K Number
K210502

Validate with FDA (Live)

Date Cleared
2021-08-27

(186 days)

Product Code
Regulation Number
882.5600
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
  2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
  3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50cc for age <80 years, 0-20cc for age ≥80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
Device Description

The Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever delivery wire has a hydrophilic coating on the distal 101cm length to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A Torque Device is provided with the Retriever to facilitate manipulation and retrieval. The Retriever comes preloaded in an insertion tool to introduce the Retriever into a Microcatheter.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Trevo NXT ProVue Retriever, based on the provided text:

Acceptance Criteria and Reported Device Performance

Note: The document primarily focuses on establishing substantial equivalence to a predicate device, specifically regarding a change in labeling for compatibility with smaller aspiration catheters. It doesn't explicitly state quantitative acceptance criteria for primary clinical endpoints of the device itself in the context of a de novo submission. Instead, it demonstrates through comparative data that the modified use case (with smaller catheters) maintains equivalent performance to the established use case.

Acceptance Criterion (Implicit)Reported Device Performance
Successful Clot Retrieval and Recanalization (Simulated Use)Simulated Use testing met acceptance criteria.
Acceptable Particulate GenerationAll test samples met the applicable user needs and acceptance criteria for particulate characterization when used with AXS Vecta 46 Intermediate Catheter.
Equivalence in Functional Outcome (Modified vs. Predicate Use)Regression analysis of RWD indicated that using Trevo with 0.046in (1.17mm) ID aspiration catheters is equivalent to using Trevo with larger aspiration catheters with similar odds of good functional outcome (Odds Ratio: 0.72; 95% CI: 0.44-1.17).
Equivalence in Safety Endpoints (Modified vs. Predicate Use)Regression analysis of RWD indicated similar odds for multiple safety endpoints when using Trevo with 0.046in (1.17mm) ID aspiration catheters compared to larger ones.
Equivalence in Revascularization Rates (Modified vs. Predicate)Registry data confirmed equivalence of revascularization rates for eTICI>2c, with a 90% CI for the difference in means of severity-adjusted posterior probabilities from 0.003 to 0.009 (0.30% to 0.90%). This was within the equivalence margin of ±5%.
Equivalence in 90-day Modified Rankin Scale (MRS) (Modified vs. Predicate)Registry data confirmed equivalence for 90-day MRS, with a 90% CI for the difference from -0.050 to 0.042 (-5.0% to 4.2%). This was within the equivalence margin of ±5%.

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set A (Simulated Use & Particulate Characterization): Sample sizes are not explicitly stated, but these were bench tests performed by Stryker Neurovascular.
    • Test Set B (Animal Study): Sample size for the animal study is not specified.
    • Test Set C (Real-World Data - RWD):
      • An administrative claims database (details regarding country/scope not specified).
      • Post-market registries: "The Trevo Retriever Registry" and "The ASSIST Registry" (details regarding country/scope not specified).
      • Retrospective data, as it's an "analysis of Real-World Data (RWD) from an administrative claims database, post-market registries...as well as a review of relevant literature."
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • The document does not specify the number or qualifications of experts for establishing ground truth for any of the clinical/RWD test sets. The RWD analysis likely used existing clinical outcomes as "ground truth." For bench and animal studies, "ground truth" would be determined by the study design and measurements.
  3. Adjudication Method for the Test Set:

    • The document does not describe any adjudication method for the RWD or registry data.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study involving human readers or AI assistance in that direct context was described. The RWD analysis was focused on the device's performance with different aspiration catheter sizes, not on human-AI interaction.
  5. Standalone (Algorithm Only) Performance:

    • Not applicable. The device is a physical neurovascular mechanical thrombectomy device, not an algorithm. The "performance data" refers to the device's efficacy and safety in its intended use.
  6. Type of Ground Truth Used:

    • Bench Testing: Internal performance criteria and measurements.
    • Animal Study: Evaluation of vascular trauma, recanalization, and distal emboli based on experimental observations.
    • Real-World Data (RWD):
      • Patient outcomes data from an administrative claims database.
      • Revascularization rates (eTICI>2c) and 90-day Modified Rankin Scale (MRS) from post-market registries and relevant literature. This would be considered outcomes data.
  7. Sample Size for the Training Set:

    • Not applicable. The device is a physical medical device, not an AI/ML algorithm that requires a "training set" in the computational sense. The document refers to testing data to support a change in labeling for compatibility.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no "training set" for this physical device.

{0}------------------------------------------------

August 27, 2021

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Stryker Neurovascular Rebecca Rosman Senior Staff Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, California 94538

Re: K210502

Trade/Device Name: Trevo NXT ProVue Retriever Regulation Number: 21 CFR 882.5600 Regulation Name: Neurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment Regulatory Class: Class II Product Code: POL, NRY Dated: July 22, 2021 Received: July 26, 2021

Dear Rebecca Rosman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng, Ph.D. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210502

Device Name Trevo NXTTM ProVue Retriever

Indications for Use (Describe)

  1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.

  2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

  3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50cc for age <80 years, 0-20cc for age ≥80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K210502

Introduction:

According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for a determination of substantial equivalence.

Submitter Name, Address and Content:

Submitter:Stryker Neurovascular47900 Bayside ParkwayFremont, CA 94538-6515(FDA Registration Number: 3008853977)
Contact:Ashley TwittyManager, Regulatory AffairsPhone: 602-621-3089Fax: 510-413-2588Email: ashley.twitty@stryker.com

August 23, 2021

Device Name and Classification:

Date Prepared:

Trade/Proprietary Name:Trevo NXTTM ProVue Retriever
Common Name:Trevo Retriever
Classification Name:Neurovascular Mechanical Thrombectomy Device for Acute IschemicStroke Treatment, 21 CFR 882.5600, Class IIPercutaneous Catheter, 21CFR 870.1250 – Class II
Product Code:POL. NRY

{4}------------------------------------------------

Legally Marketed Predicate Device

Name of Predicate DeviceName ofManufacturer510(k)Number
Trevo NXT™ ProVueRetrieverStrykerNeurovascularK203219

Device Description

The Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever delivery wire has a hydrophilic coating on the distal 101cm length to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A Torque Device is provided with the Retriever to facilitate manipulation and retrieval. The Retriever comes preloaded in an insertion tool to introduce the Retriever into a Microcatheter.

Indications for Use

    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
    1. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50cc for age <80 years, 0-20cc for age ≥80 years). Endovascular therapy with the device

{5}------------------------------------------------

should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.

Technological Characteristics and Product Feature Comparison

Stryker Neurovascular has demonstrated the Trevo NXT ProVue Retriever with the modified IFU is substantially equivalent to the Predicate device, Trevo NXT ProVue Retriever (K203219) based on the same indications for use, device design, materials, manufacturing, packaging and sterilization methods. A comparison of the Subject device with the Predicate device is summarized in Table 1 below.

FeaturePredicate DeviceTrevo NXT ProVue RetrieverSubject DeviceTrevo NXT ProVue Retriever
RegulationNumber21 CFR 882.5600 21 CFR 870.1250Same
RegulationNameNeurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment Percutaneous CatheterSame
ClassificationClass IISame
Product CodePOL NRYSame
Intended UseNeurovascular mechanical thrombectomy device for acute ischemic stroke treatment used in the treatment of acute ischemic stroke to improve clinical outcomes.Same
Indications forUse1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activatorSame
FeaturePredicate DeviceTrevo NXT ProVue RetrieverSubject DeviceTrevo NXT ProVue Retriever
(IV t-PA) or who fail IV t-PA therapyare candidates for treatment.3. The Trevo Retriever is indicated for useto restore blood flow in theneurovasculature by removing thrombusfor the treatment of acute ischemicstroke to reduce disability in patientswith a persistent, proximal anteriorcirculation, large vessel occlusion of theinternal carotid artery (ICA) or middlecerebral artery (MCA)-M1 segmentswith smaller core infarcts (0-50cc for age<80 years, 0-20cc for age ≥80 years).Endovascular therapy with the deviceshould start within 6-24 hours of timelast seen well in patients who areineligible for intravenous tissueplasminogen activator (IV t-PA) or whofail IV t-PA therapy.
TargetPopulationPatients experiencing acute ischemic strokeSame
AnatomicalSitesNeurovasculatureSame
TECHNOLOGICAL CHARACTERISTICS
DeviceDescriptionThe Retriever consists of a flexible, taperedcore wire with a shaped section at the distalend. Platinum markers at the distal end allowfluoroscopic visualization. In addition, theshaped section is also radiopaque. Retrieverdimensions are indicated on product label.The Retriever delivery wire has ahydrophilic coating on the distal 101cm toreduce friction during use. The Retriever hasa shaft marker to indicate proximity ofRetriever tip relative to Microcatheter tip. Atorque device is provided with the Retrieverto facilitate manipulation and retrieval. TheRetriever comes preloaded in an insertiontool to introduce the Retriever into aMicrocatheter.Same
Principle ofOperationThe Trevo Retriever is delivered to thethrombus using a microcatheter. TheMicrocatheter is then retracted to deploy theshaped section of the Retriever. If using anAspiration Catheter, remove Microcatheter.Advance Aspiration Catheter over proximalsection of Retriever while applyingaspiration using a 60 mL syringe or anaspiration pump. The Retriever withMicrocatheter or Aspiration Catheter areSame

Table 1. Product Feature Comparison of Subject Device to Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

FeaturePredicate DeviceTrevo NXT ProVue RetrieverSubject DeviceTrevo NXT ProVue Retriever
pulled back to capture the thrombus. TheRetriever, thrombus, and Microcatheter orAspiration Catheter are removed as a unitfrom the body.Same
ProceduralStepsAspirationSourceSyringe, Aspiration pumpSame
AccessoryDevicesInsertion tool and torque device providedwithin product packageSame
Sizes3x32mm4x28mm4x41mm6x37mmSame
CompatibilityRetrieverSize Trevo Pro14MicrocatheterInner Diameter0.017 inches(0.43 mm) Trevo Trak 21MicrocatheterInner Diameter0.021 inches(0.53 mm) Trevo Pro18MicrocatheterInner Diameter0.021 inches(0.53 mm) Excelsior®XT-27®Microcatheter(REF XT275081)Inner diameter0.027 inches(0.69 mm) RecommendedMinimumVessel ID (mm) Trevo NXT3x32 ✓ ✓ ✓ ✓ Trevo NXT4x28 ✓ ✓ ✓ 2.5 Trevo NXT4x41 ✓ ✓ ✓ ✓ Trevo NXT6x37 ✓ ✓ ✓ Compatibility of the Retriever with othermicrocatheters has not been established.Performance ofthe Retriever device may be impacted if adifferent microcatheter is used.Balloon Guide Catheters and AspirationCatheters (commercially availableaspiration catheters with minimum innerdiameter 0.058 inches (1.47mm)) arerecommended for use during thrombusremoval procedures.RetrieverSize Trevo Pro14MicrocatheterInner Diameter0.017 inches(0.43 mm) Trevo Trak 21MicrocatheterInner Diameter0.021 inches(0.53 mm) Trevo Pro18MicrocatheterInner Diameter0.021 inches(0.53 mm) Excelsior®XT-27®Microcatheter(REF XT275081)Inner diameter0.027 inches(0.69 mm) RecommendedMinimumVessel ID (mm) Trevo NXT3x32 ✓ ✓ ✓ ✓ Trevo NXT4x28 ✓ ✓ ✓ 2.5 Trevo NXT4x41 ✓ ✓ ✓ ✓ Trevo NXT6x37 ✓ ✓ ✓ Compatibility of the Retriever with othermicrocatheters has not been established.Performance ofthe Retriever device may be impacted if adifferent microcatheter is used.Balloon Guide Catheters and AspirationCatheters (commercially availableaspiration catheters with minimum innerdiameter 0.046 inches (1.17mm)) arerecommended for use during thrombusremoval procedures.
MATERIALS
Core WireNitinol (nickel titanium alloy)Same
Shaped SectionNitinolSame
Distal CoilPlatinum/TungstenSame
Shaped SectionRadiopaqueWirePlatinum/TungstenSame
Mid Coil304 Stainless SteelSame
Proximal CoilPebaxSame
SolderGold/TinSame

{8}------------------------------------------------

FeaturePredicate DeviceTrevo NXT ProVue RetrieverSubject DeviceTrevo NXT ProVue Retriever
HydrophilicCoatingSodium hyaluronate mixtureSame
DIMENSIONAL DRAWING
DimensionalDrawingAll Retriever sizes:Image: Dimensional Drawing with labels A, B, C, D, E, and FAll Retriever sizes:Same
Overall Length(A)200cmSame
Total ShapedSection Length(B)32, 36, 40, 44mmSame
Full DiameterLength (C)21, 25, 30, 35mmSame
Shaped SectionDiameter (D)3, 4, 6mmSame
Delivery CoreWire OuterDiameter (E)0.015, 0.019"Same
Cell CoverageLength (F)28, 32, 37, 41mmSame
PACKAGING
Materials andConfigurationPolyethylene Hoop, polycarbonate mountingcard, Tyvek/Film Pouch, Chipboard cartonSame
SterilizationMethod100% EtOSame
How SuppliedSterile/Single UseSame

The differences between the devices are not critical as demonstrated above and through the testing referenced below.

Risk Assessment

Risk assessment of the Trevo NXT ProVue Retriever has been conducted in accordance with EN ISO 14971. Stryker Neurovascular has determined that the labeling changes to the Trevo NXT ProVue Retriever raise no new questions of safety or effectiveness. Results of testing are appropriate for determining that the Trevo NXT ProVue Retriever with the modified IFU is substantially equivalent to the legally marketed Predicate device.

{9}------------------------------------------------

Testing Summary

There are no changes to the device intended use or indications for use statement. Other than the proposed labeling changes regarding compatibility, there are no changes in the device design, materials, manufacturing, packaging and sterilization methods; therefore, biocompatibility data, bench performance data, sterilization and stability data from the Predicate device (K203219) are directly applicable. Relevant testing data supporting the Subject device are summarized as follows.

Performance Data – Bench Testing

Stryker Neurovascular performed the following non-clinical bench test to assess the usability of the Trevo Retriever with 0.046 in. ID aspiration catheters. The additional bench testing is summarized in Table 2 below.

TestTest Method SummaryConclusions
Simulated UseSimulated Use testing utilized aneurovascular model to assess thedevice's ability to retrieve the clot andachieve recanalization.Simulated Use testingmet acceptance criteria.
ParticulateCharacterizationPurpose: The purpose of this test was todocument and assess the particulatematter generated from the delivery ofneuro-interventional devices (includingthe Trevo NXT ProVue Retriever)through the inner lumen of the AXSVecta 46 Intermediate Catheter.Method: Particulate testing wasconducted with AXS Vecta 46 and TrevoNXT ProVue Retriever in a simulateduse condition based on the device IFUand a clinically relevant tortuous model.All test samples met theapplicable user needsand acceptance criteria.

Table 2. Performance Data - Design Verification Bench Testing

Performance Data – Animal

{10}------------------------------------------------

To support the labeling modification, Stryker Neurovascular leveraged data from an animal study that was conducted in compliance with applicable requirements in the GLP regulation (21 CFR Part 58) to evaluate vascular trauma, recanalization and/or distal emboli of the proposed devices for performing a combined neurothrombectomy procedure.

Performance Data – Clinical

To support the labeling modifications, Stryker Neurovascular conducted analyses of Real-World Data (RWD) from an administrative claims database, post-market registries (The Trevo Retriever Registry and The ASSIST Registry), as well as a review of relevant literature to assess safety and effectiveness of the recommended use of the Trevo with aspiration catheters with a minimum inner diameter (ID) of 0.046in (1.17mm).

Regression analysis of RWD indicated that using Trevo with smaller aspiration catheters is equivalent to using Trevo with larger aspiration catheters with similar odds of good functional outcome (0.72; 95% CI: 0.44-1.17) and similar odds for multiple safety endpoints. Further analysis of registry data confirmed equivalence of revascularization rates and clinical outcomes independent of catheter size with the 90% CI for the difference in means of the severity adjusted posterior probabilities for eTICI>2c to be from 0.003 to 0.009 (.30% to 0.90%). For 90-day MRS, the CI was from -0.050 to 0.042 (-5.0% to 4.2%). In both cases, the upper and lower bounds of the intervals were within the equivalence margin of ±5%.

Shelf Life Testing

Shelf life testing previously conducted for the Trevo NXT ProVue Retriever was leveraged to support the changes to the device and can be found in K192207. Shelf life testing was not performed on the Subject device since there was no impact to device material, design, or safety and efficacy as a result of the labeling changes. As with the Predicate device, the Subject device is labeled with a 2-year shelf life.

Sterilization

Sterilization evaluation previously conducted for the Trevo NXT ProVue Retriever was leveraged to support the labeling changes to the device and can be found in K192207. The Trevo

{11}------------------------------------------------

NXT ProVue Retrievers are sterilized with 100% Ethylene Oxide and provided sterility assurance level (SAL) of 10 6 has been demonstrated. The Trevo NXT ProVue Retrievers meet EO residuals per EN ISO 10993-7 for limited contacting, externally communicated devices. The Trevo NXT ProVue Retrievers are for single use only.

Biocompatibilitv

Biocompatibility testing previously conducted for the Trevo NXT ProVue Retriever was leveraged to support the labeling changes to the device and can be found in K192207. The results of biocompatibility testing, and biological safety evaluation of the Trevo NXT ProVue Retrievers demonstrate that the devices meet biological safety requirements per ISO 10993-1 for externally communicating medical devices with circulating blood contact for less than 24 hours. The Trevo NXT ProVue Retrievers are considered to have no residual risk of biological hazards. Also, the devices and its packaging do not contain detectable latex. Therefore, the Trevo NXT ProVue Retriever devices, accessories, and primary packaging are considered biocompatible for their intended use.

Conclusion

Based on the conclusions drawn from risk assessment, bench testing results, Real World Data from an Administrative Claims, and an analysis of relevant clinical literature summarized above, the Subject device demonstrates substantial equivalence to the legally marketed Predicate device (K203219).

§ 882.5600 Neurovascular mechanical thrombectomy device for acute ischemic stroke treatment.

(a)
Identification. A neurovascular mechanical thrombectomy device for acute ischemic stroke treatment is a prescription device used in the treatment of acute ischemic stroke to improve clinical outcomes. The device is delivered into the neurovasculature with an endovascular approach, mechanically removes thrombus from the body, and restores blood flow in the neurovasculature.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including:
(i) Mechanical testing to demonstrate the device can withstand anticipated tensile, torsional, and compressive forces.
(ii) Mechanical testing to evaluate the radial forces exerted by the device.
(iii) Non-clinical testing to verify the dimensions of the device.
(iv) Non-clinical testing must demonstrate the device can be delivered to the target location in the neurovasculature and retrieve simulated thrombus under simulated use conditions.
(v) Non-clinical testing must demonstrate the device is radiopaque and can be visualized.
(vi) Non-clinical testing must evaluate the coating integrity and particulates under simulated use conditions.
(vii) Animal testing must evaluate the safety of the device, including damage to the vessels or tissue under anticipated use conditions.
(3) Performance data must support the sterility and pyrogenicity of the patient contacting components of the device.
(4) Performance data must support the shelf-life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf-life.
(5) Clinical performance testing of the device must demonstrate the device performs as intended for use in the treatment of acute ischemic stroke and must capture any adverse events associated with the device and procedure.
(6) The labeling must include:
(i) Information on the specific patient population for which the device is intended for use in the treatment of acute ischemic stroke, including but not limited to, specifying time from symptom onset, vessels or location of the neurovasculature that can be accessed for treatment, and limitations on core infarct size.
(ii) Detailed instructions on proper device preparation and use for thrombus retrieval from the neurovasculature.
(iii) A summary of the clinical testing results, including a detailed summary of the device- and procedure-related complications and adverse events.
(iv) A shelf life.