K Number
K210279

Validate with FDA (Live)

Device Name
Enbio S
Manufacturer
Date Cleared
2021-10-29

(270 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.

Device Description

The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It has a hermetically sealed, heated chamber made from aluminum, with two heaters sink inside chamber walls. Inside this chamber, the sterilized load is placed on a special perforated tray. After closing the chamber, the user selects the appropriate sterilization program through the TFT touch screen. The actual sterilization phase starts after the pre-vacuum phase. The aluminum steam generator produces superheated steam and applies it inside the chamber. That steam penetrates the sterilized instruments. The set temperature is maintained inside the chamber depending on the selected sterilization cycle (121C° or 134°C), during a specified time (30 minutes or 4 minutes). After that time all the steam accumulated inside the chamber is pumped out and the drying cycle begins. When sterilization is finished, device displays to the user that process is completed, and that the load is sterile.

AI/ML Overview

This document is a 510(k) summary for the Enbio S steam sterilizer. It outlines the acceptance criteria and the non-clinical bench testing conducted to prove that the device meets these criteria. Since this is a sterilizer, the relevant acceptance criteria and studies revolve around its ability to achieve sterilization.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodPurposeAcceptance CriteriaReported Device Performance
ANSI AAMI ST55:2016 Vacuum TestVerify air removal performanceAverage leak rate of 1 millimeter of mercury (mmHg) (0.13 kPa) (0.019 psia) per min or lessPass
Bowie Dick TestVerify air removal performanceThe Bowie-Dick test indicator sheet shall show a uniform color change; i.e., the color in the center should be the same as that at the outer edges.Pass
ANSI AAMI ST55:2016 Empty Chamber Study (134°C in 4 min & 121°C in 30 min)Verify pressure and temperature – to ensure that the sterilizer is capable of providing steady-state thermal and pressure conditions within the chamber+3°C/-0°C, ±0.3 barPass
ANSI AAMI ST55:2016 Full Cycle Study (134°C in 4 min & 121°C in 30 min)Verify pressure and temperature - to ensure that the sterilizer is capable of providing steady-state thermal and pressure conditions+3°C/-0°C, ±0.3 barPass
ANSI AAMI ST55:2016 Half Cycle Study (134°C in 4 min & 121°C in 30 min)Verify pressure and temperature – to ensure that the sterilizer is capable of providing steady-state thermal and pressure conditions during the cycle+3°C/-0°C, ±0.3 barPass
ANSI AAMI ST55:2016 Full Cycle Biological Indicators (134°C in 4 min)To verify biological sterilization performanceThe tested cycle has a 10^-6 SAL (Sterility Assurance Level)Pass
ANSI AAMI ST55:2016 Full Cycle Biological Indicators (121°C in 30 min)To verify biological sterilization performanceThe tested cycle has a 10^-6 SAL (Sterility Assurance Level)Pass
ANSI AAMI ST55:2016 Half Cycle Biological Indicators (134°C in 2 min)To ensure the efficacy of the equipment and the lethality of the recommended processing parameters by biological challengeThe tested cycle has a 10^-6 SAL (Sterility Assurance Level)Pass
ANSI AAMI ST55:2016 Half Cycle Biological Indicators (121°C in 15 min)To ensure the efficacy of the equipment and the lethality of the recommended processing parameters by biological challengeThe tested cycle has a 10^-6 SAL (Sterility Assurance Level)Pass
ANSI AAMI ST55:2016 Half Cycle Study / Textile PCD (134°C in 2 min)Verify half-cycle sterilization of BI inside textile PCDThe tested cycle has a 10^-6 SAL (Sterility Assurance Level)Pass
ANSI AAMI ST55:2016 Half Cycle Study / Textile PCD (121°C in 15 min)Verify half-cycle sterilization of BI inside textile PCDThe tested cycle has a 10^-6 SAL (Sterility Assurance Level)Pass
IEC 61010-1, 61010-2Verify electrical safetyMeets specifications of standardPass
IEC 60601-1-2, IEC 61326-1, 47 CFR 15 Subpart BVerify electromagnetic compatibilityMeets specifications of standardsPass

2. Sample Size Used for the Test Set and Data Provenance:

The document describes non-clinical bench (performance) testing, not a test set in the sense of patient data or algorithm performance. Therefore, typical "sample size" and "data provenance" (country of origin, retrospective/prospective) don't directly apply here.

The tests conducted are physical performance tests on the sterilizer as a device. The "sample size" would refer to the number of sterilization cycles performed or units tested to demonstrate consistency and adherence to parameters. This specific detail is not provided in text, but it's implied that sufficient tests were run to determine a "Pass" for each criterion.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not applicable for the given document. The "ground truth" for a steam sterilizer's performance is established by well-defined physical and biological standards (e.g., temperature/pressure tolerances, biological indicator kill rates). There isn't a need for human expert consensus in the same way there would be for a diagnostic imaging AI, for example.

4. Adjudication Method for the Test Set:

This information is not applicable. Adjudication methods like 2+1 or 3+1 are used for expert review of human-interpreted data (e.g., medical images). The tests performed on the sterilizer yield objective, measurable results (e.g., temperature readings, biological indicator color changes, leak rates) which do not require human adjudication in this context.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The submission does not contain any data from clinical testing." MRMC studies are typically for evaluating the impact of an AI on human reader performance for diagnostic tasks. This device is a mechanical sterilizer, not a diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, in a sense, a "standalone" evaluation was done. The tests performed are directly on the performance of the sterilizer itself, without human interaction impacting the sterilization efficacy. For example, the biological indicator tests verify the sterilizer's ability to kill microorganisms on its own. While a human initiates the cycle, the effectiveness of the sterilization process is tested independently.

7. The Type of Ground Truth Used:

The ground truth used for these tests is based on established scientific and regulatory standards for sterilization, specifically:

  • Physical Parameters: Defined temperature and pressure ranges, leak rates, and cycle times as specified in ANSI AAMI ST55:2016.
  • Biological Efficacy: The 10^-6 Sterility Assurance Level (SAL), verified using biological indicators (BIs) which contain highly resistant bacterial spores (e.g., Geobacillus stearothermophilus). A successful sterilization run means the BIs show no growth, indicating a 1 in a million chance of a non-sterile item.
  • Electrical Safety & EMC: Compliance with international safety and electromagnetic compatibility standards (IEC, CFR).

8. The Sample Size for the Training Set:

This information is not applicable. This device is a traditional medical device (steam sterilizer), not an AI/Machine Learning algorithm. Therefore, there is no "training set" in the context of data used to train an algorithm.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 29, 2021

Enbio Group AG Lukasz Rogowski Corporate Quality Manager Eichengasse 3 Oensingen, CH-4702 Switzerland

Re: K210279

Trade/Device Name: Enbio S Regulation Number: 21 CFR 880.6880 Regulation Name: Steam sterilizer Regulatory Class: Class II Product Code: FLE Dated: September 24, 2021 Received: September 27, 2021

Dear Lukasz Rogowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Assistant Director THT4B1:Sterility Devices Team DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K210279

Device Name Enbio S

Indications for Use (Describe)

The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.

Please refer to the table below for program name, load description, sterilization temperature, exposure time, drying time and maximum load.

ProgramNameLoad DescriptionSterilizationTemperatureSterilizationTimeDryingTimeMaximumLoad
134°Csolid objects, small porousobjects, simple objectsrecessed, narrow-clearanceitems, dental handpieces,and textiles; wrapped andunwrapped134°C (273°F)4 minutes3 minutes0.5 Kg/1.1 lbs
121°Csolid objects, small porousobjects, simple objectsrecessed, narrow-clearanceitems, dental handpieces,textiles, and plastics;wrapped and unwrapped121°C (250°F)30 minutes5 minutes0.5 Kg/1.1 lbs

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K210279

Sponsor/ Applicant 1.

Enbio Group AG Eichengasse 3 CH-4702 Oensingen, Switzerland

Mr. Lukasz Rogowski Corporate Quality Manager Email: lukasz.rogowski@enbio.com Phone: +48 605 058 629

Summary Preparation Date: October 29, 2021

2. Device

Trade NameEnbio S
ClassificationClass 2
Classification NameSteam Sterilizer
Product CodeFLE
Regulation Number21 CFR 880.6880
Review PanelGeneral Hospital

3. Predicate Device

Statclave G4 Chamber Autoclave (K190062)

4. Device Description

The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It has a hermetically sealed, heated chamber made from aluminum, with two heaters sink inside chamber walls. Inside this chamber, the sterilized load is placed on a special perforated tray. After closing the chamber, the user selects the appropriate sterilization program through the TFT touch screen.

{4}------------------------------------------------

The actual sterilization phase starts after the pre-vacuum phase. The aluminum steam generator produces superheated steam and applies it inside the chamber. That steam penetrates the sterilized instruments. The set temperature is maintained inside the chamber depending on the selected sterilization cycle (121C° or 134°C), during a specified time (30 minutes or 4 minutes). After that time all the steam accumulated inside the chamber is pumped out and the drying cycle begins. When sterilization is finished, device displays to the user that process is completed, and that the load is sterile.

5. Indications for Use

The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.

Please refer to the table below for program name, load description, sterilization temperature, exposure time, drying time and maximum load.

ProgramNameLoad DescriptionSterilizationTemperatureSterilizationTimeDryingTimeMaximumLoad
134°Csolid objects, small porousobjects, simple objectsrecessed, narrow-clearanceitems, dental handpieces,and textiles; wrapped andunwrapped134°C (273°F)4 minutes3 minutes0.5 Kg/1.1 lbs
121°Csolid objects, small porousobjects, simple objectsrecessed, narrow-clearanceitems, dental handpieces,textiles, and plastics;wrapped and unwrapped121°C (250°F)30 minutes5 minutes0.5 Kg/1.1 lbs

{5}------------------------------------------------

Technological Characteristics Comparision Table 6.

Provided below is a technological comparison of the subject device with the predicate device.

Subject DevicePredicate DeviceComparison
Trade NameEnbio SSTATCLAVE G4 ChamberAutoclave (K190062)
SubmitterEnbio Group AGSciCan Ltd.-
Product CodeFLEFLESame
Regulation Number21 CFR 880.688021 CFR 880.6880Same
Device ClassClass 2Class 2Same
Prescription / Over-The-Counter UseOver-The-CounterOver-The-CounterSame
Intended UseThe Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.The STATCLAVE G4 is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.Same
Water tankExternalInternal reservoirDifferent
Sterilization Chamber Volume2.7 L26 LDifferent
Sterilization Chamber Dimensions292 x 192 x 45 mm(L x W x H)Diameter: 280 mmDepth: 381 mmDifferent
Device Dimensions (L x W x H)561 x 252 x 162 mm635 x 450 x 495 mmDifferent
Weight15 kg (approximately)62.7kg (approximately)Different
Power Rating110-120 V, 60Hz, 15A120V, 60Hz, 12 ASimilar
WirelessTransmissionCapabilityNoYesDifferent
USB PortYesYesSame
Sterility and Shelf-lifeNot provided sterile.No shelf-life claimedNot provided sterile.No shelf-life claimedSame

{6}------------------------------------------------

Non-clinical Bench (Performance) testing 7.

Provided below is the non-clinical testing to demonstrated the subject device meets the specification and accpetance criteria found in the Standards and test method listed below.

Test MethodPurposeAcceptance criteriaResult
ANSI AAMI ST55:2016Vacuum TestVerify air removalperformanceAverage leak rate of 1millimeter of mercury(mmHg) (0.13 kPa) (0.019psia) per min or lessPass
Bowie Dick TestVerify air removalperformanceThe Bowie-Dick testindicator sheet shall showa uniform color change;i.e., the color in the centershould be the same asthat at the outer edges.Pass
ANSI AAMI ST55:2016Empty Chamber Study• 134°C (273°F) in 4 minutesVerify pressure andtemperature – to ensurethat the sterilizer iscapable of providingsteady-state thermaland pressure conditionswithin the chamber+3°C/-0°C±0.3 barPass
ANSI AAMI ST55:2016Empty Chamber Study• 121°C (250°F) in 30 minutesVerify pressure andtemperature - to ensurethat the sterilizer iscapable of providingsteady-state thermaland pressure conditionswithin the chamberPass
ANSI AAMI ST55:2016Full Cycle Study• 134°C (273°F) in 4minutesVerify pressure andtemperature - to ensurethat the sterilizer iscapable of providingsteady-state thermaland pressure conditions+3°C/-0°C±0.3 barPass
ANSI AAMI ST55:2016Full Cycle Study• 121°C (250°F) in 30minutesVerify pressure andtemperature - to ensurethat the sterilizer iscapable of providingsteady-state thermaland pressure conditionsduring the cyclePass
ANSI AAMI ST55:2016Half Cycle Study• 134°C (273°F) in 4minutesVerify pressure andtemperature – to ensurethat the sterilizer iscapable of providing+3°C/-0°C±0.3 barPass
ANSI AAMI ST55:2016Half Cycle Study• 121°C (250°F) in 30 minutessteady-state thermaland pressure conditionsduring the cyclePass
ANSI AAMI ST55:2016Full Cycle BiologicalIndicators• 134°C (273°F) in 4 minutesTo verify biologicalsterilization performanceThe tested cycle has a10-6 SALPass
ANSI AAMI ST55:2016Full Cycle BiologicalIndicators• 121°C (250°F) in 30 minutesPass
ANSI AAMI ST55:2016Half Cycle BiologicalIndicators• 134°C (273°F) in 2 minutesTo ensure the efficacy ofthe equipment and thelethality of therecommendedprocessing parametersby biological challengeThe tested cycle has a10-6 SALPass
ANSI AAMI ST55:2016Half Cycle BiologicalIndicators• 121°C (250°F) in 15 minutesPass
ANSI AAMI ST55:2016Half Cycle Study /Textile PCD• 134°C (273°F) in 2 minutesVerify half-cyclesterilization of BI insidetextile PCDThe tested cycle has a10-6 SALPass
ANSI AAMI ST55:2016Half Cycle Study /Textile PCD• 121°C (250°F) in 15 minutesPass
IEC 61010-1, 61010-2Verify electrical safetyMeets specifications ofstandardPass
IEC 60601-1-2,IEC 61326-1,47 CFR 15 Subpart B,Verify electromagneticcompatibilityMeets specifications ofstandardsPass

{7}------------------------------------------------

8. Clinical Testing

The submission does not contain any data from clinical testing.

9. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the Enbio S is as safe, as effective, and performs as well as or better than the legally marketed predicate device, STATCLAVE G4 Chamber Autoclave (K190062).

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).