AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

GPC Variable Angle Locking Two Column Volar Distal Radius Plating System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones. The system is indicated for use in adult patients only. All implants are for single use only.

Device Description

GPC Variable Angle Locking Two Column Volar Distal Radius Plates are machined metal plates designed for fixation of various fracture modes of the distal portion of the Variable Angle Locking Two Column Volar Distal Radius Plates allow fragment-specific fracture fixation by providing the flexibility to lock screws in the head or shaft of the plate in trajectories that can diverge from the central axis of the plate hole. The variable angle locking mechanism improves the plate-screw construct stability. GPC Variable Angle Locking Two Column Volar Distal Radius Plates are manufactured using titanium material for surgical implant applications.

Following are the types of the GPC Variable Angle Locking Two Column Volar Distal Radius Plates:

  • Variable Angle Locking Two-Column Volar Distal Radius Plate, Width 22 mm, 6 Head Holes (with option of Left and Right Variants)
  • Variable Angle Locking Two-Column Volar Distal Radius Plate, Width 25.5 mm, 7 Head Holes (with option of left and right variants)
  • Variable Angle Locking Two Column Volar Distal Radius Plate- Narrow, Width 19.5 mm, 6 Head Holes (with option of left and right variants)

Corresponding Screws for fixation:

  • Variable Angle Locking Screw 2.4 mm, Star drive Lengths: 6mm to 30mm in 2mm increments
  • FixLOCK Self-Tapping Screw 2.4 mm, Star drive Length 6mm to 30mm in 2mm increments
  • Self-Tapping Cortex Screw 2.7 mm, Star drive Length: 12mm to 24mm

These implants are supplied non-sterile, the products have to be sterilized prior to use.

AI/ML Overview

The provided document describes a 510(k) premarket notification for the GPC 2.4 mm Variable Angle Locking Two Column Volar Distal Radius Plates. This notification aims to demonstrate substantial equivalence to legally marketed predicate devices, rather than establishing de novo safety and effectiveness. Therefore, the information provided focuses on the comparison to the predicate device and non-clinical testing against established standards, not on a study proving the device meets specific acceptance criteria in a clinical context with human users or AI.

However, based on the provided text, we can extract details related to the "device performance" as demonstrated through non-clinical testing.

Here's an analysis based on the information provided, framed as if the non-clinical tests are the acceptance criteria for substantial equivalence:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" with numerical targets for the mechanical tests. Instead, it states that the device "Conforms" to the specified ASTM standards, implying successful performance according to the methodology and requirements of those standards.

Acceptance Criteria (Implied by standard conformance)Reported Device Performance
For Plates (ASTM F382):
Static Four Point Bend TestConforms
Bending Fatigue PropertiesConforms
For Bone Screws (ASTM F543):
Torsional PropertiesConforms
Driving TorqueConforms
Pull-out TestConforms
General Characteristics compared to predicate:
Regulatory ClassificationEquivalent (Same Classification and Regulation Number)
Indications for useEquivalent (Similar intended use)
MaterialEquivalent (Both use material for surgical implant application)
Performance StandardsEquivalent (Same performance standards)
SterilizationEquivalent (Same method of sterilization)
Dimensional VerificationEquivalent (Similar dimensions, minor differences raise no safety concerns)

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify the sample size (number of plates or screws) used for each mechanical test.
  • Data Provenance: The tests are non-clinical, conducted by the manufacturer, GPC Medical Ltd. The country of origin of the data would be India, where the manufacturer is located. The tests are prospective in the sense that they were conducted specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable to the provided context as the "ground truth" here is established by engineering standards and direct physical measurement of mechanical properties, not by human expert interpretation. There were no human readers or expert consensus involved in determining if the device "conforms" to a bending fatigue test.

4. Adjudication method for the test set

This question is not applicable. Adjudication methods like 2+1 or 3+1 are used for clinical studies with expert reviewers establishing ground truth, which is not the case here. The "adjudication" for mechanical tests involves comparing measured values against specified thresholds within the ASTM standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. The device is a physical medical implant (bone plate and screws), not an AI software/algorithm. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable, as the device is not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical performance tests, the "ground truth" is defined by the specifications and methodologies of the applicable ASTM standards (ASTM F382 for plates and ASTM F543 for screws). The device's performance is compared directly against these engineering standards. For the comparison of characteristics, the "ground truth" is the specifications and characteristics of the identified predicate devices.

8. The sample size for the training set

This question is not applicable. This is a medical implant, not a machine learning model, so there is no "training set."

9. How the ground truth for the training set was established

This question is not applicable, as the device is not a machine learning model.

{0}------------------------------------------------

January 7, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

GPC Medical Limited Vikas Narang Director GPC Square, M-Block, DDA LSC, Vikas Puri, New Delhi New Delhi, Delhi 110018 India

Re: K210198

Trade/Device Name: GPC 2.4 mm Variable Angle Locking Two Column Volar Distal Radius Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS. HWC Dated: December 6, 2021 Received: December 6, 2021

Dear Vikas Narang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210198

Device Name

Indications for Use (Describe)

Type of Use (Select one or both , as applicable)
----------------------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

C Medical Ltd

K210198 Page 1 of 4

Premarket Notification 510(k) Summary as required by Section 807.92

Submitter's Name:GPC Medical Ltd.
Address:Office : GPC Square, M-Block, DDA, LSC, Vikas Puri New Delhi,110018 INDIA.Factory : Plot C-3, Sec B-1, UPSIDC, Trans Delhi Signature City,Loni, Ghaziabad, UP 201102
Contact Person Name:Mr. Vikas Narang
Title:Director
Phone Number:+91-9810638797
Dated:25-05-2021

Proprietary Name:

  • Name of the Device: GPC Variable Angle Locking Two Column Volar Distal Radius Plating . System
    Common or Usual Name:

  • Orthopaedic Bone Plates

  • . Orthopaedic Bone Screws

Classification Name:

  • PLATE,FIXATION,BONE
  • SCREW,FIXATION, BONE .

Product Code: HRS (Primary), HWC

Device Class: Il

  • Review Pane: Orthopaedic
    Regulation Number: 21CFR888.3030 (Primary) and 21CFR888.3040

Device Details:

  • GPC Variable Angle Locking Two Column Volar Distal Radius Plates are manufactured in various ● shapes and sizes. The GPC Variable Angle Locking Two Column Volar Distal Radius Plates are fixed with corresponding sizes of locking screws.

{4}------------------------------------------------

cal Ltc

A3) Identification of the Predicate Device

Following are the predicate device 510(k) with which we are declaring substantial equivalence:

Primary predicate: Synthes (USA) 2.4mim VA-LCP Two-Column Narrow Volar Distal (K092556)

Additional Predicate: Synthes 2.4mm Variable Angle LCR Volar Rim Distal Radius System (K110125)

Reference Device: Synthes 2.4 mm VA-LCP Two-Column Volar Distal Radius Plate (K083694)

A4). A description of the device that is the subject of the premarket notification submission, such as might be found in the labeling or promotional material for the device

GPC Variable Angle Locking Two Column Volar Distal Radius Plates are machined metal plates designed for fixation of various fracture modes of the distal portion of the Variable Angle Locking Two Column Volar Distal Radius Plates allow fragment-specific fracture fixation by providing the flexibility to lock screws in the head or shaft of the plate in trajectories that can diverge from the central axis of the plate hole. The variable angle locking mechanism improves the plate-screw construct stability. GPC Variable Angle Locking Two Column Volar Distal Radius Plates are manufactured using titanium material for surgical implant applications.

Following are the types of the GPC Variable Angle Locking Two Column Volar Distal Radius Plates:

  • Variable Angle Locking Two-Column Volar Distal Radius Plate, Width 22 mm, 6 Head Holes (with option of ● Left and Right Variants)
  • Variable Angle Locking Two-Column Volar Distal Radius Plate, Width 25.5 mm, 7 Head Holes (with option of left and right variants)
  • . Variable Angle Locking Two Column Volar Distal Radius Plate- Narrow, Width 19.5 mm, 6 Head Holes (with option of left and right variants)

Corresponding Screws for fixation:

  • Variable Angle Locking Screw 2.4 mm, Star drive Lengths: 6mm to 30mm in 2mm increments
  • FixLOCK Self-Tapping Screw 2.4 mm, Star drive Length 6mm to 30mm in 2mm increments
  • Self-Tapping Cortex Screw 2.7 mm, Star drive Length: 12mm to 24mm

These implants are supplied non-sterile, the products have to be sterilized prior to use.

A5). Indications for Use:

GPC Variable Angle Locking Two Column Volar Distal Radius Plating System is provided non-sterile.

GPC Variable Angle Locking Two Column Volar Distal Radius Plating System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones. The system is indicated for use in adult patients only. All implants are for single use only.

{5}------------------------------------------------

C Medical Ltd

A6). Summary of Technological Characteristics as compared to the predicate devices:

A comparison between the GPC Variable Angle Locking Two Column Volar Distal Radius Plating System

and predicate devices has been performed which has resulted in demonstration of similarities in dimensional and performance criteria.

Following is the summary of parameters in which the comparison has been verified:

S. No.CharacteristicsPredicate Device Versus New Device (GPC Variable Angle Locking Two Column Volar Distal Radius Plates)Remarks on Equivalence
01Regulatory ClassificationSame Classification and Regulation NumberEquivalent
02Indications for useSimilar intended use in New Device and Predicate deviceEquivalent
03MaterialBoth the predicate and subject devices use material meant for surgical implant applicationEquivalent
04Performance StandardsSame performance standards used in both New Device as well as predicate deviceEquivalent
05SterilizationSame method of sterilization used in both New Device as well as Predicate deviceEquivalent
06Dimensional VerificationSimilar dimensions found in both New Device as well as Predicate device The minor differences that exist, do not raise any safety concerns.Equivalent

b1).Discussion on the non-clinical testing performed

The device performance of GPC Variable Angle Locking Two Column Volar Distal Radius Plating System has been demonstrated against following applicable standards

For GPC Variable Angle Locking Two Column Volar Distal Radius Plates:

  • The testing performed as per ASTMF382 ●
    Static Four Point Bend Test: Conforms Bending Fatique Properties: Conforms

For Bone Screws:

The Testing Performed As per ASTMF 543: Torsional Properties: Conforms,

{6}------------------------------------------------

C Medical Ltd

Driving Torque : Conforms, Pull-out Test: Conforms

b2). Discussion on the clinical evaluation referenced and relied up on:

GPC Variable Angle Locking Two Column Volar Distal Radius Plating System is of similar design and pattern as well as similar intended use as that of predicate devices which are in market already and in clinical use. Therefore, Clinical information was not necessary to demonstrate substantial equivalence.

CONCLUSION:

From the available data we can justify that the GPC Variable Angle Locking Two Column Volar Distal Radius Plating System is as safe, as effective and perform as same indications for use as that of already marketed predicate devices identified in a 3. of 510(k) summary.

Hence these devices can be considered safe and effective for their intended use.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.