K Number
K210487

Validate with FDA (Live)

Date Cleared
2021-03-16

(25 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The subject device is a Blue color and Flat Pleated type mask, utilizing the Ear Loops for wearing, and has Nose Piece design for fitting the facemask around the nose. The subject device is manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filters. The subject device is held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece contained in the subject device is in the layers of facemask to allow the user to fit the facemask around their nose, which is made of metal core plastic (aluminum metal core inside of polypropylene). The subject device is sold non-sterile and is intended to be single use, disposable device.

AI/ML Overview

The provided document describes the performance of a MIPLNI YK-244 Disposable Surgical Mask based on non-clinical performance testing. It does not contain information about a study proving that an AI device meets acceptance criteria. Therefore, most of the requested information regarding AI device studies cannot be extracted from this document.

However, I can provide the acceptance criteria and reported device performance for the surgical mask.

1. Table of Acceptance Criteria and Reported Device Performance (for the surgical mask):

Test ItemStandardsAcceptance CriteriaReported Device Performance (Results)
Dimensions-Mask Layer: Length: 17.5cm±0.5cm, Width: 9.5cm±0.5cmPass
Nose Piece: 3mm x 105 mm x 0.5mmPass
Ear loop: 2.8mm in diameter, 195mm in lengthPass
ASTM F2100 LevelASTM F2100-19Level 3Pass
Fluid Resistance PerformanceASTM F186232 out of 32 pass at 160 mmHgPass LOT #YKWK20201108: 32/32 at 160 mmHgLOT #YKWK20201111: 32/32 at 160 mmHgLOT #YKWK20201120: 32/32 at 160 mmHg
Particulate Filtration Efficiency (PFE)ASTM F2299≥ 98%Pass LOT #YKWK20201108: 32/32 99.9% ±0.1%LOT #YKWK20201111: 32/32 99.9% ±0.1%LOT #YKWK20201120: 32/32 99.9% ±0.1%
Bacterial Filtration Efficiency (BFE)ASTM F2101≥ 98%Pass LOT #YKWK20201108: 32/32 99.9%LOT #YKWK20201111: 32/32 99.9%LOT #YKWK20201120: 32/32 99.9%
Differential Pressure (Delta P)EN 14683:2019+AC:2019 Annex C< 5.0 mmH2O/cm²Pass LOT #YKWK20201108: < 3.6 mm H2O /cm² (20-22°C and 55-62% RH, Flow Rate: 8 L/min)LOT #YKWK20201111: < 4.0 mm H2O / cm² (20-22°C and 55-62% RH, Flow Rate: 8 L/min)LOT #YKWK20201120: < 4.0 mm H2O /cm² (20-22°C and 55-62% RH, Flow Rate: 8 L/min)
Flammability16 CFR 1610Class 1Pass LOT #YKWK20201108: IBELOT #YKWK20201111: IBELOT #YKWK20201120: IBE
In Vitro CytotoxicityISO 10993-5Non-cytotoxicPass
Skin IrritationISO 10993-10Non-irritatingPass
Skin SensitizationISO 10993-10Non-sensitizingPass

Since this document pertains to a surgical mask and not an AI device, the following points are not applicable and are not found in the provided text:

  • 2. Sample sized used for the test set and the data provenance (for an AI device)
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (for an AI device)
  • 4. Adjudication method (for an AI device)
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
  • 7. The type of ground truth used (for an AI device)
  • 8. The sample size for the training set (for an AI device)
  • 9. How the ground truth for the training set was established (for an AI device)

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Hunan Yuankang Biological Technology Co., Ltd % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K210487

Trade/Device Name: MIPLNI YK-244 Disposable Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 16, 2021 Received: February 19, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210487

Device Name MIPLNI YK-244 Disposable Surgical Mask

Indications for Use (Describe)

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)Own-Use Production of CFR 221 Subject Cigarettes Contract Production of CFR 221 Subject Cigarettes
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Administrative Information

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

Assigned 510(k) Number: K210487

Date of Preparation: 2021/3/8

The submitter's information is listed in the Table below:

Submitter Information
EstablishmentHunan Yuankang Biological Technology Co., LtdHeping Team, Sanlian Community, Xidu TownHengyang, Hunan, CN 421000
Registration Number3016752849
ContactKylin Rui ChenDirector, Business OperationsHeping Team, Sanlian Community, Xidu TownHengyang, Hunan, CN 421000Phone: +86 13560755999 / 400-6886-658Fax: +86-734-6856608Email: kylin.chen@yuankangbio.com
Submission CorrespondentKelvin Fei ChuRegulatory ConsultantKMax Medtech Consulting3109 Butte St,Santa Clara, CA 95051Phone: 612-203-5245Fax: +86-734-6856608Email: kelvin.chu@kmaxmedtech.com

Table Submitter information

The subject device's information is listed in the Table below:

Subject Device Information
Trade NameMIPLNI YK-244 Disposable Surgical Mask

{4}------------------------------------------------

Subject Device Information
Common NameSurgical Face Mask
Product CodeFXX
ClassificationClass II
Classification NameMask, Surgical
CFR Section #878.4040
Review PanelGeneral Hospital

Table Subject device information

The predicate device's information is listed in the Table below:

Predicate Device Information
Predicate 510(k) #K153496
Predicate Trade NameDisposable Surgical Face Mask
Predicate ManufacturerXiantao Rayxin Medical Products Co., ltd
Predicate Product CodeFXX

Table Predicate device information

Intended Use / Indications for Use

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The subject device has been designed following the recommendations provided in Surgical Masks - Premarket Notification [510(k)] Submissions Guidance for Industry and FDA Staff (2004).

The subject device is a Blue color and Flat Pleated type mask, utilizing the Ear Loops for wearing, and has Nose Piece design for fitting the facemask around the nose.

The subject device is manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filters.

The subject device is held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex.

{5}------------------------------------------------

The nose piece contained in the subject device is in the layers of facemask to allow the user to fit the facemask around their nose, which is made of metal core plastic (aluminum metal core inside of polypropylene).

The subject device is sold non-sterile and is intended to be single use, disposable device.

Summary of Non-Clinical Performance Testing

Non-clinical performance tests have been conducted to verify that the subject device meets all design requirements following standards and items recommended in Surgical Masks - Premarket Notification [510(k)] Submissions Guidance for Industry and FDA Staff (2004).

  • . ASTM F2100-19, Standard Specification for Performance of Materials Used in Medical Face Masks
  • ASTM F1862-13, Standard Test Method for Resistance of Medical Face Masks to ● Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity).
  • ASTM F2299-03, Stand Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres.
  • ASTM F2101-19, Standard Test Method for Evaluating the Bacterial Filtration Efficiency ● (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
  • . EN 14683:2019+AC:2019 Medical face masks - Requirements and test methods Annex C Method for determination of breathability (differential pressure)
  • 16 CFR 1610, Standard for The Flammability of Clothing Textiles. ●
  • ISO 10993-5: 2009 Biological Evaluation of Medical Devices Part 5: Tests For In Vitro ● Cytotoxicity
  • . ISO 10993-10: 2010 Biological Evaluation of Medical Devices – Part 10: Tests For Irritation And Skin Sensitization
  • Bench Testing for the performance of Dimensions ●

The summary of non-clinical performance testing is listed below in Table.

Test ItemStandardsAcceptance CriteriaResults
Dimensions-Mask Layer:Length: 17.5cm±0.5cmWidth: 9.5cm±0.5cmNose Piece:3mm x 105 mm xPass
Test ItemStandardsAcceptance CriteriaResults
0.5mmEar loop:2.8mm in diameter,195mm in length.
ASTM F2100 LevelASTM F2100-19Level 3Pass
Fluid ResistancePerformanceASTM F186232 out of 32 passat 160 mmHgPassLOT#YKWK20201108:32/32 at 160 mmHgLOT#YKWK20201111:32/32 at 160 mmHgLOT#YKWK20201120:32/32 at 160 mmHg
ParticulateFiltration Efficiency(PFE)ASTM F2299≥ 98%PassLOT#YKWK20201108:32/32 99.9% ±0.1%LOT#YKWK20201111:32/32 99.9% ±0.1%LOT#YKWK20201120:32/32 99.9% ±0.1%
Bacterial FiltrationEfficiency(BFE)ASTM F2101≥ 98%PassLOT#YKWK20201108:32/32 99.9%LOT#YKWK20201111:32/32 99.9%LOT#YKWK20201120:32/32 99.9%
Test ItemStandardsAcceptance CriteriaResults
DifferentialPressure(Delta P)EN14683:2019+AC:2019Annex C< 5.0 mmH2O/cm²PassLOT #YKWK20201108:32/32 < 3.6 mm H2O /cm²(20-22°C and 55-62% RH, Flow Rate: 8 L/min)LOT #YKWK20201111:32/32 < 4.0 mm H2O / cm²(20-22°C and 55-62% RH, Flow Rate: 8 L/min)LOT #YKWK20201120:32/32 < 4.0 mm H2O /cm²(20-22°C and 55-62% RH, Flow Rate: 8 L/min)
Flammability16 CFR 1610Class 1PassLOT #YKWK20201108:32/32 IBELOT #YKWK20201111:32/32 IBELOT #YKWK20201120:32/32 IBE
In VitroCytotoxicityISO 10993-5Non-cytotoxicPass
Skin IrritationISO 10993-10Non-irritatingPass
Test ItemStandardsAcceptance CriteriaResults
Skin SensitizationISO 10993-10Non-sensitizingPass
Table Summary of non-clinical performance testing

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Summary of Clinical Performance Test

No clinical study is included in this submission.

Comparison of Technological Characteristics with Predicate Device

A summary of the technological characteristics of the subject device as compared to the predicate device is listed below in Table.

DeviceSubject DevicePredicate DeviceComparison
510 (k) #K210487K153496-
ManufacturerHunan Yuankang BiologicalTechnology Co., LtdXiantao Rayxin MedicalProducts Co., ltd.-
Product NameDisposable Surgical Face MaskDisposable Surgical Face MaskSame
ClassificationClass II Device, FXX(21 CFR878.4040)Class II Device, FXX(21 CFR878.4040)Same
Intended useThe Disposable Surgical FaceMasks are intended to be wornto protect both the patient andhealthcare personnel fromtransfer of microorganisms,body fluids and particulatematerial. These face masks areintended for use in infectioncontrol practices to reduce thepotential exposure to blood andbody fluids. This is a single use,disposable device(s), providednon-sterile.The Disposable Surgical FaceMasks are intended to be wornto protect both the patient andhealthcare personnel fromtransfer of microorganisms,body fluids and particulatematerial. These face masks areintended for use in infectioncontrol practices to reduce thepotential exposure to blood andbody fluids. This is a singleuse, disposable device(s),provided non-sterile.Same
Material & Design
Outer facinglayerSpun-bond polypropyleneSpun-bond polypropyleneSame

{9}------------------------------------------------

DeviceSubject DevicePredicate DeviceComparison
Middle layerMelt blown polypropylene filterMelt blown polypropylene filterSame
Inner facing layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Nose pieceMetal core plastic(aluminum metal core inside of polypropylene)Malleable aluminum wireSimilar
Ear loopsSpandexPolyesterSimilar
Design featuresColor: BlueEar loopsColor: BlueEar Loops or Tie-OnSimilar
Mask StyleFlat PleatedFlat PleatedSame
Specification andDimensionLength: 17.5cm±0.5cmWidth: 9.5cm±0.5cmLength: 17.5cm±1cmWidth: 9.5cm±1cmSimilar
Usage
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
Performance Testing
ASTM F2100LevelLevel 3Level 2Better
32 out of 32 passat 160 mmHg(ASTM F1862)32 out of 32 passat 120 mmHg(ASTM F1862)Better
Fluid ResistancePerformanceLOT #YKWK20201108: 32/32at 160 mmHgLOT #YKWK20201111: 32/32at 160 mmHgLOT #YKWK20201120: 32/32at 160 mmHg
DeviceSubject DevicePredicate DeviceComparison
ParticulateFiltrationEfficiency (PFE)≥ 98%(ASTM F2299)≥ 98%(ASTM F2299)Same
LOT #YKWK20201108: 32/3299.9% ±0.1%
LOT #YKWK20201111: 32/3299.9% ±0.1%
LOT #YKWK20201120: 32/3299.9% ±0.1%
BacterialFiltrationEfficiency(BFE)≥ 98%(ASTM F2101)≥ 98%(ASTM F2101)Same
LOT #YKWK20201108: 32/3299.9%
LOT #YKWK20201111: 32/3299.9%
LOT #YKWK20201120: 32/3299.9%
DifferentialPressure(Delta P)< 5.0 mmH2O/cm2(EN 14683:2019+AC:2019Annex C)< 5.0 mmH2O/cm2(MIL-M-36954C)Same
LOT #YKWK20201108: 32/32< 3.6 mm H2O / cm2(20-22°C and 55-62% RH, FlowRate: 8 L/min)
LOT #YKWK20201111: 32/32< 4.0 mm H2O / cm2(20-22°C and 55-62% RH, FlowRate: 8 L/min)
LOT #YKWK20201120: 32/32< 4.0 mm H2O / cm2(20-22°C and 55-62% RH, FlowRate: 8 L/min)
DeviceSubject DevicePredicate DeviceComparison
FlammabilityClass 1(16 CFR 1610)Class 1(16 CFR 1610)Same
LOT #YKWK20201108: 32/32IBE
LOT #YKWK20201111: 32/32IBE
LOT #YKWK20201120: 32/32IBE
Biocompatibility Testing
In VitroCytotoxicityUnder the conditions of thestudy, the subject device extractwas determined to be non-cytotoxic. (ISO 10993-5)Under the conditions of thestudy, the predicate deviceextract was determined to benon-cytotoxic. (ISO 10993-5)Same
Skin IrritationUnder the conditions of thestudy, the subject device non-polar and polar extracts weredetermined to be non-irritating.(ISO 10993-10)Under the conditions of thestudy, the predicate device non-polar and polar extracts weredetermined to be non-irritating.(ISO 10993-10)Same
Skin SensitizationUnder the conditions of thestudy, the subject device non-polar and polar extracts weredetermined to be non-sensitizing. (ISO 10993-10)Under the conditions of thestudy, the predicate device non-polar and polar extracts weredetermined to be non-sensitizing. (ISO 10993-10)Same

{10}------------------------------------------------

{11}------------------------------------------------

Table Comparison of the subject device and the predicate device

Substantially Equivalent (SE) Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as the legally marketed predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.