K Number
K210567

Validate with FDA (Live)

Date Cleared
2022-09-30

(581 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The K500 Ultrasonic Surgical System is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The scalpel can be used as an adjunct or substitute for electrosurgery, laser and steel scalpels.

Device Description

Ultrasonic Surgical System consists of a generator, a foot switch, a scalpel and a hand piece. In addition, a cart is provided as an optional component. The electrical energy supplied by the generator will be converted into mechanical motion by the transducer in hand piece, which will drive the scalpel vibrate longitudinally. The foot switch is used to control the activation of energy output. The scalpels are used to cut and coagulate soft bodily tissues and structures in many surgery procedures.

AI/ML Overview

The provided text describes a 510(k) submission for the K500 Ultrasonic Surgical System, asserting its substantial equivalence to predicate devices. It does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML-driven device.

The study described is a non-clinical performance comparison test to ensure that the proposed device, an ultrasonic surgical system, meets its design specifications and is substantially equivalent to a predicate device. This is a traditional medical device submission, not specifically an AI/ML device submission that would involve human-in-the-loop performance or standalone algorithm performance testing.

Therefore, many of the requested categories related to AI/ML device performance and ground truth establishment cannot be fulfilled from the provided document.

Here's a breakdown of the available information based on the provided text, acknowledging that it's for a traditional medical device and not an AI/ML device performance study:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in a quantitative table for device performance in the way one would for an AI/ML model (e.g., sensitivity, specificity thresholds). Instead, it describes performance comparison tests to demonstrate substantial equivalence to a predicate device.

Table 1: Non-Clinical Performance Testing and Outcomes

Test CategorySpecific TestPurposeOutcome/Finding
Functional PerformanceVessel burst pressureTo ensure effective sealing of blood vessels.Not explicitly detailed, but implied to meet design specifications and be comparable to the predicate.
Thermal spreadTo assess minimal thermal injury.Not explicitly detailed, but implied to meet design specifications and be comparable to the predicate.
Acute animal vessel sealing studyTo evaluate immediate vessel sealing capabilities.No significant difference found compared to the predicate device despite slight tip diameter difference.
Chronic animal survival studyTo assess long-term effects and safety in a living organism.Not explicitly detailed, but implied to meet design specifications and be comparable to the predicate.
BiocompatibilitySkin sensitization (ISO 10993-10)To evaluate potential for allergic reactions.Pass
Intracutaneous reactivity (ISO 10993-10)To evaluate local tissue compatibility.Pass
Cytotoxicity (ISO 10993-5)To assess potential toxicity to cells.Pass
Pyrogenicity (ISO 10993-11)To detect fever-inducing substances.Pass
Systemic toxicity (ISO 10993-11)To evaluate potential for adverse effects on organ systems.Pass
Electrical Safety & EMCAAMI/ANSI ES 60601-1, IEC 60601-2-18, IEC 60601-1-2To ensure electrical safety and electromagnetic compatibility.Complies with relevant standards (AAMI/ANSI/ES 60601-1, IEC 60601-2-18, IEC 60601-1-2).
Sterility AssuranceGB18280:2007 (IDT ISO 11137: 2006)To ensure adequate sterilization (for sterile components).10-6 SAL (Sterility Assurance Level)
Seal StrengthASTM F88/F88M-15Standard method for seal strength of flexible barrier materials (likely for packaging).Complies with the standard.
Software Verification/ValidationSoftware V&V testingTo ensure software functions as intended and is safe.Conducted and documentation provided in accordance with FDA guidance for software in medical devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified for any of the non-clinical tests (e.g., number of vessels, number of animals).
  • Data Provenance: Not specified. "Animal study" mentioned without country or specific details. The manufacturer is Miconvey Technologies Co., Ltd. from Chongqing, China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the study described is a non-clinical performance and biocompatibility study for a surgical device, not a diagnostic or AI/ML study requiring expert ground truth for interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for this type of non-clinical device performance study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device or a diagnostic device where human readers would be involved in interpreting data assisted by AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device. The "device" is an ultrasonic surgical system, a physical tool.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical performance tests:

  • Performance Metrics: Measured directly (e.g., burst pressure, thermal spread, physical dimensions) against established engineering specifications and comparison to the predicate device's known performance.
  • Biocompatibility: Assessed against ISO standards (e.g., presence/absence of irritation, cytotoxicity).
  • Animal Studies: Observed physiological effects and outcomes (e.g., vessel sealing success, survival).

8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

September 30, 2022

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

Miconvey Technologies Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China

Re: K210567

Trade/Device Name: Ultrasonic Surgical System Regulatory Class: Unclassified Product Code: LFL Dated: August 29, 2022 Received: August 29, 2022

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin K. Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Druq Administration

Indications for Use

510(k) Number (if known) K210567

Device Name K500 Ultrasonic Surgical System

Indications for Use (Describe)

The K500 Ultrasonic Surgical System is indicated for soft tissue incisions when bleeding control and minimal injury are desired. The scalpel can be used as an adjunct or substitute for electrosurgery, laser and steel scalpels.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K210567 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

    1. Date of Preparation: 09/26/2022
    1. Sponsor Identification

Miconvey Technologies Co., Ltd.

No.16 Fangzheng Avenue, Beibei District, Chongqing, 400714, China

Establishment Registration Number: Not yet registered

Contact Person: Kang Li Position: General Manager Tel: +86-23-63171596 Fax: +86-23-63171516 Email: kang.li@miconvey.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: K500 Ultrasonic Surgical System Common Name: Ultrasonic Surgical Generator and Accessory Model: K500

Regulatory Information: Classification Name: Instrument, Ultrasonic Surgical Classification: Unclassified; Pre-amendment Product Code: LFL; Regulation Number: N.A. Review Panel: General& Plastic Surgery;

Indication for Use Statement:

The K500 Ultrasonic Surgical System is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The scalpel can be used as an adjunct or substitute for electrosurgery, laser and steel scalpels.

Device Description

Ultrasonic Surgical System consists of a generator, a foot switch, a scalpel and a hand piece. In addition, a cart is provided as an optional component. The electrical energy supplied by the generator will be converted into mechanical motion by the transducer in hand piece, which will drive the scalpel vibrate longitudinally. The foot switch is used to control the activation of energy output. The scalpels are used to cut and coagulate soft bodily tissues and structures in many surgery procedures.

5. Identification of Predicate Devices

Primary Predicate Device

510(k) Number: K002906 Product Name: Ultracision Harmonic Scalpel System Model Name: GEN04 Manufacturer: Ethicon Endo-Surgery, LLC

Secondary Predicate Device

510(k) Number: K042777 Product Name: Harmonic ACETM Curved Shears with Hand Control

{5}------------------------------------------------

Manufacturer: Ethicon Endo-Surgery, LLC

6. Summary of Technological Characteristics

ItemProposed DevicePrimary Predicate DeviceRemark
Product CodeLFLLFLSame
ClassificationUnclassifiedUnclassifiedSame
Regulation NumberN.A.N.A.Same
Intended UseThe K500 Ultrasonic Surgical System is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The scalpel can be used as an adjunct or substitute for electrosurgery, laser and steel scalpels.The Ultracision Harmonic Scalpel System is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The Ultracision Harmonic Scalpel System instruments can be used as an adjunct or substitute for electrosurgery, laser and steel scalpels.Same
ConfigurationGeneratorGeneratorSame
FootswitchFoot SwitchSame
Hand pieceHand pieceSame
ScalpelN.A.Different
SterileGenerator- NonsterileGenerator- NonsterileSame
Handpiece- Nonsterile. Sterilization required prior to use.Handpiece- Nonsterile. Sterilization required prior to use.Same
Scalpel-Provided sterileN.A.Different
BiocompatibilityNo Patient Contact MaterialNo Patient Contact MaterialSame
Electrical SafetyComply with AAMI/ANSI/ ES60601-1Comply with IEC 60601-1Same
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Same
Table 2 Comparison of Technology Characteristics- Ultrasonic Surgical System-Scalpel
ItemProposed DeviceSecondary Predicate DeviceRemark
Product CodeLFLK042777LFLSame
ClassificationUnclassifiedUnclassifiedSame
Regulation NumberN.A. Pre-AmendmentN.A. Pre-AmendmentSame
Intended UseThe K500 Ultrasonic Surgical System is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The scalpel can be used as an adjunct or substitute for electrosurgery, laser and steel scalpels.The Harmonic ACETM Curved Shears with Hand Control is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic and other open and endoscopic procedures.Same
Shaft length14cm, 23cm and 36cm23cm and 36cmDifferent
Pin hole to proximal end18.85mm18.85mmSame
Pin hole diameter1.8mm1.8mmSame
Tip diameter2.14mm2.16mmSimilar
Jaw diameter5.45mm5.45mmSame
Rotary head length2.9mm2.9mmSame
Handle length36.9mm37mmSimilar
Handle width30.05mm30mmSimilar
Closing trigger widthUpper width: 3.2mmLower width: 7mmUpper width: 3.2mmLower width: 7mmSame
Trigger button length2.2mm2.2mmSame
Torque wrench diameter8.15mm8.15mmSame
Shaft diameter3.42mm3.42mmSame
Blade shapeCurvedCurvedSame
Energy TypeUltrasonic and MechanicalUltrasonic and MechanicalSame
Sterility10-6 SAL10-6 SALSame
Single UseSingle UseSingle UseSame
BiocompatibilityPassPassSame
Electrical SafetyComply with AAMI/ANSI/ES60601-1Comply with IEC 60601-1Same
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Same

Table 1 Comparison of Technology Characteristics- Ultrasonic Surgical System

{6}------------------------------------------------

Table 2 Comparison of Technology Characteristics- Ultrasonic Surgical System-Scalpel

{7}------------------------------------------------

Different-Shaft Length

The proposed scalpel is available in three shaft length 14cm, 23cm and two shaft length models are covered in predicate device. The different shaft length does not affect the indication of soft tissue incisions.

Similar-Tip Diameter

The tip diameter for the proposed device is different from predicate device. In addition, this difference is very slight and animal study has been conducted on the subject device and predicate device. The animal study result showed there were no significant difference.

Similar-Handle Length and Handle Width

The handle length and width for the proposed device is different from predicate device. In addition, this difference is very slight and does not affect intended use.

    1. Software
      Software verification and validation testing were conducted, and documentation was provided in accordance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
    1. Summary of Non-clinical Performance Testing
      Nonclinical performance comparison tests were conducted to verify that the proposed device met all design specifications. The tests performed on the subject device included testing and evaluation for vessel burst pressure, thermal spread, acute animal vessel sealing study, and chronic animal survival study.

The biocompatibility tests were performed on the proposed device. Based on the contact level and contact duration, the proposed device was tested for skin sensitization (ISO 10993-10), intracutaneous reactivity (ISO 10993-10), cytotoxicity (ISO 10993-5), pyrogenicity (ISO 10993-11) and system toxicity (ISO 10993-11).

Electrical safety and EMC test was conducted on the proposed device per AAMI/ANSI ES 60601-1, IEC 60601-2-18 and IEC 60601-1-2.

{8}------------------------------------------------

The following testing demonstrated that the subject device complies with the following standards:

  • AAMI/ANSI/ ES 60601-1:2005/(R)2012 and A1: 2012, C1:2009(R)2012 and A2:2010(R)2012(Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;

  • IEC 60601-2-18:2009 Medical electric equipment- Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment

  • A IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests;
  • ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity;

  • A ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization;
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity A
  • ASTM F88/F88M-15 Standard method for seal strength of flexible barrier materials

  • GB18280:2007 (IDT ISO 11137: 2006)

    1. Clinical Test Conclusion

No clinical study is included in this submission.

    1. Conclusion
      The conclusions drawn from the subject device indications for use, technological characteristics and performance testing demonstrates that the subject device is as safe and effective as and is substantially equivalent to the legally marketed predicate devices.

N/A