K Number
K210110

Validate with FDA (Live)

Device Name
Guidion
Date Cleared
2021-03-31

(71 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Guidion catheters are intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement and exchange of guide wires and other interventional devices.

Device Description

The Guidion catheter is a single lumen rapid exchange catheter offered in sizes being compatible with 5F; 6F; 7F and 8F guide catheters and may be placed over an guide wire. The 150cm long device has a stainless steel shaft section. The stainless steel shaft is followed distally by a 25cm lumen section.

The Guidion catheter has a radiopaque distal end which enables visibility while using standard fluoroscopic methods. The device has two positioning marks located at 95cm and 105cm from the distal tip, respectively.

The Guidion catheter is delivered through a guiding catheter resulting in an inner diameter that is approximately 1 French smaller than the guide catheter. The Guidion catheter has a proximal hub which indicates guide catheter compatibility.

AI/ML Overview

The provided document is a 510(k) summary for the Guidion Catheter, a medical device. It does not describe an AI/ML powered device, nor does it detail a study involving AI for diagnostic assistance. Therefore, it does not contain the information required to answer the prompt regarding acceptance criteria and a study proving an AI device meets those criteria.

The document discusses the substantial equivalence of the Guidion Catheter to a predicate device based on performance and biocompatibility tests. It lists the types of tests performed (e.g., kink resistance, radiopacity, bond strength, biocompatibility), but it does not provide specific acceptance criteria or detailed results of these tests. It simply states that "The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues."

Therefore, I cannot provide information for the following points as they are not present in the given text:

  1. A table of acceptance criteria and the reported device performance
  2. Sample sized used for the test set and the data provenance
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
  4. Adjudication method for the test set
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
  7. The type of ground truth used
  8. The sample size for the training set
  9. How the ground truth for the training set was established

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 31, 2021

IMDS Operations B.V. Edwin Schulting CEO Ceintuurbaan Noord 150 Roden, Drenthe 9301 NZ Netherlands

Re: K210110

Trade/Device Name: Guidion Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: January 14, 2021 Received: January 19, 2021

Dear Edwin Schulting:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Interventional Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210110

Device Name Guidion Catheter

Indications for Use (Describe)

Guidion catheters are intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement and exchange of guide wires and other interventional devices.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 201 Subpart D)☐ Over-The-Counter Use (21 CFR 201 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

|| | Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K210110

Image /page/3/Picture/1 description: The image contains the logo for IMDS. The logo consists of a stylized atom symbol on the left, with a dark teal circle connected to a light teal chain of circles. To the right of the atom symbol is the text "IMDS", with the "I" and "DS" in dark teal and the "M" in light teal.

510(k) Summary

Date Prepared: March 22nd , 2021

Submitter's Name / Contact Person Manufacturer IMDS Operations BV Ceintuurbaan Noord 150 9301 NZ Roden . The Netherlands Establishment Registration #3007740583

Contact Person Florence Wagter

Director of Quality and Regulatory Tel: 0031651453880 Fax: 0031508200231

General Information

Trade Name Common/ Usual Name Classification Name Product Code Predicate Device Reference Device

Guidion Catheter Rapid Exchange Guide Extension Catheter Catheter, percutaneous DOY K172090, Guideliner V3 Catheter, Vascular Solutions, Inc. None

Device Description

The Guidion catheter is a single lumen rapid exchange catheter offered in sizes being compatible with 5F; 6F; 7F and 8F guide catheters and may be placed over an guide wire. The 150cm long device has a stainless steel shaft section. The stainless steel shaft is followed distally by a 25cm lumen section.

The Guidion catheter has a radiopaque distal end which enables visibility while using standard fluoroscopic methods. The device has two positioning marks located at 95cm and 105cm from the distal tip, respectively.

The Guidion catheter is delivered through a guiding catheter resulting in an inner diameter that is approximately 1 French smaller than the guide catheter. The Guidion catheter has a proximal hub which indicates guide catheter compatibility.

Intended Use

Guidion catheters are intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement and exchange of guide wires and other interventional devices.

K210110Page 1 of 3
------------------------

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for IMDS. The logo consists of a blue molecular structure on the left and the letters "IMDS" in black on the right. The molecular structure is made up of a black circle and several blue circles connected by lines.

Technological Characteristics Comparison

The Guidion is similar in design to the predicate device and both are Rapid Exchange Guide Extension Catheters intended to be used to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement and exchange of guide wires and other interventional devices.

#ItemGuidioncatheterGuideliner V3Catheter
Model number (s)G50F25150,G60F25150,G70F25150 &G80F251505569, 5571, 5572& 5573
1Type clinically basedGuide ExtensioncatheterGuide Extensioncatheter
2Design constructionRx, single lumenRx, single lumen
2Distal shaft materialStainless steel coilingembedded in apolymer shaftStainless steel coilingembedded in apolymer shaft
3Proximal shaftStainless steelStainless steel
4Distal radiopague markers11
5Radiopaque tip/ marker materialTPE/ TungstenPlatinum/ Iridium
6Effective length150 (cm)150 (cm)
7Length distal shaft25 (cm)25 (cm)
8Required guide catheter ID5F 1.42 (mm)6F 1.78 (mm)7F 1.98 (mm)8F 2.24 (mm)5F 1.42 (mm)6F 1.78 (mm)7F 1.98 (mm)8F 2.24 (mm)
9Tip design shapeStraightStraight
10Inner diameter distal shaft5F 1.04 (mm)6F 1.42 (mm)7F 1.57 (mm)8F 1.80 (mm)5F 1.17 (mm)6F 1.42 (mm)7F 1.57 (mm)8F 1.80 (mm)
11Product coatingHydrophilicSilicon

With the exception of

dimensional, material and package configuration differences, the Guidion is similar in design and technological characteristics to the predicate device. The dimensional, material and package configuration differences were successfully evaluated in performance tests.

K210110Page 2 of 3
------------------------

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for IMDS. The logo consists of a blue molecular structure on the left and the text "IMDS" on the right. The "I" is dark green, while the "MDS" is blue.

Substantial Equivalence and Summary of Studies

The technological differences between the subject and predicate devices have been evaluated through performance and biocompatibility tests and results did not raise new questions of safety or effectiveness. The Guidion Rapid Exchange Guide Extension catheter is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use.

The device design has been verified through the following tests:

    1. Kind resistance/ flexibility
    1. Radiopacity
    1. Catheter bond strength
    1. Crossing profile
    1. Catheter preparation, deployment & retraction
    1. Hub bond strength
    1. Proximal shaft depth marker position
    1. Torque strength
    1. Effective length
    1. Length distal shaft
    1. Inner diameter
    1. Coating Particulate Evaluation
    1. Coating integrity
    1. Packaging integrity

Device samples passed the following biocompatibility tests performed in accordance with ISO 10993-1:

  • . Cytotoxicity
  • Sensitization
  • Irritation .
  • Systemic toxicity ●
  • Pyrogenicity ●
  • Hemocompatibility

The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the Guidion Rapid Exchange Guide Extension catheter is substantially equivalent to the predicate device.

K210110Page 3 of 6
------------------------

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).