K Number
K210429

Validate with FDA (Live)

Date Cleared
2021-06-02

(110 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided sterile and non-sterile.

Device Description

The Disposable Surgical Face Masks are composed of mask body, nose clip and ear loop. The mask is composed of three layers: the inner and outer layers are made of spun-bond nonwoven fabric, and the middle layer is made of melt blown non-woven fabric. The nose clip is made of polyethylene and iron material, ear loop is made of nylon and spandex material. The size of the disposable surgical mask is 17.5*9.5cm with tolerance±5%cm, the length of the ear loop is ≥ 10.0cm. The outer layer of disposable surgical mask will be provided in blue, the inner layer of the disposable surgical mask will be provided in white, and it will be provided with sterile and is intended to be single use, disposable devices.

AI/ML Overview

This is a 510(k) premarket notification for a Disposable Surgical Face Mask, which is a Class II medical device. The document summarizes the performance testing and substantial equivalence to a predicate device.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodologyPurposeAcceptance CriteriaReported Device PerformanceResult
Functional TestsASTM F2100 Level 3 Requirements
Fluid Resistance (ASTM F1862)To demonstrate the functionality of the subject device.29 out of 32 pass at 160 mmHg for Level 332 out of 32 per lot pass at 160 mmHg (3 non-consecutive lots tested for sterile and non-sterile)PASS
Particulate Filtration Efficiency (ASTM F2299)To demonstrate the functionality of the subject device.≥ 98%Lot1: 99.41% (before sterile), 99.50% (sterile), 99.44% (after aging)Lot2: 99.36% (before sterile), 99.26% (sterile), 99.18% (after aging)Lot3: 99.36% (before sterile), 99.29% (sterile), 99.27% (after aging)PASS
Bacterial Filtration Efficiency (ASTM F2101)To demonstrate the functionality of the subject device.≥ 98%Lot1: 99.77% (before sterile), 99.79% (sterile), 99.74% (after aging)Lot2: 99.80% (before sterile), 99.83% (sterile), 99.77% (after aging)Lot3: 99.80% (before sterile), 99.78% (sterile), 99.76% (after aging)PASS
Differential Pressure (EN 14683)To demonstrate the functionality of the subject device.< 6.0 mmH2O/cm²Lot1: 4.52 (before sterile), 4.52 (sterile), 4.63 (after aging)Lot2: 4.51 (before sterile), 4.54 (sterile), 4.57 (after aging)Lot3: 4.50 (before sterile), 4.50 (sterile), 4.53 (after aging)PASS
Flammability (16 CFR 1610)To demonstrate the functionality of the subject device.Class 1All lots (before sterile, sterile, after aging) for all three lots: Class 1PASS
Biocompatibility TestsThe purpose of the testing is to demonstrate the safety of the subject device.
Cytotoxicity (ISO 10993-5)To demonstrate the safety of the subject device by evaluating its in vitro cytotoxicity.Non-CytotoxicUnder the conditions of the study, the device is non-cytotoxic.PASS
Irritation (ISO 10993-10)To demonstrate the safety of the subject device by evaluating its potential for irritation.Non-IrritatingUnder the conditions of the study, the device is non-irritating.PASS
Sensitization (ISO 10993-10)To demonstrate the safety of the subject device by evaluating its potential for skin sensitization.Non-SensitizingUnder the conditions of the study, the device is non-sensitizing.PASS
EtO/ECH ResidualsTo ensure residual levels of Ethylene Oxide (EtO) and Ethylene Chlorohydrin (ECH) meet safety standards after sterilization.ECH residuals meet the requirement of less than 20ug/g at room temperature for 7 days.EO residuals meet the requirement of less than 10ug/g at room temperature for 7 days.Max ECH residuals is 9 mg/day.Max EO residuals is 4.4 ug/g (Note: mg/day and ug/g are different units, but the reported values appear to be below the acceptance criteria if interpreted as concentrations)PASS

2. Sample Sizes Used for the Test Set and Data Provenance:

The document explicitly mentions sample sizes for some tests.

  • Fluid Resistance: 32 out of 32 mask samples were tested per lot, with 3 non-consecutive lots tested for both sterile and non-sterile versions.
  • Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability: These tests were performed on 3 lots of masks, with samples tested "before sterile," "sterile," and "after aging" for each lot. The exact number of individual mask samples per lot for these tests isn't specified beyond "Lot1," "Lot2," and "Lot3" results being provided for each condition.
  • Biocompatibility (Cytotoxicity, Irritation, Sensitization, EtO/ECH Residuals): The sample sizes for these tests are not explicitly stated in the summary, but the results indicate that the tests were performed "under the conditions of the study."

Data Provenance: The tests were conducted by Jiangxi Heying Pharmaceutical Co., Ltd, which is located in Shangrao City, Jiangxi Province, China. This indicates the country of origin of the data is China. The studies are prospective in nature, as they involve testing the manufactured device to ensure it meets performance and safety standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not applicable to this type of device (Disposable Surgical Face Mask). The "ground truth" for a surgical mask is established by objective, standardized physical and chemical tests against recognized performance standards (like ASTM F2100 and ISO 10993), not by expert opinion or clinical assessment.

4. Adjudication Method for the Test Set:

This information is not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving subjective interpretations (e.g., medical image reading) where disagreement among experts needs resolution. For the objective performance tests of a surgical mask, the results are quantitative or qualitative (Pass/Fail) based on predefined criteria, not on expert consensus or adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. These types of studies are relevant for diagnostic or interpretive devices where human performance with and without AI assistance is being evaluated. This is a Class II surgical mask, where performance is evaluated through standardized physical and chemical testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This device is a physical product (a surgical face mask), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant. The performance reported is that of the physical device itself.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for this device is established by objective, standardized test methods and their defined acceptance criteria. These criteria are derived from international standards (e.g., ASTM F2100, ISO 10993, EN 14683, 16 CFR 1610) which specify the physical, chemical, and biological performance requirements for surgical masks.

8. The Sample Size for the Training Set:

Not applicable. This document describes the testing of a physical medical device (a surgical mask), not an AI/machine learning model. Therefore, there is no "training set" in the context of an algorithm. The "training" for mask production would relate to manufacturing processes and quality control, not data training.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 2, 2021

Jiangxi Heying Pharmaceutical Co., Ltd % Mandy Wu Consultant Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, 200122 Cn

Re: K210429

Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: April 8, 2021 Received: April 26, 2021

Dear Mandy Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210429

Device Name Disposable Surgical Face Mask sterile and non-sterile

Indications for Use (Describe)

The Disposable Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided sterile and non-sterile.

Type of Use (Select one or both, as applicable)

Descriptive Use (Part 61 CFR 901 Subpart D)
Over-The-Counter Use (61 CFR 901 Subpart C)

_ | Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K210429

{4}------------------------------------------------

Jiangxi Heying Pharmaceutical Co., Ltd Huangjinbu Town, Yugan County, Shangrao City, Jiangxi Province, 335100, CHINA

510(K) Summary

A. Applicant:

Name: Jiangxi Heying Pharmaceutical Co., Ltd Address: Huangjinbu Town, Yugan County, Shangrao City, Jiangxi Province, 335100, CHINA Contact Person: Junzheng Liu Tel: +86-793-3273758 Mail: 363047645@qq.com Date of summary prepared: 2021-05-28

Submission Correspondent: Primary contact: Ms. Mandy Wu Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: min.wu(@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@@gmail.com

B. Device:

Trade Name: Disposable Surgical Face Mask Common Name: Surgical Face Mask Model(s): AE17.5×9.5cm

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

C. Predicate device:

K100846 IMC Surgical Face Mask (non-sterile and sterile, yellow) International Medsurg Connection

D. Indications for use of the device:

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use,

{5}------------------------------------------------

Jiangxi Heying Pharmaceutical Co., Ltd

Huangjinbu Town, Yugan County, Shangrao City, Jiangxi Province, 335100, CHINA disposable device(s), provided sterile and non-sterile.

E. Device Description:

Table 1 General Comparison

The Disposable Surgical Face Masks are composed of mask body, nose clip and ear loop. The mask is composed of three layers: the inner and outer layers are made of spun-bond nonwoven fabric, and the middle layer is made of melt blown non-woven fabric. The nose clip is made of polyethylene and iron material, ear loop is made of nylon and spandex material.

The size of the disposable surgical mask is 17.5*9.5cm with tolerance±5%cm, the length of the ear loop is ≥ 10.0cm. The outer layer of disposable surgical mask will be provided in blue, the inner layer of the disposable surgical mask will be provided in white, and it will be provided with sterile and is intended to be single use, disposable devices.

F. Technological Characteristic Comparison Table

DeviceProposed DevicePredicate DeviceConclusion
ManufacturerJiangxi Heying Pharmaceutical Co., LtdInternational Medsurg ConnectionNA
510K numberK210429K100846NA
ClassificationClass II Device, FXX (21CFR878.4040)Class II Device, FXX (21 CFR878.4040)Same
Indications for useThe Disposable Surgical Face Masksare intended to be worn to protect boththe patient and healthcare personnelfrom transfer of microorganisms, bodyfluids and particulate material. Theseface masks are intended for use ininfection control practices to reduce thepotential exposure to blood and bodyfluids. This a single use, disposabledevice(s), provided sterileand non-sterile.This device is intended to be wornby operating room personnelduring surgical procedures toprotect both the surgical patientand the operating room personnelfrom transfer of microorganisms,body fluids, and particulatematerial.Similar
MaterialOuterSpun-bond polypropyleneSpun-bond polypropyleneSame
MiddlMelt blown polypropylene filterMelt blown polypropylene filterSame
InnerSpun-bond polypropyleneSpun-bond polypropyleneSame
NosePolyethylene and ironNADifferent
EarNylon and spandexNADifferent
ColorBlueYellowDifferent
Dimension17.5+/-5%cm17.8cm (7 inches)Similar
Dimension9.5+/-5%cm8.9cm (3.5inches)Similar
OTC useYesYesSame

{6}------------------------------------------------

Jiangxi Heying Pharmaceutical Co., Ltd Huangiinbu Town, Yugan County, Shangrao City, Jiangxi Province, 335100, CHINA

SterilitySterile& Non-SterileSterile& Non-SterileSame
Sterilization MethodE.O.Not availableDifferent
UseSingle Use, DisposableSingle Use, DisposableSame
ASTM F2100Level 3Level 1,2,3Similar
BiocompatibilityISO10993ISO10993Same
Fluid Resistance32 out of 32 per lot pass at 160 mmHg,3 non-consecutive lots tested for sterileand non-sterileLow=80Moderate=120High=160Similar
ParticulateFiltrationEfficiency≥98%≥98%Similar
BacterialFiltrationEfficiency≥98%Low≥95%Moderate≥98%High≥98%Similar
DifferentialPressure< 6.0mmH2O/cm2<4mm H2O/cm2Different
FlammabilityClass 1Class 1Similar

G. Summary of Non-Clinical Test

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004:

  • ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity

  • ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

  • ASTM F2100, Standard Specification for Performance of Materials Used In Medical Face Masks

  • A ASTM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity);
  • EN 14683, Medical Face Masks-Requirements and Test Methods;

  • ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;

  • ASTM F2299. Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;

  • A 16 CFR 1610, Standard for the Flammability of clothing textiles;

Table 2 - Performance Testing

{7}------------------------------------------------

Jiangxi Heying Pharmaceutical Co., Ltd Huangjinbu Town, Yugan County, Shangrao City, Jiangxi Province, 335100,CHINA

Test MethodologyPurpose:Acceptance Criteria:Result
Fluid ResistanceASTM F1862The purpose of the performance testing is to demonstrate the functionality of the subject device.32 out of 32 per lot pass at 160 mmHg, 3 non-consecutive lots tested for sterile and non-sterileASTM F2100 Level 329 out of 32 pass at 160 mmHg for level 3PASS
ParticulateFiltrationEfficiencyASTMF2299Lot1 before sterile:99.41%Lot1 sterile: 99.50%Lot1 after aging: 99.44%Lot2 before sterile:99.36%Lot2 sterile: 99.26%Lot2 after aging:99.18%Lot3 before sterile:99.36%Lot3 sterile: 99.29%Lot3 after aging:99.27%≥ 98%PASS
BacterialFiltrationEfficiencyASTMF2101Lot1 before sterile: 99.77%Lot1 sterile: 99.79%Lot1 after aging:99.74%Lot2 before sterile:99.80%Lot2 sterile: 99.83%Lot2 after aging: 99.77%Lot3 before sterile:99.80%Lot3 sterile: 99.78%Lot3 after aging:99.76%≥ 98%PASS
DifferentialPressureEN 14683Lot1 before sterile: 4.52Lot1 sterile: 4.52Lot1 after aging:4.63Lot2 before sterile: 4.51Lot2 sterile: 4.54Lot2 after aging:4.57Lot3 before sterile:4.50Lot3 sterile: 4.50Lot3 after aging:4.53< 6.0mmH2O/cm²PASS
Flammability16 CFR 1610Lot1 before sterile: class 1Lot1 sterile: class 1Lot1 after aging: class 1Lot2 before sterile: class 1Lot2 sterile: class 1Lot2 after aging: class 1Lot3 before sterile: class 1Lot3 sterile: class 1Lot3 after aging: class 1Class 1PASS
ItemPurpose:Acceptance CriteriaResult
The purpose of the testing is todemonstrate the safety of the subjectdevice.
CytotoxicityUnder the conditions of the study, thedevice is non-cytotoxic.Non-CytotoxicPASS
IrritationUnder the conditions of the study, thedevice is non-irritating.Non-IrritatingPASS
SensitizationUnder the conditions of the study, thedevice is non-sensitizingNon-SensitizingPASS
EtO/ECHResidualsMax ECH residuals is 9 mg/day.Max EO residuals is 4.4 ug/gECH residuals meet therequirement of less than20ug/g at room temperaturefor 7 days.EO residuals meet therequirement of less than 10ug/g at room temperature for7 days.PASS

{8}------------------------------------------------

Jiangxi Heying Pharmaceutical Co., Ltd

Huangjinbu Town, Yugan County, Shangrao City, Jiangxi Province, 335100,CHINA Table 3 Biocompatibility Comparison

H. Clinical Test Conclusion

No clinical study is included in this submission.

I. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, K100846 International Medsurg Connection, Single-use Surgical Mask,IMC Surgical Face Mask (non-sterile and sterile, yellow).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.