K Number
K210485

Validate with FDA (Live)

Date Cleared
2021-05-13

(83 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EG Family: This instrument is intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

EC Family: This instrument is intended to be used with a PENTAX Video Processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

Device Description

The PENTAX Medical Video Upper G.I. scopes (EG Family) and Colonoscopes (EC Family) are used with Video Processors. They are composed of three main components: Insertion Portion, Control Body and PVE Connector. The Insertion Portion is inserted into the body cavity of patient. The Insertion Portion includes the Distal End and Bending Section. The Objective Lens, Light Guide, and Instrument Channel are located on the Distal End of the Insertion. The Distal End also contains an Air/Water Nozzle and Water Jet Nozzle. This Air/Water Feeding System is used to deliver the air and water to the Objective Lens. The Control Body is held by the user's hand. The Control Body includes the Angulation Control lever used to operate the endoscope angulation; Suction Cylinder and Suction Nipple for suctioning fluid and air in the body cavity; Remote Button used to operate the function of video processor; and Instrument Channel Inlet used to insert endotherapy devices, such as biopsy forceps, into the body cavity. The Magnification Control Lever is used to magnify the image on the video monitor, as necessary. As this magnification function is performed electrically, focus and depth of field are not changed. The PVE Connector is connected to the Video Processor via an Electrical Contacts. The PVE Connector includes the Electrical Contacts and Light Guide Plug. The Light Guide Plug is connected to the Light Source inside the Video Processor. The Light Guide of the Distal End is used to illuminate the body cavity by light that is carried through the Light Carrying Bundle. The Light Carrying Bundle guides the light from Light Guide Plug that is connected to the Light Source. The CCD built into the Distal End receives reflected light (image data) from the body cavity, and sends the image data to the Video Processor through the video cable. The image data are converted into the image signal by the Video Processor, and the image inside the body cavity is displayed on the Monitor. The PENTAX Medical Video Upper G.I. scopes and Colonoscopes are reusable semi-critical devices. Since they are packaged non-sterile, they must be high-level disinfected or sterilized before initial use. Prior to each subsequent procedure, they must be subjected to an appropriate cleaning and either high-level disinfection or sterilization processes.

AI/ML Overview

This document describes a 510(k) premarket notification for PENTAX Medical Video Upper GI Scopes (EG Family) and PENTAX Medical Video Colonoscopes (EC Family). The submission focuses on changes to reprocessing instructions, not the devices themselves.

Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

  1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of acceptance criteria or performance metrics for the device itself in terms of diagnostic accuracy, sensitivity, specificity, or similar clinical performance measures. This 510(k) submission is for a modification to the reprocessing instructions for existing, already cleared devices (K131902 and K131855). Therefore, the "acceptance criteria" here relate to the reprocessing instructions being safe and effective.

    The primary "performance" reported is related to human factors and the safety/effectiveness of the reprocessing process as modified.

    Acceptance Criterion (Implicit for Reprocessing)Reported Performance (from Human Factors Study)
    Reprocessing instructions are effectiveA Summative Human Factors Study was conducted to validate the effectiveness and risk management measures for reprocessing.
    Reprocessing instructions are safeThe study aimed to ensure reprocessing can be performed in a safe manner. Although use errors, close calls, and difficulties were observed, "it is estimated that the risks identified, and the root causes assigned to those risks can be reduced and/or mitigated via additional modifications of the rIFU design and the further refinement of the training materials."
  2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions a "Summative Human Factors Study." While the number of participants (test set size) is not explicitly stated, it refers to "the same group of stakeholders." The provenance of this data (e.g., country of origin) is not provided, but such studies are typically conducted with users representative of the intended clinical environment. The study is implicitly prospective, as human factor studies involve observing users interacting with the device/instructions.

  3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    This information is not applicable in the context of this 510(k) submission. The ground truth for a human factors study on reprocessing focuses on whether users can correctly and safely follow the instructions. This isn't a diagnostic accuracy study requiring expert adjudication of clinical findings. The "experts" involved would likely be human factors professionals and potentially subject matter experts in endoscope reprocessing, but their number and specific qualifications are not detailed.

  4. Adjudication Method for the Test Set:

    This is not a clinical study involving diagnostic outputs requiring adjudication. In a human factors study, observations of user performance (following instructions, committing errors) are typically recorded and analyzed by human factors specialists. There is no mention of a traditional adjudication method (like 2+1, 3+1 consensus) for this type of test.

  5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This submission is about reprocessing instructions, not a diagnostic AI device to be used by human readers for clinical diagnosis.

  6. Standalone (Algorithm Only Without Human-in-the-loop Performance) Study:

    No, a standalone algorithm study was not done. This submission is for endoscopes and their reprocessing instructions, not an AI algorithm.

  7. Type of Ground Truth Used:

    For the human factors study, the "ground truth" would be the correct and safe steps for reprocessing the endoscopes, as defined by the manufacturer's validated procedures. The performance of the participants is compared against these predetermined correct steps. It is not clinical ground truth derived from pathology or patient outcomes.

  8. Sample Size for the Training Set:

    This information is not applicable. This is not an AI/machine learning submission with training data.

  9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no training set for an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

May 13, 2021

PENTAX of America, Inc. William Goeller VP Quality Assurance and Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782

Re: K210485

Trade/Device Name: PENTAX Medical Video Upper GI Scopes (EG Family), PENTAX Medical Video Colonoscopes (EC Family) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FDS, FDF Dated: February 17, 2021 Received: February 19, 2021

Dear William Goeller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director. DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210485

Device Name

PENTAX Medical Video Upper GI Scopes (EG Family) PENTAX Medical Video Colonoscopes (EC Family)

Indications for Use (Describe)

EG Family:

This instrument is intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

EC Family:

This instrument is intended to be used with a PENTAX Video Processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PENTAX Medical Video Upper G.I. scopes (EG Family) and Colonoscopes (EC Family)- 5 rIFUs Traditional 510(k) Submission

510(k) Summary

SUBMITTER l.

PENTAX of America, Inc., HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782 Phone: 201-571-2318 Ext 2318 Fax: 201-571-2340

Contact: William Goeller Date Prepared: February 17, 2021

II. DEVICE

The purpose of this 510(k) is to obtain clearance for the simplified reprocessing instructions which are separated into (5) rIFU's to eliminate the need for reprocessing personnel to identify what configuration endoscope they are working with.

There are no other changes to the reprocessing instructions for use, nor are there any changes to the design, intended use, or indications for use of the EG Family and EC Family of endoscopes.

Table 5.1: Regulatory Classification of PENTAX Medical Video Upper GI Scopes.
---------------------------------------------------------------------------------
Device NamePENTAX Video Upper GI Scopes (EG Family)
510(k) NumberK131902
Common NameGastroscope And Accessories, Flexible/Rigid
Classification NameEndoscope and accessories
Regulation No.876.1500
Device Class2
Product CodeFDS
Classification PanelGastroenterology/ Urology

Table 5.2: Regulatory Classification of PENTAX Medical Video Colonoscopes

Device NamePENTAX Video Colonoscopes (EC Family)
510(k) NumberK131855
Common NameColonoscopes And Accessories
Classification NameEndoscope and accessories
Regulation No.876.1500

{4}------------------------------------------------

PENTAX Medical Video Upper G.I. scopes (EG Family) and Colonoscopes (EC Family)- 5 rIFUs Traditional 510(k) Submission

Device Class2
Product CodeFDF
Classification PanelGastroenterology/ Urology

III. PREDICATE DEVICES

The predicate devices for this submission are also the subject of this submission, the PENTAX Medical Video Upper G.I. scopes (EG Family) (K131902) and PENTAX Medical Video Colonoscopes (EC Family) (K131855). The predicate devices are identical to the devices subject to this 510(k); the only change relates to the reprocessing instructions that are now broken down into 5 separate manuals for each endoscope type.

IV. DEVICE DESCRIPTION

The subject devices are identical to the predicate devices. There are no changes to the design, specifications, or technological characteristics of the PENTAX Medical Video Upper GI Scopes and PENTAX Medical Video Colonoscopes as described in K131902 and K131855 respectively. The associated model numbers for the PENTAX Medical Video Upper GI Scopes are provided in Table 5.2.

PENTAX Video Upper G.I. scopes (EG Family)
EG-2990iEG-1690KEG-2790KEG-3490KEG27-i10
EG-2790iEG-2490KEG-2990KEG-3890TKEG29-i10

Table 5.2: Model Numbers Associated with the PENTAX Medical Video Upper GJ Scopes

The associated model numbers for the PENTAX Medical Video Colonoscopes are provided in Table 5.3.

Table 5.3: Model Numbers Associated with the PENTAX Medical Video Colonoscopes
PENTAX Video Colonoscopes (EC Family)
EC-3890TLKEC-3490LiEC-3490TLiEC-3890LKEC34-i10L
EC-2990LiEC-3890LiEC-3490LKEC-3890LZiEC38-i10L

The PENTAX Medical Video Upper G.I. scopes (EG Family) and Colonoscopes (EC Family) are used with Video Processors. They are composed of three main components: Insertion Portion, Control Body and PVE Connector.

The Insertion Portion is inserted into the body cavity of patient. The Insertion Portion includes the Distal End and Bending Section. The Objective Lens, Light Guide, and Instrument Channel are located on the Distal End of the Insertion. The Distal End also contains an Air/Water

{5}------------------------------------------------

PENTAX Medical Video Upper G.I. scopes (EG Family) and Colonoscopes (EC Family)- 5 rlFUs Traditional 510(k) Submission

Nozzle and Water Jet Nozzle. This Air/Water Feeding System is used to deliver the air and water to the Objective Lens.

The Control Body is held by the user's hand. The Control Body includes the Angulation Control lever used to operate the endoscope angulation; Suction Cylinder and Suction Nipple for suctioning fluid and air in the body cavity; Remote Button used to operate the function of video processor; and Instrument Channel Inlet used to insert endotherapy devices, such as biopsy forceps, into the body cavity.

The Magnification Control Lever is used to magnify the image on the video monitor, as necessary. As this magnification function is performed electrically, focus and depth of field are not changed.

The PVE Connector is connected to the Video Processor via an Electrical Contacts. The PVE Connector includes the Electrical Contacts and Light Guide Plug. The Light Guide Plug is connected to the Light Source inside the Video Processor. The Light Guide of the Distal End is used to illuminate the body cavity by light that is carried through the Light Carrying Bundle. The Light Carrying Bundle guides the light from Light Guide Plug that is connected to the Light Source. The CCD built into the Distal End receives reflected light (image data) from the body cavity, and sends the image data to the Video Processor through the video cable. The image data are converted into the image signal by the Video Processor, and the image inside the body cavity is displayed on the Monitor.

The PENTAX Medical Video Upper G.I. scopes and Colonoscopes are reusable semi-critical devices. Since they are packaged non-sterile, they must be high-level disinfected or sterilized before initial use. Prior to each subsequent procedure, they must be subjected to an appropriate cleaning and either high-level disinfection or sterilization processes.

V. INDICATIONS FOR USE

There are no changes to the Indications for Use for the subject devices. The sole purpose of this 510(k) is to simplify the reprocessing instructions by breaking two rIFU's into (5) to eliminate the need for reprocessing personnel to identify what configuration endoscope they are working with.

The PENTAX Medical Video Upper Gl Scopes (EG Family) are intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

The PENTAX Medical Video Colonoscopes (EC Family) are intended to be used with a PENTAX Video Processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

{6}------------------------------------------------

VI. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE

The subject devices are identical to the predicate devices. The subject devices have the same fundamental technology and operating principles of the predicate devices, including the same intended use and design technological characteristics, such as Insertion Portion, Control Body and fiberoptic illumination. There are no differences in specifications, including, but not limited to, the depth of field, distal end width, insertion tube width, instrument channel width, and total length. The sole difference between the subject and predicate device is that the reprocessing instructions have been simplified by breaking down rIFUs into (5) separate groups. This modification does not impact the intended use and does not raise different questions of safety and effectiveness.

VII. PERFORMANCE DATA

The optimization of the reprocessing instructions for the PENTAX Medical Video Upper GI scopes and Colonoscopes do not affect the biocompatibility, electrical safety, electromagnetic compatibility, software verification and validation, or performance testing for the scopes.

VIII. HUMAN FACTORS

PENTAX Medical conducted a Summative Human Factors Study to validate the effectiveness and the risk management measures that are in place for the reprocessing of the EG29-i10 Video Gastroscope with a water jet channel. An ancillary study of the EC-3890TLK limited to its second therapeutic channel was carried out with the same group of stakeholders.

The combination of these 2 scopes provides adequate Human Factors evaluation of all design features of EG/EC families related to reprocessing.

As a result of testing, use errors, close calls and difficulties were observed to have occurred in some critical tasks. However, it is estimated that the risks identified, and the root causes assigned to those risks can be reduced and/or mitigated via additional modifications of the rIFU design and the further refinement of the training materials.

VIII. CONCLUSION

The subject devices are identical to the predicate devices. The subject devices have the same intended use and technological characteristics as the predicate devices. There are no changes to the design of the subject devices.

The data submitted support the separation of reprocessing instructions into 5 groups for the PENTAX Medical Video Upper G.I. scopes and Colonoscopes demonstrate that the scopes can be reprocessed in a safe and effective manner.

The data provided in this 510(k) Premarket Notification support the equivalence of the subject and predicate devices.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.