K Number
K210666

Validate with FDA (Live)

Device Name
Chest-CAD
Date Cleared
2021-07-20

(137 days)

Product Code
Regulation Number
892.2070
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Chest-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies using machine learning techniques to identify, categorize, and highlight suspicious regions of interest (ROI). Any suspicious ROI identified by Chest-CAD is assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. The device is intended for use as a concurrent reading aid for physicians. Chest-CAD is indicated for adults only.

Device Description

Chest-CAD is a computer-assisted detection (CADe) software device designed to assist physicians in identifying suspicious regions of interest (ROIs) in adult chest X-rays. Suspicious ROIs identified by Chest-CAD are assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. Chest-CAD detects suspicious ROIs by analyzing radiographs using deep learning algorithms for computer vision and provides relevant annotations to assist physicians with their interpretations.

For each image within a study, Chest-CAD generates a DICOM Presentation State file (output overlay). If any suspicious ROI is detected by Chest-CAD in the study, the output overlay for all images includes the text "ROI(s) Detected:" followed by a list of the category/categories for which suspicious ROI(s) were found, such as "Lungs, Bones". In addition, if suspicious ROI(s) are detected in the image, bounding boxes surrounding each detected suspicious ROI are included in the output overlay. If no suspicious ROI is detected by Chest-CAD in the study, the output overlay for each image will include the text "No ROI(s) Detected" and no bounding boxes will be included. Regardless of whether a suspicious ROI is detected, the overlay includes text identifying the X-ray study as analyzed by Chest-CAD and a customer configurable message containing a link to or instructions for users to access labeling. The Chest-CAD overlay can be toggled on or off by the physician within their Picture Archiving and Communication System (PACS) viewer, allowing for concurrent review of the X-ray study.

AI/ML Overview

Acceptance Criteria and Device Performance for Chest-CAD (K210666)

The Chest-CAD device is a computer-assisted detection (CADe) software that analyzes chest radiographs to identify, categorize, and highlight suspicious regions of interest (ROI). It is intended as a concurrent reading aid for physicians and is indicated for adults only.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for Chest-CAD were established through both a standalone performance assessment and a multi-reader, multi-case (MRMC) comparative effectiveness study. No explicit numerical acceptance criteria were stated as targets for the standalone performance metrics (sensitivity, specificity, AUC). Instead, the performance demonstrated in the standalone test formed a basis for comparison and validation. For the MRMC study, the primary objective was to demonstrate superiority of aided reading over unaided reading in terms of AUC.

Table 1: Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Internal/Implicit)Standalone Device PerformanceMRMC Study Performance (Aided vs. Unaided)
Standalone Performance
Overall SensitivityHigh sensitivity is expected0.908 (95% CI: 0.905, 0.911)N/A
Overall SpecificityHigh specificity is expected0.887 (95% CI: 0.885, 0.889)N/A
Overall AUCHigh AUC is expected0.976 (95% CI: 0.975, 0.976)N/A
Clinical Performance (MRMC Study)Aided vs. Unaided
Reader AUCAided AUC > Unaided AUCN/AAided: 0.894 (95% CI: 0.879, 0.909) Unaided: 0.836 (95% CI: 0.816, 0.856)
Reader SensitivityAided Sensitivity > Unaided SensitivityN/AAided: 0.856 (95% CI: 0.850, 0.862) Unaided: 0.757 (95% CI: 0.750, 0.764)
Reader SpecificityAided Specificity > Unaided SpecificityN/AAided: 0.870 (95% CI: 0.866, 0.873) Unaided: 0.843 (95% CI: 0.839, 0.847)

2. Sample Sizes and Data Provenance

Standalone Test Set:

  • Sample Size: 20,000 chest radiograph cases.
  • Data Provenance: From 12 hospitals, outpatient centers, and specialty centers in the United States. The study was retrospective.

MRMC Study Test Set:

  • Sample Size: 238 cases.
  • Data Provenance: From 9 hospitals, outpatient centers, and specialty centers in the United States. The study was retrospective.

3. Number and Qualifications of Experts for Ground Truth (Test Set)

  • Standalone Test Set: Not explicitly stated for the standalone test set. The document indicates "Suspicious ROIs were manually annotated and categorized by board-certified radiologists before the images were used for benchmarking." This implies that the ground truth for standalone testing was established by expert radiologists, but the number and specific qualifications (years of experience) are not detailed.
  • MRMC Study Test Set: "Each case was previously evaluated by a panel of U.S. board-certified radiologists who assigned a ground truth binary label indicating the presence or absence of a suspicious ROI for each Chest-CAD category." The exact number of experts in the panel is not specified, nor is their specific level of experience (e.g., 10 years).

4. Adjudication Method (Test Set)

  • The document implies a consensus-based adjudication method for the ground truth of the MRMC study cases, stating that "a panel of U.S. board-certified radiologists who assigned a ground truth binary label..." This strongly suggests that multiple radiologists reviewed the cases and reached a consensus for the ground truth. It does not explicitly state a 2+1 or 3+1 method but refers to a "panel" establishing the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Yes, a fully-crossed MRMC retrospective reader study was conducted.
  • Effect Size of Human Reader Improvement with AI vs. Without AI Assistance:
    • Reader AUC: Improved from 0.836 (Unaided) to 0.894 (Aided). Effect size = 0.058 increase.
    • Reader Sensitivity: Improved from 0.757 (Unaided) to 0.856 (Aided). Effect size = 0.099 increase.
    • Reader Specificity: Improved from 0.843 (Unaided) to 0.870 (Aided). Effect size = 0.027 increase.
      The study concluded that the accuracy of readers was superior when aided by Chest-CAD.

6. Standalone (Algorithm Only) Performance Study

  • Yes, a standalone performance assessment was conducted.
  • Reported Performance:
    • Overall Sensitivity: 0.908 (95% CI: 0.905, 0.911)
    • Overall Specificity: 0.887 (95% CI: 0.885, 0.889)
    • Overall AUC: 0.976 (95% CI: 0.975, 0.976)
    • AUC by Category: Ranged from 0.921 (Mediastinum/Hila) to 0.994 (Hardware).
    • Sensitivity by Category: Ranged from 0.854 (Bones) to 0.967 (Hardware).
    • Specificity by Category: Ranged from 0.830 (Mediastinum/Hila) to 0.960 (Hardware).

7. Type of Ground Truth Used (Test Set)

  • Expert Consensus: For both the standalone testing and the MRMC study, the ground truth was established by board-certified radiologists who manually annotated and assigned binary labels for suspicious ROIs.

8. Sample Size for the Training Set

  • The document does not explicitly state the sample size for the training set. It refers to "deep learning algorithms for computer vision" but does not detail the dataset used for training or validation of these algorithms.

9. How the Ground Truth for the Training Set was Established

  • The document does not explicitly state how the ground truth for the training set was established. It only mentions that the device uses "machine learning techniques" and "deep learning algorithms." Typically, for such devices, the training set ground truth would also be established by expert radiologists, usually through a process similar to or more extensive than that used for the test sets. However, this is not detailed in the provided text.

{0}------------------------------------------------

July 20, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle emblem, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" in a larger font size and a blue square behind it.

Imagen Technologies, Inc % Robert Lindsey, Ph.D. Chief Science Officer 151 West 26th Street, 10th Floor NEW YORK NY 10001

Re: K210666

Trade/Device Name: Chest-CAD Regulation Number: 21 CFR 892.2070 Regulation Name: Medical image analyzer Regulatory Class: Class II Product Code: MYN Dated: June 9, 2021 Received: June 10, 2021

Dear Dr. Lindsey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210666

Device Name Chest-CAD

Indications for Use (Describe)

Chest-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies using machine learning techniques to identify, categorize, and highlight suspicious regions of interest (ROI). Any suspicious ROI identified by Chest-CAD is assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. The device is intended for use as a concurrent reading aid for physicians. Chest-CAD is indicated for adults only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) (and 21 CFR 807.92) the 510(k) Summary for Chest-CAD is provided below.

SUBMITTER 1.

Applicant:Imagen Technologies, Inc.151 West 26th Street, Suite 1001New York, NY 10001
Contact and PrimaryCorrespondent:Robert Lindsey, Ph.D.Chief Science OfficerImagen Technologies, Inc.151 West 26th Street, Suite 1001New York, NY 10001917-830-4721rob@imagen.ai
Secondary Correspondent:Becky DittyConsultantBiologics Consulting1555 King St., Suite 300Alexandria, VA 22314269-888-2516bditty@biologicsconsulting.com
Date Prepared:July 12th, 2021

2. DEVICE

Device Trade Name:Chest-CAD
Device Common Name orClassification Name:Medical Image Analyzer
Regulation21 CFR 892.2070
Regulatory Class:II
Product Code:MYN

{4}------------------------------------------------

PREDICATE DEVICE 3.

On January 22, 2020, FDA published the final rule down-classifying medical image analyzers (product code MYN) from Class III to Class II. Therefore, Riverain Technologies' RapidScreen™ RS-2000 (P000041) has been identified as the predicate device for Chest-CAD.

DEVICE DESCRIPTION 4.

Chest-CAD is a computer-assisted detection (CADe) software device designed to assist physicians in identifying suspicious regions of interest (ROIs) in adult chest X-rays. Suspicious ROIs identified by Chest-CAD are assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. Chest-CAD detects suspicious ROIs by analyzing radiographs using deep learning algorithms for computer vision and provides relevant annotations to assist physicians with their interpretations.

For each image within a study, Chest-CAD generates a DICOM Presentation State file (output overlay). If any suspicious ROI is detected by Chest-CAD in the study, the output overlay for all images includes the text "ROI(s) Detected:" followed by a list of the category/categories for which suspicious ROI(s) were found, such as "Lungs, Bones". In addition, if suspicious ROI(s) are detected in the image, bounding boxes surrounding each detected suspicious ROI are included in the output overlay. If no suspicious ROI is detected by Chest-CAD in the study, the output overlay for each image will include the text "No ROI(s) Detected" and no bounding boxes will be included. Regardless of whether a suspicious ROI is detected, the overlay includes text identifying the X-ray study as analyzed by Chest-CAD and a customer configurable message containing a link to or instructions for users to access labeling. The Chest-CAD overlay can be toggled on or off by the physician within their Picture Archiving and Communication System (PACS) viewer, allowing for concurrent review of the X-ray study.

ട്. INTENDED USE/INDICATIONS FOR USE

Chest-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies using machine learning techniques to identify, categorize, and highlight suspicious regions of interest (ROI). Any suspicious ROI identified by Chest-CAD is assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. The device is intended for use as a concurrent reading aid for physicians. Chest-CAD is indicated for adults only.

SUBSTANTIAL EQUIVALENCE 6.

Comparison of Indications

The predicate device for Chest-CAD (Riverain Technologies' RapidScreen™ RS-2000) has the following FDA-approved Indications for Use:

The RapidScreen™ RS-2000 is a computer-aided detection (CAD) system intended to identify and mark regions of interest (ROIs) on digitized frontal chest radiographs. It identifies

{5}------------------------------------------------

features associated with solitary pulmonary nodules from 9 to 30 mm in size, which could represent early-stage lung cancer. The device is intended for use as an aid only after the physician has performed an initial interpretation of the radiograph.

RapidScreen™ RS-2000 and Chest-CAD both analyze chest radiographs, and both identify regions of interest (ROI) in the chest. Chest-CAD detects ROIs and assigns each ROI to one of eight categories compared to RapidScreen™ RS-2000 that detects ROIs and assigns ROIs to a single category (i.e., features associated with pulmonary nodules). RapidScreen™ RS-2000 is indicated for use as a second read, while Chest-CAD is indicated for use as a concurrent read. However, both devices are only intended as an aid to the physician and not intended to replace the diagnosis by the physician. The differences in Indications for Use do not constitute a new intended use, as both devices are intended to assist physicians by identifying and marking ROIs in chest radiographs.

Technological Comparisons

Table 1 provides a comparison of the Technological Characteristics of Chest-CAD to the predicate RapidScreen™ RS-2000.

Proposed DevicePredicate
NumberK210666P000041
ApplicantImagen TechnologiesRiverain Medical Group
Device NameChest-CADRapidScreen™ RS-2000
Classification Regulation892.2070892.2070
Product CodeMYNMYN
Image ModalityX-rayX-ray
Study TypeChestChest
Clinical OutputIdentify and mark regions ofinterest (ROIs) on chestradiographsIdentify and mark regions ofinterest (ROIs) on chestradiographs
Clinical FindingIdentified ROIs are assigned toone of the following categories:Cardiac, Mediastinum/Hila,Lungs, Pleura, Bones, SoftTissues, Hardware, or OtherIdentified ROIs are assigned to asingle category (i.e., featuresassociated with solitarypulmonary nodules from 9 to 30mm in size)
Intended UsersPhysicianPhysician
Intended User WorkflowDevice intended for use as areading aid for physiciansinterpreting chest radiographsDevice intended for use as areading aid for physiciansinterpreting chest radiographs
Patient PopulationAdults with Chest RadiographsAdults with Chest Radiographs
Algorithm MethodologyArtificial Neural NetworksArtificial Neural Networks
Table 1:Technological Comparison
------------------------------------

{6}------------------------------------------------

Proposed DevicePredicate
PlatformSecure cloud-based processingand delivery of chest radiographsSecure on-premise processingand delivery of chest radiographs
Image SourceDigital X-rayFilm X-ray
Image ViewingImage displayed on PACS systemImage displayed on videomonitor

Chest-CAD's intended end-users, imaging modality, output display on X-ray studies, and assistive functionality during chest radiograph interpretation workflows are similar to those of RapidScreen™ RS-2000. Chest-CAD differs from RapidScreen™ RS-2000 in that Chest-CAD detects ROIs and assigns each ROI to one of eight categories compared to RapidScreen™ RS-2000 that detects ROIs and assigns ROIs to a single category (i.e., features associated with pulmonary nodules). Chest-CAD operates on digital X-rays from a DICOM node, whereas RapidScreen™ RS-2000 operates on digitized X-ray films. RapidScreen™ RS-2000 was approved when digital X-rays were not standard of care, however, the Riverain device was approved by FDA to process digital X-rays in P000041/S001. The fundamental purpose of both devices is to identify ROIs on chest X-rays for further consideration by the physician, and these differences in technological characteristics do not raise different concerns of safety and effectiveness.

7. PERFORMANCE DATA

Biocompatibility Testing

There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.

Electrical Safety and Electromagnetic Compatibility (EMC)

The subject device is a software-only device. Therefore, electrical safety and EMC testing was not necessary to establish the substantial equivalence of this device.

Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software level of concern for Chest-CAD is Moderate, since a malfunction of, or a latent design flaw in, the software device may lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.

Bench Testing

Imagen conducted a standalone performance assessment on 20,000 chest radiograph cases from 12 hospitals, outpatient centers, and specialty centers in the United States representative of the

{7}------------------------------------------------

intended use population. The results of the standalone testing demonstrated that Chest-CAD detects suspicious ROIs with high sensitivity (0.908; 95% Wilson's Confidence Interval: 0.905, 0.911), high specificity (0.887; 95% Wilson's Confidence Interval: 0.885, 0.889), and high Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve (0.976, 95% Bootstrap Confidence Interval: 0.975, 0.976).

The AUC of the ROC curve was also estimated for each Chest-CAD category and Figure 1 shows AUCs remained high across the eight categories (further detail described in Table 2). The highest AUCs of the ROC curve were for Hardware (0.994) and the lowest were for Mediastinum/Hila (0.921). Sensitivity and specificity were calculated for each of the Chest-CAD categories. As shown in Table 3, sensitivity was highest for Hardware (0.967) and was lowest for Bones (0.854). Specificity was highest for Hardware (0.960) and lowest for Mediastinum/Hila (0.830). The Free-Response ROC (FROC) curve was also estimated for each Chest-CAD category and Figure 2 shows the box-level sensitivity versus the false positives per image. The FROC curves terminate at the device's box-level sensitivity for each category due to the cascaded nature of the Chest-CAD predictions.

Image /page/7/Figure/4 description: The image is a plot of sensitivity vs 1 - specificity for different body parts. The plot shows the ROC curves for hardware, soft tissues, pleura, lungs, cardiac, other, bones, and mediastinum/hila. The area under the curve (AUC) is shown in parentheses for each body part, with hardware having the highest AUC of 0.994 and mediastinum/hila having the lowest AUC of 0.921.

Chest-CAD ROC Curve by Category Figure 1:

{8}------------------------------------------------

CategoryAUC95% Bootstrap CI
Cardiac0.9610.959, 0.963
Mediastinum/Hila0.9210.918, 0.924
Lungs0.9670.966, 0.969
Pleura0.9730.972, 0.975
Bones0.9300.926, 0.934
Soft Tissues0.9810.977, 0.985
Hardware0.9940.994, 0.995
Other0.9530.950, 0.957

Table 2: AUC of the ROC Curve for Chest-CAD Model Predictions by Category

Abbreviations: AUC= Area Under the Curve; CI= Confidence Interval.

Table 3: Sensitivity and Specificity for Chest-CAD Model Predictions by Category

CategorySensitivitySpecificity
95%Wilson's CI95%Wilson's CI
Cardiac0.889(0.881, 0.897)0.892(0.887, 0.897)
Mediastinum/Hila0.856(0.844, 0.867)0.830(0.824, 0.835)
Lungs0.888(0.882, 0.893)0.915(0.908, 0.921)
Pleura0.919(0.912, 0.925)0.899(0.894, 0.904)
Bones0.854(0.838, 0.868)0.856(0.850, 0.861)
Soft Tissues0.938(0.916, 0.955)0.919(0.916, 0.923)
Hardware0.967(0.963, 0.970)0.960(0.956, 0.964)
Other0.906(0.889, 0.920)0.872(0.867, 0.877)

Abbreviations: CI= Confidence Interval.

{9}------------------------------------------------

Image /page/9/Figure/2 description: This image is a plot of box-level sensitivity versus false positives per image. There are eight different lines plotted on the graph, each representing a different anatomical structure: Hardware, Lungs, Cardiac, Mediastinum/Hila, Soft Tissues, Pleura, Other, and Bones. The x-axis represents the number of false positives per image, ranging from 0.0 to 0.8. The y-axis represents the box-level sensitivity, ranging from 0.0 to 0.8.

Chest-CAD Free-Response ROC (FROC) Curve by Category Figure 2:

Animal Testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

Clinical Data

Imagen conducted a fully-crossed multiple reader, multiple case (MRMC) retrospective reader study to determine the impact of Chest-CAD on reader performance in detecting suspicious ROIs in chest radiograph cases. The primary objective of this study was to determine whether the accuracy of readers aided by Chest-CAD ("Aided") was superior to the accuracy of readers when unaided by Chest-CAD ("Unaided") as determined by the case-level, across-category aggregate Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve.

24 clinical readers each evaluated 238 cases in Chest-CAD's Indications for Use under both Aided and Unaided conditions. The cases were from 9 hospitals, outpatient centers, and specialty centers in the United States. Each case was previously evaluated by a panel of U.S. boardcertified radiologists who assigned a ground truth binary label indicating the presence or absence of a suspicious ROI for each Chest-CAD category. The MRMC study consisted of two independent reading sessions separated by a washout period of at least 28 days in order to avoid memory bias. For each case, each reader was required to provide a binary determination of the presence or absence of a suspicious ROI for each category and to provide a confidence score representing their certainty.

{10}------------------------------------------------

The results of the study found that the accuracy of readers in the intended use population was superior when Aided by Chest-CAD than when Unaided by Chest-CAD, as measured by the task of suspicious ROI detection using the AUC of the ROC curve as calculated by the Dorfman, Berbaum, and Metz (DBM) modeling approach.

Image /page/10/Figure/3 description: The image is a plot of sensitivity vs 1-specificity. There are two curves plotted on the graph, one for aided and one for unaided. The aided curve is in blue and the unaided curve is in red. The aided curve is generally higher than the unaided curve, indicating that the aided modality has better performance.

Figure 3: Clinical Reader Study Results - Aided and Unaided ROC Curves

In particular, the study results demonstrated improvements when Aided versus Unaided:

  • When calculated using Wilcoxon rank-sum scores with bootstrap confidence intervals as outlined in Beiden et al. 20001, reader AUC estimates improved from 0.836 (95% Bootstrap CI: 0.816, 0.856) to 0.894 (95% Bootstrap CI: 0.879, 0.909).
  • Reader sensitivity improved from 0.757 (95% Wilson's CI: 0.750, 0.764) to 0.856 (95% Wilson's CI: 0.850, 0.862).
  • Reader specificity improved from 0.843 (95% Wilson's CI: 0.839, 0.847) to 0.870 (95% ● Wilson's CI: 0.866, 0.873).

'Beiden, S.V., Wagner, R.F., & Campbell, G. (2000). Components-of-variance models and multiple-bootstrap experiments: An alternative method for random-effects, receiver operating characteristic analysis. Academic Radiology, 7, p.341-p.349.

{11}------------------------------------------------

CONCLUSION 8.

The conclusions drawn from the standalone and clinical studies demonstrate that Chest-CAD is as safe, as effective, and performs as well as RapidScreen™ RS-2000. The special controls for the Medical Image Analyzer (CADe) 21 CFR 892.2070 regulation are satisfied by demonstrating effectiveness of the device in both the standalone testing and the clinical testing, showing superiority of Aided versus Unaided reads in the clinical testing, and communicating testing results in the labeling. Chest-CAD's intended end-users, imaging modality, output display on Xray studies, and assistive functionality during chest radiograph interpretation workflows are similar to those of RapidScreen™ RS-2000. The technological differences identified and discussed in Section 6 do not raise different concerns of safety and effectiveness. Thus, Chest-CAD is substantially equivalent to RapidScreen™ RS-2000 for the intended use of computerassisted detection.

N/A