K Number
K210645

Validate with FDA (Live)

Device Name
RayStation 10.1
Date Cleared
2021-06-29

(118 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

The system functionality can be configured based on user needs.

Device Description

RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines.

RayStation consists of multiple applications:

  • The main RayStation application is used for treatment planning.
  • . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
  • . The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.
  • o The RayCommand application is used for treatment session management including treatment preparation and sending instructions to the treatment delivery devices.
AI/ML Overview

The provided text details the 510(k) summary for RayStation 10.1, a software system for radiation therapy and medical oncology. The document indicates that the determination of substantial equivalence to the primary predicate device (RayStation 9.1) is not based on an assessment of non-clinical performance data. Instead, it relies on the entire system verification and validation specifications and reports.

However, the document does describe the performance data for several new features and explicitly states that these features have been "successfully validated for accuracy in clinically relevant settings according to specification" or "successfully validated according to specification." While these statements imply acceptance criteria were met, the specific numerical acceptance criteria and the reported device performance values are not explicitly provided in a comparative table format within the given text.

Therefore, the following response will extract the implied acceptance criteria and reported performance from the descriptions provided, and note where specific numerical values are absent.


Acceptance Criteria and Device Performance Study for RayStation 10.1

The provided document, K210645 for RayStation 10.1, indicates that the determination of substantial equivalence to the primary predicate device (RayStation 9.1) is not based on non-clinical performance data directly comparing the existing features of 10.1 to 9.1. Instead, it relies on the comprehensive system verification and validation reports.

However, for the new features introduced in RayStation 10.1, the document states that these features have undergone validation and met their respective specifications. While specific, quantifiable acceptance criteria and reported performance values are not presented in a direct comparative table within the provided text, the descriptions imply the following:

1. A table of acceptance criteria and the reported device performance

FeatureImplied Acceptance Criteria (from text)Reported Device Performance (from text)
Brachytherapy TG43 Dose CalculationAccurately models output from single and combined brachytherapy sources in clinical plans. All doses reported as dose-to-water (DWw).Successfully validated for accuracy in clinically relevant settings according to specification.
Medical Oncology Dose Calculation FunctionsAppropriate for supporting medical oncology planning workflows when used by qualified users according to IFU.Validated to be appropriate for supporting medical oncology planning workflows.
Proton Ocular Treatment Dose CalculationAccurately models proton dose calculation for ocular treatments using the single scattering (SS) delivery technique (modeled as double scattering).Successfully validated for accuracy in clinically relevant settings according to specification.
Robust Planning of Organ MotionCorrectly generates deformed image sets to simulate organ motion and uses them for robust planning against intra-fractional or inter-fractional organ motion.Successfully validated according to specification.

Note: The provided text does not contain specific numerical acceptance criteria (e.g., "accuracy within X%") or quantitative reported performance data for any of these features. The reported performance essentially states that the criteria were "met" or "validated."

2. Sample size used for the test set and the data provenance

The document indicates that RayStation 10.1's test specification is a further developed version of RayStation 9.1's, supported by requirements specification. The verification activities included "User validation in cooperation with cancer clinics." However, no specific sample sizes for test sets (e.g., number of patient cases) or data provenance (e.g., country of origin, retrospective/prospective nature) are provided in the given text for any of the validations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The text mentions "User validation in cooperation with cancer clinics" but does not specify the number of experts, their qualifications, or how ground truth was established for the "test set" (if a distinct clinical test set was used for ground truth establishment).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

No information on adjudication methods is provided in the supplied text.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The text does not mention any multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss human readers improving with or without AI assistance. The device is a treatment planning system, and while it states it "proposes treatment plans" based on user input, it does not describe AI-assisted diagnostic or interpretation tasks. It explicitly states, "Related to machine learning, there is no change compared to the primary predicate device." suggesting limited or no direct machine learning components in the new features where such a study would typically be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The validations described for brachytherapy, medical oncology, proton ocular treatment, and robust planning of organ motion appear to be standalone algorithm performance assessments against defined specifications. These validations verify the accuracy and appropriateness of the software's calculations and functionalities independently, assuming "intended qualified user" interaction for medical oncology, but not as part of a human-in-the-loop performance study.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the dose calculation features (brachytherapy, proton ocular treatment), the "ground truth" implicitly refers to theoretical models and established physical principles (e.g., accurate modeling of TG43 formalism, proton dose calculations) as compared against the output of the software. For medical oncology functions, the "ground truth" for validation appears to be whether the functions are "appropriate" for planning workflows, likely assessed against clinical guidelines or expert workflows. For robust planning of organ motion, the ground truth relates to the correct generation and application of deformed image sets according to specifications. The document does not explicitly state that ground truth was established through pathology or outcomes data.

8. The sample size for the training set

The document refers to the system as "built on the same software platform" and "developed under the same quality system, by the same development teams." It mentions that "related to machine learning, there is no change compared to the primary predicate device." Given this, and the nature of treatment planning software, the concept of a "training set" in the context of machine learning (e.g., for image classification or prediction models) is not directly applicable or discussed for the validations mentioned. The system's development would involve software engineering and clinical validation rather than machine learning training sets for the described functionalities.

9. How the ground truth for the training set was established

As there is no mention of a training set or machine learning components for the new features, information on how its ground truth was established is not provided.

{0}------------------------------------------------

June 29, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

RaySearch Laboratories AB (publ) % Mr. David Hedfors Quality and Regulatory Affairs Director Sveavägen 44 Stockholm, 111 34 SWEDEN

Re: K210645

Trade/Device Name: RayStation 10.1 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: May 25, 2021 Received: May 28, 2021

Dear Mr. Hedfors:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210645

Device Name RayStation 10.1

Indications for Use (Describe)

RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

The system functionality can be configured based on user needs.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Document ID and TitleVersion:
RSL-D-RS-10.1 510(k) Summary RayStation 10.11.0

510(k) Summary 1.

1.1 510(k) owner

RaySearch Laboratories AB (publ) Sveavägen 44 111 34 Stockholm Sweden

Tel: +46 8 510 530 00

1.2 Contact person

David Hedfors Quality and Regulatory Affairs Director RaySearch Laboratories AB (publ) Email: quality(@raysearchlabs.com +46 722 366 110 Tel:

1.3 Preparation date

February 26th, 2021

1.4 Trade name

The trade name is RayStation.

The trade name and version number are written together, i.e. "RayStation 10.1" to easily distinguish the submitted device from the primary predicate device RayStation 9.1.

The marketing name is RayStation 10B.

1.5 Common name

Radiation therapy treatment planning system

1.6 Classification name

Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUI)

1.7 Predicate devices

  • 1.7.1 Primary predicate device
    K200569 RayStation 9.1

1.7.2 Secondary predicate devices

K203172 MOSAIQ Oncology Information System K181145 Eclipse Treatment Planning System

1.8 Device description

RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines.

RayStation consists of multiple applications:

  • The main RayStation application is used for treatment planning.
  • . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
  • . The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.
  • o The RayCommand application is used for treatment session management including treatment preparation and sending instructions to the treatment delivery devices.

{4}------------------------------------------------

Document ID and TitleVersion:
RSL-D-RS-10.1 510(k) Summary RayStation 10.11.0

The device to be marketed, RayStation version 10.1, marketing name "RayStation 10B", contains modified features compared to version 9.1. The main changes are:

  • . the addition of medical oncology and corresponding update of the intended use. This feature is included in the intended use of the MOSAIQ OIS secondary predicate device.
  • . the addition of brachy planning. This feature is included in the intended use of the Eclipse TPS secondary predicate device.
    • RayStation 10.1 is compatible with the Bebig SagiNova system afterloader. ●
  • the addition of eye treatment planning (ocular planning) with protons. This feature is included in the ● intended use of the Eclipse TPS secondary predicate device.
  • the extension of the robust planning to take organ motion into account and
  • the extension of the treatment session management, from the RayTreat module in RayStation 9.1 to a ● new module called RayCommand. Like RayTreat, RayCommand is not intended for real-time interaction with the delivery device and does not contain any emergency shutdown functionality.

1.9 Intended use

RavStation is a software system for radiation therapy and medical oncology. Based on user inout. RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

The system functionality can be configured based on user needs.

1.10 Technological characteristics summary

The technological characteristics are the same for RayStation 10.1 as for the primary predicate device RavStation 9.1. Both versions are built on the same software platform and share design to a high degree. Both versions have been developed under the same quality system, by the same development teams, meeting the same requirements for safety and effectiveness.

Comparing RayStation 10.1 with RayStation 9.1, the newer version includes usability improvements as well as new features.

The most significant new feature is the addition of medical oncology and the corresponding update of the intended use. For this feature the K203172 MOSAIQ OIS is used as a secondary predicate device. Its 510(k) Summary describes this feature as "Generate and keep medication formulary lists and calculate applicable medication dosages for medical oncology", and this is the same as in RayStation 10.1.

Other main changes are the additions of brachytherapy planning, aka ocular planning. For these features the K181145 Eclipse Treatment Planning System is used as a secondary predicate device. Its 510(k) Summary describes this feature as "Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irrachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye." This is the same features as in RayStation 10.1.

Other significant changes are:

  • the extension of the robust planning to take organ motion into account and ●
  • the extension of the treatment session management, from the RayTreat module in RayStation 9.1 to a ● new module called RayCommand. Like RayTreat, RayCommand is not intended for real-time interaction with the delivery device and does not contain any emergency shutdown functionality.

Related to machine learning, there is no change compared to the primary predicate device.

1.11 Assessment of non-clinical performance data

The test specification of RayStation 10.1 is a further developed version of the test specification of RayStation 9.1. This is supported by the requirements specification, for which the same is true. The successful verification and validation of RayStation 10.1 therefore support the substantial equivalence of the above RayStation versions.

{5}------------------------------------------------

Document ID and TitleVersion:
RSL-D-RS-10.1 510(k) Summary RayStation 10.11.0

1.11.1 Brachytherapy TG43 dose calculation performance data

RayStation 10.1 accurately models the output from single sources and from sources combined into brachytherapy treatment setups in clinical plans. Since the TG43 dose computation formalism treats the full patient geometry as water, no dedicated test cases for validation of heterogenous geometries are included in the specification. This also means that all doses are reported as dose-to-water with radiation transport in water, DWw.

The brachy TG43 dose computation algorithm in RayStation10B has been successfully validated for accuracy in clinically relevant settings according to specification.

1.11.2 Medical oncology dose calculation performance data

Medical oncology planning in RayStation supports dose calculation based on patient weight, patient body surface area, and flat-rate dose specification. These units of calculation are used in the majority of medical oncology regimens and are the available options in other systems.

Numerous publications describe features of computerized treatment planning for medical oncology that reduce medication errors and improve workflows. The RayStation medical oncology functions have been validated to be appropriate, when used by an intended qualified user according to the Instructions for Use, for supporting medical oncology planning workflows.

1.11.3 Proton ocular treatment dose calculation performance data

The validation covers ocular treatments using the single scattering (SS) delivery technique. Since ocular treatments require small field sizes, achieved by using a field-specific block aperture, the generated Spread Out Bragg Peaks (SOBPs) are approximately laterally uniform. Hence, the SS delivery technique is modelled in RayStation as the double scattering (DS) treatment technique. A range compensator is not used.

The proton dose computation for proton ocular treatments in RayStation 10.1 has been successfully validated for accuracy in clinically relevant settings according to specification.

1.11.4 Robust planning of organ motion performance data

The robustness of a plan can be evaluated by taking uncertainties such as CT inaccuracy and setup errors into account. Robustness handling in RayStation 10.1 has been expanded to also take the uncertainty of the patient's internal organ motion into account. Based on user defined criteria the system computes a set of scenario doses for evaluation.

Deformed image sets can be generated to simulate organ motion without taking multiple CT/CBCT/MR scans of the patient. Structures that are not considered to move can also be used as input to the deformation by being specified as fixed. When the new anatomy has been defined, all other structures will be mapped deformably to the generated deformed image sets.

The generated motion image sets may be used both for evaluation and as input when planning robustly against intra-fractional or inter-fractional organ motion.

The robust planning of organ motion in RayStation 10.1 has been successfully validated according to specification.

1.12 Test conclusion

The determination of substantial equivalence compared to the primary predicate device is not based on an assessment of non-clinical performance data. However, the entire system verification and validation specifications and reports are included in the submission as required by a software device of major concern.

A number of different types of verification activities have been performed:

  • . System Tests of RayStation
  • . Risk analysis-based tests for use error mitigation verification
  • Unit and subsystem testing for low-level testing
  • Dose engine validation including internal testing ●

{6}------------------------------------------------

  • User validation in cooperation with cancer clinics ●
  • Reviews of design, code and Master Labeling .

The data obtained from the verification show that system tests, unit and subsystem tests have passed, and the validations been completed successfully. The reviews of design, code and labeling are also passed.

From the successful verification and validation activities, including the brachytherapy TG43, the medical oncology and the proton ocular treatment dose calculations and the robust planning of organ motion performance data, the conclusion can be drawn that RayStation 10.1 have met specifications and is as safe, as effective and performs as well as or better than the legally marketed primary predicate device.

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.