K Number
K210088

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-04-13

(90 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

1688 4K Camera System with Advanced Imaging Modality (AIM):
The 1688 Video Camera is indicated for use in general laparoscopy, ear endoscopy, sinuscopy, sinuscopy and plastic surgery whenever a laparoscope/ arthroscope is indicated for use. The 1688 Video Camera is indicated for use in adults and pediatric patients aged one month or older.
A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the 1688 Video Camera are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.
L11 LED Light Source with Advanced Imaging Modality (AIM) and SafeLight Cable:
Upon intravenous administration of SPY AGENTTM GREEN (indocyanine green for injection, USP), the L11 LED Light Source with AIM and SafeLightTM Cable is used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L11 LED Light Source with AIM and SafeLight Cable enable surgeons to perform minimally invasive surgery using endoscope visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month or older, and visualization of at least one of the major extra-hepatic bile ducts (cystic duct common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging.
Fluorescence imaging of biliary ducts with the L11 LED Light Source with AIM and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Upon interstitial administration of SPY AGENT GREEN, the L11 LED Light Source with AIM and SafeLight Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
The L11 LED Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.
Advanced Imaging Modality (AIM) HD Autoclavable Laparoscope:
The AIM HD Autoclavable Laparoscope is intended to be used for gynecological and general procedures that clinicians deem appropriate for adults and pediatric patients aged one month or older, when the dimensions of the AIM HD Autoclavable Laparoscope are appropriate for the patient size and anatomy.

Device Description

The AIM (Advanced Imaging Modality) System is an endoscopic real-time 4K visible white light and near-infrared light illumination and imaging system. The AIM (Advanced Imaging Modality) System includes the following components: (1) A Camera System for processing near-infrared and visible light images; (2) A Light Source and SafeLight Cable for emitting light within the visible light as well as near-infrared light spectrum; (3) A Laparoscope for visible light and near-infrared light illumination and imaging; (4) SPY AGENT GREEN (indocyanine green for injection, USP) an optical imaging agent used for fluorescence imaging.

AI/ML Overview

The provided text is a 510(k) summary for the Stryker AIM (Advanced Imaging Modality) System, detailing its substantial equivalence to predicate devices. While it lists indications for use, technological characteristics, and non-clinical testing performed, it does not contain the specific information required to answer your request about acceptance criteria and a study proving the device meets those criteria.

Specifically, the document lacks:

  • A table of acceptance criteria and reported device performance.
  • Details on sample size, data provenance (e.g., country, retrospective/prospective) for a test set. The clinical performance data section only mentions "Published clinical literature was provided," which is not a description of a specific study and its test set.
  • Number of experts and their qualifications for ground truth establishment.
  • Adjudication method for a test set.
  • Information on a multi-reader multi-case (MRMC) comparative effectiveness study or its effect size.
  • Information on standalone (algorithm-only) performance.
  • Specific type of ground truth used (expert consensus, pathology, outcomes data).
  • Sample size for the training set.
  • How ground truth for the training set was established.

The document primarily focuses on demonstrating substantial equivalence through comparison of technical specifications and indications for use with predicate devices, along with standard non-clinical testing (biocompatibility, cleaning/sterilization validation) and reference to published clinical literature, rather than detailing a specific performance study against pre-defined acceptance criteria for a new AI/imaging modality.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

Stryker Jessie Duong Manager, Regulatory Affairs 5900 Optical Court San Jose. California 95138

April 13, 2021

Re: K210088

Trade/Device Name: AIM (Advanced Imaging Modality) System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, OWN, FCS, FCW Dated: March 17, 2021 Received: March 18, 2021

Dear Jessie Duong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@;tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/6 description: The image shows the name "Neil R.P." in a large, bold, sans-serif font. The text is black and stands out against a background that appears to be a light blue or white. The letters are clearly legible and well-spaced.

Ogden

Digitally signed by Neil R.P. Ogden Date: 2021.04.13 11:30:32 -04'00'

Neil R.P. Ogden Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210088

Device Name AIM (Advanced Imaging Modality) System

Indications for Use (Describe)

1688 4K Camera System with Advanced Imaging Modality (AIM):

The 1688 Video Camera is indicated for use in general laparoscopy, ear endoscopy, sinuscopy, sinuscopy and plastic surgery whenever a laparoscope/ arthroscope is indicated for use. The 1688 Video Camera is indicated for use in adults and pediatric patients aged one month or older.

A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.

The users of the 1688 Video Camera are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

L11 LED Light Source with Advanced Imaging Modality (AIM) and SafeLight Cable:

Upon intravenous administration of SPY AGENTTM GREEN (indocyanine green for injection, USP), the L11 LED Light Source with AIM and SafeLightTM Cable is used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L11 LED Light Source with AIM and SafeLight Cable enable surgeons to perform minimally invasive surgery using endoscope visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month or older, and visualization of at least one of the major extra-hepatic bile ducts (cystic duct common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging.

Fluorescence imaging of biliary ducts with the L11 LED Light Source with AIM and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Upon interstitial administration of SPY AGENT GREEN, the L11 LED Light Source with AIM and SafeLight Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

The L11 LED Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

Advanced Imaging Modality (AIM) HD Autoclavable Laparoscope:

The AIM HD Autoclavable Laparoscope is intended to be used for gynecological and general procedures that clinicians deem appropriate for adults and pediatric patients aged one month or older, when the dimensions of the AIM HD Autoclavable Laparoscope are appropriate for the patient size and anatomy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

This 510(k) summary is prepared in accordance with the requirements of 21 C.F.R Part 807.92.

Submitter:

Applicant:Stryker Endoscopy5900 Optical CourtSan Jose, CA 95138
Contact Person:Jessie DuongManager, Regulatory AffairsPhone: (408) 754-2077Email: jessie.duong@stryker.com
Date Prepared:April 8, 2021

Subject Device:

The subject device is the AIM (Advanced Imaging Modality) System specifically the following system components:

Name of Device:1688 4K Camera System with Advanced Imaging Modality (AIM)
Common orUsual Name3-chip Video Camera
ClassificationName:Laparoscope, General and Plastic Surgery (21 C.F.R. §876.1500)
Regulatory Class:II
Product Code:GCJ
510(k) ReviewPanel:General & Plastic Surgery
Name of Device:L11 LED Light Source with Advanced Imaging Modality (AIM) and SafeLight Cable
Common or Usual NameLight Source, Illuminator
Classification Name:Confocal Optical Imaging1 (21 C.F.R. §876.1500)
Fiberoptic light ureteral catheter2 (21 C.F.R. §876.4020)
Light Source, Fiberoptic, Routine3 (21 C.F.R. §876.4020)
Regulatory Class:II
Product Code:OWN1
FCS2
FCW3
510(k) Review Panel:General & Plastic Surgery1
Gastroenterology/ Urology2,3

1When used for assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging

2When used to transilluminate the ureter during open or laparoscopic surgical procedures

3When used to provide standard endoscopic visible light to support real-time endoscopic visible imaging.

{5}------------------------------------------------

Name of DeviceAdvanced Imaging Modality (AIM) HD Autoclavable Laparoscope
Common orUsual NameLaparoscope
ClassificationNameLaparoscope, General and Plastic Surgery (21 C.F.R. §876.1500)
Regulatory ClassClass II
Product CodeGCJ
510(k) ReviewPanel:General & Plastic Surgery

Predicate Device(s):

Primary Predicate - Advanced Imaging Modality System (1688 4K Camera System with Advanced Imaging Modality, L11 LED Light Source with Advanced Imaging Modality), cleared via K202592.

Reference Predicate - Stryker® IRF Light Source and Safelight Cable, cleared via K142310

NOTE: The predicate device has not been subject to a design-related recall.

Device Description:

The AIM (Advanced Imaging Modality) System is an endoscopic real-time 4K visible white light and near-infrared light illumination and imaging system. The AIM (Advanced Imaging Modality) System includes the following components: (1) A Camera System for processing near-infrared and visible light images; (2) A Light Source and SafeLight Cable for emitting light within the visible light as well as near-infrared light spectrum; (3) A Laparoscope for visible light and near-infrared light illumination and imaging; (4) SPY AGENT GREEN (indocyanine green for injection, USP) an optical imaging agent used for fluorescence imaging.

Indications for Use:

Subject DevicePredicate Device
AIM System (1688, L11)AIM System (1688, L11)
This SubmissionK202592
Intended Use:Intended Use:
Endoscopic white light and near-infrared illumination andimaging during endoscopic procedures.Same as subject device
Indications for Use:Indications for Use:
L11 LED Light Source with Advanced Imaging Modality(AIM) and SafeLight CableL11 LED Light Source with AIM and SafeLight Cable
Upon intravenous administration of SPY AGENTTM GREEN(indocyanine green for injection, USP), the L11 LED LightSource with AIM and SafeLightTM Cable is used with SPYAGENT GREEN to provide real-time endoscopic visible andnear infrared fluorescence imaging. The L11 LED LightUpon intravenous administration of SPY AGENTTM GREEN(Indocyanine green for injection, USP), the L11 LED LightSource and SafeLightTM Cable is used with SPY AGENTGREEN to provide real-time endoscopic visible and nearinfrared fluorescence imaging. The L11 LED Light Source andSafeLight Cable enable surgeons to perform minimally
Subject DevicePredicate Device
AIM System (1688, L11)AIM System (1688, L11)
This SubmissionK202592
Source with AIM and SafeLight Cable enable surgeons toperform minimally invasive surgery using endoscope visuallight as well as visual assessment of vessels, blood flow andrelated tissue perfusion in adults and pediatric patients agedone month or older, and visualization of at least one of themajor extra-hepatic bile ducts (cystic duct, common bile ductand common hepatic duct) in adults and pediatric patients 12 to17 years of age, using near-infrared imaging.invasive surgery using standard endoscope visual light as wellas visual assessment of vessels, blood flow and related tissueperfusion, and at least one of the major extra-hepatic bile ducts(cystic duct, common bile duct and common hepatic duct),using near-infrared imaging.
Fluorescence imaging of biliary ducts with the L11 LED LightSource with AIM and SafeLight Cable is intended for use withstandard-of-care white light and, when indicated,intraoperative cholangiography. The devices are not intendedfor standalone use for biliary duct visualization.Fluorescence imaging of biliary ducts with the L11 LED LightSource and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperativecholangiography. The devices are not intended for standaloneuse for biliary duct visualization.
Upon interstitial administration of SPY AGENT GREEN, theL11 LED Light Source with AIM and SafeLight Cable is usedto perform intraoperative fluorescence imaging andvisualization of the lymphatic system, including lymphaticUpon interstitial administration of SPY AGENT GREEN, theL11 LED Light Source and SafeLightTM Cable is used toperform intraoperative fluorescence imaging and visualizationof the lymphatic system, including lymphatic vessels andlymph nodes.
vessels and lymph nodes.The L11 LED Light Source is also intended to transilluminatethe ureter during open or laparoscopic surgical procedures.The L11 LED Light Source is also intended to transilluminatethe ureter during open or laparoscopic surgical procedures.
1688 4K Camera System with Advanced Imaging Modality(AIM)The 1688 Video Camera is indicated for use in generallaparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopyand plastic surgery whenever a laparoscope/ endoscope/arthroscope is indicated for use. The 1688 Video Camera is1688 4K Camera System with Advanced Imaging ModalityThe 1688 Video Camera is indicated for use in generallaparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopyand plastic surgery whenever a laparoscope/ endoscope/arthroscope is indicated for use.
indicated for use in adults and pediatric patients aged onemonth or older.A few examples of the more common endoscope surgeries areLaparoscopic cholecystectomy, Laparoscopic hernia repair,Laparoscopic appendectomy, Laparoscopic pelvic lymph node
A few examples of the more common endoscope surgeries areLaparoscopic cholecystectomy, Laparoscopic hernia repair,Laparoscopic appendectomy, Laparoscopic pelvic lymph nodedetection, Laparoscopically assisted hysterectomy,Laparoscopic and thorascopic anterior spinal fusion, Anteriorcruciate ligament reconstruction, Knee arthroscopy, Small jointarthroscopy, Decompression fixation, Wedge resection, Lungbiopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis,Internal mammary artery dissection for coronary artery bypass,Coronary artery bypass grafting where endoscopicvisualization is indicated and Examination of the evacuatedcardiac chamber during performance of valve replacement.detection, Laparoscopically assisted hysterectomy,Laparoscopic and thorascopic anterior spinal fusion, Anteriorcruciate ligament reconstruction, Knee arthroscopy, Small jointarthroscopy, Decompression fixation, Wedge resection, Lungbiopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis,Internal mammary artery dissection for coronary artery bypass,Coronary artery bypass grafting where endoscopicvisualization is indicated and Examination of the evacuatedcardiac chamber during performance of valve replacement.The users of the 1688 Video Camera are general surgeons,gynecologists, cardiac surgeons, thoracic surgeons, plasticsurgeons, orthopedic surgeons, ENT surgeons and urologists.
The users of the 1688 Video Camera are general and pediatricsurgeons, gynecologists, cardiac surgeons, thoracic surgeons,plastic surgeons, orthopedic surgeons, ENT surgeons andurologists.

{6}------------------------------------------------

stryker

{7}------------------------------------------------

stryker

Subject DevicePredicate Device
AIM System (AIM HD Autoclavable Laparoscope)Infrared Fluorescence System (IRF Laparoscope)
This SubmissionK142310
Intended Use:Intended Use:
Endoscopic white light and near-infrared illumination andimaging during endoscopic procedures.Same as subject device
Indications for Use:Indications for Use:
The AIM HD Autoclavable Laparoscope is intended to be usedfor gynecological and general procedures that clinicians deemappropriate for adults and pediatric patients aged one month orolder, when the dimensions of the AIM HD AutoclavableLaparoscope are appropriate for the patient size and anatomy.The Stryker® IRF Laparoscope is intended to be used forgynecological and general procedures that clinicians deemappropriate for the patient.

Comparison of Technological Characteristics with the Predicate Device:

Subject DevicePredicate Device
ItemAIM System (1688, L11)This SubmissionAIM System (1688, L11)K202592
ManufacturerStrykerSame as subject device
Imaging ModesWhite LightNear-infrared - fluorescenceNear-infrared - transilluminationSame as subject device
Device SystemComponentsCamera SystemLight Source and Light CableLaparoscopesIRIS Ureteral KitSame as subject device
Imaging AgentSPY AGENT GREENSame as subject device
Principles of OperationsVia an optical endoscope and coupler, light is projected from alight source onto one or more complementary metal oxidesemiconductor image sensors which acquire a continuousstream of image data. The image data is processed to provide avideo stream that is then sent to a display for viewing.Same as subject device
Safety StandardsIEC 60601-1IEC 60601-2-18IEC 60601-1-2IEC 60825-1Same as subject device
Camera SystemImage Processing/ Video OutputResolutionSame as subject device
Digital4K (up to 3840 x 2160)Same as subject device
Light SourceLight Source/ LaserSame as subject device
RGB LEDsInfrared LaserSame as subject device
Infrared WavelengthsSame as subject device
806nm (used for NIRfluorescence)830nm (used for NIRtransillumination)Same as subject device
Laser Safety ClassificationSame as subject device
Class 1MSame as subject device

{8}------------------------------------------------

Subject DevicePredicate Device
ItemAIM System(AIM HD Autoclavable Laparoscope)This SubmissionInfrared Fluorescence System (IRF Laparoscope)K142310
ManufacturerStrykerSame as subject device
Endoscope TypeRigid rod lensSame as subject device
Outer Diameter5.4mm, 10mm10mm
Working Length300 mm for 5.4mm diameter laparoscope330 mm for 10mm diameter laparoscope330 mm for 10mm diameter laparoscope
Direction of View0°, 30°Same as subject device
Transmission SpectrumVisible and Near-InfraredSame as subject device
BiocompatibilityISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-11Same as subject device
How ProvidedNon-SterileSame as subject device
ProcessingCleaning, Disinfection (optional), SterilizationSame as subject device

Non-Clinical Testing

The following non-clinical testing were performed and data/test reports were provided to support substantial equivalence.

  • Cytotoxicity (ISO 10993-5, MEM Elution Method)
  • Sensitization (ISO 10993-10, Guinea Pig Maximization Test)
  • Irritation or Intracutaneous Reactivity (ISO 10993-10, Intracutaneous Injection in Rabbits)
  • Acute Systemic Toxicity (ISO 10993-11, Injection in Mice)
  • Material Mediated Pyrogenicity (ISO 10993-11, Injection in Rabbits)
  • Cleaning, disinfection, and sterilization validations

Clinical Performance Data

Published clinical literature was provided to support the safety and effectiveness of the AIM System when used in the pediatric patient population.

Conclusions:

The modified AIM System is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate device. There are no different questions of safety and/or effectiveness introduced by the modified AIM System when used in the pediatric patient population as instructed during endoscopic imaging using visible white light or near-infrared imaging utilizing SPY AGENT GREEN.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.