin sensitizer under the conditions of testPass
ISO 10993-11BiocompatibilityAcute SystemicToxicityDid not reveal systemic toxicity under the conditions of test.Pass

{10}------------------------------------------------

GAMMEX® Latex Standard, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 µg/dm² per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

Standard/Test MethodPurposeAcceptance CriteriaResults
ASTM D 3577ASTM D 5151Freedom From HolesMeet requirement inspection level G-I, AQL 1.5Pass
ASTM D 3577DimensionSizeLength, min. mmThickness, min. mmWidth, ± 6 mm5.5245706.00.10766.5837.0897.5958.01028.51089.0265114Pass
ASTM D 3577,Type IPhysical PropertiesBefore AgingAfter Accelerated AgingPass
TensileStrengthUltimateElongationStress at 500%ElongationTensileStrengthUltimateElongation
24 MPa min750% min5.5 MPa max18 MPa min560% min
ASTM D 3577ASTM D 6124Powder ContentNot more than 2mg per glovePass
ASTM D 3577ASTM D 5712Protein ContentThis glove contains 50 µg/dm² or less per glove of total aqueousextractable protein of total water extractable protein per gramPass
ISO10993-10BiocompatibilityPrimary Skin IrritationTestDermal SensitizationTestNot a skin irritant under the conditions of test.Not a skin sensitizer under the conditions of testPass
ISO 10993-11BiocompatibilityAcute SystemicToxicityDid not reveal systemic toxicity under the conditions of test.Pass

Summary of Clinical Testing 9.0

Clinical testing is not applicable to this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device in 510(K) K210724, the ENCORE® Latex Acclaim, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 µg/dm² per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs and GAMMEX®Latex Standard, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 µg/dm² per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs, proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K103714.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).