K Number
K210047

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-08-27

(231 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stiletto Extended Dwell Catheter is inserted into a patient's vascular system for short-term use (<28 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Stiletto Extended Dwell Catheter is suitable for use with power injectors.

Device Description

The Stiletto Extended Dwell Catheter (EDC) is a sterile, single use device designed to allow users to sample blood and/or administer fluids over a short duration (<28 days) through a peripherally inserted catheter. The Stiletto EDC utilizes an introducer needle and sheathing cannula to support advancement of a single lumen, power-injectable, catheter into peripheral vasculature. Unique catheter tip geometry is employed that works in conjunction with a sheathing cannula to provide support as the catheter advances into the vessel, providing a reliable method of insertion without need of a guidewire. The catheter, hub and insertion components are integrated into a single housing to enhance ease of use.

The Stiletto EDC will be available in 8 cm and 10 cm lengths with a 20 gauge catheter, and 10 cm and 12 cm lengths with an 18 gauge catheter.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the "Stiletto Extended Dwell Catheter" based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Standard / RequirementTest NameAcceptance Criteria (Implied)Reported Device Performance
ISO 7864Needle Surface and SharpnessPassedPassed
ISO 9626Needle Surface and SharpnessPassedPassed
ISO 10555-1Needle Surface and SharpnessPassedPassed
Internal RequirementDimensions – Needle Effective LengthPassedPassed
ISO 10555-5Catheter to Needle Bevel Heel Lie DistancePassedPassed
ISO 10555-1RadiopacityPassedPassed
ASTM F640Needle Cap Removal ForcePassedPassed
Internal RequirementNeedle Cap Removal ForcePassedPassed
ISO 10555-5Flashback DetectionPassedPassed
ANSI/AAMI HE75Assembly Force/Safety Activation ForcePassedPassed
ISO 23908Assembly Force/Safety Activation ForcePassedPassed
Internal RequirementInsertion Tool Withdrawal ForcePassedPassed
Internal RequirementCatheter Kink DiameterPassedPassed
FDA Guidance DocumentDimensions – Catheter Effective LengthPassedPassed
ISO 10555-5Dimensions – Catheter ODPassedPassed
ISO 10555-1Ink PermanencePassedPassed
ISO 10555-1LeakPassedPassed
Internal RequirementPump FlowPassedPassed
Internal RequirementAssembly Tensile – Cannula – PEEK to Stainless SteelPassedPassed
ISO 23908Needle Safety Override ForcePassedPassed
ANSI/AAMI HE75Needle Safety Override ForcePassedPassed
ISO 7864Assembly Tensile – Needle to Top CasePassedPassed
ISO 9626Needle Tubing StiffnessPassedPassed
Internal RequirementAssembly Tensile – Cannula – Stainless Steel to Thumb SlidePassedPassed
ISO 9626Cannula Resistance to BreakagePassedPassed
ISO 9626Needle Resistance to BreakagePassedPassed
ISO 10555-5Needle LooseningPassedPassed
FDA Guidance DocumentPriming VolumePassedPassed
Internal RequirementGravity FlowPassedPassed
Internal RequirementCatheter CollapsePassedPassed
Internal RequirementDimensions – Catheter Outer Diameter (Optical Micrometer)PassedPassed
ISO 10555-1Assembly Tensile – Catheter Shaft to Core HubPassedPassed
ISO 10555-1Catheter Shaft TensilePassedPassed
ISO 10555-1Catheter BurstPassedPassed
ISO 80369Luer/Hub Evaluations Per ISO 80369PassedPassed
Biocompatibility
CytotoxicityMEM ElutionNon-cytotoxicNon-cytotoxic
SensitizationMagnusson-Kligman MaximizationNon-sensitizerNon-sensitizer
IrritationIntracutaneous ReactivityNon-irritantNon-irritant
Systemic Toxicity (Acute)Systemic InjectionNon-toxicNon-toxic
PyrigenicityMaterial Mediated PyrogenNon-pyrogenicNon-pyrogenic
HemolysisHemolysis (indirect)Non-hemolyticNon-hemolytic
HemolysisHemolysis (direct)Non-hemolyticNon-hemolytic
HemocompatibilityPartial Thromboplastin TimeMinimal ActivatorMinimal Activator
HemocompatibilityComplement ActivationSimilar to Comparison DeviceSimilar to Comparison Device
HemocompatibilityDog ThrombogenicityEquivocalEquivocal
ImplantationImplantationNon-irritantNon-irritant (with rationalization from previous testing)
Toxicity (various) and CarcinogenicityChemical evaluation of leachates and Toxicological Risk AssessmentMOS > 1 for all compounds and populationsMOS > 1 for all compounds and populations
Particulate MatterSizing and Counting Particulate Matter: Light Obscuration MethodAcceptable per USP <788> and Legally Marketed Comparative DeviceAcceptable per USP <788> and Legally Marketed Comparative Device

2. Sample Size Used for the Test Set and the Data Provenance

The provided text details bench testing and biocompatibility testing. It does not describe a clinical test set in the traditional sense of patient data.

  • Bench Testing: The specific sample sizes for each bench test are not provided in the document. The provenance is implied to be laboratory testing of the device itself.
  • Biocompatibility Testing: The sample sizes for these tests are not explicitly stated, but they would typically involve material samples of the device components. The provenance is implied to be laboratory testing in accordance with ISO 10993 standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. The study relies on objective bench test measurements and established biological safety protocols, not subjective expert assessment of a test set in the context of diagnostic or interpretive performance.

4. Adjudication Method for the Test Set

Not applicable. There is no expert adjudication method described as the performance is based on objective measurements against standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This is a medical device (catheter) and the provided documentation focuses on its physical and biological performance, not an AI or imaging diagnostic tool. Therefore, an MRMC study is not relevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Not applicable. This document is about a physical medical device (catheter), not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established by:

  • International Standards: Adherence to various ISO and ASTM standards (e.g., ISO 10555-1, ISO 10555-5, ISO 9626, ISO 7864, ISO 80369, ASTM F640, ANSI/AAMI HE75, ISO 23908).
  • FDA Guidance Documents: Compliance with specific FDA guidance.
  • Internal Requirements: Company-defined performance specifications.
  • Biocompatibility Standards: Adherence to ISO 10993 series and USP <788> for particulate matter.
  • Predicate Device Comparison: Performance is often benchmarked against the "PowerGlide Pro™ Midline Catheter" (K162377) to demonstrate substantial equivalence, particularly in terms of flow rates and other functional characteristics.

8. The Sample Size for the Training Set

Not applicable. There is no mention of a training set as this is a physical medical device, not a machine learning model.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set involved. The evaluation is based on direct physical and biological testing of the device itself.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 27, 2021

Piper Access, LLC Jay Muse President and CEO 6030 W. Harold Gatty Dr. Salt Lake City, Utah 84116

Re: K210047

Trade/Device Name: Stiletto Extended Dwell Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: PND Dated: January 7, 2021 Received: January 8, 2021

Dear Jay Muse:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210047

Device Name

Stiletto Extended Dwell Catheter

Indications for Use (Describe)

The Stiletto Extended Dwell Catheter is inserted into a patient's vascular system for short-term use (<28 days) to sample blood or administer fluids intravenously. These catheters may patient population with consideration given to adequacy of vascular anatomy and appropriateness of the Stiletto Extended Dwell Catheter is suitable for use with power injectors.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Piper Access. The logo consists of a stylized, curved line drawing above the company name. The word "PIPER" is in gray, while "ACCESS" is in orange, with a trademark symbol next to it.

510(k) Summary (21 CFR 807.92(c))

SUBMITTER I.

Submitter Name:Piper Access, LLC
Address:6030 West Harold Gatty Dr.Salt Lake City, UT 84116
Telephone:801-210-2886
Contact Person:Jay Muse
Email of Contact:jay.muse@piperaccess.com
Date Prepared:26 August 2021

II. DEVICE

Trade Name:Stiletto Extended Dwell Catheter
Common/Usual Name:Midline Catheter
Regulation Name:Intravascular catheter
Regulation Number:21 CFR 880.5200
Product Code:PND
Device Class:Class II
Review Panel:General Hospital

III. PREDICATE DEVICE

Predicate Name:PowerGlide Pro™ Midline Catheter
Common/Usual Name:Intravascular Catheter, Therapeutic
Short-Term Less Than 30 Days
Classification Name:Intravascular catheter
Premarket Notification:K162377
Product Code:FOZ
Manufacturer:Bard Access Systems, Inc.

DEVICE DESCRIPTION IV.

The Stiletto Extended Dwell Catheter (EDC) is a sterile, single use device designed to allow users to sample blood and/or administer fluids over a short duration (<28 days) through a peripherally inserted catheter. The Stiletto EDC utilizes an introducer needle and sheathing cannula to support advancement of a single lumen, power-injectable, catheter into peripheral vasculature. Unique catheter tip geometry is employed that works in conjunction with a sheathing cannula to provide support as the catheter advances into the vessel, providing a reliable method of insertion without need of a guidewire. The

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Piper Access. The logo consists of a stylized bird-like graphic above the text "PIPERACCESS™". The graphic is a curved line, with the top portion in orange and the bottom portion in gray. The text "PIPERACCESS" is in a sans-serif font, with "PIPER" in gray and "ACCESS" in orange, and the trademark symbol is in black.

catheter, hub and insertion components are integrated into a single housing to enhance ease of use.

The Stiletto EDC will be available in 8 cm and 10 cm lengths with a 20 gauge catheter, and 10 cm and 12 cm lengths with an 18 gauge catheter.

INDICATIONS FOR USE V.

The Stiletto Extended Dwell Catheter is inserted into a patient's vascular system for short-term use (<28 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Stiletto Extended Dwell Catheter is suitable for use with power injectors.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

The following table provides a comparison of technological characteristics basis for a substantial equivalence determination utilizing the PowerGlide Pro™ Midline Catheter as the predicate device.

AttributePredicate DeviceSubject Device
Device NamePowerGlide ProTMMidline CatheterStiletto Extended DwellCatheterSubstantiallyEquivalent orIdentical toPredicate
Variants• 18 Gauge Catheter• 20 Gauge Catheter• 22 Gauge Catheter• 18 Gauge Catheter• 20 Gauge CatheterDifferent
AttributePredicate DeviceSubject DeviceSubstantiallyEquivalent orIdentical toPredicate
Device NamePowerGlide Pro™Midline CatheterStiletto Extended DwellCatheter
Indications forUseThe PowerGlide Pro™Midline Catheter isintended to be inserted inthe patient's vascularsystem for short term use(less than 30 days) tosample blood oradminister fluidsintravenously. Thesecatheters may be used forany patient populationwith consideration givento adequacy of vascularanatomy andappropriateness of theprocedure. ThePowerGlide Pro™Midline Catheter issuitable for use withpower injectors.The Stiletto ExtendedDwell Catheter isinserted into a patient'svascular system for short-term use (<28 days) tosample blood oradminister fluidsintravenously. Thesecatheters may be used forany patient populationwith consideration givento adequacy of vascularanatomy andappropriateness of theprocedure. The StilettoExtended Dwell Catheteris suitable for use withpower injectors.Different
Duration of UseShort Term (less than 30days)Short Term (less than 28days)Different
Insertion SitePeripheralPeripheralIdentical
InsertionMethodPercutaneous, over aguidewirePercutaneous, over aneedle (utilizing asheathing cannula)Different
Specific DrugUseNoneNoneIdentical
LumensSingle LumenSingle LumenIdentical
AttributePredicate DeviceSubject DeviceSubstantiallyEquivalent orIdentical toPredicate
Device NamePowerGlide Pro™Midline CatheterStiletto Extended DwellCatheterSubstantiallyEquivalent orIdentical toPredicate
CatheterDimensions18 Gauge CatheterLength: 8 and 10 cm20 Gauge CatheterLength: 8 and 10 cm22 Gauge CatheterLength: 8 cm18 Gauge CatheterLength: 10 and 12 cm20 Gauge CatheterLength: 8 and 10 cmDifferent
NeedleDimensions18 Gauge DeviceNeedle OD: 21 gauge20 Gauge DeviceNeedle OD: 22 gauge22 Gauge DeviceNeedle OD: 24 gauge18 Gauge DeviceNeedle OD: 22 gauge20 Gauge DeviceNeedle OD: 24 gaugeDifferent
Catheter Base /Needle BaseMaterialsShaft tubing:PolyurethaneLuer hub: PolyurethaneNeedle: Stainless SteelShaft tubing:PolyurethaneLuer hub: PolyurethaneNeedle: Stainless SteelIdentical
Power InjectionMaximum FlowRate18 Gauge Catheter:7mL/sec20 Gauge Catheter:5mL/sec22 Gauge Catheter:2mL/sec18 Gauge Catheter:7mL/sec20 Gauge Catheter:5mL/secDifferent
Needle BevelB BevelB BevelIdentical
Single-UseYesYesIdentical
TargetPopulationAny patient populationwith consideration givento adequacy of vascularanatomy andappropriateness of theprocedureAny patient populationwith consideration givento adequacy of vascularanatomy andappropriateness of theprocedureIdentical
AttributePredicate DeviceSubject Device
Device NamePowerGlide Pro™Midline CatheterStiletto Extended DwellCatheterSubstantiallyEquivalent orIdentical toPredicate
Shelf Life24 Months60 Days:20G-10cm6 Months:18G - 10cm18G - 12cm20G-8cmDifferent
PackagingDevice provided in asterile pouch. Entirepouch is EO sterilizedDevice provided in asterile pouch. Entirepouch is EO sterilizedIdentical
SterilizationProvided Sterile (EO)Provided Sterile (EO)Identical
BiocompatibilityISO 10993 CompliantISO 10993 CompliantIdentical

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a curved line that is partially orange and partially gray. The orange portion of the line is on the left side of the image, and the gray portion is on the right. The line curves upwards and then downwards, creating a wave-like shape. The line appears to be a drawing or illustration.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a curved line that is partially orange and partially gray. The line starts as gray, curves upwards and to the left, and then transitions to orange as it curves back down and to the right. The remaining portion of the line is gray and curves to the right.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a curved line that resembles a swan's neck. The line is mostly gray, but the top portion of the curve is orange. The line is smooth and continuous, and it appears to be drawn with a single stroke.

Discussion

  • . Trade Name: Both trade names are appropriate for identifying the subject and predicate devices. The trade names are therefore substantially equivalent.
  • Product Code, Duration of Use: The PND Product Code is specific to midline . catheters, which was determined by the Agency to be more appropriate for the subject device than the more general FOZ Product Code. The PND Product Code specifies short term use as <28 days, rather than <30 days for FOZ. Therefore, the duration of use for the subject device was adjusted accordingly. The slightly shorter patient contact time does not introduce any new risks to safety or efficacy for the subject device, as verified through biocompatibility testing per ISO 10993 series of applicable standards, and is therefore substantially equivalent to the predicate device.
  • Variants: The subject device size variants are bracketed by the predicate device . size variants. All device variants were thoroughly tested in their worst case configurations via benchtop, simulated use/human factors, and biocompatibility testing (See Section VII), and no new risks of safety or efficacy were identified. The subject device variants are therefore substantially equivalent to the predicate device variants.
  • . Indications for Use: The indications for use are identical with the exception of the duration of use. The slightly shorter patient contact time does not introduce

{8}------------------------------------------------

Image /page/8/Picture/8 description: The image shows the logo for Piper Access. The logo features a stylized, curved line drawing above the company name. The word "PIPER" is in a light blue color, while "ACCESS" is in orange, and there is a trademark symbol to the right of the word ACCESS.

any new risks to safety or efficacy for the subject device, as discussed above. The subject device is therefore substantially equivalent to the predicate device.

  • Insertion Method: The use of a sheathing cannula to insert the subject devices . has been evaluated through both validation and verification testing. No new concerns of safety or efficacy were identified in either validation or verification testing. The subject device is therefore substantially equivalent to the predicate device.
  • . Catheter Dimensions: Outer diameters of 18G and 20G catheters were verified to be compliant with ISO 10555-5 and applicable color coding. With respect to other catheter dimensions, the subject device catheters underwent benchtop verification testing in their worst-case dimensional configurations (See Section VII for a comprehensive list) and were shown to comply with applicable performance criteria specified in ISO 10555-1 and ISO 10555-5. All predetermined acceptance criteria were met, and the slight variations in dimensions introduced no new concerns of safety or efficacy. The subject device is therefore substantially equivalent to the predicate device.
  • Needle Dimensions: The needle dimensions for the two subject device offerings . are bracketed by the predicate device offerings. Furthermore, engineering evaluations, including FMEA, dimensional, benchtop, and simulated use/human factors testing, established that the needle dimensions of the subject device are adequate for the intended use of the device. Needles were inspected and verified to be compliant with needle requirements per ISO 10555-5 and applicable portions of ISO 9626 and ISO 7864, including criteria for needle point. The subject device needle dimensions are therefore substantially equivalent to the predicate device needle dimensions.
  • . Power Injection Maximum Flow Rate: The maximum flow rate for the two comparable device sizes between the subject device and predicate devices are identical. The subject device catheters were evaluated in compliance with ISO 10555-1 for power injection and maximum flow rates and comply with all associated criteria. The predicate device offers a smaller device outside of the subject device offerings which is not applicable for comparison. The subject device power injection maximum flow rates are therefore substantially equivalent to the comparable size offerings of the predicate device.
  • . Shelf Life: The subject device and predicate device are both verified through their respective labeled shelf lives without performance degradation by benchtop evaluations following shelf life conditioning. Testing following shelf life conditioning was performed on appropriate characteristics that have the potential to be affected by time and adequately assess the potential for performance degradation. Shelf life testing raised no new concerns of safety or efficacy for the subject device, and applicable acceptance criteria were met following conditioning. The subject device and predicate device are therefore substantially equivalent for their respective labeled shelf lives.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Piper Access. The logo consists of a stylized line drawing above the company name. The line drawing is a curved line that is orange at the top and transitions to gray at the bottom. The company name, "PIPERACCESS," is written in a sans-serif font, with "PIPER" in gray and "ACCESS" in orange, followed by a trademark symbol.

VII. PERFORMANCE DATA

Bench Testing

Bench tests are listed below in Table 2.

StandardTest NameResult
ISO 7864
ISO 9626Needle Surface and SharpnessPassed
ISO 10555-1
Internal RequirementDimensions – Needle Effective LengthPassed
ISO 10555-5Catheter to Needle Bevel Heel Lie DistancePassed
ISO 10555-1RadiopacityPassed
ASTM F640
Internal RequirementNeedle Cap Removal ForcePassed
ISO 10555-5Flashback DetectionPassed

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a curved line that resembles a handwritten letter. The line is mostly gray, but a portion of the upper curve is colored orange. The line appears to be drawn with a smooth, continuous stroke, creating a flowing and elegant shape.

StandardTest NameResult
ANSI/AAMI HE75ISO 23908Assembly Force/Safety Activation ForcePassed
Internal RequirementInsertion Tool Withdrawal ForcePassed
Internal RequirementCatheter Kink DiameterPassed
FDA GuidanceDocumentDimensions – Catheter Effective LengthPassed
ISO 10555-5Dimensions – Catheter ODPassed
ISO 10555-1Ink PermanencePassed
ISO 10555-1LeakPassed
Internal RequirementPump FlowPassed
Internal RequirementAssembly Tensile – Cannula – PEEK toStainless SteelPassed
ISO 23908ANSI/AAMI HE75Needle Safety Override ForcePassed
ISO 7864Assembly Tensile – Needle to Top CasePassed
ISO 9626Needle Tubing StiffnessPassed
Internal RequirementAssembly Tensile – Cannula – Stainless Steelto Thumb SlidePassed
ISO 9626Cannula Resistance to BreakagePassed
ISO 9626Needle Resistance to BreakagePassed
ISO 10555-5Needle LooseningPassed
FDA GuidanceDocumentPriming VolumePassed
Internal RequirementGravity FlowPassed
Internal RequirementCatheter CollapsePassed
Internal RequirementDimensions – Catheter Outer Diameter (OpticalMicrometer)Passed
ISO 10555-1Assembly Tensile – Catheter Shaft to Core HubPassed
ISO 10555-1Catheter Shaft TensilePassed
ISO 10555-1Catheter BurstPassed
ISO 80369Luer/Hub Evaluations Per ISO 80369Passed

Clinical Testing

No Clinical Testing was performed.

Biocompatibility

Biological EffectMitigation StepResult
CytotoxicityMEM ElutionNon-cytotoxic
SensitizationMagnusson-KligmanMaximizationNon-sensitizer
IrritationIntracutaneous ReactivityNon-irritant
Systemic Toxicity (Acute)Systemic InjectionNon-toxic

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows two curved lines. The line on the left is orange and the line on the right is gray. The lines are curved in opposite directions.

Biological EffectMitigation StepResult
Material Mediated PyrogenNon-pyrogenic
HemocompatibilityHemolysis (indirect)Non-hemolytic
Hemolysis (direct)Non-hemolytic
Partial Thromboplastin TimeMinimal Activator
Complement ActivationSimilar toComparison Device
Dog ThrombogenicityEquivocal
ImplantationRationalization for evaluationwith previous testing onrepresentative test sample (SeeM114-006)Non-irritant
Acute/Subacute/SubchronicToxicityGenotoxicityChronic ToxicityCarcinogenicityAssessed through chemicalevaluation of leachates and aToxicological Risk AssessmentMOS > 1 for allcompounds andpopulations
Particulate MatterSizing and Counting ParticulateMatter: Light ObscurationMethodAcceptable per USP<788> and LegallyMarketedComparative Device

VIII. CONCLUSION

Through performance bench testing results, it has been demonstrated that the subject device is substantially equivalent to the predicate device, PowerGlide Pro™ Midline Catheter (K162377)with respect to the intended use, target populations, treatment method, and technological characteristics. The identified differences between the devices do not raise any new questions of safety or effectiveness.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).